13 October 2017 Analysis of red blood cell parameters by Talbot-projected fringes
Author Affiliations +
Abstract
Red blood cell (RBC) anomalies are significant symptoms for identification of health disorders and several blood diseases, which involve the modification of the parameters and biophysical characteristics of such cells. The aim of this study is to measure the three-dimensional phase information of healthy RBCs and their parameters, such as cell diameter, thickness, and hemoglobin (Hb) content, using Talbot-projected fringes. The Talbot image of linear grating is projected onto an RBC slide. The deformed grating lines due to the shape and refractive index of RBCs are recorded by a CCD camera through a 20× microscope objective. Hilbert transform is used to extract the phase image from the deformed projected grating lines. Experimentally calculated values of diameter ( 8.2  μm), thickness ( 2.7  μm), and Hb content ( 28.7  pg/cell) are well within the limits available in the literature. The proposed system is robust and user-friendly and performs the imaging of RBCs with high axial and lateral resolution ( 2.19  μm).

1.

Introduction

In recent decades, several optical quantitative phase imaging techniques have been developed for biological cells, including human red blood cell (RBC) imaging. These techniques are spatial phase-shifting (e.g., Ref. 1), Fourier phase microscopy (e.g., Refs. 2 and 3), Hilbert phase microscopy (e.g., Ref. 4), digital holographic microscopy (e.g., Ref. 5), diffraction phase microscopy (e.g., Ref. 6), etc. In these methods, coherent, monochromatic light source such as lasers are used to measure the properties of biological cells. Because different properties of the biological cells, for example, phase, refractive index, and dry mass, are wavelength-dependent,7,8 some different interferometric or microscopic techniques, such as spectroscopic phase microscopy (e.g., Ref. 9), spectroscopic diffraction phase microscopy (e.g., Ref. 10), quantitative dispersion microscopy (e.g., Ref. 11), quantitative phase spectroscopy (e.g., Ref. 12), and dynamic spectroscopic phase microscopy (e.g., Ref. 13), etc., have been developed. In these techniques, multiple wavelength lasers, color filter wheels, and a broadband source with acousto-optic tunable filters are used to filter the multiple wavelengths and to determine the wavelength-dependent properties of biological cells. However, all of these techniques are costly, require complicated setups to align, need more optical components, and involve many mechanical moving parts. Partially coherent light (temporally low-coherent) sources such as a white light source,14 superluminescent diode,15 and light-emitting device8 are also used for biological cell imaging. Although these (spectrally broad band) sources reduce the unwanted speckle noise, they also require a dispersion compensation mechanism for the dispersion correction. Furthermore, it is not a good choice to use a spectrally broad light source if the object or medium has an inhomogeneous spectral response.16

The fringe projection is a noncontact technique that can be performed either with coherent or incoherent light, by illuminating the object with a fringe pattern generated by a binary or sinusoidal grating. The advantages of using a binary fringe pattern are—(1) simple: the coding and decoding algorithms are very simple, (2) fast: because the processing algorithm is very simple, it can reach very fast processing speeds, and (3) robust: since only two levels are used, it is very robust to the noise.17 The projection of a light pattern with a regular structure is a noncontact technique and easy to implement for measuring the three-dimensional (3-D) profile of a biological cell, which is decoded from the deformed fringe pattern recorded on the object.18 The digital fringe projection technique based on the Lau effect, Moiré method, Talbot effect has been used for surface profiling and contouring of reflecting objects.1720 If the Talbot image of grating is projected on the object, the surface information is depth encoded in a deformed fringe pattern by a carrier frequency corresponding to the grating frequency.21 This depth encoded shape information is obtained by the phase extraction and image processing techniques.2123 The Talbot-projected fringe technique has been used for profiling an object from the macroscopic (0.5 mm)22 to microscopic (11  μm)21 scales. Rodriguez-Vera et al.21 demonstrated the 3-D contouring of diffuse objects for shape measurement using Talbot fringe projection. Microscopic shape and deformation measurement, calibration of microtopographic measurement system, and vibration analysis at microscale, etc. have also been investigated by the Talbot fringe projection method.22,23

In this paper, the microscopic 3-D imaging of RBCs is demonstrated using the Talbot-projected fringes in transmission mode. The phase information is extracted by Hilbert transform (HT). HT is simple to implement, and phase information can be extracted from deformed fringe patterns with a phase measurement accuracy of λ/25.24 It requires less computational time, as compared with other phase extraction methods, such as Fourier transform, phase-shifting, etc.

