Open Access
1 August 2016 Dual-modulation, dual-wavelength, optical polarimetry system for glucose monitoring
Author Affiliations +
Abstract
A dual modulation optical polarimetry system utilizing both laser intensity and polarization modulation was designed, built, and tested. The system was designed to reduce complexity and enhance the speed in order to facilitate the reduction of motion-induced time-varying birefringence, which is one of the major limitations to the realization of polarimetry for glucose monitoring in the eye. The high-speed less complex technique was tested using in vitro phantom studies with and without motion artifact introduced. The glucose concentration ranged from 0 to 600  mg/dl and the glucose measurements demonstrated a standard error of prediction to within 8.1  mg/dl without motion and to within 13.9  mg/dl with motion. Our feedback control systems took less than 10 ms to reach stabilization, which is adequately fast to eliminate the effect of time-varying birefringence. The results indicate that this new optical polarimetric approach has improved the speed and reduced the complexity, showing the potential for it to be used for noninvasive glucose measurements.

1.

Introduction

Diabetes afflicts an estimated 347 million people worldwide, and nearly 29.1 million people in the United States.1,2 Intensive management of blood sugars is an effective way to prevent or at least slow the progression of diabetic complications and this management requires frequent monitoring of glucose levels followed by mitigation via insulin or lifestyle changes.3 Currently, the most common commercial methods for glucose monitoring are invasive and require a blood sample from the patients’ finger or forearm each time a reading is needed. In some cases, patients use continuous glucose monitoring (CGM) approaches, which require an indwelling probe, typically inserted in the abdomen. Although the CGM approaches provide continuous data, they require frequent calibration; the probes need to be replaced typically every 3 and 7 days depending on the brand.4 Both of the finger/forearm stick and CGM approaches can be painful, uncomfortable, and can lead to infection for the end user.5 Therefore, development of a noninvasive sensor can overcome these problems, greatly enhance the experience for the end user, and potentially facilitate better management of diabetes.

Over the past few decades, many companies and universities have and are attempting to quantify glucose using various noninvasive optical methods, which include, but are not limited to Raman spectroscopy,6,7 near-infrared spectroscopy,8,9 optical coherence tomography,1012 photoacoustic spectroscopy,13 fluorescence,1416 and optical polarimetry.17,18 Each of these approaches has their strengths and drawbacks as covered not only in the above articles but also in several good review articles on the topic4,1922 and so we will not go into detail here on each of them but rather focus briefly on the polarimetric approach.

Optical polarimetry applied to glucose detection relies on optical activity or rather chirality of the glucose molecule, which rotates the polarization plane of a linearly polarized light. The amount of rotation is proportional to specific rotation, the concentration of the optically active compound, and the length of the sample. The following equation describes this interaction between an optical active molecule with polarized light, namely

Eq. (1)

[α]λ=αLC,
in which [α]λ is the specific rotation of the optically active molecule at a given wavelength (λ), α is the observed rotation, C is the sample concentration, and L is the sample path length.

Optical polarimetry for glucose monitoring has been thoroughly discussed in the literature.1732 Briefly, polarimetry was first used in the food industry in 195623 and was first proposed as a possible method for noninvasive glucose monitoring across the aqueous humor of the eye in 1982.17 The advantages that were shown over the years were that the aqueous humor glucose is correlated with blood glucose, it is a clear medium with negligible loss of polarization due to scattering effects, had the submillidegree sensitivity to measure the small rotations, and had the potential to overcome motion artifact.2431

A closed-loop dual-wavelength polarimetric system utilizing a real-time feedback controller was designed and used for both in vitro and in vivo rabbit studies.2931 The approach was sufficiently fast enough to show some success in measuring glucose concentrations in the presence of motion-induced birefringence but the overall response time of this system was roughly 300 ms, which was limited by the Faraday modulator’s (FM) modulation frequency. A second system was developed using ferrite-based FM to decrease the stabilization time.32 Although using the new ferrite FM provided improved system response speed, it still required the more complex two detectors for differentiating the two wavelength signals and suffered from electromagnetic interference and 1/f noise.

To date, these studies have been performed in vitro or using an eye coupling device on anesthetized rabbits with minimal motion and, although the results are promising, by enhancing the system speed and reducing the system complexity, a more effective and efficient system is anticipated that will provide near real time monitoring needed to overcome the noise from time-varying birefringence expected in further in vivo studies.

2.

Materials and Methods

2.1.

