Open Access
5 April 2016 Measuring reactive hyperemia in the lower limb using near-infrared spectroscopy
Author Affiliations +
Abstract
Near-infrared spectroscopy (NIRS) has been used to measure reactive hyperemia following a vascular occlusion. However, the procedures and methods of analysis used have varied. The purpose of the present study is to identify reproducible methods for measuring reactive hyperemia using HbO2 NIRS signals in the calf and foot. Healthy participants (10 male, 10 female) aged 19 to 28 years performed one of two tests: reproducibility trials or elevation protocol (30 and 60 cm limb elevation above the heart). The time to 50% reperfusion (T1/2) and the second (R2q) quartile rates of reperfusion were found to be the most reproducible parameters (coefficient of variation=7.12 to 14.1%). The time to 95% reperfusion (T95) was 12.7% more reproducible on average than the previously reported parameter of time to peak hyperemia. Measures of reperfusion time and rate slowed with increasing limb elevation. Correlations were identified between the calf and foot in the measurements of R2q (R2=0.713, p=0.021), T1/2 (R2=0.673, p=0.033), and T95 (R2=0.792, p=0.006). Half and 95% recovery times and second and third quartile rates expressed good reproducibility and sensitivity to change with reduced perfusion pressure. NIRS measures of reactive hyperemia have the potential to evaluate microvascular perfusion in clinical populations.

1.

Introduction

Near-infrared spectroscopy (NIRS) has been well established as a method of measuring blood flow, oxygen saturation, and oxidative capacity in skeletal muscle and cerebral tissue.111 NIRS measures of perfusion have also been used to identify impaired microvascular perfusion in clinical populations with conditions such as peripheral vascular disease, diabetes, and hypovolemia.1218 Specifically, continuous-wave NIRS (CW-NIRS) and frequency-domain NIRS devices have been used to noninvasively measure microvasculature function by observing the kinetics of NIRS signals during reactive hyperemia following a vascular occlusion.5,12,16,1924 NIRS measures of reactive hyperemia correlate well with other measures of tissue perfusion, including conduit artery blood flow kinetics, transcutaneous oximetry, and plethysmography.12,13,20,2527 Additionally, studies have found baseline oxygen saturation and reperfusion rates during reactive hyperemia to be predictors of mortality in critically ill patients.2831

Patients with peripheral vascular disease and other pathologies experience impaired circulation in the lower extremities, particularly in the calf and foot.18,3238 CW-NIRS devices have been used to study recovery kinetics during reactive hyperemia in the lower limbs of these patient populations.12,18,23 The CW-NIRS probe is typically placed over the calf muscle or foot pad, and reperfusion is observed following 5 min of ischemia. Recovery kinetics have been characterized by reperfusion rates and times of the hemoglobin (HbO2), deoxygenated hemoglobin (HHb), and oxygen saturation signals; however, HbO2 is consistently reported in studies measuring reactive hyperemia in patient populations.12,18,39

A limitation to the previous studies is the use of different methods of analysis, and very few studies have reported the reproducibility of their measurements. Only one previous study has reported the reproducibility of CW-NIRS HbO2 kinetics in the lower limb, and that study measured only reactive hyperemia in the foot of six subjects.24 Another recent study reported the reproducibility of CW-NIRS HbO2 kinetics in the forearm, but the sensitivity to perfusion pressure of the parameters was not investigated.5 The purpose of the present study was to identify the most reproducible methods for measuring reactive hyperemia using CW-NIRS and to characterize reactive hyperemia in two different tissue types (in calf muscle and in the foot pad). We quantified the reproducibility of reperfusion times and reperfusion rates of HbO2 recovery kinetics as measured by CW-NIRS. Furthermore, we hypothesized that reperfusion rates and times of HbO2 would change with perfusion pressure.

2.

Materials and Methods

2.1.

Participants

Twenty participants (10 male, 10 female) aged 19 to 28 years performed one of two tests: reproducibility trials or elevation protocol. The study was conducted with the approval of the institutional review board at the University of Georgia (Athens, Georgia), and all subjects gave written, informed consent before testing. Participant characteristics are presented in Table 1.

Table 1

Characteristics of study participants.

ReproducibilityElevation protocol
N1010
Gender (M/F)5/55/5
Age (year)22.3 (2.4)21.5 (1.4)
Height (cm)172.1 (7.3)174.9 (7.8)
Weight (kg)71.1 (7.6)72.9 (14.8)
Body mass index (kgm2)24.1 (3.1)23.8 (4.1)

2.2.

Near-Infrared Spectroscopy

Reactive hyperemia was measured in the left medial gastrocnemius (calf) and the plantar midline of the left foot using CW-NIRS (Oxymon Mk III, Artinis Medical Systems). One transmitter and two receivers were placed at each measurement location. At the calf muscle, the interoptode distance was set to measure NIRS signals from a tissue depth of at least twice that of the adipose tissue thickness (ATT). The distance between the NIRS transmitter and receivers (interoptode distance) was adjusted (25 to 55 mm) according to each individual’s respective ATT. ATT was measured using ultrasound (LOGIQ, GE HealthCare) as previously described.40 The two receivers were always separated by a distance greater than 10 mm. For the plantar foot measurements, a custom-made rubber bracket was used to secure the transmitter and receivers to the foot, and the interoptode distances were constant at 35 and 45 mm for each participant. The optode distance of 35 to 45 mm was selected to provide measures of oxygen kinetics at a depth similar to previous studies measuring reactive hyperemia in the foot.12,18,24 NIRS measurements were digitally recorded in real time throughout the duration of the protocol at an acquisition frequency of 10 Hz.

