Open Access
10 August 2016 Miniature standoff Raman probe for neurosurgical applications
Oliver A. C. Stevens, Joanne Hutchings, William Gray, Rosa Louise Vincent, John C. Day
Author Affiliations +
Abstract
Removal of intrinsic brain tumors is a delicate process, where a high degree of specificity is required to remove all of the tumor tissue without damaging healthy brain. The accuracy of this process can be greatly enhanced by intraoperative guidance. Optical biopsies using Raman spectroscopy are a minimally invasive and lower-cost alternative to current guidance methods. A miniature Raman probe for performing optical biopsies of human brain tissue is presented. The probe allows sampling inside a conventional stereotactic brain biopsy system: a needle of length 200 mm and inner diameter of 1.8 mm. By employing a miniature stand-off Raman design, the probe removes the need for any additional components to be inserted into the brain. Additionally, the probe achieves a very low internal silica background while maintaining good collection of Raman signal. To illustrate this, the probe is compared with a Raman probe that uses a pair of optical fibers for collection. The miniature stand-off Raman probe is shown to collect a comparable number of Raman scattered photons, but the Raman signal to background ratio is improved by a factor of five at Raman shifts below ∼500  cm−1. The probe’s suitability for use on tissue is demonstrated by discriminating between different types of healthy porcine brain tissue.

1.

Introduction

Removal of intrinsic brain tumors, particularly gliomas, is a challenging process, and requires careful consideration to selectively and completely remove tumor tissue, thus improving outcome without damage to healthy brain.1 Surgical guidance is currently assisted by a variety of imaging techniques, such as pre- or intraoperative magnetic resonance imaging (MRI), computerized tomography (CT), intraoperative ultrasound, 5-aminolevulinic acid (5-ALA) fluorescent imaging,2 and reflectance spectroscopy,3 many of which have been shown to improve patient prognosis4,5 by ensuring maximal resection of radiologically identified tumor. Optical biopsies using Raman spectroscopy have been demonstrated to be a less invasive and potentially more effective alternative to traditional biopsy for a number of different applications,68 and potentially offer a reliable and lower cost method of identifying residual tumor intraoperatively, as well as defining apparent tumor edges.9 This is especially relevant to glial tumors in which edges are invasive and indistinct. In addition, an alternative to traditional biopsy is particularly important for pontine tumors due to their location in the brain stem, where any removal of tissue carries a risk of irreversible damage to the patient.

To perform “optical biopsies” in vivo, small Raman spectroscopy probes are typically required to efficiently collect a spectrum in situ. A variety of such probes for different applications have been constructed, and shown to be effective in providing spectra information of clinical value.1019

Optical fibers are an efficient method of guiding light where physical dimensions are constrained; for this reason, the majority of compact Raman probes employ fibers to carry the laser light to the sample and the Raman scattered photons back to the spectrometer. However, these fibers impart a background Raman spectrum of silica onto the spectrum of the sample, and hence Raman background caused by silica is an issue that affects most Raman probes.20 This silica background is shown in Fig. 1. The silica background is present at all Raman shifts, but is particularly prevalent at low wavenumbers. While background subtraction can be performed after collection of the spectra, this masks the problem rather than truly removing it, and care must be used in its application to prevent artefacts that confound statistical analysis occurring on the spectra. Additionally, as the total number of background counts is increased for a given amount of Raman signal, the shot noise is also increased, leading to a reduced signal-to-noise ratio (SNR). This problem can be avoided by restricting analysis to higher wavenumber regions (above 800  cm1),21,22 away from the silica fluorescence, but this risks losing some valuable chemical information at lower wavenumbers. Brain tissue in particular contains peaks potentially useful for classification at Raman shifts of 400  cm1,23 and possibly lower.