Optical Polarimetry System and In Vitro Test Materials

The optical configuration of our experiment is shown in Fig. 1 and includes a 635-nm laser diode with maximum power of 7 mW (Power Technology Inc., Little Rock, Arkansas) as well as a green laser diode with a maximum power of 22 mW of power at a wavelength of 515 nm (Oxxius S.A. Lannion, France). Both lasers set to output 3 mW and the power at the sample was even less and, for any human studies, would be less than the FDA standards for maximum power through the eye.33

Fig. 1

Block diagram of dual-modulation, dual-wavelength, optical polarimetry system.

JBO_21_8_087001_f001.png

The lasers were modulated using synchronous sinusoidal signals and linear amplifiers (carrier frequencies: fc1=45  kHz, fc2=82  kHz). The sinusoidal signals were generated from synchronous sine wave generator programmed in LabVIEW 10.0 (National Instruments, Austin, Texas) and implemented through a PCI-6713 card (National Instruments, Austin, Texas). The modulated beams from both sources were linearly polarized in the horizontal direction employing Glan-Thompson 100,000:1 linear polarizers (P) (Newport, Irvine, California). The individual beams propagated through respective in-house-built Faraday compensators (FCs) that serve as rotation compensators in order to achieve closed-loop feedback control. These compensators were made of 1-cm-long terbium–gallium–garnet (TGG) crystals (Deltronic Crystal Inc., Dover, New Jersey), which were inside bobbins wrapped with electrical magnetically coated coils of wire. TGG crystal was employed since it provided a high Verdet-constant to get the desired optical rotation for the generated field without losing a significant amount of optical signal transmitted through it.

The two beams were made coincident by a beam splitter/combiner. The amplitude of each beam was modulated by a ferrite-based FM at frequency of fm=8.75  kHz. The coil around the ferrite core was powered with an audio amplifier (Radio Shack, Fort Worth, Texas) connected in series with a 0.18-μF capacitor in order to achieve resonance at the modulation frequency. The modulation frequency signal was derived from the same synchronous sine wave generator which was used to power the lasers.

Following the ferrite-based FM, the light beam proceeded through a rectangular sample cell (Starna Cells Incorporated, Atascadero California) constructed of birefringent material with a path length of 1 cm. Following the sample was a second Glan-Thompson polarizer, known as the analyzer, which was oriented perpendicular to the initial polarizer in the layout. Since the combined light beams at the two wavelengths each had separate carrier modulated frequencies (fc1, fc2), detection could be performed using a single biased photo-diode light detector (Thor Labs, Inc., Newton, New Jersey), which converted the optical signals into electrical signals.

The electrical signals were amplified by a wide-bandwidth transimpedance amplifier (CVI Melles Griot, Albuquerque, New Mexico) and fed into two lock-in amplifiers (Stanford Research Systems, Sunnyvale, California). The DC output voltages from the lock-in amplifiers served as the inputs to proportional-integral-derivative (PID) controllers programmed in LabVIEW (10.0, 32 bit, National Instruments, Austin, Texas) using a field programmable gate array, which provided real-time, closed-loop, feedback. The outputs from PID controllers were used to drive the respective FCs via linear driver circuitry. This circuitry provided sufficient current to each FC to produce rotations originating from the combination of the glucose and induced birefringence changes with motion artifact.

The in vitro studies were performed using the birefringent test cell, which was stationary or moved up and down and was filled with glucose-doped water solutions in the physiological range of 0 to 600  mg/dl. All samples were created from a 1000-mg/dl stock glucose solution, which was made by dissolving 1.0 g of d-glucose (EM Science) into 100-ml deionized water and was mixed with a magnetic stirrer for 8 h in order to completely dissolve the solute. The stock solution was left undisturbed for at least 3 h to allow for mutarotation before being used.

2.2.

Mathematical Description of the System

Without laser modulation, the intensity detected by the detector for each wavelength in the closed-loop system is derived by Jones vectors and matrices and can be described as2729

Eq. (2)

IE2=(ϕ2+θm22)+2ϕθmsin(2πfm)tθm2cos(4πfm)t,
where I is the intensity of the detected signal, ϕ is the rotation due to the optically active sample, θm is the modulation depth of the FM, fm is the modulation frequency from the Faraday rotator, and t is time.