2.3.

Vascular Occlusions

Vascular occlusions were performed on the left lower extremity with the participant in the supine position and the foot transmitter positioned 2 cm above the level of the heart. A blood pressure cuff (Hokanson, 20c, Bellevue, Washington) was placed proximal to the knee and rapidly inflated to 250 to 300 mm Hg using a rapid cuff inflation system (Hokanson, E20, Bellevue). The cuff was inflated for 5 min, and HbO2 signals were monitored during this period to ensure that the arterial occlusion was maintained throughout the test. Following the 5-min arterial occlusion, the cuff was rapidly deflated and reactive hyperemia was observed until oxygen levels returned to baseline. A representative test result is shown in Fig. 1.

Fig. 1

Representative data (a) showing the oxygenated hemoglobin/myoglobin (HbO2) NIRS signal during arterial occlusion and reperfusion. (b) Measures of reactive hyperemia. Broken lines indicate temporal reperfusion parameters: time to 50% magnitude (T1/2), time to 95% magnitude (T95), and time to the peak hyperemic signal (T100). Solid arrow indicates the second quartile rate (R2q).

JBO_21_9_091302_f001.png

2.4.

Experimental Protocols

The first aim of the study was to assess the test-retest reliability of reactive hyperemia kinetics in the calf muscle and the foot by performing two vascular occlusions at a baseline elevation of 2 cm above the level of the heart. The tests were performed 5  min apart on the same day by the same tester. The second aim of this study was to assess the influence of reduced perfusion pressure on reactive hyperemia kinetics in the foot. A model of reduced perfusion pressure was achieved by conducting the elevation protocol consisting of three separate vascular occlusions performed at three different limb elevations (baseline, 30 cm, and 60 cm). All elevations were calculated as the elevation of the NIRS transmitter above the level of the heart. Ankle segmental pressures were measured at baseline and at each level of elevation. Limb elevation was achieved by supporting the heel of the foot with padded rubber blocks to provide comfort and prevent the restriction of blood flow throughout the elevation protocol. HbO2 signals were monitored after each test to ensure that oxygen levels returned to baseline before proceeding to the next test.

2.5.

Data Analysis

The oxygenated hemoglobin signal (HbO2) was selected as a measure of wash-in kinetics of oxygen reperfusion during reactive hyperemia.5,18,39,41 The raw data collected from the NIRS device were exported and analyzed using custom-written routines in MATLAB® R2014b (MathWorks Inc.). Four reperfusion times of the HbO2 signal were calculated: T1/2 as the time from release of the cuff to return to 50% magnitude, T50100 as the time from 50% magnitude to the maximal hyperemic signal, T95 as the time from release of the cuff to 95% of the magnitude, and T100 as the time from release of the cuff to the peak hyperemic signal (Fig. 1). A series of reperfusion slopes of the HbO2 signal were calculated, including the slope of the complete recovery response and slopes of segments of the recovery response (increments of 25, 10, and 5%). Only the most reproducible incremental rates were reported. To control for differences in the range of reperfusion and in NIRS signal calibration between tests, all slope measurements were normalized to the range of reperfusion (end occlusion to peak hyperemia) and expressed as a percent of the range.

Statistical analysis was performed using IBM SPSS Statistics 22 (IBM®, Armonk, New York).

Differences between measurements were identified using one-way analysis of variance (ANOVA) comparing means of measures at different anatomical locations and between channels at the same anatomical locations. One-way repeated measures ANOVA was performed to identify differences in reactive hyperemia and ankle blood pressure during the elevation protocol. Bonferroni corrections for multiple comparisons were performed to identify differences from baseline at each elevation. Data reported at means (±standard deviation) unless otherwise specified. Significance was accepted at p<0.05 for all comparisons.

3.

Results

3.1.

Reproducibility

Test-retest reproducibility was performed on 10 participants. One-way repeated measures ANOVA indicated no significant differences in the means of any measures between the two trials. Interoptode distance had no significant effect on measures of reperfusion time or rate of reperfusion at the calf, and the signal from the receiver with the smaller interoptode distance consistently produced stronger signal-to-noise values. Therefore, the signal from the receiver with the smaller interoptode distance was selected for further analysis.

Overall, measurements from the first half of the reactive hyperemia response were most reproducible (Table 2). T1/2 was the most reliable measure of reperfusion time in the calf and foot (CVcalf=9.68±4.19, CVfoot=14.1±8.61), and the second quartile rate (R2q) was consistently the most reliable measure of rate at both measurement sites [coefficient of variation(CV)calf=7.12±3.95, CVfoot=8.86±6.80]. Measures of second half reperfusion time (T50100) were 25% more variable and significantly slower in the calf (p<0.01) and foot (p<0.01) compared to the measure of T1/2 (Table 2). The highest variability was found in the last 5% of the reperfusion curve. Furthermore, the measure of T95 (CVcalf=15.8±9.12, CVfoot=17.0±6.18) was found to be more reproducible than the measure of T100 (CVcalf=33.3±9.46, CVfoot=25.8±9.65) at both measurement sites (Table 2).

Table 2

Coefficients of variation (%) of measures of reperfusion.

