Open Access
5 June 2015 Estimation and imaging of breast lesions using a two-layer tissue structure by ultrasound-guided optical tomography
Yan Xu, Quing Zhu
Author Affiliations +
Abstract
A new two-step estimation and imaging method is developed for a two-layer breast tissue structure consisting of a breast tissue layer and a chest wall underneath. First, a smaller probe with shorter distance source-detector pairs was used to collect the reflected light mainly from the breast tissue layer. Then, a larger probe with 9×14 source-detector pairs and a centrally located ultrasound transducer was used to collect reflected light from the two-layer tissue structure. The data collected from the smaller probe were used to estimate breast tissue optical properties. With more accurate estimation of the average breast tissue properties, the second layer properties can be assessed from data obtained from the larger probe. Using this approach, the unknown variables have been reduced from four to two and the estimated bulk tissue optical properties are more accurate and robust. In addition, a two-step reconstruction using a genetic algorithm and conjugate gradient method is implemented to simultaneously reconstruct the absorption and reduced scattering maps of targets inside a two-layer tissue structure. Simulations and phantom experiments have been performed to validate the new reconstruction method, and a clinical example is given to demonstrate the feasibility of this approach.

1.

Introduction

Diffuse optical tomography (DOT) in the near-infrared (NIR) spectrum is a promising noninvasive approach for functional breast imaging.115 However, the resolution and lesion localization are poor due to intense light scattering in tissue. Additionally, the inverse problem of DOT is ill-posed and the solution is not unique. To improve the target quantification accuracy of DOT, several research groups have explored the use of a priori information provided by MRI, x-ray, and ultrasound (US).4,712 Our group has pioneered the use of US-guided DOT by deploying many source and detection fibers around a US transducer located in the middle of a hand-held probe. In general, a semi-infinite model is used as a reasonable approximation for calibration and weight matrix calculation when the patient is in a supine position. However, when the breast tissue thickness is <2cm, the chest wall underneath the tissue layer needs to be accounted for. In this case, a two-layer tissue structure is a more precise model for more accurate imaging reconstruction.1619

In the past, we have introduced several approaches to account for chest wall effects. For example, we have implemented an analytical solution for a two-layer tissue structure with and without a tilted interface20 and developed a finite-element method to handle the reference mismatch problem due to the presence of the chest wall.21 However, these approaches require the fitting of amplitude and phase profiles of reflected light to estimate four parameters of background tissue absorption and reduced scattering coefficients of both layers. The errors of the estimated second layer absorption and reduced scattering coefficients are much larger than that of the first layer. In this paper, we introduce a new approach by utilizing a small probe with several short distance source-detector pairs to estimate the first layer absorption and reduced scattering coefficients. The estimated first layer properties are used to fit the amplitude and phase profiles of reflected light measured from a larger probe with both short and longer source-detection separations and to estimate the second layer properties. Using this approach, the unknown variables have been reduced from four to two and the results are more accurate and robust. In addition, we have implemented a two-step reconstruction using a genetic algorithm (GA) and conjugate gradient (CG) method22 to simultaneously reconstruct the absorption and reduced scattering maps of a target inside a two-layer tissue structure. Simulations and phantom experiments have been performed to validate the new reconstruction method, and a clinical example is given to demonstrate the feasibility of this approach.

2.

Methods

2.1.

Two-Layer Tissue Model

Figure 1 illustrates the structure of the two-layer tissue model. A cylinder with a 10-cm diameter was used to model a homogeneous medium, which had two layers of different absorption and reduced scattering coefficients. The first layer simulates the breast tissue layer with <2.5cm thickness and the second layer represents the chest wall of 7 cm depth. To estimate the optical properties of the two-layer medium, a four-parameter analytical fitting method was developed in our previous work.17 Briefly, by solving the two-layer diffusion equation, the measured fluence rate at the tissue surface is related to the absorption and reduced scattering coefficients of the two tissue layers.

Eq. (1)

ϕ(r,ω)=f(μa1,μs1,μa2,μs2),
where ϕ(r,ω) is the fluence rate and μa1, μs1, μa2, and μs2 are absorption and reduced scattering coefficients of the two tissue layers. A nonlinear optimization method based on Nelder-Mead simplex algorithm20,23 was used to simultaneously fit the amplitude and phase and estimate the background properties of the two layers. To improve the accuracy of the estimation, this method was modified to two-parameter fitting by taking the first layer μa1 and μs1 fitted from a smaller probe as known background values. The smaller probe consists of three sources and five detectors with source-detector separations ranging from 1.5 to 3.5 cm as shown in Fig. 2(a). The number of sources and detectors of the smaller probe was optimized based on the accuracy of the estimated optical properties of intralipid solution with different concentrations.

Fig. 1

A two-layer tissue model. h1 is the first layer thickness, h2 is the second layer thickness, and D1 is the diameter of the medium. D1 is typically 10cm.