Fig. 1

Silica background present in most Raman probes. Spectrum collected from the tip of an optical fiber imaged on a desktop Raman microsope (Renishaw RM1000, 633 nm, 5  mW, 1 s). The background is present throughout all Raman shifts but is particularly large below 700  cm1

JBO_21_8_087002_f001.png

There are a number of techniques that can be used to reduce the background in a probe while still meeting the requirements for small dimensions. Some groups have used hollow-core waveguides to reduce the silica signal,24,25 but as these use a metal coating to guide the light rather than total internal reflection with an associated attenuation of the signal, resulting in long (>100  s) acquisition times.

Hollow-cored photonic crystal fibers (PCF) have been used26,27 to replace the excitation and collection fibers, but these are expensive and difficult to work with due to small core diameters and limitations of the range of wavelengths they will accept.

In this paper, we demonstrate a miniature Raman probe design that uses an alternative approach: reducing the background by avoiding the use of fibers in the main section of the probe altogether, thus reducing the internal Raman background but also eliminating the need for additional components to be inserted into the brain.

2.

Design and Construction

The probe is designed to collect Raman scattered light from within a standard stereotactic brain biopsy needle (AD-TECH medical). A picture of the assembled probe is shown in Fig. 2. The stereotactic needle is 200-mm long and 2-mm diameter.

Fig. 2

Operational assembled probe pointed at an IR viewing card.

JBO_21_8_087002_f002.png

The optical configuration of the probe is shown in Fig. 3. The probe employs a miniaturized version of “standoff” Raman probe, replacing the optical fibers with a collimated beam to direct the light down the biopsy needle. It uses a single- or few-mode fiber connected to an aspheric lens (Thor Labs model C430TME-B) for collimation of the excitation laser. This serves to reduce divergence within the probe length, and produce the desired 1.8-mm spot at the sample.

Fig. 3

Optical layout of the probe. The design requires no components in the biopsy needle, greatly increasing medical compatability and reducing Raman background.

JBO_21_8_087002_f003.png

Throughput of a probe is defined largely by its optical geometry, or etendue. This is a measure of the total light flux that can pass through an optical system. Etendue is conserved in any optical system, so the element of the design with the smallest etendue limits the overall light throughput; for this probe, this is the hollow needle section, and for many other probes, it is the optical fibers.

The etendue (G) captures this total flux by summing the total area (A) and range of solid angles (θ) that light can pass through in an optical system.

This is represented below. The elemental etendue d2G of a system is the product of the elements of area dA and dΘ. The total etendue is found by integrating this over all angles and area28

Eq. (1)

d2G=dAdΘ,

Eq. (2)

G=dAdΘ.

For a conical beam of light area S bounded by angle θ, this becomes

Eq. (3)

G=πSsin2θ.

For an optical fiber, the small angle approximation can be used [for small numerical apertures (NA)] to simplify the expression to

Eq. (4)

G=πA(NA)2,
where NA is the numerical aperture of the fiber and A is the area of the fiber core.

For the tube in the miniature standoff probe, we must find the amount of light that can pass from one end of the tube to the other. To collect light from an uncoated tube as used in a biopsy needle, there is a negligible component from reflections on the tube walls (waveguiding). In this case, the geometric etendue (G) is found by the “view factor” from a disk of area A representing the entrance of the tube, to another identical disk at the other end. This is expressed as follows:29

Eq. (5)

G=πAF12,
where F12 is the view factor between the ends of the needle. This is represented as

Eq. (6)

F12=1+14r2+12r2,
where

Eq. (7)

r=d2l,
where d is the diameter of the needle and l is the length of the needle.

Example numerical etendues for common fiber diameters and the miniature standoff probe are shown in Table 1.

Table 1

Example etendues. The miniature standoff probe has a slightly better throughput compared to a 40-μm fiber, but lower throughput than larger fibers.

ProbeEtendue (m2 sr)
Mini standoff0.161×109
40-μm fiber0.157×109
65-μm fiber0.416×109
105-μm fiber1.09×109
200-μm fiber (needle probe)3.95×109

Table 2

Comparison of quantitative peak to background and peak to noise ratios, at 30-s collection time and 20-mW laser power. The Raman peak (measured at 385  cm−1) to background is improved by a factor of around five, and peak to noise is similar between the two probes. Despite a lower throughput, performance is maintained due to greater collection area and a lower internal silica background.