2.

Theory

The imaging of a periodic object without any imaging device is known as the Talbot effect.2527 For plane wave illumination of a grating, behind periodic objects, at a distance ZT=ka2/λ(k=2,4,6,8,), Fresnel diffraction gives well-defined images of periodic objects, known as Fourier images or Talbot images/self-images, whereas the intermediate intensity distribution appearing between the Fourier image planes is a Fresnel image.28,29 If these Talbot fringes (self-image) are projected on an RBC slide, these fringes are deformed according to the surface profile and refractive index of RBC.30 These deformed Talbot-projected fringes or codified optical signals are recorded by the CCD camera and used to decodify or extract the phase information21 of RBCs.

The diffraction intensity distribution of the grating G1 at the Talbot plane, after passing through the object (RBCs) is given as31,32

Eq. (1)

I(x)=A(x)+Bcos[2πmxa+ϕ(x)],
where A(x) is the background intensity distribution, B is a constant, m is the grating order, a is the period of grating, and ϕ is the phase delay due to the presence of transparent phase object (RBCs).

The object phase is extracted using HT. HT is a linear operator that takes a function and produces a function with a shift of π/2 in the same spatial domain. The imaginary part of the analytical signal can be practically generated by passing the original signal through a “Hilbert transform.” A real function and its Hilbert-transformed function are related to each other in such a way that they together create an analytical signal, which can be written with an amplitude and phase, and the derivative of the phase can be identified as the instantaneous frequency.4,32,33

The background intensity must be eliminated prior to the application of HT for retrieving the phase. To isolate the sinusoidal term and to eliminate the background intensity, Eq. (1) is highpass filtered in the spatial domain. The filtered image can be written as

Eq. (2)

I(x)=Bcos[2πmxa+ϕ(x)].

Now the HT of the filtered image is performed to get the complex analytical signal associated with the real function I(x). The complex analytical signal Z(x) is given by4,33

Eq. (3)

Z(x)=12{I(x)+iHT[I(x)]},
where HT is the Hilbert transform.

The phase associated with Z(x) can be extracted as

Eq. (4)

Δϕ=arctan{HT[I(x)]I(x)}.

The complex analytical signal Z(x) exhibits rapid phase modulation with spatial frequency 1/a; thus, Δϕ is strongly wrapped. Finally, the phase ϕ(x) associated with the presence of the object is extracted as4,32,33

Eq. (5)

ϕ(x)=Δϕq(x),where  q(x)=2πma.

The wrapped phase, Δϕ, is unwrapped using the Goldstein phase unwrapping algorithm.34

3.

Experimental Setup

Figure 1 shows the schematic diagram of the experimental setup. A light beam from a He–Ne laser of wavelength 632.8 nm is expanded using a 40× microscope objective (NA=0.65) by Melles Griot (model No. 160/017 40/0.65) and filtered by a pinhole of 5-μm diameter. A collimating lens of diameter 50 mm and focal length 250 mm is employed to illuminate the grating G1 (300  lines/mm) by a collimated beam. The RBC slide is placed at Talbot distance (ZT) from the grating G1 (1.62 cm), so the Talbot-projected fringes (self-image of grating) lie on the RBC slide surface.

Fig. 1

Schematic diagram of the experimental setup.

JBO_22_10_106009_f001.png

To increase the lateral resolution for RBC imaging, a microscope objective (20×, NA=0.40) is placed such that the RBC slide with Talbot-projected fringe image should be at the focus of the microscope objective. A CCD camera is placed at a distance of 3.24 cm from the microscope objective to capture the maximum field of view.

For preparing a blood slide for RBC imaging (RBC slide), the blood samples were collected from the Indian Institute of Technology (IIT) Delhi, Hospital Pathology Lab, and approved by the responsible ethics committee of IIT Delhi, Hospital. The blood is diluted by mixing 1 ml of blood with 10 ml of saline solution (NaCl) in a test tube. One drop of this diluted blood is placed at one end of a slide and is dispersed using another slide over the slide’s length. The aim is to get a region, called a monolayer, in which the cells are spaced far enough apart to be differentiated.