JBO_20_6_066002_f001.png

Fig. 2

Probes used in simulations and experiments. (a) Smaller probe with short distance source-detector pairs (1.5 to 3.5 cm) for estimation of first layer optical properties. (b) Larger probe with a central slot used for ultrasound (US) transducer.

JBO_20_6_066002_f002.png

2.2.

Two-Step Reconstruction with GA

Recently, we have developed a two-step reconstruction method using GA followed by the CG method for simultaneous image reconstruction of absorption and scattering maps of targets inside a semi-infinite medium.22 Briefly, GA was used to obtain an initial estimate of the optical properties of a lesion. Because coregistered US images were used to locate the target approximately in the middle of the probe, the target center and its radius in the z direction were estimated from US images, and its radius in spatial x and y dimensions (rx ry) was iteratively estimated by GA. First, an initial population is generated, formed by certain individuals while each member is a possible solution of the optimization problem and can be represented by a vector of real numbers, for example, [μam μsm rxry], where μam and μsm are unknown maximum target absorption and reduced scattering coefficients. These individuals then undergo a set of genetic operations—selection, crossover, and mutation—in order to promote the population evolution. The error of each individual is defined as

Eq. (2)

Er=UscWX2,

Eq. (3)

W=[WA,WS],

Eq. (4)

X=[vδμadvvδDdv]=[v(μaμa0)dvv(μs1μs01)/3dv],
where Usc is the measurement, WA and WS are weight matrices for the absorption and reduced scattering coefficients, respectively, and (μa, μs) and (μa0, μs0) are the lesion and background absorption and reduced scattering coefficients, respectively. The fitness of each individual is defined as 1/Er. When the improvement in the fitness of members is <106, or the generation number exceeds 100, the generation of the populations is stopped and the selected values for the unknown parameters of the model are obtained. After that, the CG reconstruction method with the initial guess obtained from GA is applied to obtain the absorption and scattering distributions. The GA method is modified for use in the two-layer structure; the weight matrix W in Eq. (3) is calculated as
WA=G(rvj,rdi)Φ0(rsi,rvj)/DlandWS=G(rvk,rdi)Φ0(rsi,rvk)/Dl,
where Dl is the diffusion coefficient and equals D1 or D2, depending on the voxel location in the layered medium; rv represents the voxel location, rs is the source position, and rd is the detector position. In Eq. (4), μa0, μs0 are modified to μa01, μs01 and μa02, μs02, which are the background optical properties of the first and the second layer obtained from the two-parameter fitting, respectively. After the initial values were obtained from GA, they were applied to the CG reconstruction algorithm, which computes the final distribution of the lesion.

2.3.

Simulation and Experiments

In simulations, a commercial finite-element package COMSOL was employed to solve the forward diffusion equation in the frequency domain.21,24 A modulation frequency of 140 MHz was used in all simulations. A cylinder of 10 cm diameter consisting of two layers was used to model the two-layer tissue medium (Fig. 1), and 9 sources and 14 detectors were distributed on the surface in reflection geometry as shown in Fig. 2(b). The first layer thickness varied from 1.0 to 2.5 cm, and the second layer thickness was fixed to 7 cm.

In imaging reconstruction, a target of a 1.0-cm diameter sphere was embedded in the first layer. The corresponding reference data were generated from the same two-layer structure without the target inside and the perturbation was the difference between the target and the reference data sets.

In phantom experiments, a frequency domain system consisting of 14 parallel detectors and 4 laser diodes of wavelengths 740, 780, 808, and 830 nm was used to acquire data. Each laser diode was sequentially switched to nine positions on the probe. The central slot on the probe shown in Fig. 2(b) was used for the US transducer, and the source and detector fibers were distributed on both sides. Figure 2(a) shows the small probe with three central sources and five detectors to estimate the first layer background optical properties. The small probe was designed by starting a populated probe of 9 sources and 14 detectors with source and detector distances ranging from 1.5 to 4.5 cm. Then different combinations of sources and detectors were tested with an intralipid solution to select the minimal number of sources and detectors to obtain the same estimated properties as the populated probe.

The first layer of the phantom was made of an intralipid solution with drops of Indian ink. The absorption and reduced scattering coefficients were calibrated with the frequency domain system. Soft phantoms made of polyvinyl chloride-plastisol (PVCP) solution were placed underneath the solution as the second layer. To validate the simulations, experiments were made at similar optical properties and layer thicknesses. To control the properties of the first layer, the smaller probe was used to calibrate and adjust the amount of Indian ink and intralipid until reaching the required absorption and reduced scattering coefficients. Phantoms of the second layer were made of PVCP solution, which was a white opaque solution and became translucent when it was heated.25 When the solution was gradually heated, the Indian ink and titanium dioxide powder were added to control the optical absorption and reduced scattering coefficients of the phantom. The heated solution was poured into a 10-cm-diameter bowl and solidified after cooling for several hours.

In imaging experiments, a target of a 1.0-cm diameter sphere made of PVCP was embedded in the first layer. The corresponding reference data were generated from the same two-layer structure without a target inside and the perturbation was the difference between the target and the reference data sets.