ProbePeak/backgroundPeak/noise
Mini standoff1.7531.9
Needle probe0.3531.1

Thus by using the full area of the tip, the total light throughput (etendue) is comparable to a 40-μm fiber of NA 0.2, despite the reduced numerical aperture of 0.005. The design thus has the advantage of employing no optics at the distal end; the only component at the tip is the steel biopsy needle, which is already approved for medical use. There is an overall light throughput penalty compared to a fiber-based probe in a similar package (equipped with 200-μm fibers), however, the reduced unwanted Raman background results in a similar SNR to a comparable needle probe (Table 2).

The probe operates at an excitation wavelength of 830 nm to minimize the sample fluorescence.

A dichroic beamsplitter (Semrock, Inc. model LPD01-830RU-25) is used to reflect the laser light down the needle, but allow the longer Raman wavelengths to return toward the spectrometer. An additional long wavelength pass filter (Semrock, Inc. model LPD02-830RU-25) further removes unwanted elastically scattered light from the Raman signal before the collected light is passed into another lens (Thor Labs model F220FC-780), which focuses light into a 100-μm diameter fiber. A laser cleanup filter is (Semrock, Inc. model LL01-830-12.5) also fitted after the laser collimator.

The probe is attached to a Raman explorer spectrometer (Headwall Photonics) and iDus 420BR-DD (Andor Technology) camera controlled with Solis software (Andor Technology).

3.

Testing and Results

3.1.

Polytetrafluoroethylene (Teflon)

The probe was tested using a polytetrafluoroethylene (PTFE) reference sample, and compared to a fiber-based needle probe. Fiber background is sample dependent, with reflective samples causing more laser radiation to return into the probe and hence generate significantly more internal Raman background than transmissive (e.g., liquid) samples. PTFE was chosen in this case as it represents a compromise between these two extremes. The needle probe is constructed from low-OH silica fibers of numerical aperture 0.22- and 200-μm diameter. Full details of this probe are described by Day and Stone.30 Both probes were connected to the same Raman explorer spectrometer (Headwall Photonics) and iDus 420BR-DD (Andor Technology) camera, via a 1.5-m fiber patch cable. As this patch cable is behind the dichroic filters, it has negligible impact on the internal silica Raman signal generated within the probes. Laser power at the sample was identical between probes (20 mW, 830 nm at the sample), and acquisition time was set to 30 s for both probes.

Figure 4 shows a PTFE spectrum taken with the miniature standoff Raman probe, a fiber-based needle probe of the same length (20 cm), as well as reference spectrum acquired using a Renishaw RM1000 Raman microscope (5  mW, 633 nm, 10-s integration time, 5× objective). In addition, a spectrum from a commercially available Raman probe (BAC101, B&Wtek, Inc.) is included for reference, although this probe is a much larger handheld Raman probe and would be unsuitable for this application. The spectra are scaled by the principal peak at 385  cm1, which represents the collection performance of desired Raman signal from the sample for each of the probes. The background from the silica of the fiber causes significant additional background at low wavenumbers in the needle probe and the commercial BAC101 probe. In contrast, the miniature standoff Raman probe has a relatively flat background. Peak to background between the needle probe and miniature standoff probe is improved by a factor of five at 385  cm1. The SNR, calculated from the ratio of the height of the 385  cm1 peak to the standard deviation of the baseline at 1400  cm1, is similar between the two probes (Table 2).

Fig. 4

Raman spectra of the miniature standoff probe and comparable needle probe. Sample laser power was 20 mW, and collection time 30 s. Specta scaled by Raman peak at 385  cm1 and smoothed with Savitsky–Golay noise filter. The miniature standoff probe has significantly lower background at low wavenumbers than either of the other Raman probes, despite comparable desired Raman signal collection performance. A spectrum from a BAC101 (B&Wtek, Inc.) handheld Raman probe, and a Raman microscope (RM1000, Renishaw Plc.) are also included for reference.