For the dual-modulation, dual-wavelength system, the lasers were modulated by sinusoidal signals at frequencies of fc1=45  kHz, fc2=82  kHz, respectively, and were also synchronously modulated by the Faraday rotator at a frequency of fm=8.75  kHz. Due to the crossed-polarized nature of the alignment and according matrices for the optical components presented in the block diagram in Fig. 1, the detected signal of each wavelength can be represented as

Eq. (3)

IiVDCi(ϕi2+θm22)+(ϕi2+θm22)Vcisin2πfcit+ϕiθmVcicos2π(fcifm)tϕiθmVcicos2π(fci+fm)tθm2Vci4sin2π(fci2fm)tθm2Vci4sin2π(fci+2fm)t+2VDCiϕiθmsin(2πfmt)VDCiθm22cos(4πfmt),
(see Appendix for derivation) where Vci represents the amplitude of the sinusoidal signal at each wavelength, fci is the carrier frequency of each of the laser modulation signals, and the subscript, i, is either 1 or 2, depending on whether it is the first or second laser modulation wavelength. Thus, the major components of the detected signal at each wavelength include fc1,fc2, 2fm, fc1±2fm, fc2±2fm and noise. Equation (3) describes the detected intensity of the light from each wavelength at any instance in time.

In the presence of an optical active element such as glucose, each wavelength of light experiences rotation, ϕi, which is a constant and depending on the concentration of sample and the wavelength of light used, a different sideband frequency component fc1±fm, fc2±fm was observed, which had a proportional increase with the increase in the rotation, ϕi. Two lock-in amplifiers were used to lock into the detected signals at the frequency of interest, namely, fc1+fm and fc2fm to extract the information carried by each wavelength, which in turn was processed to obtain the concentration of the glucose in the sample. Since the lock-in amplifiers locked in to the sum of the modulation frequency and one of the carrier frequencies, phase-sensitive detection was used. In order to measure the amplitude of the signals at a specified reference frequency, the frequency and phase stability had to be maintained to prevent signal drift, and hence a synchronous sinusoidal signal generator programmed in LabVIEW 10.0 and implemented through PCI-6713 was used to generate the carrier signals, modulation signal, and reference signals.

2.3.

Theoretical Description for Overcoming Motion Artifact

Mathematically, as noted above, the operation of each of the single wavelengths is based on Eq. (3). From Eq. (3), it is evident that with any optically active sample the signal has DC terms and, at each wavelength, has major frequency components at fc1,fc2,2fm,fc1±2fm,fc2±2fm and without an optically active sample there are no components at fc1±fm,fc2±fm, because the rotation, ϕi, due to glucose is zero. So when an optically active element such as glucose is present the rotation, ϕi, will cause the intensity of detected signal to vary at the frequencies of fc1±fm,fc2±fm and the signals at these frequencies proportionally increase with the increase in the rotation, ϕi, depending on the concentration of sample and the wavelength (i=1 or i=2) of light used. So locking into one of sideband frequency signals provides glucose concentration information carried by each wavelength.

In a closed-loop system, if there is minimal birefringence (i.e., not enough to make the polarization become circular) and the birefringence of the sensing site did not change, the glucose concentration could be extracted from the feedback voltage to the FCs from either wavelength signal, since each compensates for the rotation in polarization due to glucose. In this case, ϕgi, (rotation at wavelength i due to glucose) is equal to ϕfi (rotation at wavelength i fed back to the FC), although the rotation is different for each wavelength and follows Drudes law. However, when motion artifact is introduced into the system, ϕi becomes a function of birefringence, glucose, and the feedback of the Faraday for a given wavelength: ϕi=ϕgi+ϕbiϕfi, where, ϕbi is the rotation induced by the birefringence from each wavelength, but the relationship between ϕb1 and ϕb2 are fixed as long as the wavelengths are known because birefringence is a function of both the wavelength and birefringence |neno| for a fixed path length through sample. Birefringence for an individual wavelength can be expressed as ϕb,i(t)=2πλi|neno|L(t), which is constant with wavelength at a given position, a behavior attributed to the form birefringence of the cornea.3436

For the case of minimal, nonvarying birefringence (without motion), the relation between the feedback voltage and glucose concentration is linear, a linear regression is then used to predicted glucose concentrations, and a predicted concentration model for individual wavelength takes the form: Glucosepred=slope*[V(t)]+b, where V(t) is the compensation voltage signal applied to the FC for each wavelength used to null the detected signal via a closed-loop PID controller, Glucosepred is the predicted concentration of the sample, slope and b are the slope and intercept of calibration line determined by the regression model for each wavelength. As stated, since the rotation due to glucose varies with wavelength following Drudes equation, the values for each are not identical. For the dual-wavelength system, a multiple regression model was used to determine the glucose concentration and the model had the form: Glucosepred=slope1*[V1(t)]+slope2*[V2(t)]+b, where V1(t) and V2(t) are the feedback voltage obtained to nullify the system for the corresponding dual-wavelength system and slope1, slope2, and b are the calibration coefficients calculated by multiple linear regression (MLR) model.30,31 This MLR model has been shown previously to minimize the effect of time-varying birefringence caused by motion using the dual-wavelength polarimetric system due to the difference in the signal from glucose rotation with wavelength (Drudes equation) and the relative common mode noise in each wavelength from time-varying birefringence.