Clinical experiments were performed with a system of similar electronic and optical design as that used for phantom experiments. The study protocol was approved by the UConn Health Center institution review board. All patients who participated in our study signed the informed consent. The patients’ data were taken at the lesion area and the contralateral breast of the same quadrant as the lesion. Contralateral data set was used to estimate background optical properties for weight matrix computation. The perturbations computed between lesion data and contralateral data were used for imaging reconstruction.

3.

Results

3.1.

Estimation of Optical Properties of the Two-Layer Structure

3.1.1.

Simulations

Simulations were performed using the two-layer model shown in Fig. 1. In simulations, optical properties of the first layer were assumed as known values, and the second layer properties were estimated from the two-parameter fitting described in Sec. 2.1. Three sets of typical breast tissue optical properties were used as the first layer properties of (1) μa=0.02cm1 and μs=4cm1; (2) μa=0.05cm1 and μs=6cm1; and (3) μa=0.05cm1 and μs=9cm1. Two sets of second layer optical properties of (1) μa=0.1cm1 and μs=6cm1 and (2) μa=0.2cm1 and μs=6cm1 were used as the estimate of chest wall properties (see Table 1). These values were obtained from our clinical studies of several hundred patients.8,11 Simulations were performed for first layer thicknesses of 1.0, 1.5, 2.0, and 2.5 cm, respectively. Figure 3 shows one example of estimated second layer μa and μs versus iterations. The first layer properties used in simulations were true values given in set #1 (Table 1), and the fitted results of the second layer were converged to μa=0.12cm1 and μs=6.7cm1. The true values of the second layer were μa=0.1cm1 and μs=6cm1. The average accuracy of the fitted second layer properties for each set of first layer properties is given in Table 1. The average was obtained from results of different first layer thicknesses, however, with the same first layer optical properties. On average, the fitting errors of μa and μs are <5 and 23%, respectively, for the first set of second layer properties and the corresponding errors are <17 and 42% for the second set of second layer properties (see Fig. 4). In general, the error is larger when the second layer μa is higher and the mismatch of μs between two layers is larger.

Table 1

Simulation results of accuracy of fitted second layer properties. The first layer properties are the true values given in the first column.

First layer (cm−1)Second layer (cm−1)
1. μa=0.1, μs′=6.02. μa=0.2, μs′=6.0
1. μa=0.02, μs=4.0μaμs97.9% 76.3%85.0% 63.2%
2. μa=0.05, μs=6.0μaμs97.5% 93.4%86.3% 66.1%
3. μa=0.05, μs=9.0μaμs90.0% 61.3%77.7% 43.4%

Fig. 3

Simulation results of the fitted second layer optical properties versus iterations. (a) Fitted second layer absorption coefficient versus iterations and (b) fitted reduced scattering coefficient versus iterations. The dashed lines are the calibrated values. The first layer optical properties of μa=0.02cm1 and μs=4.0cm1 were used as the initial values. The layer thickness was 1.5 cm. The fitted second layer properties were μa=0.12cm1 and μs=6.7cm1 .

JBO_20_6_066002_f003.png

Fig. 4

Simulation results of the estimated second layer properties versus different thickness of the first layer. Data estimated from three sets of first layer properties and the first set of second layer properties (Table 1) were plotted in black and those from the second set of second layer properties were plotted in gray. (a) The fitted second layer absorption coefficients and (b) the fitted second layer reduced scattering coefficients. The dashed lines are the true values.

JBO_20_6_066002_f004.png

3.1.2.

Phantom experiments

To validate the simulation results, experiments were conducted. Optical properties of the first layer were measured from the smaller probe, and the second layer properties were estimated from the two-parameter fitting approach. The calibrated intralipid solution was used as the first layer and the soft phantom was placed underneath. The thickness of the first layer was defined by the coregistered US image. The first layer had the same calibrated optical properties as the simulation, and the two phantoms used for the second layer had similar values as simulations of μa=0.1cm1 and μs=9.0cm1 and μa=0.2cm1 and μs=8.0cm1, respectively. We varied the first layer depths from 1.0 to 2.5 cm, and the results have shown that the estimation of the first layer was more accurate when the first layer was thicker (Fig. 5). For example, for the same combination of the first layer of μa=0.02cm1 and μs=4.0cm1 and the second layer of μa=0.1cm1 and μs=9.0cm1, the estimated μa for the first layer were 0.06, 0.05, 0.03, and 0.023cm1 for the layer thicknesses of 1.0, 1.5, 2.0, and 2.5 cm, respectively. After the estimation of the first layer, the large probe with the US transducer located in the central slot was used to collect the data of the homogenous layers. The estimation from the smaller probe was used as the initial values for the first layer, and the properties of the second layer were fitted by the two-parameter fitting method. Figure 6 shows an example of one fitting result of the second layer μa and μs' when the first layer of calibrated μa=0.02cm1 and μs=4.0cm1 and the second layer of μa=0.1cm1 and μs=9.0cm1 were used. The thickness of the first layer was 1.5 cm. The initial values of the first layer estimated from the smaller probe were μa=0.025cm1 and μs=5.0cm1, and the fitted results of the second layer were μa=0.103cm1 and μs=8.9cm1. The accuracy of the fitted optical properties of the second layer for each set of the experiments is given in Table 2. For each set of calibrated first layer optical properties, eight experiments of different first layer thickness and second layer properties were conducted, and the ranges of fitted optical properties obtained from the smaller probe are given in the first column. On average, the fitting error of the second layer was <15.6% for μa and 35.3% for μs when the first set of the second layer properties was used, and the error was <18.1% for μa and 37.2% for μs when the second set of the second layer properties was used (see Fig. 7). The 15.7% fitting error for the first set of lower second layer μa compared with 18.1% error of the second set of higher second layer μa is in the same trend as that observed in simulations. A much smaller difference was obtained for the fitted second layer μs 35.3% compared with 37.2%.