JBO_21_8_087002_f004.png

3.2.

Use on Porcine Brain

To demonstrate the effectiveness of the probe on biological samples, it was used to distinguish between different types of healthy brain tissue. Spectra were acquired from porcine brain tissue (obtained from pigs destined for human consumption). One hundred and twenty spectra were acquired of white matter, gray matter, and blood vessels (identified by eye) from a pig brain. A laser excitation of 20 mW at the sample was used, and an acquisition time of 60 s. Laser power was restricted to 20 mW due to limitations in available equipement. These spectra are shown with background subtraction applied in Fig. 5, and without subtraction in Fig. 6. The sample fluorescence has been removed from these spectra using a gradient domain approach (outlined in Zhao et al.31) for clarity. These spectra exhibit the expected Raman bands from brain tissue, previously reported in the literature,23 with cholesterol (430, 705, 1441  cm1), fatty acid (1130, 1298, 1441  cm1), ceramide (1671  cm1), and sphingomyelin (875  cm1) Raman peaks all visible. It is also immediately apparent that there are discriminatory differences below 700  cm1, a region where the silica background of the needle probe is very prevalent, and hence this spectral information would typically be discarded.

Fig. 5

Raman spectra of porcine brain tissue (60 s, 20 mW, background subtracted). Spectra are offset for illustrative purposes. Clear differences can be seen between the tissue types across the spectral range.

JBO_21_8_087002_f005.png

Fig. 6

Raman spectra of porcine brain tissue, as Fig. 5 without background subtraction.

JBO_21_8_087002_f006.png

4.

Summary and Conclusions

A miniature Raman probe without fibers in the distal portion of the probe was demonstrated. It employs an approach not previously demonstrated in a Raman probe of this size. By using a collimated beam instead of optical fibers, some throughput is exchanged for a lower background, which can be useful in certain applications. An example of this is brain tissue where there are Raman signatures present at low wavenumber shifts. While this tradeoff will not be advantageous in all applications, this design offers another option for probe designers in applications where silica background is problematic, or where it is necessary to collect signal from within a needle without optical components at the tip. Additionally, the large spot size at the tip results in a low power density reducing the potential for harm to tissue. Flexibility is not a requirement for use with a stereotactic brain biopsy system, and the miniature standoff design is not flexible as a result; this could be a disadvantage in other applications. The lack of silica fibers greatly increases the medical compatibility of the probe, requiring no additional components to be inserted into the patient. The probe was compared to a silica fiber-based needle probe of similar dimensions, and shown to have similar signal-to-noise performance but much lower background at low wavenumbers, which allows features below 600  cm1 to be more readily investigated. The suitability of the probe for classification of brain tissue was also demonstrated using porcine tissue. It is anticipated that the probe will be used in the future for optical biopsy of human brain tissue for identification of brain tumor tissue.

Acknowledgments

The authors would like to thank Severnside Alliance for Translational Research (SATRE) Bio-E initiative and EPSRC for funding this research. Original data for this project is available at http://dx.doi.org/10.5523/bris.zwbmwp6hzes41dhyv2xvz14i1.

References

1. 

M. Glas et al., “Residual tumor cells are unique cellular targets in glioblastoma,” Ann. Neurol., 68 264 –249 (2010). http://dx.doi.org/10.1002/ana.22036 ANNED3 0364-5134 Google Scholar

2. 

W. Stummer et al., “Fluorescence-guided surgery with 5-aminolevulinic acid for resection of malignant glioma: a randomised controlled multicentre phase III trial,” Lancet Oncol., 7 392 –401 (2006). http://dx.doi.org/10.1016/S1470-2045(06)70665-9 LOANBN 1470-2045 Google Scholar

3. 

S. A. Toms et al., “Intraoperative optical spectroscopy identifies infiltrating glioma margins with high sensitivity,” Neurosurgery, 57 382 –391 (2005). http://dx.doi.org/10.1227/01.NEU.000176855.39826.2D NEQUEB Google Scholar

4. 