When a prepared RBC slide is inserted at the Talbot plane, the Talbot-projected fringes get distorted. These distorted Talbot-projected fringes, having the phase, refractive index, and thickness information of RBCs, are recorded by a CCD camera (Make—Lumenera Corporation, Model—Infinity3-1M). The pixel size on the CCD sensor is 6.45  μm×6.45  μm. The total number of pixels is 1392×1040, and the sensor chip dimension is 2/3.

4.

Results

HT is used for the evaluation of the phase map from the captured image.

Initially, the experiment is carried out using a USAF resolution test chart and polystyrene beads to know the resolution of the setup. A small portion of the USAF chart is captured by a CCD camera using Talbot-projected fringes and a self-imaging effect to know the lateral resolution of setup. No other imaging optics are used for recording the image. Figure 2(a) shows the recorded image of the USAF resolution test chart with Talbot-projected fringes. The phase image of the USAF resolution test chart is reconstructed using HT. From the reconstructed wrapped phase image of the test chart, the sixth element of the seventh group can be easily seen in Fig. 2(b). This element corresponds to 228.1  line pairs/mm (i.e., the width of one line is 2.19  μm in the USAF resolution test target). From the line profile of the unwrapped phase image shown in Fig. 2(c), drawn along line AB marked in the Fig. 2(b), the observed value of one pixel is 0.4  μm, which is used for the determination of the lateral dimension of object RBCs under study.

Fig. 2

USAF resolution test chart. (a) Recorded image, (b) reconstructed wrapped phase image, and (c) line profile along line AB on unwrapped phase image.

JBO_22_10_106009_f002.png

To study the accuracy of the system in retrieving the phase map and profiles, measurements are performed on polystyrene microbeads of a known diameter of 10  μm. The polystyrene beads are diluted in an index matching oil medium and placed on the glass slide to prepare the polystyrene beads sample slide. Figure 3(a) shows the recorded image of polystyrene beads, and Fig. 3(b) shows the reconstructed 3-D unwrapped phase corresponding to Fig. 3(a).

Fig. 3

Polystyrene beads. (a) Recorded image and (b) 3-D unwrapped phase.

JBO_22_10_106009_f003.png

From the 3-D unwrapped phase map shown in Fig. 3(b), the phase of selected polystyrene bead is 3.8 rad. The size of the polystyrene bead is calculated by ϕ(x)=(2π/λ)(npnm)d, where ϕ(x) is the phase in radian, d is the size of the polystyrene bead, np is the refractive index of the polystyrene bead (1.59), and nm is the refractive index of the index matching oil medium (1.55).

The calculated size of the polystyrene bead is 9.57  μm. The size of the polystyrene bead, according to the specification by the supplier (10  μm) is closely related to the calculated value.

Figure 4(a) shows the 3-D thickness map of a polystyrene bead, and Fig. 4(b) shows the thickness profile along line AB as shown in Fig. 4(a).

Fig. 4

(a) 3-D thickness map of polystyrene bead and (b) thickness profile along line AB.

JBO_22_10_106009_f004.png

To check the consistency of the results, the same procedure was repeated with various polystyrene beads of the same size. Figure 5 shows the unwrapped phase map and the calculated thickness profile of selected polystyrene beads. The calculated size of different polystyrene beads as shown in Fig. 5 is 10.9, 9.4, 10.3, 10.1, and 9.7  μm. The experimentally calculated size of polystyrene beads shows a standard deviation (SD) of 0.32  μm from the mean value of 9.96  μm. The results are quite consistent with the size of these beads given by the supplier. The percentage error between the experimentally calculated size of polystyrene beads (average size 9.96  μm with SD 0.32) and the supplier specification (10  μm) is 0.4%.

Fig. 5

3-D unwrapped phase map and the calculated thickness profile corresponding to selected beads.

JBO_22_10_106009_f005.png

A preliminary experiment is conducted on RBCs to ascertain the phase map and the viability of this technique for biological cells. Figure 6(a) shows the recorded image of the RBCs. The phase information is obtained by the application of HT using Eq. (4). Figure 6(b) shows the wrapped phase, and Fig. 6(c) shows the 3-D unwrapped phase map corresponding to Fig. 6(a). The obtained phase map is wrapped between 0 and 2π. This 2π phase discontinuity is removed using the Goldstein phase unwrapping algorithm.34 Figures 7(a) and 7(b) show the two-dimensional (2-D) unwrapped phase map and 3-D unwrapped phase map of selected RBCs from Fig. 6(b) (shown in circle), respectively.