Fig. 5

Experimental results of estimated properties of the first layer versus layer thickness. (a) The estimated absorption coefficient and (b) the estimated reduced scattering coefficient versus the layer thickness of 1.0, 1.5, 2.0, and 2.5 cm, respectively. The dashed lines are the calibrated values of the first layer optical properties.

JBO_20_6_066002_f005.png

Fig. 6

Phantom experiment of fitted second layer properties versus iterations. (a) Fitted absorption coefficient versus iterations and (b) fitted reduced scattering coefficient versus iterations. The dashed lines are the true values. First layer properties estimated from the smaller probe were μa=0.025cm1 and μs=5.0cm1, and the layer thickness is 1.5 cm. The second layer of μa=0.1cm1 and μs=9.0cm1 was used. The fitted results of the second layer were μa=0.103cm1 and μs=8.9cm1 .

JBO_20_6_066002_f006.png

Table 2

Phantom results of accuracy of fitted second layer properties. The first layer properties are calibrated or true values given in the first column. The ranges of fitted properties from the smaller probe were given in the first column as well.

First layer (unit: cm−1)Second layer (unit: cm−1)
1. μa=0.1, μs′=9.02. μa=0.2, μs′=8.0
1. μa=0.02, μs=4.0 Fitted μa (range: 0.018 to 0.035cm1) Fitted μs (range: 3.2 to 5.0cm1)μaμs88.3% 75.3%80.0% 70.0%
2. μa=0.05, μs=6.0 Fitted μa (range: 0.045 to 0.058cm1) Fitted μs (range: 5.3 to 8.2cm1)μaμs'85.0% 56.9%83.3% 53.8%
3. μa=0.05, μs=9.0 Fitted μa (range: 0.048 to 0.060cm1) Fitted μs (range: 7.7 to 11.0cm1)μaμs'80.0% 61.9%82.5% 64.6%

Fig. 7

Fitted second layer properties versus different thickness of the first layer obtained from all sets of phantom experiments. (a) Fitted absorption coefficients and (b) fitted reduced scattering coefficients as functions of first layer thickness of 1.0, 1.5, 2.0, and 2.5 cm. Black bars were fitted optical properties corresponding to the first set of second layer properties given in Table 2, and gray bars represent fitted optical properties corresponding to the second set of second layer properties given in the same table. The dashed lines are the calibrated values.

JBO_20_6_066002_f007.png

To compare with the previous four-parameter fitting method reported in Ref. 17, four parameters (μa and μs of the two layers) were fitted from the data acquired from the large probe. The calibrated two-layer phantom optical properties were μa=0.02cm1 and μs=7.0cm1 for the first layer and μa=0.1cm1 and μs=9.0cm1 for the second layer. As shown in Fig. 8, the fitted results using the four parameters converge to first layer properties of μa=0.023cm1 and μs=6.9cm1, and second layer properties of μa=0.11cm1 and μs=4.1cm1. The absorption coefficient of the second layer is reasonable, but the scattering coefficient is far lower than the true value. When the new method was used, the first layer properties were first estimated by the smaller probe, which were μa=0.025cm1 and μs=7.2cm1. Using this result as the known parameters, the second layer properties were estimated as μa=0.11cm1 and μs=9.1cm1 (Fig. 9). One can see that the new method can obtain a much higher fitting accuracy for the second layer reduced scattering coefficient than that of the four-parameter fitting. Additionally, the new method converges within 70 iterations, which is 3.5 times faster than that of the four-parameter fitting method.

Fig. 8

Phantom results of fitted properties of both layers versus iterations using the four-parameter fitting. (a) and (c) fitted absorption coefficient versus iterations of the first layer and the second layer, respectively. (b) and (d) fitted reduced scattering coefficient versus iterations of the first layer and the second layer, respectively. The dashed lines are the calibrated values. The first layer thickness is 1.5 cm. The calibrated values for first layer were μa=0.02cm1 and μs=7cm1, and those for the second layer were μa=0.1cm1 and μs=9cm1 .