T100T95T1/2T50−100R1qR2qR3qR4q
Calf33.3 (9.46)15.8 (9.12)9.68 (4.19)42.4 (13.3)10.6 (6.13)7.12 (3.95)8.35 (6.38)15.23 (17.6)
Foot25.8 (9.65)17.0 (6.18)14.1 (8.61)31.10 (13.8)15.82 (8.54)8.86 (6.80)11.75 (11.9)30.35 (18.8)

Selected analysis of HHb recovery kinetics was also performed. HHb measures of T1/2 (CVcalf=9.43±7.31, CVfoot=13.4±8.92) and T95 (CVcalf=11.8±10.9, CVfoot=23.5±18.4) expressed similar reproducibility compared to HbO2. While HHb reperfusion times of T1/2 (calf=10.3±3.35  s, foot=23.9±6.75  s) and T95 (calf=28.3±7.7  s, foot=54.2±16.4  s) were significantly slower compared to HbO2, the measurements consistently correlated with HbO2 measures at the calf (T1/2: R2=0.88, p<0.01; T95: R2=0.91, p<0.01) and foot (T1/2: R2=0.93, p<0.01; T95: R2=0.75, p<0.01).

3.2.

Anatomical Comparisons of Reactive Hyperemia

At baseline, measures of HbO2 reperfusion time and rate were significantly slower in the foot compared to the calf (Fig. 2). However, correlational analysis did identify significant relationships between the calf and foot in the measures of T1/2 (R2=0.673, p=0.033), T95 (R2=0.792, p=0.006), and R2q (R2=0.713, p=0.021). No correlational relationships were identified in the measures of total rate or T100 between the two anatomical locations.

Fig. 2

Perfusion time and rate in foot compared to calf. (a) to (c) Values are means±SD; temporal reperfusion parameters: time to 50% magnitude (T1/2), time to 95% magnitude (T95), and time to the peak hyperemic signal (T100) at the foot and the calf. * indicates significantly different from calf. (d) to (f) Correlational analysis of temporal measures at the foot and the calf. Significance accepted at p<0.05.

JBO_21_9_091302_f002.png

3.3.

Reperfusion Parameters During Elevation

Elevation significantly reduced ankle systolic blood pressure from baseline (108±10.9  mmHg) by 16.0 mm Hg at 30 cm (p<0.01) and 28.9 mm Hg at 60 cm (p<0.01) on average. The temporal measures of T1/2 and T95 were significantly increased and the measure of second quartile rate was significantly decreased at 30 and 60 cm elevation (Fig. 3). Significant correlations between the most reproducible time and rate measurements were consistently identified at all levels of elevation (Fig. 4). HHb measures of T1/2 and T95 were also significantly slower at each level of limb elevation.

Fig. 3

Values are means±SD; measures of (a) time to 50% magnitude (T1/2), (b) time to 95% magnitude (T95), and (c) time to the peak hyperemic signal (T100) in the foot at three difference levels of elevation. * indicates significant difference from 2 cm; # indicates significant difference from 2 and 30 cm. Significance accepted at p<0.05.

JBO_21_9_091302_f003.png

Fig. 4

Correlational analysis between measures of time to 50% magnitude (T1/2) and second quartile rate (R2q) in the foot at (a) 2 cm, (b) 30 cm, and (c) 60 cm of limb elevation. Significance accepted at p<0.05.

JBO_21_9_091302_f004.png

4.

Discussion

This study systematically examined reperfusion kinetics of HbO2 as measured by CW-NIRS to determine the most reproducible measures of reactive hyperemia. Our results identified the previously reported parameter of T1/2 as the most reproducible measure of reperfusion time and the novel measure of R2q as the most reliable measure of rate.14,24,42 T95 was also found to be a more reproducible measure of reperfusion time compared to the commonly reported parameter of time to peak hyperemia.5,12,18,23,24 The present study further demonstrated the sensitivity to perfusion pressure of each reproducible parameter, suggesting that either time or rate variables can be used to characterize reactive hyperemia in healthy and diseased populations.

4.1.

Reproducibility

Various recovery times have been used to characterize reactive hyperemia using NIRS, but few studies have assessed the reproducibility of these measures.5,12,18,23,24,43 The present study found values and CVs of total HbO2 reperfusion times similar to those previously reported.5,24 In comparison, we found the temporal measures of T95 and T1/2 to be more reproducible parameters than the previously reported parameter of T100.18,23,24 By excluding the highly variable top 5% of the reperfusion response in the calculation of T95, the reliability of measuring reperfusion time increased by 17% in the calf and 8% in the foot when comparing T95 to T100. T95 has been reported once before, but the reproducibility was not defined.12 Our results suggest that T95 may serve as a more reliable, and therefore clinically relevant, measure of the total reperfusion time than measures that include 100% of the recovery period. A significant correlation between T1/2 and T95 was also identified at the calf (R2=0.75, p=0.013) and foot (R2=0.879, p=0.001), indicating that similar information can be obtained from either measurement.

Recovery rates of NIRS oxygen signals have also been used to characterize reactive hyperemia, and we found the variability of the total rate of reperfusion to be similar to the CV values previously reported (CVcalf=12.5±10.7, CVfoot=17.1±10.9).5,12,21,23,24,31,4345 Alternatively, our results identified R2q as consistently the most reproducible measure of rate at both measurement sites. Furthermore, the R3q measure was identified as the most reproducible second half rate with a CV of 8.35 and 11.75% in the calf and foot, respectively. Reperfusion rates derived from the first and second half of the reperfusion curve may have different physiological mechanisms associated with them, and the present results identify R2q and R3q as reproducible markers of primary and secondary reperfusion rates, respectively. We also found significant correlations of measures of R2q with measures of T1/2 at both measurement sites, indicating that the measures of rate and time may be used interchangeably. However, ATT has not been found to influence temporal parameters of reactive hyperemia, so T1/2 and T95 could potentially be more reliable measures of reperfusion in populations with high ATT compared to measures of rate.19,20

4.2.