3.

Results and Discussion

A series of experiments was performed on the new dual-modulation, dual-wavelength system in order to assess its time response and determine its sensitivity in vitro across the physiological range of glucose both without and with the presence of time-varying birefringence.

3.1.

Speed Determination of the System

As mentioned earlier, a previous attempt to minimize the effect of time-varying corneal birefringence in the eye that used a dual-wavelength system was only modulated at 1.09 kHz and required 100 ms time constant setting on the lock-in amplifiers, resulting in around 300 ms response time to stabilize the system.29 With our new system the time response was measured by visualizing the response speed of the feedback voltage produced by the control system. One widely used measure of the speed of response is the 10% to 90% rise time or the amount of time the system takes to go from 10% to 90% of the steady-state. For the response speed studies, 300  μs was the lock-in amplifier time constant that was used. In Fig. 2, the feedback voltages as a function of time for the single wavelength systems (515 and 635 nm, respectively) are depicted. The plot indicates that the PID control system can reach stability in less than 10 ms, which is 30 times faster compared with the real-time, dual-wavelength approach presented by Malik and Coté.29 This shows that the laser intensity modulation combined with Faraday modulation and an adaptive PID controller can improve the response speed of the system.

Fig. 2

Stabilization of 515 and 635 nm PID control system without motion.

JBO_21_8_087001_f002.png

3.2.

System Performance without Motion

Although it has been shown in the previous section that the new system can provide enhancement in speed, the performance in terms of sensitivity to glucose measurements over the same range as shown previously but requiring less time needs to be determined. The performance of the system was first tested and evaluated for glucose-doped water solutions in the motionless system. In particular, a 1-cm rectangular quartz cuvette that allowed for sample to be easily pipetted into was mounted in a fixed holder. The experiments were performed for individual glucose samples between the concentration range of 0 to 600  mg/dl. The sample cell was in a fixed position for each sample in the motionless system.

The motionless system was calibrated, and then the glucose concentrations were calculated using a simple linear regression model of the data acquired without motion.37 In each experiment, three runs of data were recorded for the two wavelengths, 635 and 515 nm, respectively. As it was shown in Sec. 2, the relation between the measured signal in volts and sample concentration is linear, so least-squares linear regression was used for the validation of the model. The predicted versus actual glucose concentrations were plotted for the single wavelength system as shown in Figs. 3 and 4, respectively. It can be observed that the plots indicate the dependency of glucose on one wavelength only; the slopes of this fitted line are slope1 = 2591.12, slope2 = 2605.89, slope3 = 2574.22 for the green (515 nm) light and for the red (635 nm) light are slope1 = 2609.30, slope2 = 2570.36, slope3 = 2598.10, which remained relatively constant. One can see that a high degree of linearity exists given the correlation coefficients for the green (515 nm) light are r1=0.9990, r2=0.9993, r3=0.9993 and for the red (635 nm) light are r1=0.9985, r2=0.9993, r3=0.9990, respectively. Further, the average standard error of prediction (SEP) for the validation of the glucose-doped water was 6.48  mg/dl for the green light and 7.18  mg/dl for the red light. The feedback voltages for both wavelengths were applied in an MLR model, which was used to predict the glucose concentration, and Fig. 5 shows the estimated glucose concentration as a function of the actual glucose concentration when both wavelengths are taken into account. The slopes of MLR model are slope11=2091.5, slope12 = 748.47, slope21=1463.52, slope22 = 1664.27, slope31=1703.72, and slope32 = 1362.41 for the different runs. Using the MLR model, the information from both wavelengths was combined to improve the results without motion and indeed one can see that a high degree of linearity still exists for all three cases since each correlation coefficient exceeds 0.9996 and the SEP is slightly less than using each wavelength alone ranging from 4.23 to 5.63  mg/dl, with no outliers present in any of three data runs.

Fig. 3

Predicted glucose concentration as a function of actual glucose concentration for the 515-nm laser without motion using a single linear regression model.

JBO_21_8_087001_f003.png

Fig. 4

Predicted glucose concentration as a function of actual glucose concentration for the 635-nm laser wavelength without motion using a single linear regression model.

JBO_21_8_087001_f004.png

Fig. 5

Predicted glucose concentration as a function of actual glucose concentration for the combined 635 and 515 nm laser wavelengths without motion using an MLR model.

JBO_21_8_087001_f005.png

3.3.