JBO_20_6_066002_f008.png

Fig. 9

Phantom results of fitted properties of the second layer versus iterations using the calibrated values of the first layer. (a) Fitted absorption coefficient versus iterations and (b) fitted reduced scattering coefficient versus iterations. The dashed lines are the calibrated values. The first layer properties of μa=0.02cm1 and μs=7.0cm1 and layer thickness of 1.5 cm were used. The estimated first layer properties by the smaller probe were μa=0.025cm1 and μs=7.2cm1. The fitted second layer properties were μa=0.11cm1 and μs=9.1cm1.

JBO_20_6_066002_f009.png

3.2.

Target Imaging Using Two-Step Reconstruction Method

3.2.1.

Simulations

In simulations, targets of 1.0 cm diameter were located at 1.0 cm depth (center location) in the first layer of thicknesses 1.5 and 2.0 cm. Because the 1.0 cm first layer thickness was too shallow to embed the 1 cm target and 2.5 cm thickness was deep enough to neglect the chest wall effect, we did not include them in the simulation and phantom experiments. In this data set, we simulated targets of absorption coefficient μa=0.16cm1 and different scattering coefficients of μs=6.0cm1 and μs=15.0cm1. The two-layer structures with optical properties used are given in Table 1. The GA fitted and the GA fitted and CG reconstructed maximum absorption and reduced scattering coefficients of the targets obtained from all sets of simulations were plotted in Fig. 10. The GA fitted parameters were applied as the initial values to the CG-based reconstruction method. The target with μa=0.16cm1 and μs=6.0cm1 embedded in the two-layer structure with the first layer properties of μa=0.02cm1 and μs=4.0cm1 and the second layer properties of μa=0.1cm1 and μs=6.0cm1 was taken as an example. The target center depth was 1.0 cm and the thickness of the first layer was 1.5 cm. The GA fitted initial parameters were μa=0.147cm1 and μs=4.86cm1, and the CG reconstructed values were μa=0.154cm1 (96%) and μs=6.5cm1 (108.3%). The absorption and scattering distributions are shown in Fig. 11. On average, the accuracy of the reconstructed μa for the targets was >88.3%. The accuracy of reconstructed μs was 106% using calibrated μs=6.0cm1 and 76.7% using calibrated μs=15.0cm1, respectively.

Fig. 10

Genetic algorithm (GA) fitted (black) and GA fitted and conjugate gradient (CG) reconstructed (gray) maximum (a) absorption coefficient and (b) reduced scattering coefficient of the simulated targets located in 1.0 cm center depth obtained from all sets of optical properties of the two-layer structure with 1.5 and 2.0 cm first layer thickness.

JBO_20_6_066002_f010.png

Fig. 11

Simulated (a) absorption and (b) reduced scattering coefficient maps of a target with μa=0.1cm1 and μs=6.0cm1 located in the 1.0 cm center depth of the first layer. The first layer thickness was 1.5 cm and properties were μa=0.02cm1 and μs=4.0cm1. The second layer properties were μa=0.1cm1 and μs=6.0cm1. Tomography images are shown in six slices at different depths from 0.5 to 3 cm with 0.5 cm increment in depth.

JBO_20_6_066002_f011.png

3.2.2.

Phantom experiments

Phantom experiments were performed to validate the simulations. Two targets, one with μa=0.18cm1 and μs=11cm1 and another one with μa=0.14cm1 and μs=15cm1, were tested in the two-layer turbid medium. Just as with the simulations, the initial values of μa and μs of the targets were fitted by GA algorithm, and these values were applied to the two-layer CG-based reconstruction method. Figure 12 shows the GA fitted initial and GA fitted and CG reconstructed properties. On average, the accuracy of the reconstructed μa for the target of μa=0.18cm1 was 105.7%, and for μa=0.14cm1, it was 95.6%, and that of the reconstructed μs was 101.5% using calibrated μs=11cm1 and 108.6% using calibrated μs=15cm1. Thus, the average error of reconstructed target μa is <6% and μs is <9% under various first layer thicknesses and first layer and second layer properties.

Fig. 12

GA fitted (black) and GA fitted and CG reconstructed (gray) maximum (a) absorption coefficient and (b) reduced scattering coefficient of the phantom targets located in 1.0 cm depth obtained from all sets of optical properties of the two-layer structure with 1.5 and 2.0 cm first layer thickness.

JBO_20_6_066002_f012.png

Figure 13 shows the reconstructed images of an example with the target μa=0.18cm1 and μs=11cm1 embedded in a 1.5-cm-thick intralipid solution of μa=0.02cm1 and μs=4.0cm1 using a phantom of μa=0.2cm1 and μs=8.0cm1 as the second layer underneath. The fitted first layer properties from the smaller probe were μa=0.029cm1 and μs=3.87cm1. The GA fitted initial parameters were μa=0.21cm1 and μs=10.07cm1, and the CG reconstructed values were μa=0.204cm1 (113%) and μs=10.38cm1 (94.4%). Figure 13(a) is the coregistered US image, Fig. 13(b) is the target absorption map, and Fig. 13(c) is the scattering map.