Perfusion Pressure Sensitivity

We found that the reproducible reperfusion times and rates were also robustly related to perfusion pressures associated with limb elevation. Although a recent study demonstrated the utility of changing limb elevation using NIRS, the study only evaluated resting muscle oxygen saturation.46 In our study, increases in reperfusion time and decreases in rates are consistent with perfusion parameters reported for patients with cardiovascular and metabolic disease.14,15,17,30,47 Specifically, we observed a 2.2-fold increase in reperfusion time with 60 cm elevation, which was comparable to the magnitude of changes seen in patients with mild-to-moderate peripheral arterial disease.12 The results of the elevation protocol support the use of the proposed measures in populations that experience decreased blood flow as an index of perfusion impairment.

4.3.

Anatomical Differences

We found that our measures of reperfusion time and rate were consistently shorter and faster in the calf muscles compared to the foot tissue, suggesting that the microvasculature of the calf muscle has a faster responsiveness to hypoxic stimuli. The differences in reperfusion between the calf and the foot are likely a result of differences in tissue composition. The signal from the CW-NIRS device at the calf measurement site is assumed to reflect the microvasculature of predominantly skeletal muscle tissue, which may have higher vascular reactivity compared to the heterogeneous composition of fascia, tendon, and muscle under the probe placed on the foot pad. While the values of reperfusion times in the calf and foot measured in the present study are similar to the times previously reported, not all studies have found differences in recovery rates between the calf and foot.23,24,39 Despite the observed differences, we found correlational relationships in both time and rate parameters between the two measurement sites, indicating preserved relative reperfusion kinetics in the proximal and distal lower limb.

4.4.

Methodology

Reactive hyperemia measured by various methodologies has been used to characterize the magnitude of diseases as well as to predict future health outcomes.4852 The reactive hyperemia measured by NIRS is largely mediated by the same mechanisms that drive reactive hyperemia in larger resistance arteries; however, NIRS measures are specific to oxygen delivery in the microvasculature of the skeletal muscle.1,7,8,20,26,5355 Other methodologies such as laser Doppler have been used to study microvascular perfusion, but these measurements represent superficial skin blood flow and not skeletal muscle microvascular beds.56,57 Magnetic resonance imaging (MRI) studies using muscle BOLD technology have reported microvascular perfusion parameters similar to NIRS measures in this study, but MRI methodology is far more expensive and methodologically involved.5860 NIRS technology provides a much simpler and less expensive alternative in assessing the endothelial function of the microvasculature. Specifically, NIRS measures of reactive hyperemia may be particularly beneficial in studying vascular pathology and intervention outcomes in populations with impairments in nutritive flow.1,17,49,61

The present study reported HbO2 as a measure of wash-in kinetics during reactive hyperemia. Several previous studies using CW-NIRS to measure reactive hyperemia have also reported parameters of HHb and oxygen saturation.5,23 The slower HHb reperfusion times measured in the present study suggest that these signals may be influenced by accumulations of blood in the tissue. Interestingly, parameters of HHb had similar reproducibility and expressed sensitivity to changes in perfusion pressure at 30 and 60 cm of elevation.

4.5.

Limitations

A key to NIRS measurements of the microvascular is accounting for tissue heterogeneity over the sampling site as the intensity of the HbO2 signal may be influenced by optical measurement calibration and tissue composition.7,26 The present study minimized the potential influence of varying ATT on the rate measurements by normalizing all HbO2 signals to their respective ranges of reperfusion.40 Some of the variations in the finding of studies using NIRS to measure reactive hyperemia kinetics could be a result of using un-normalized NIRS signals and different lengths of occlusion time.43,44,62 While some studies have used occlusion lengths shorter and longer than 5 min, reactive hyperemia is typically examined following a 5 min occlusion as longer or shorter durations of ischemia may alter recovery times and rates.12,15,19,63,64 Vascular occlusions up to 5 min have been shown not to result in significant depletion of phosphocreatine stores, which can potentially influence oxygen kinetics during recovery.65 A plateau in the oxygen signal may occur in longer occlusion lengths and could indicate the use of phosphocreatine stores. Furthermore, the 5 min occlusion used in the present study did not appear to result in any long-term changes to the vascular reactivity as indicated by the reproducibility of our results in sequential trials. Future studies should investigate the reproducibility of reperfusion kinetics following exercise. Although the two groups were not significantly different in age, BMI, or disease status, it should be noted that the reproducibility and elevation protocol were measured in two separate cohorts. The use of young, healthy participants may also be a potential limitation to our study, and future studies will need to examine the reproducibility and utility of these measurements in diseased/injured populations.

5.

Conclusion

This study employed a systematic analysis of reperfusion kinetics during reactive hyperemia to develop reproducible, standardized, and physiologically relevant measures of reactive hyperemia as assessed by NIRS. Furthermore, this study demonstrated the ability of the NIRS measures of reactive hyperemia to detect changes in perfusion pressure in the lower limb, revealing potential clinical applications. The capacity of NIRS to consistently measure reactive hyperemia in the presence of impaired blood flow is imperative to the use of this methodology in diseased populations and its implementation in clinical practice. Future studies should seek to further validate the application of the measures presented in this study and explore the applicability to diseased populations.