System Performance with Motion from Time-Varying Birefringence

The experiments were conducted to test the performance of the system for glucose solution in the sample cell while introducing motion into the system. For these experiments, the test cell was mounted on a programmable translation stage (Thorlabs, Newtown, New Jersey) and the stage was controlled by a T-Cube DC Servo Motor Controller (Thorlabs, Newtown, New Jersey) and allowed to move up and down to simulate the birefringence changes of the cornea due to motion artifact. The speed and distance of the translation of the motor stage were controlled by a visual basic program.

The system with motion was calibrated and the glucose concentrations were extracted using a linear regression model. This was done for glucose concentrations from 0 to 600  mg/dl.

In order to compare the results of the single wavelength closed-loop and dual-wavelength closed-loop system, actual versus predicted glucose concentration for glucose-doped water coupled with a varying birefringence signal using the single wavelength closed-loop system for each wavelength (515 and 635 nm) is plotted in Figs. 6 and 7. As shown in Figs. 6 and 7, the glucose concentrations cannot be predicted very well using a single closed-loop system, since the change in the birefringence of the sample masks the optical rotation due to glucose in the sample. Thus, the single wavelength model has an SEP over 100  mg/dl in the presence of birefringence changes due to motion.

Fig. 6

Actual versus predicted glucose concentration for glucose-doped water in the presence of time varying birefringence due to motion using a single wavelength closed-loop system for 515 nm.

JBO_21_8_087001_f006.png

Fig. 7

Actual versus predicted glucose concentration for glucose-doped water in the presence of time varying birefringence due to motion using a single wavelength closed-loop system for 635 nm.

JBO_21_8_087001_f007.png

The predicted concentrations results for the dual-wavelength experiments using MLR calibration are shown in Fig. 8. The predicted calibration results when both wavelengths are used to compensate for the correlated noise due to motion showed a high degree of linearity with all correlation coefficients exceeding 0.9972 and the standard errors of prediction of 13.0, 13.7, and 13.9  mg/dl as shown in Fig. 8. No outliers were present in any of the data sets. Although the SEP was greater than the motionless system, it was still as good as the previous slower system, and a very simple sample was still within the range needed for noninvasive glucose monitoring.

Fig. 8

Predicted glucose concentration as a function of actual glucose concentration with motion due to time-varying birefringence calculated using an MLR regression model for 635 and 515 nm wavelengths.

JBO_21_8_087001_f008.png

4.

Conclusions

In summary, a dual-modulation, dual-wavelength, near real-time closed-loop polarimeter system for glucose monitoring was presented that utilized the combination of laser modulation and Faraday polarization modulation for each wavelength simultaneously. This approach was shown to increase the speed over previous systems by 30 times with a stability of less than 10 ms. Further, the system complexity was reduced by removing one detector from the previous system configuration. Lastly, the in vitro glucose measurement results with and without motion demonstrate the SEP of the system to be within 8.07 and 13.9  mg/dl, respectively, indicating equivalent sensitivity to the previous slower systems in overcoming the noise due to motion artifact with less complexity, showing the potential for its utility as a noninvasive glucose monitoring system. In the future, the steps needed to show this approach has clinical viability are to first perform human subject tests to assess the motion artifact is removed in vivo with light that can be repeatedly coupled through the eye within and between subjects without the use of an external coupler and, second, to assess the potential time delay from blood to aqueous humor in a human.

Acknowledgments

The authors wish to acknowledge the financial support from the program of China Scholarships Council (CSC). The authors would like to thank Danial Grundan, former MS student in the Optical Biosensing Laboratory at Texas A&M, and Qiu Qi, professor of School of Photoelectronic Information at the University of Electronic Science and Technology of China, for their assistance and helpful technical discussions. As per SPIE guidelines, the authors acknowledge the SPIE Photonics West proceedings number 9715-29, in which part of this work was presented in San Francisco, California, on February 16, 2016.

Appendices

Appendix:

Appendix for Eq. (3)

The use of normalized Jones vectors and matrices provides a convenient mechanism to model the detected optical signal. Each matrix represents the operation that a specific optical component performs on the electrical field vector. The system of matrices that represents the optical train is shown in Fig. 1