Fig. 6

(a) Recorded image, (b) 2-D wrapped phase of RBC, and (c) 3-D unwrapped phase of RBC.

JBO_22_10_106009_f006.png

Fig. 7

Phase image: (a) 2-D unwrapped phase of selected portion and (b) 3-D unwrapped phase of RBC.

JBO_22_10_106009_f007.png

Figure 8(a) shows the unwrapped phase map of a single RBC selected from Fig. 7(b), and Fig. 8(b) shows the phase profile with respect to position in μm (RBC diameter), obtained by drawing a line profile on phase map along line AB. The experimentally calculated value of the selected RBC diameter is 8.2  μm.

Fig. 8

(a) 3-D unwrapped phase of single RBC and (b) 2-D unwrapped phase profile.

JBO_22_10_106009_f008.png

The experiment was repeated for several RBCs. The unwrapped phase map of RBCs as shown in Fig. 6(c) was used to calculate the diameter of various RBCs. Figure (9) shows the unwrapped phase map and diameter of RBCs obtained by drawing a line profile on the phase image of various selected RBCs. The calculated diameter of different RBCs as shown in Fig. 9 is 6.8, 7.3, 8.1, 8.3, and 7.8  μm.

Fig. 9

3-D unwrapped phase map and obtained diameter of RBCs.

JBO_22_10_106009_f009.png

The mean value of the experimentally calculated diameter is found to be 7.6  μm with SD 1.21  μm, which is in good agreement with the value available in the literature.35,36

The phase information retrieved by the proposed method can be expeditiously translated into thickness information, which in turn can be utilized to measure other relevant morphological parameters, such as cell shape and volume. As we know, the refractive index of biological cells is a vital parameter for determining the thickness and hemoglobin (Hb) concentration in RBC. The thickness of the human RBC can be calculated using the following equation:

Eq. (6)

t=λ2π(nRBCnm)ϕ(x),
where λ is the wavelength of laser light, ϕ(x) is the optical phase, nRBC is the refractive index of a healthy RBC, nm is the refractive index of medium, and t is the geometrical thickness of RBC. The refractive index of saline water for He–Ne laser wavelengths (632.8 nm) is 1.331, and the value of the refractive index of a healthy RBC is 1.394, as reported in the literature.37,38 The calculated value of the RBC thickness is 2.7  μm, which is in good agreement with the value available in the literature.35,39 Figure 10(a) shows the 3-D thickness map of RBC, and Fig. 10(b) shows the thickness profile along line AB as shown in Fig. 10(a).

Fig. 10

(a) 3-D thickness map of RBC and (b) thickness profile along line AB.

JBO_22_10_106009_f010.png

Hb content measures the amount of protein present in one RBC. For the measurement of the Hb content, the measured 2-D phase at the normal angle was used. The Hb content of an RBC is given as40,41

Eq. (7)

Hb content=λ2παϕ(x),
where α is the refractive index increment (assuming average α=0.2  ml/g)4042 and ϕ(x) is the 2-D optical phase integrated over the entire cell area. The term ϕ(x) is calculated by multiplying the phase ϕ(x) to the projected cell area (shape of the RBC is considered circular), so fewer errors are involved in assessing the phase value over a small area.40 The calculated mean value of the Hb content is 28.7  pg/cell.40,41 Figure 11 shows the 2-D and 3-D Hb contents of one RBC.

Fig. 11

Hemoglobin content of RBC: (a) 2-D and (b) 3-D.

JBO_22_10_106009_f011.png

In addition, the Hb concentration in an RBC can also be obtained from the Hb content divided by the cellular volume.

5.

Conclusion

We have presented a simple and low-cost, quantitative phase measurement technique for 3-D microscopic imaging of RBCs using the Talbot-projected fringes of a linear grating. The present method will open possibilities for diagnosing diseases as well as studying the pathophysiology of human RBCs. In addition, the stability of the proposed system from environmental perturbation due to its common-path geometry as well as its compactness and portability give it a high potential for clinical applications.