N. Haberland et al., “Neuronavigation in surgery of intracranial and spinal tumors,” J. Cancer Res. Clin. Oncol., 126 529 –541 (2000). http://dx.doi.org/10.1007/s004320000122 JCROD7 1432-1335 Google Scholar

5. 

N. S. Litofsky et al., “Image-guided resection of high-grade glioma: patient selection factors and outcome,” Neurosurg. Focus, 20 E16 (2006). http://dx.doi.org/10.3171/foc.2006.20.4.10 Google Scholar

6. 

D. I. Ellis et al., “Illuminating disease and enlightening biomedicine: Raman spectroscopy as a diagnostic tool,” Analyst, 138 3871 –3884 (2013). http://dx.doi.org/10.1039/c3an00698k ANALAO 0003-2654 Google Scholar

7. 

A. Nijssen et al., “Towards oncological application of Raman spectroscopy,” J. Biophotonics, 2 29 –36 (2009). http://dx.doi.org/10.1002/jbio.v2:1/2 Google Scholar

8. 

I. Barman et al., “Application of Raman spectroscopy to identify microcalcifications and underlying breast lesions at stereotactic core needle biopsy,” Cancer Res., 73 (11), 3206 –3215 (2013). http://dx.doi.org/10.1158/0008-5472.CAN-12-2313 CNREA8 0008-5472 Google Scholar

9. 

M. Jermyn et al., “Intraoperative brain cancer detection with Raman spectroscopy in humans,” Sci. Transl. Med., 7 (274), 274ra19 (2015). http://dx.doi.org/10.1126/scitranslmed.aaa2384 Google Scholar

10. 

K. A. Esmonde-White et al., “Fiber-optic Raman spectroscopy of joint tissues,” Analyst, 136 1675 –1685 (2011). http://dx.doi.org/10.1039/c0an00824a ANALAO 0003-2654 Google Scholar

11. 

J. C. C. Day et al., “A miniature confocal Raman probe for endoscopic use,” Phys. Med. Biol., 54 7077 –7087 (2009). http://dx.doi.org/10.1088/0031-9155/54/23/003 PHMBA7 0031-9155 Google Scholar

12. 

M. Kirsch et al., “Raman spectroscopic imaging for in vivo detection of cerebral brain metastases,” Anal. Bioanal.Chem., 398 1707 –1713 (2010). http://dx.doi.org/10.1007/s00216-010-4116-7 ABCNBP 1618-2642 Google Scholar

13. 

L. M. Almond et al., “Assessment of a custom-built Raman spectroscopic probe for diagnosis of early oesophageal neoplasia,” J. Biomed. Opt., 17 081421 (2012). http://dx.doi.org/10.1117/1.JBO.17.8.081421 JBOPFO 1083-3668 Google Scholar

14. 

L. M. Almond et al., “Development of fibre-optic Raman probes for in vivo diagnosis of upper gastrointestinal cancers,” Spectrosc. Eur., 23 (3), 3 –6 (2011). SPEUEF 0966-0941 Google Scholar

15. 

A. T. Harris et al., “Raman spectroscopy in head and neck cancer,” Head Neck Oncol, 2 26 (2010). http://dx.doi.org/10.1186/1758-3284-2-26 Google Scholar

16. 

C. Krafft et al., “Diagnosis and screening of cancer tissues by fiber-optic probe Raman spectroscopy,” Biomed. Spectrosc. Imaging, 1 39 –55 (2012). http://dx.doi.org/10.3233/BSI-2012-0004 Google Scholar

17. 

J. Mo, W. Zheng and Z. Huang, “Fiber-optic Raman probe couples ball lens for depth-selected Raman measurements of epithelial tissue,” Biomed. Opt. Express, 1 17 –30 (2010). http://dx.doi.org/10.1364/BOE.1.000017 BOEICL 2156-7085 Google Scholar

18. 