References

1. 

K. K. McCully and T. Hamaoka, “Near-infrared spectroscopy: what can it tell us about oxygen saturation in skeletal muscle?,” Exerc. Sport Sci. Rev., 28 (3), 123 –127 (2000). ESSRB8 0091-6331 Google Scholar

2. 

T. E. Ryan et al., “A cross-validation of near-infrared spectroscopy measurements of skeletal muscle oxidative capacity with phosphorus magnetic resonance spectroscopy,” J. Appl. Physiol., 115 (12), 1757 –1766 (2013). http://dx.doi.org/10.1152/japplphysiol.00835.2013 Google Scholar

3. 

T. E. Ryan et al., “Activity-induced changes in skeletal muscle metabolism measured with optical spectroscopy,” Med. Sci. Sports Exerc., 45 (12), 2346 –2352 (2013). http://dx.doi.org/10.1249/MSS.0b013e31829a726a Google Scholar

4. 

W. M. Southern et al., “Reproducibility of near-infrared spectroscopy measurements of oxidative function and postexercise recovery kinetics in the medial gastrocnemius muscle,” Appl. Physiol. Nutr. Metab., 39 (5), 521 –529 (2014). http://dx.doi.org/10.1139/apnm-2013-0347 Google Scholar

5. 

S. Lacroix et al., “Reproducibility of near-infrared spectroscopy parameters measured during brachial artery occlusion and reactive hyperemia in healthy men,” J. Biomed. Opt., 17 (7), 077010 (2012). http://dx.doi.org/10.1117/1.JBO.17.7.077010 JBOPFO 1083-3668 Google Scholar

6. 

D. M. Mancini et al., “Validation of near-infrared spectroscopy in humans,” J. Appl. Physiol., 77 (6), 2740 –2747 (1994). Google Scholar

7. 

T. Hamaoka et al., “The use of muscle near-infrared spectroscopy in sport, health and medical sciences: recent developments,” Philos. Trans. A Math. Phys. Eng. Sci., 369 (1955), 4591 –4604 (2011). http://dx.doi.org/10.1098/rsta.2011.0298 Google Scholar

8. 

M. Ferrari, M. Muthalib and V. Quaresima, “The use of near-infrared spectroscopy in understanding skeletal muscle physiology: recent developments,” Philos. Trans. A Math. Phys. Eng. Sci., 369 (1955), 4577 –4590 (2011). http://dx.doi.org/10.1098/rsta.2011.0230 Google Scholar

9. 

R. A. De Blasi et al., “Muscle oxygenation by fast near infrared spectrophotometry (NIRS) in ischemic forearm,” Adv. Exp. Med. Biol., 316 163 –172 (1992). http://dx.doi.org/10.1007/978-1-4615-3404-4_19 AEMBAP 0065-2598 Google Scholar

10. 

R. A. De Blasi et al., “Noninvasive measurement of forearm blood flow and oxygen consumption by near-infrared spectroscopy,” J. Appl. Physiol., 76 (3), 1388 –1393 (1994). Google Scholar

11. 

M. Ferrari and V. Quaresima, “A brief review on the history of human functional near-infrared spectroscopy (fNIRS) development and fields of application,” Neuroimage, 63 (2), 921 –935 (2012). http://dx.doi.org/10.1016/j.neuroimage.2012.03.049 NEIMEF 1053-8119 Google Scholar

12. 

R. Kragelj et al., “Parameters of postocclusive reactive hyperemia measured by near infrared spectroscopy in patients with peripheral vascular disease and in healthy volunteers,” Ann. Biomed. Eng., 29 (4), 311 –320 (2001). http://dx.doi.org/10.1114/1.1359451 ABMECF 0090-6964 Google Scholar

13. 

A. Lima et al., “The relation of near-infrared spectroscopy with changes in peripheral circulation in critically ill patients,” Crit. Care Med., 39 (7), 1649 –1654 (2011). http://dx.doi.org/10.1097/CCM.0b013e3182186675 CCMDC7 0090-3493 Google Scholar

14. 

K. K. McCully et al., “Identification of peripheral vascular disease in elderly subjects using optical spectroscopy,” J. Gerontol. A Biol. Sci. Med. Sci., 52 (3), B159 –B165 (1997). http://dx.doi.org/10.1093/gerona/52A.3.B159 Google Scholar

15. 

K. K. McCully and B. H. Natelson, “Impaired oxygen delivery to muscle in chronic fatigue syndrome,” Clin. Sci., 97 (5), 603 –608 (1999). http://dx.doi.org/10.1042/cs0970603 Google Scholar

16. 

K. C. Doerschug et al., “Impairments in microvascular reactivity are related to organ failure in human sepsis,” Am. J. Physiol. Heart Circ. Physiol., 293 (2), H1065 –H1071 (2007). http://dx.doi.org/10.1152/ajpheart.01237.2006 Google Scholar

17. 

A. Mizuno, J. Isobe and K. Shima, “Vascular dysfunction detected by a simplified venous occlusion test in NIDDM patients,” Diabetes Care, 16 (7), 1019 –1021 (1993). http://dx.doi.org/10.2337/diacare.16.7.1019 DICAD2 0149-5992 Google Scholar

18. 

T. Jarm et al., “Postocclusive reactive hyperemia in healthy volunteers and patients with peripheral vascular disease measured by three noninvasive methods,” Adv. Exp. Med. Biol., 530 661 –669 (2003). http://dx.doi.org/10.1007/978-1-4615-0075-9_66 AEMBAP 0065-2598 Google Scholar

19. 