E=[Ex(t)Ey(t)]=[1000]*[cos(ϕf)sin(ϕf)sin(ϕf)cos(ϕf)]*[cos(ϕg)sin(ϕg)sin(ϕg)cos(ϕg)]*[cos[θmsin(2πfmt)]sin[θmsin(2πfmt)]sin[θmsin(2πfmt)]cos[θmsin(2πfmt)]]*[0001][0VDC+Vcsin2πfct]=[VDC+Vcsin2πfct*sin[θmsin(2πfmt)+ϕgϕf]0].
The intensity (I) on the photo-detector at any instance in time is proportional to the square of the electric field. If we assume that sin(θmsin(2πfmt))θmsin(2πfmt) for θmsin(2πfmt)1, then the detected intensity can be represented as
IE2={VDC+Vcsin2πfct*sin[θmsin(2πfmt)+ϕgϕf]}2=(VDC+Vcsin2πfct)*{sin[θmsin(2πfmt)+ϕgϕf]}2=(VDC+Vcsin2πfct)*{θmsin(2πfmt)+ϕ}2=(VDC+Vcsin2πfct)*{θm2sin2(2πfmt)+2ϕθmsin(2πfmt)+ϕ2}=(VDC+Vcsin2πfct)*{θm2[1cos(2*2πfmt)]2+2ϕθmsin(2πfmt)+ϕ2}=VDC(ϕ2+θm22)+(ϕ2+θm22)Vcsin2πfct+ϕθmVccos2π(fcfm)tϕθmVccos2π(fc+fm)tθm2Vc4sin2π(fc2fm)tθm2Vc4sin2π(fc+2fm)t+2VDCϕθmsin(2πfmt)VDCθm22cos(4πfmt).

References

1. 

A. Alwan, Global Status Report on Noncommunicable Diseases 2010, World Health Organization, WHO Press, Geneva (2011). Google Scholar

2. 

Centers for Disease Control and Prevention. National Diabetes Statistics Report: Estimates of Diabetes and Its Burden in the United States, U.S. Department of Health and Human Services, Atlanta, Georgia (2014). Google Scholar

3. 

T. D. Control, and Complications Trial Research Group, “The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus,” N. Engl. J. Med., 329 977 –986 (1993). http://dx.doi.org/10.1056/NEJM199309303291401 NEJMAG 0028-4793 Google Scholar

4. 

S. K. Vashist, “Continuous glucose monitoring systems: a review,” Diagnostics, 3 385 –412 (2013). http://dx.doi.org/10.3390/diagnostics3040385 Google Scholar

5. 

T. Koschinsky and L. Heinemann, “Sensors for glucose monitoring: technical and clinical aspects,” Diabetes Metab. Res. Rev., 17 113 –123 (2001). http://dx.doi.org/10.1002/(ISSN)1520-7560 Google Scholar

6. 

A. M. Enejder et al., “Raman spectroscopy for noninvasive glucose measurements,” J. Biomed. Opt., 10 (3), 031114 (2005). http://dx.doi.org/10.1117/1.1920212 JBOPFO 1083-3668 Google Scholar

7. 

M. J. Scholtes-Timmerman et al., “Raman spectroscopy as a promising tool for noninvasive point-of-care glucose monitoring,” J. Diabetes Sci. Technol., 8 (5), 974 –979 (2014). http://dx.doi.org/10.1177/1932296814543104 Google Scholar

8. 

J. Yadav et al., “Prospects and limitations of non-invasive blood glucose monitoring using near-infrared spectroscopy,” Biomed. Signal Process. Control, 18 214 –227 (2015). http://dx.doi.org/10.1016/j.bspc.2015.01.005 Google Scholar

9. 

M. Goodarzi et al., “Multivariate calibration of NIR spectroscopic sensors for continuous glucose monitoring,” TrAC, Trends Anal. Chem., 67 147 –158 (2015). http://dx.doi.org/10.1016/j.trac.2014.12.005 TTAEDJ 0165-9936 Google Scholar

10. 

K. V. Larin et al., “Specificity of noninvasive blood glucose sensing using optical coherence tomography technique: a pilot study,” Phys. Med. Biol., 48 (10), 1371 (2003). http://dx.doi.org/10.1088/0031-9155/48/10/310 PHMBA7 0031-9155 Google Scholar

11. 

K. V. Larin, M. S. Eledrisi and M. Motamedi, “Noninvasive blood glucose monitoring with optical coherence tomography,” Diabetes Care, 25 (12), 2263 –2267 (2002). http://dx.doi.org/10.2337/diacare.25.12.2263 DICAD2 0149-5992 Google Scholar

12. 

H. Ullah et al., “Autocorrelation optical coherence tomography for glucose quantification in blood,” Laser Phys. Lett., 12 (12), 125602 (2015). http://dx.doi.org/10.1088/1612-2011/12/12/125602 1612-2011 Google Scholar

13. 