Fig. 13

(a) Coregistered US image and CG reconstructed (b) absorption and (c) reduced scattering coefficient maps of a target with μa=0.18cm1 and μs=11.0cm1 located at the 1.0 cm depth of the first layer. The fitted first layer properties estimated from the smaller probe were μa=0.029cm1 and μs=3.87cm1. The first layer thickness was 1.5 cm with optical properties of μa=0.02cm1 and μs=4.0cm1, and the second layer properties were μa=0.2cm1 and μs=8.0cm1.

JBO_20_6_066002_f013.png

3.3.

Clinical Example

A 37-year-old patient who had an advanced cancer with mixed ductal and lobular features was imaged with an NIR system of 9 sources and 14 detectors.26 A smaller probe was constructed by using a hard sponge with a center source to replace the US transducer slot and four detectors on the imaging probe to estimate the first layer properties on the contralateral normal breast. As shown in Fig. 14(a) US image, the malignant lesion was located on top of the chest wall marked by the white arrows. The center of the lesion is 1.0cm and the chest wall is 1.5cm in depth. First, the breast tissue background properties of contralateral side were estimated as μa=0.04cm1 and μs=4.8cm1 using the center source and the four detectors as an equivalent smaller probe. Then the properties of the chest wall (second layer) were fitted as μa=0.15cm1 and μs=11.0cm1 using the data acquired from the entire probe at the contralateral side. After that, using the two-layer structure and the estimated layer properties as the background, the GA fitted initial values of the lesion were μa=0.19cm1 and μs=8.6cm1 at 780 nm. These initial values were applied to the CG-based reconstruction method to obtain the absorption and scattering maps of the lesion at 780 nm, as shown in Figs. 14(b) and 14(c). The maximum reconstructed μa is 0.197cm1 and μs is 7.9cm1. Absorption and scattering maps of the other three wavelengths were similarly reconstructed with maximum μa and μs of 0.157 and 7.4cm1, 0.198 and 9.2cm1, and 0.233 and 9.2cm1, at 740, 808, and 830 nm, respectively. Figures 14(d)14(f) show the total hemoglobin, oxyhemoglobin, and deoxyhemoglobin maps calculated from the absorption maps of four wavelengths. The maximum concentrations of total hemoglobin and oxyhemoglobin were 100.3 and 81.0μmol/l, which were at the level of the malignant category based on the threshold obtained from >200 patients.8,11

Fig. 14

A clinical example of an advanced carcinoma. (a) Coregistered US image of the cancer located at 1.0cm depth. The chest wall is marked by the white arrows. CG reconstructed (b) absorption and (c) reduced scattering coefficient maps at 780 nm. (d) Total hemoglobin, (e) oxyhemoglobin, and (f) deoxyhemoglobin maps calculated from the absorption maps of four wavelengths of 740, 780, 808, and 830 nm.

JBO_20_6_066002_f014.png

4.

Discussion and Summary

In this paper, we have introduced a new estimation and imaging procedure suitable for imaging breast lesions presented inside a two-layer tissue structure. As validated by simulations and phantom experiments, the new procedure is more accurate and robust in recovering lesion optical properties. A clinical example is given to demonstrate the feasibility of this approach. In clinical studies, the chest wall presented in <2.0 to 2.5 cm deep, causes a fitted average breast tissue absorption higher than it should be if a semi-infinite tissue structure is used. As a result, the weight matrix computed from the estimated background properties is smaller, and the reconstructed lesion absorption is higher as compared with using accurate background properties. This could lead to false positives for cancer diagnosis if the chest wall is not accounted for.

Our new estimation and imaging procedure does not account for the errors introduced by chest-wall depth and position mismatch between lesion and reference sites. In practice, we have used real-time coregistered US to guide the probe position by matching the chest-wall depths at both lesion and reference sites. This effort minimizes the difference in chest-wall positions at both sites and produces the best imaging results.21 Therefore, the operators’ knowledge and skills are important. For the reported in vivo experiment using the new procedure, an extra step was needed to unplug the US probe and place its replacement probe with one central source when imaging normal contralateral side of the breast. This procedure adds complexity and time to the overall imaging protocol. In future clinical studies, a dual probe system is ideal for clinical operation and will be investigated in the near future.

A two-layer tissue structure is a simple approximation to model and estimate the influence of the chest wall upon the measured photon-density waves when patients are studied in reflection geometry. The interface between the breast tissue and the chest wall is not always flat and can be more complex, which may need additional effort on modeling. In Ref. 27, the possible distortion and mismatch were discussed extensively. Here, we have focused on the new estimation and imaging method for the two-layer tissue structure. However, our method can be extended to two-layer models with complex boundaries.