C. M. Bopp, D. K. Townsend and T. J. Barstow, “Characterizing near-infrared spectroscopy responses to forearm post-occlusive reactive hyperemia in healthy subjects,” Eur. J. Appl. Physiol., 111 (11), 2753 –2761 (2011). http://dx.doi.org/10.1007/s00421-011-1898-z EJAPFN 1439-6319 Google Scholar

20. 

C. M. Bopp et al., “Relationship between brachial artery blood flow and total [hemoglobin + myoglobin] during post-occlusive reactive hyperemia,” Microvasc. Res., 91 37 –43 (2014). http://dx.doi.org/10.1016/j.mvr.2013.10.004 MIVRA6 0026-2862 Google Scholar

21. 

G. Zamparini et al., “Noninvasive assessment of peripheral microcirculation by near-infrared spectroscopy: a comparative study in healthy smoking and nonsmoking volunteers,” J. Clin. Monit. Comput., 29 (5), 555 –559 (2015). http://dx.doi.org/10.1007/s10877-014-9631-1 Google Scholar

22. 

C. Mayeur et al., “Comparison of four different vascular occlusion tests for assessing reactive hyperemia using near-infrared spectroscopy,” Crit. Care Med., 39 (4), 695 –701 (2011). http://dx.doi.org/10.1097/CCM.0b013e318206d256 CCMDC7 0090-3493 Google Scholar

23. 

J. L. Fellahi et al., “Lower limb peripheral NIRS parameters during a vascular occlusion test: an experimental study in healthy volunteers,” Ann. Fr. Anesth. Reanim., 33 (1), e9 –e14 (2014). http://dx.doi.org/10.1016/j.annfar.2013.11.014 Google Scholar

24. 

R. Kragelj, T. Jarm and D. Miklavcic, “Reproducibility of parameters of postocclusive reactive hyperemia measured by near infrared spectroscopy and transcutaneous oximetry,” Ann. Biomed. Eng., 28 (2), 168 –173 (2000). http://dx.doi.org/10.1114/1.241 ABMECF 0090-6964 Google Scholar

25. 

N. C. Woinarski et al., “Near-infrared spectroscopy of the thenar eminence to estimate forearm blood flow,” Crit. Care Resusc., 15 (4), 323 –326 (2013). Google Scholar

26. 

M. C. Van Beekvelt et al., “Performance of near-infrared spectroscopy in measuring local O(2) consumption and blood flow in skeletal muscle,” J. Appl. Physiol., 90 (2), 511 –519 (2001). Google Scholar

27. 

F. Harel et al., “Arterial flow measurements during reactive hyperemia using NIRS,” Physiol. Meas., 29 (9), 1033 –1040 (2008). http://dx.doi.org/10.1088/0967-3334/29/9/003 PMEAE3 0967-3334 Google Scholar

28. 

G. Colin et al., “Masseter tissue oxygen saturation predicts normal central venous oxygen saturation during early goal-directed therapy and predicts mortality in patients with severe sepsis,” Crit. Care Med., 40 (2), 435 –440 (2012). http://dx.doi.org/10.1097/CCM.0b013e3182329645 CCMDC7 0090-3493 Google Scholar

29. 

H. Ait-Oufella et al., “Knee area tissue oxygen saturation is predictive of 14-day mortality in septic shock,” Intensive Care Med., 38 (6), 976 –983 (2012). http://dx.doi.org/10.1007/s00134-012-2555-7 ICMED9 0342-4642 Google Scholar

30. 

M. Leone et al., “Oxygen tissue saturation is lower in nonsurvivors than in survivors after early resuscitation of septic shock,” Anesthesiology, 111 (2), 366 –371 (2009). http://dx.doi.org/10.1097/ALN.0b013e3181aae72d ANESAV 0003-3022 Google Scholar

31. 

J. Duret et al., “Skeletal muscle oxygenation in severe trauma patients during haemorrhagic shock resuscitation,” Crit. Care, 19 (1), 141 (2015). http://dx.doi.org/10.1186/s13054-015-0854-4 Google Scholar

32. 

J. Duan et al., “Ultrasonography of lower limb vascular angiopathy and plaque formation in type 2 diabetes patients and finding its relevance to the carotid atherosclerotic formation,” Pak. J. Med. Sci., 30 (1), 54 –58 (2014). http://dx.doi.org/10.12669/pjms.301.3907 Google Scholar

33. 

A. L. Carrington et al., “Peripheral vascular and nerve function associated with lower limb amputation in people with and without diabetes,” Clin. Sci., 101 (3), 261 –266 (2001). http://dx.doi.org/10.1042/cs1010261 Google Scholar

34. 

C. M. Akbari and F. W. LoGerfo, “Diabetes and peripheral vascular disease,” J. Vasc. Surg., 30 (2), 373 –384 (1999). http://dx.doi.org/10.1016/S0741-5214(99)70154-0 Google Scholar

35. 

I. Faris and H. Duncan, “Vascular disease and vascular function in the lower limb in diabetes,” Diabetes Res., 1 (4), 171 –177 (1984). Google Scholar

36. 

M. Kabbani et al., “Impact of diabetes and peripheral arterial occlusive disease on the functional microcirculation at the plantar foot,” Plast. Reconstr. Surg. Glob. Open, 1 (7), e48 (2013). http://dx.doi.org/10.1097/GOX.0b013e3182a4b9cb Google Scholar

37. 