M. A. Pleitez et al., “In vivo noninvasive monitoring of glucose concentration in human epidermis by mid-infrared pulsed photoacoustic spectroscopy,” Anal. Chem., 85 (2), 1013 –1020 (2013). http://dx.doi.org/10.1021/ac302841f ANCHAM 0003-2700 Google Scholar

14. 

J. C. Pickup et al., “Fluorescence based glucose sensors,” Biosens. Bioelectron., 20 (12), 2555 –2565 (2005). http://dx.doi.org/10.1016/j.bios.2004.10.002 BBIOE4 0956-5663 Google Scholar

15. 

J. C. Pickup et al., “In vivo glucose monitoring: the clinical reality and the promise,” Biosens. Bioelectron., 20 (10), 1897 –1902 (2005). http://dx.doi.org/10.1016/j.bios.2004.08.016 BBIOE4 0956-5663 Google Scholar

16. 

J. C. Pickup et al., “Fluorescence based glucose sensors,” Biosens. Bioelectron., 20 2555 –2565 (2005). http://dx.doi.org/10.1016/j.bios.2004.10.002 BBIOE4 0956-5663 Google Scholar

17. 

B. Rabinovitch, W. F. March and R. L. Adams, “Noninvasive glucose monitoring of the aqueous humor of the eye: part I. measurement of very small optical rotation,” Diabetes Care, 5 (3), 254 –258 (1982). http://dx.doi.org/10.2337/diacare.5.3.254 DICAD2 0149-5992 Google Scholar

18. 

W. F. March, B. Rabinovitch and R. L. Adams, “Noninvasive glucose monitoring of the aqueous humor of the eye: part II. Animal studies and the scleral lens,” Diabetes Care, 5 (3), 259 –265 (1982). http://dx.doi.org/10.2337/diacare.5.3.259 DICAD2 0149-5992 Google Scholar

19. 

A. Tura, A. Maran and G. Pacini, “Non-invasive glucose monitoring: assessment of technologies and devices according to quantitative criteria,” Diabetes Res. Clin. Pract., 77 (1), 16 –40 (2007). http://dx.doi.org/10.1016/j.diabres.2006.10.027 DRCPE9 0168-8227 Google Scholar

20. 

C. E. Ferrante do Amaral and B. Wolf, “Current development in non-invasive glucose monitoring,” Med. Eng. Phys., 30 (5), 541 –549 (2008). http://dx.doi.org/10.1016/j.medengphy.2007.06.003 Google Scholar

21. 

S. K. Vashistr, “Non-invasive glucose monitoring technology in diabetes management: a review,” Anal. Chim. Acta, 750 16 –27 (2012). http://dx.doi.org/10.1016/j.aca.2012.03.043 ACACAM 0003-2670 Google Scholar

22. 

J. L. Smith, “The pursuit of noninvasive glucose: ‘Hunting the deceitful turkey’,” (2016) http://www.mendosa.com/The%20Pursuit%20of%20Noninvsive%20Glucose,%20Fourth%20Edition.pdf July ). 2016). Google Scholar

23. 

E. J. Gillham, “Photoelectric polarimeter using the Faraday effect,” Nature, 178 1412 –1413 (1956). http://dx.doi.org/10.1038/1781412b0 Google Scholar

24. 

G. L. Coté, M. D. Fox and R. B. Northrop, “Noninvasive optical polarimetric glucose sensing using a true phase measurement technique,” IEEE Trans. Biomed. Eng., 39 (7), 752 –756 (1992). http://dx.doi.org/10.1109/10.142650 IEBEAX 0018-9294 Google Scholar

25. 

Jr. M. J. Goetz, “Microdegree polarimetry for glucose detection,” Storrs, Connecticut (1992). Google Scholar

26. 

Jr. M. J. Goetz, M. D. Fox and R. B. Northrop, “Microdegree polarimetry using a diode laser for glucose detection,” in Proc. 18th IEEE Annual Northeast Bioengineering Conf., 97 –98 (1992). http://dx.doi.org/10.1109/NEBC.1992.285971 Google Scholar

27. 

B. D. Cameron and G. L. Coté, “Noninvasive glucose sensing utilizing a digital closed-loop polarimetric approach,” IEEE Trans. Biomed. Eng., 44 1221 –1227 (1997). http://dx.doi.org/10.1109/10.649993 IEBEAX 0018-9294 Google Scholar

28. 

Q. Wan, G. L. Coté and J. B. Dixon, “Dual-wavelength polarimetry for monitoring glucose in the presence of varying birefringence,” J. Biomed. Opt., 10 (2), 024029 (2005). http://dx.doi.org/10.1117/1.1891175 JBOPFO 1083-3668 Google Scholar

29. 