In summary, a new estimation and imaging procedure is investigated and it provides more accurate estimation and reconstruction of lesions presented in the layered structure. Simulations and phantom experiments have been performed to validate the new reconstruction method, and a clinical example is given to demonstrate the feasibility of this approach. A dual probe system will be investigated for easy and robust clinical operation in the near future.

Acknowledgments

The authors thank Dr. Susan Tannenbaum at UCONN Health Center for helping with patient protocol and recruiting. This research is supported by the National Institutes of Health (R01EB002136) and Seery Foundation.

References

1. 

B. J. Tromberg et al., “Imaging in breast cancer: diffuse optics in breast cancer: detecting tumors in pre-menopausal women and monitoring neoadjuvant chemotherapy,” Breast Cancer Res., 7 (6), 279 –285 (2005). BCTRD6 Google Scholar

2. 

B. Chance et al., “Breast cancer detection based on incremental biochemical and physiological properties of breast cancers: a six-year, two-site study,” Acad. Radiol., 12 (8), 925 –933 (2005). http://dx.doi.org/10.1016/j.acra.2005.04.016 1076-6332 Google Scholar

3. 

L. S. Fournier et al., “Dynamic optical breast imaging: a novel technique to detect and characterize tumor vessels,” Eur. J. Radiol., 69 (1), 43 –49 (2009). http://dx.doi.org/10.1016/j.ejrad.2008.07.038 EJRADR 0720-048X Google Scholar

4. 

B. Brooksby et al., “Imaging breast adipose and fibroglandular tissue molecular signatures by using hybrid MRI-guided near-infrared spectral tomography,” Proc. Natl. Acad. Sci. U.S.A., 103 8828 –8833 (2006). http://dx.doi.org/10.1073/pnas.0509636103 PNASA6 0027-8424 Google Scholar

5. 

R. Choe et al., “Differentiation of benign and malignant breast tumors by in-vivo three-dimensional parallel-plate diffuse optical tomography,” J. Biomed. Opt., 14 (2), 024020 (2009). http://dx.doi.org/10.1117/1.3103325 JBOPFO 1083-3668 Google Scholar

6. 

H. R. Ghadyani et al., “Characterizing accuracy of total hemoglobin recovery using contrast-detail analysis in 3D image-guided near infraredspectroscopy with the boundaryelement method,” Opt. Express, 18 (15), 15917 –15935 (2010). http://dx.doi.org/10.1364/OE.18.015917 OPEXFF 1094-4087 Google Scholar

7. 

Q. Zhu et al., “Breast cancer: assessing response to neoadjuvant chemotherapy by using US-guided near-infrared tomography,” Radiology, 266 (2), 433 –442 (2013). http://dx.doi.org/10.1148/radiol.12112415 RADLAX 0033-8419 Google Scholar

8. 

Q. Zhu et al., “The potential role of optical tomography with ultrasound localization in assisting ultrasound diagnosis of early-stage invasive breast cancers,” Radiology, 256 (2), 367 –378 (2010). http://dx.doi.org/10.1148/radiol.10091237 RADLAX 0033-8419 Google Scholar

9. 

Q. Zhang et al., “Coregistered tomographic x-ray and optical breast imaging: initial results,” J. Biomed. Opt., 10 (2), 024033 (2005). http://dx.doi.org/10.1117/1.1899183 JBOPFO 1083-3668 Google Scholar

10. 

Q. Fang et al., “Combined optical and x-ray tomosynthesis breast imaging,” Radiology, 258 (1), 89 –97 (2011). http://dx.doi.org/10.1148/radiol.10082176 RADLAX 0033-8419 Google Scholar

11. 

Q. Zhu et al., “Benign versus malignant breast masses: optical differentiation with US-guided optical imaging reconstruction,” Radiology, 237 (1), 57 –66 (2005). http://dx.doi.org/10.1148/radiol.2371041236 RADLAX 0033-8419 Google Scholar

12. 

Q. Fang et al., “Combined optical imaging and mammography of the healthy breast: optical contrast derived from breast structure and compression,” IEEE Trans. Med. Imaging, 28 (1), 30 –42 (2009). http://dx.doi.org/10.1109/TMI.2008.925082 ITMID4 0278-0062 Google Scholar

13. 

R. Choe et al., “Differentiation of benign and malignant breast tumors by in-vivo three-dimensional parallel-plate diffuse optical tomography,” J. Biomed. Opt., 14 (2), 024020 (2009). http://dx.doi.org/10.1117/1.3103325 JBOPFO 1083-3668 Google Scholar

14. 

P. Taroni et al., “Noninvasive assessment of breast cancer risk using time-resolved diffuse optical spectroscopy,” J. Biomed. Opt., 15 (6), 060501 (2010). http://dx.doi.org/10.1117/1.3506043 JBOPFO 1083-3668 Google Scholar

15. 

A. Cerussi et al., “Predicting response to breast cancer neoadjuvant chemotherapy using diffuse optical spectroscopy,” Proc. Natl. Acad. Sci., 104 (10), 4014 –4019 (2007). http://dx.doi.org/10.1073/pnas.0611058104 PNASA6 0027-8424 Google Scholar

16. 