Y. Kagaya et al., “‘Real Angiosome’ assessment from peripheral tissue perfusion using tissue oxygen saturation foot-mapping in patients with critical limb ischemia,” Eur. J. Vasc. Endovasc. Surg., 47 (4), 433 –441 (2014). http://dx.doi.org/10.1016/j.ejvs.2013.11.011 Google Scholar

38. 

U. Wolf et al., “Localized irregularities in hemoglobin flow and oxygenation in calf muscle in patients with peripheral vascular disease detected with near-infrared spectrophotometry,” J. Vasc. Surg., 37 (5), 1017 –1026 (2003). http://dx.doi.org/10.1067/mva.2003.214 Google Scholar

39. 

H. M. Kooijman et al., “Near infrared spectroscopy for noninvasive assessment of claudication,” J. Surg. Res., 72 (1), 1 –7 (1997). http://dx.doi.org/10.1006/jsre.1997.5164 JSGRA2 0022-4804 Google Scholar

40. 

T. E. Ryan et al., “Noninvasive evaluation of skeletal muscle mitochondrial capacity with near-infrared spectroscopy: correcting for blood volume changes,” J. Appl. Physiol., 113 (2), 175 –183 (2012). http://dx.doi.org/10.1152/japplphysiol.00319.2012 Google Scholar

41. 

C. P. Naehle et al., “Assessment of peripheral skeletal muscle microperfusion in a porcine model of peripheral arterial stenosis by steady-state contrast-enhanced ultrasound and Doppler flow measurement,” J. Vasc. Surg., 61 (5), 1312 –1320 (2015). http://dx.doi.org/10.1016/j.jvs.2013.11.094 Google Scholar

42. 

G. Yu et al., “Time-dependent blood flow and oxygenation in human skeletal muscles measured with noninvasive near-infrared diffuse optical spectroscopies,” J. Biomed. Opt., 10 (2), 024027 (2005). http://dx.doi.org/10.1117/1.1884603 JBOPFO 1083-3668 Google Scholar

43. 

J. L. Fellahi et al., “Dynamic evaluation of near-infrared peripheral oximetry in healthy volunteers: a comparison between INVOS and EQUANOX,” J. Crit. Care, 28 (5), 881.e1 –881.e6 (2013). http://dx.doi.org/10.1016/j.jcrc.2013.05.004 JCCAER 0883-9441 Google Scholar

44. 

H. Gomez et al., “Characterization of tissue oxygen saturation and the vascular occlusion test: influence of measurement sites, probe sizes and deflation thresholds,” Crit. Care, 13 (Suppl 5), S3 (2009). http://dx.doi.org/10.1186/cc8001 Google Scholar

45. 

H. Gomez et al., “Use of non-invasive NIRS during a vascular occlusion test to assess dynamic tissue O(2) saturation response,” Intensive Care Med., 34 (9), 1600 –1607 (2008). http://dx.doi.org/10.1007/s00134-008-1145-1 ICMED9 0342-4642 Google Scholar

46. 

A. Palanca, A. Yang and J. A. Bishop, “The effects of limb elevation on muscle oxygen saturation: a near-infrared spectroscopy study in humans,” PM&R, pii 1 –4 (2015). http://dx.doi.org/10.1016/j.pmrj.2015.07.015 Google Scholar

47. 

M. E. van Genderen et al., “Peripheral perfusion index as an early predictor for central hypovolemia in awake healthy volunteers,” Anesth. Analg., 116 (2), 351 –356 (2013). http://dx.doi.org/10.1213/ANE.0b013e318274e151 Google Scholar

48. 

J. A. Vita et al., “Brachial artery vasodilator function and systemic inflammation in the Framingham Offspring Study,” Circulation, 110 (23), 3604 –3609 (2004). http://dx.doi.org/10.1161/01.CIR.0000148821.97162.5E CIRCAZ 0009-7322 Google Scholar

49. 

P. A. Gerasimchuk et al., “Endothelial dysfunction indicators in patients with diabetic foot syndrome,” Vestn Ross Akad Med Nauk, 69 (5–6), 107 –110 (2014). http://dx.doi.org/10.15690/vramn.v69i5-6.1053 Google Scholar

50. 

P. O. Bonetti et al., “Noninvasive identification of patients with early coronary atherosclerosis by assessment of digital reactive hyperemia,” J. Am. Coll. Cardiol., 44 (11), 2137 –2341 (2004). http://dx.doi.org/10.1016/j.jacc.2004.08.062 JACCDI 0735-1097 Google Scholar

51. 

P. Greenland, S. C. Smith and S. M. Grundy, “Improving coronary heart disease risk assessment in asymptomatic people: role of traditional risk factors and noninvasive cardiovascular tests,” Circulation, 104 (15), 1863 –1867 (2001). http://dx.doi.org/10.1161/hc4201.097189 CIRCAZ 0009-7322 Google Scholar

52. 

T. Suzuki et al., “Metabolic syndrome, endothelial dysfunction, and risk of cardiovascular events: the Northern Manhattan Study (NOMAS),” Am. Heart J., 156 (2), 405 –410 (2008). http://dx.doi.org/10.1016/j.ahj.2008.02.022 Google Scholar

53. 

M. L. Davis and T. J. Barstow, “Estimated contribution of hemoglobin and myoglobin to near infrared spectroscopy,” Respir. Physiol. Neurobiol., 186 (2), 180 –187 (2013). http://dx.doi.org/10.1016/j.resp.2013.01.012 RPNEAV 1569-9048 Google Scholar

54. 