A. Mahadevan-Jansen et al., “Rapid communication development of a fiber optic probe to measure NIR Raman spectra of cervical tissue in vivo,” Photochem. Photobiol., 68 (3), 427 –431 (1998). http://dx.doi.org/10.1111/j.1751-1097.1998.tb09703.x PHCBAP 0031-8655 Google Scholar

19. 

O. Stevens et al., “Developing fibre optic Raman probes for applications in clinical spectroscopy,” Chem. Soc. Rev., 45 1919 –1934 (2016). http://dx.doi.org/10.1039/C5CS00850F CSRVBR 0306-0012 Google Scholar

20. 

J. Ma and Y. S. Li, “Fiber Raman background study and its application in setting up optical fiber Raman probes,” Appl. Opt., 35 2527 –2533 (1996). http://dx.doi.org/10.1364/AO.35.002527 APOPAI 0003-6935 Google Scholar

21. 

S. Koljenović et al., “Raman spectroscopic characterization of porcine brain tissue using a single fiber-optic probe,” Anal. Chem., 79 557 –564 (2007). http://dx.doi.org/10.1021/ac0616512 ANCHAM 0003-2700 Google Scholar

22. 

L. F. Santos et al., “Fiber-optic probes for in vivo Raman spectroscopy in the high-wavenumber region,” Anal. Chem., 77 6747 –6752 (2005). http://dx.doi.org/10.1021/ac0505730 ANCHAM 0003-2700 Google Scholar

23. 

C. Krafft et al., “Near infrared Raman spectra of human brain lipids,” Spectrochim. Acta, Part A, 61 1529 –1535 (2005). http://dx.doi.org/10.1016/j.saa.2004.11.017 Google Scholar

24. 

Y. Komachi et al., “Raman probe using a single hollow waveguide,” Opt. Lett., 30 2942 –2944 (2005). http://dx.doi.org/10.1364/OL.30.002942 OPLEDP 0146-9592 Google Scholar

25. 

H. Sato et al., “Raman study of brain functions in live mice and rats: a pilot study,” Vib. Spectrosc., 50 125 –130 (2009). http://dx.doi.org/10.1016/j.vibspec.2008.09.012 VISPEK 0924-2031 Google Scholar

26. 

S. O. Konorov et al., “Hollow-core photonic crystal fiber-optic probes for Raman spectroscopy,” Opt. Lett., 31 1911 –1913 (2006). http://dx.doi.org/10.1364/OL.31.001911 OPLEDP 0146-9592 Google Scholar

27. 

P. Ghenuche et al., “Kagome hollow-core photonic crystal fiber probe for Raman spectroscopy,” Opt. Lett., 37 4371 –4373 (2012). http://dx.doi.org/10.1364/OL.37.004371 OPLEDP 0146-9592 Google Scholar

28. 

M. Bass, J. M. Enoch and W. L. Wolfe, Handbook of Optics, McGraw-Hill, Inc.(2000). Google Scholar

29. 

A. Bejan and A. D. Kraus, Heat Transfer Handbook, 1 Wiley(2003). Google Scholar

30. 

J. C. C. Day and N. Stone, “A subcutaneous Raman needle probe,” Appl. Spectrosc., 67 349 –354 (2013). http://dx.doi.org/10.1366/12-06651 APSPA4 0003-7028 Google Scholar

31. 

J. Zhao et al., “Automated autofluorescence background subtraction algorithm for biomedical Raman spectroscopy,” Appl. Spectrosc., 61 1225 –1232 (2007). http://dx.doi.org/10.1366/000370207782597003 APSPA4 0003-7028 Google Scholar

Biographies for the authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Oliver A. C. Stevens, Joanne Hutchings, William Gray, Rosa Louise Vincent, and John C. Day "Miniature standoff Raman probe for neurosurgical applications," Journal of Biomedical Optics 21(8), 087002 (10 August 2016). https://doi.org/10.1117/1.JBO.21.8.087002
Published: 10 August 2016
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications and 1 patent.
Advertisement
Advertisement
KEYWORDS
Raman spectroscopy

Optical fibers

Brain

Biopsy

Tissues

Silica

Tumors

Back to Top