M. Das and Q. Zhu, “Optical tomography for breast cancer imaging using a two-layer tissue model to include chest-wall effects,” Proc. SPIE, 6511 65112L (2007). http://dx.doi.org/10.1117/12.709732 PSISDG 0277-786X Google Scholar

17. 

C. Xu et al., “Image reconstruction method for a two-layer tissue structure accounts for chest-wall effects in breast imaging,” J. Biomed. Opt., 13 (6), 064029 (2008). http://dx.doi.org/10.1117/1.3041497 JBOPFO 1083-3668 Google Scholar

18. 

A. Kienle et al., “Noninvasive determination of the optical properties of two-layered turbid media,” Appl. Opt., 37 (4), 779 –791 (1998). http://dx.doi.org/10.1364/AO.37.000779 APOPAI 0003-6935 Google Scholar

19. 

S. H. Tseng et al., “Quantitative spectroscopy of superficial turbid media,” Opt. Lett., 30 (23), 3165 –3167 (2005). http://dx.doi.org/10.1364/OL.30.003165 OPLEDP 0146-9592 Google Scholar

20. 

M. Das, C. Xu and Q. Zhu, “Analytical solution for light propagation in a two-layer tissue structure with a tilted interface for breast imaging,” Appl. Opt., 45 (20), 5027 –5036 (2006). http://dx.doi.org/10.1364/AO.45.005027 APOPAI 0003-6935 Google Scholar

21. 

Y. Ardeshirpour, M. Huang and Q. Zhu, “Effect of the chest wall on breast lesion reconstruction,” J. Biomed. Opt., 14 (4), 044005 (2009). http://dx.doi.org/10.1117/1.3160548 JBOPFO 1083-3668 Google Scholar

22. 

B. Tavakoli and Q. Zhu, “Two-step reconstruction method using global optimization and conjugate gradient for ultrasound-guided diffuse optical tomography,” J. Biomed. Opt., 18 (1), 016006 (2013). http://dx.doi.org/10.1117/1.JBO.18.1.016006 JBOPFO 1083-3668 Google Scholar

23. 

J. C. Lagarias et al., “Convergence properties of the Nelder-Mead simplex method in low dimensions,” SIAM J. Optim., 9 (1), 112 –147 (1998). http://dx.doi.org/10.1137/S1052623496303470 SJOPE8 1095-7189 Google Scholar

24. 

M. Huang et al., “Simultaneous reconstruction of absorption and scattering maps with ultrasound localization: feasibility study using transmission geometry,” Appl. Opt., 42 (19), 4102 –4114 (2003). http://dx.doi.org/10.1364/AO.42.004102 APOPAI 0003-6935 Google Scholar

25. 

G. M. Spirou et al., “Optical and acoustic properties at 1064 nm of polyvinyl chloride-plastisol for use as a tissue phantom in biomedical optoacoustics,” Phys. Med. Biol., 50 (14), N141 (2005). http://dx.doi.org/10.1088/0031-9155/50/14/N01 PHMBA7 0031-9155 Google Scholar

26. 

C. Xu et al., “Toward miniature diffuse optical tomography system for assessing neoadjuvant chemotherapy,” (2014). Google Scholar

27. 

Y. Ardeshirpour and Q. Zhu, “Optical tomography method that accounts for tilted chest wall in breast imaging,” J. Biomed. Opt., 15 (4), 041515 (2010). http://dx.doi.org/10.1117/1.3449570 JBOPFO 1083-3668 Google Scholar

Biography

Yan Xu obtained her PhD from the Electrical Engineering Department of the University of Connecticut. She has been working on ultrasound-guided diffuse wave optical tomography for breast cancer detection and diagnosis during her PhD study at UConn. Currently, she is working as a development engineer at IBM.

Quing Zhu obtained her PhD from the University of Pennsylvania and is currently a professor in the Departments of Electrical and Biomedical Engineering at the University of Connecticut. She is a leading researcher in combining ultrasound and NIR diffuse light tomography for breast cancer diagnosis. She has also extended the technology to predict and monitor patients’ neoadjuvant chemotherapy response. In addition, she and her team have developed several multimodality techniques for ovarian cancer detection and diagnosis. She is a fellow of SPIE.

© 2015 Society of Photo-Optical Instrumentation Engineers (SPIE) 1083-3668/2015/$25.00 © 2015 SPIE
Yan Xu and Quing Zhu "Estimation and imaging of breast lesions using a two-layer tissue structure by ultrasound-guided optical tomography," Journal of Biomedical Optics 20(6), 066002 (5 June 2015). https://doi.org/10.1117/1.JBO.20.6.066002
Published: 5 June 2015
Lens.org Logo
CITATIONS
Cited by 11 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Absorption

Scattering

Breast

Optical properties

Tissue optics

Tissues

Calibration

Back to Top