T. Nakamura et al., “Flow-mediated vasodilation of a conduit artery in relation to downstream peripheral tissue blood flow during reactive hyperemia in humans,” Jpn. Circ. J., 61 (9), 772 –780 http://dx.doi.org/10.1253/jcj.61.772 JCIRA2 0047-1828 Google Scholar

55. 

M. H. Laughlin et al., “Peripheral circulation,” Compr. Physiol., 2 (1), 321 –447 (2012). http://dx.doi.org/10.1002/cphy.c100048 Google Scholar

56. 

P. Stiefel et al., “Which parameter is better to define endothelial dysfunction in a test of postocclusive hyperemia measured by laser-Doppler flowmetry?,” Coron. Artery Dis., 23 (1), 57 –61 (2012). http://dx.doi.org/10.1097/MCA.0b013e32834e4f34 CADIEX 0954-6928 Google Scholar

57. 

C. Lachenbruch et al., “Relative contributions of interface pressure, shear stress, and temperature on ischemic-induced, skin-reactive hyperemia in healthy volunteers: a repeated measures laboratory study,” Ostomy. Wound Manage., 61 (2), 16 –25 (2015). Google Scholar

58. 

R. A. Meyer et al., “BOLD MRI mapping of transient hyperemia in skeletal muscle after single contractions,” NMR Biomed., 17 (6), 392 –398 (2004). http://dx.doi.org/10.1002/nbm.893 Google Scholar

59. 

S. Kos et al., “Simultaneous dynamic blood oxygen level-dependent magnetic resonance imaging of foot and calf muscles: aging effects at ischemia and postocclusive hyperemia in healthy volunteers,” Invest. Radiol., 44 (11), 741 –747 (2009). http://dx.doi.org/10.1097/RLI.0b013e3181b248f9 INVRAV 0020-9996 Google Scholar

60. 

T. F. Towse et al., “Comparison of muscle BOLD responses to arterial occlusion at 3 and 7 Tesla,” Magn. Reson. Med., (2015). http://dx.doi.org/10.1002/mrm.25562 MRMEEN 0740-3194 Google Scholar

61. 

L. Bruyndonckx et al., “Assessment of endothelial dysfunction in childhood obesity and clinical use,” Oxid. Med. Cell. Longev., 2013 1 (2013). http://dx.doi.org/10.1155/2013/174782 Google Scholar

62. 

R. Bezemer et al., “Assessment of tissue oxygen saturation during a vascular occlusion test using near-infrared spectroscopy: the role of probe spacing and measurement site studied in healthy volunteers,” Crit. Care, 13 (Suppl 5), S4 (2009). http://dx.doi.org/10.1186/cc8002 Google Scholar

63. 

M. C. Corretti et al., “Guidelines for the ultrasound assessment of endothelial-dependent flow-mediated vasodilation of the brachial artery: a report of the International Brachial Artery Reactivity Task Force,” J. Am. Coll. Cardiol., 39 (2), 257 –265 (2002). http://dx.doi.org/10.1016/S0735-1097(01)01746-6 JACCDI 0735-1097 Google Scholar

64. 

L. Ghiadoni et al., “Evaluation of endothelial function by flow mediated dilation: methodological issues and clinical importance,” High Blood Press. Cardiovasc. Prev., 22 (1), 17 –22 (2015). http://dx.doi.org/10.1007/s40292-014-0047-2 Google Scholar

65. 

T. Hamaoka et al., “Noninvasive measures of oxidative metabolism on working human muscles by near-infrared spectroscopy,” J. Appl. Physiol., 81 (3), 1410 –1417 (1996). Google Scholar

Biography

Thomas B. Willingham is a PhD student in exercise physiology at the University of Georgia. He completed his BS in biology in 2010 and his MS in exercise physiology in 2012 at the University of Georgia. He continued his training at the Shepherd Spinal Center in Atlanta, Georgia, before returning to UGA to focus on research in neuromuscular physiology. His current interests include the development and application of noninvasive methods for measuring hemodynamics, metabolism, and muscle function.

William M. Southern is a PhD student in the exercise physiology program at the University of Georgia. He completed his BS in exercise science in 2011 at the North Greenville University and his MS in exercise physiology in 2014 at the University of Georgia. His current interests include exploration of the interaction between exercise training and skeletal muscle mitochondria in various diseases.

Kevin K. McCully is a professor in the Kinesiology Department at the University of Georgia. His research focuses on improving and extending the use of noninvasive approaches to evaluate skeletal muscle metabolism and blood flow. Current methods include near-infrared spectroscopy to measure muscle metabolism and oxygen delivery, 31P magnetic resonance spectroscopy to measure muscle metabolism, magnetic resonance imaging to study muscle composition, and Doppler ultrasound to measure arterial blood flow and arterial health.

© 2016 Society of Photo-Optical Instrumentation Engineers (SPIE) 1083-3668/2016/$25.00 © 2016 SPIE
Thomas B. B. Willingham, William M. Southern, and Kevin K. McCully "Measuring reactive hyperemia in the lower limb using near-infrared spectroscopy," Journal of Biomedical Optics 21(9), 091302 (5 April 2016). https://doi.org/10.1117/1.JBO.21.9.091302
Published: 5 April 2016
Lens.org Logo
CITATIONS
Cited by 24 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Near infrared spectroscopy

Oxygen

Time metrology

Tissues

Blood circulation

Receivers

Heart

Back to Top