B. H. Malik and G. L. Coté, “Real-time, closed-loop dual-wavelength optical polarimetry for glucose monitoring,” J. Biomed. Opt., 15 (1), 017002 (2010). http://dx.doi.org/10.1117/1.3290819 JBOPFO 1083-3668 Google Scholar

30. 

B. H. Malik, C. W. Pirnstill and G. L. Coté, “Dual wavelength polarimetric glucose sensing in the presence of birefringence and motion artifact using anterior chamber of the eye phantoms,” J. Biomed. Opt., 18 (1), 017007 (2013). http://dx.doi.org/10.1117/1.JBO.18.1.017007 Google Scholar

31. 

C. W. Pirnstill et al., “In vivo glucose monitoring using dual-wavelength polarimetry to overcome corneal birefringence in the presence of motion,” Diabetes Technol. Ther., 14 (9), 819 –827 (2012). http://dx.doi.org/10.1089/dia.2012.0070 Google Scholar

32. 

D. T. Grunden, C. W. Pirnstill and G. L. Coté, “High-speed dual-wavelength optical polarimetry for glucose sensing,” Proc. SPIE, 8951 895111 (2014). http://dx.doi.org/10.1117/12.2040239 PSISDG 0277-786X Google Scholar

33. 

Compliance Guide for Laser Products, U.S. Department of Health and Human Services Public Health Service Food and Drug Administration Center for Devices and Radiological Health, Food and Drug Administration, Rockville, Maryland (1992). Google Scholar

34. 

R. Knighton, X.-R. Huang and L. A. Cavuoto, “Corneal birefringence mapped by scanning laser polarimetry,” Opt. Express, 16 (18), 13738 –13751 (2008). http://dx.doi.org/10.1364/OE.16.013738 OPEXFF 1094-4087 Google Scholar

35. 

R. Kingston, “Spectral dependence of corneal birefringence at visible wavelengths,” Invest. Ophthalmol. Vis. Sci., 43 152 (2002). IOVSDA 0146-0404 Google Scholar

36. 

K. Irsch et al., “Modeling and minimizing interference from corneal birefringence in retinal birefringence scanning for foveal fixation detection,” Biomed. Opt. Express, 2 (7), 1955 –1968 (2011). http://dx.doi.org/10.1364/BOE.2.001955 BOEICL 2156-7085 Google Scholar

37. 

C. W. Pirnstill, D. Grunden and G. L. Coté, “Polarimetric glucose sensing in vitro: a high frequency approach,” Proc. SPIE, 8591 859101 (2013). http://dx.doi.org/10.1117/12.2004481 PSISDG 0277-786X Google Scholar

Biography

Zhen Fang Yu is pursuing her PhD at the University of Electronic Science and Technology of China, with a focus on the application of polarized light for noninvasive sensing. She was a visiting student in the optical biosensing laboratory within the Department of Biomedical Engineering at Texas A&M University for two years. During this period she worked on developing an optical polarimetry system for use in noninvasive glucose monitoring. She is a current member of SPIE.

Casey W. Pirnstill is currently employed as Research Biomedical Engineer at Air Force Research Labs Biodynamics Protection Branch at WPAFB. Primary research interests include the use of optics for medical diagnostics, optical imaging, light-tissue interaction modeling/simulation, and biomedical sensing. His research has focused on developing point-of-care devices using lasers, optics/electronics. He is a member of BMES, SPIE, AEMB, member golden key honor society, SAFE Dayton Chapter, and coauthor of 40 publications, proceedings, patents, and abstracts.

Gerard L. Coté is the director for the Center for Remote Health Technologies and Systems and holds the Charles H. & Bettye Barclay Professorship within the Department of Biomedical Engineering at Texas A&M University. He is recognized as a worldwide expert in optical sensing for diagnostic and biomedical monitoring applications. He is a fellow of four societies, coauthor of over 300 publications, proceedings, patents, and abstracts, and is cofounder of four medical device companies.

© 2016 Society of Photo-Optical Instrumentation Engineers (SPIE) 1083-3668/2016/$25.00 © 2016 SPIE
Zhen Fang Yu, Casey W. Pirnstill, and Gerard L. Coté "Dual-modulation, dual-wavelength, optical polarimetry system for glucose monitoring," Journal of Biomedical Optics 21(8), 087001 (1 August 2016). https://doi.org/10.1117/1.JBO.21.8.087001
Published: 1 August 2016
Lens.org Logo
CITATIONS
Cited by 10 scholarly publications and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Glucose

Polarimetry

Birefringence

Signal detection

Modulation

Motion models

Biomedical optics

Back to Top