Disclosures

Shilpi Agarwal, Dr. Varun Kumar, and Professor Chandra Shakher report grants from Science and Engineering Research Board (SERB), Department of Science and Technology, Government of India, during the conduct of the study. The authors have no other potential conflicts of interest to disclose.

Acknowledgments

The financial assistance from the Science and Engineering Research Board (SERB), Department of Science and Technology, Government of India, under the project Sanction Order No. SR/S2/LOP-34/2013 and dated 20 October 2014, is highly acknowledged. The authors gratefully acknowledge the comments made by the referees to improve the paper.

References

1. 

S. K. Debnath and Y. Park, “Real-time quantitative phase imaging with a spatial phase-shifting algorithm,” Opt. Lett., 36 (23), 4677 –4679 (2011). http://dx.doi.org/10.1364/OL.36.004677 OPLEDP 0146-9592 Google Scholar

2. 

G. Popescu et al., “Fourier phase microscopy for investigation of biological structures and dynamics,” Opt. Lett., 29 (21), 2503 –2505 (2004). http://dx.doi.org/10.1364/OL.29.002503 OPLEDP 0146-9592 Google Scholar

3. 

N. Lue et al., “Quantitative phase imaging of live cells using fast Fourier phase microscopy,” Appl. Opt., 46 (10), 1836 –1842 (2007). http://dx.doi.org/10.1364/AO.46.001836 APOPAI 0003-6935 Google Scholar

4. 

T. Ikeda et al., “Hilbert phase microscopy for investigating fast dynamics in transparent systems,” Opt. Lett., 30 (10), 1165 –1167 (2005). http://dx.doi.org/10.1364/OL.30.001165 OPLEDP 0146-9592 Google Scholar

5. 

B. Rappaz et al., “Measurement of the integral refractive index and dynamic cell morphometry of living cells with digital holographic microscopy,” Opt. Express, 13 (23), 9361 –9373 (2005). http://dx.doi.org/10.1364/OPEX.13.009361 OPEXFF 1094-4087 Google Scholar

6. 

G. Popescu et al., “Diffraction phase microscopy for quantifying cell structure and dynamics,” Opt. Lett., 31 (6), 775 –777 (2006). http://dx.doi.org/10.1364/OL.31.000775 OPLEDP 0146-9592 Google Scholar

7. 

K. G. Phillips, L. J. Steven and O. J. T. McCarty, “Measurement of single cell refractive index, dry mass, volume, and density using a transillumination microscope,” Phys. Rev. Lett., 109 (11), 118105 (2012). http://dx.doi.org/10.1103/PhysRevLett.109.118105 PRLTAO 0031-9007 Google Scholar

8. 

V. Dubey et al., “Multispectral quantitative phase imaging of human red blood cells using inexpensive narrowband multicolor LEDs,” Appl. Opt., 55 (10), 2521 –2525 (2016). http://dx.doi.org/10.1364/AO.55.002521 APOPAI 0003-6935 Google Scholar

9. 

Y. Park et al., “Spectroscopic phase microscopy for quantifying hemoglobin concentrations in intact red blood cells,” Opt. Lett., 34 (23), 3668 –3670 (2009). http://dx.doi.org/10.1364/OL.34.003668 OPLEDP 0146-9592 Google Scholar

10. 

B. Bhaduri et al., “Diffraction phase microscopy with white light,” Opt. Lett., 37 (6), 1094 –1096 (2012). http://dx.doi.org/10.1364/OL.37.001094 OPLEDP 0146-9592 Google Scholar

11. 

D. Fu et al., “Quantitative dispersion microscopy,” Biomed. Opt. Express, 1 (2), 347 –353 (2010). http://dx.doi.org/10.1364/BOE.1.000347 BOEICL 2156-7085 Google Scholar

12. 

M. Rinehart, Y. Zhu and A. Wax, “Quantitative phase spectroscopy,” Biomed. Opt. Express, 3 (5), 958 –965 (2012). http://dx.doi.org/10.1364/BOE.3.000958 BOEICL 2156-7085 Google Scholar

13. 

Y. Jang, J. Jang and Y. K. Park, “Dynamic spectroscopic phase microscopy for quantifying hemoglobin concentration and dynamic membrane fluctuation in red blood cells,” Opt. Express, 20 (9), 9673 –9681 (2012). http://dx.doi.org/10.1364/OE.20.009673 OPEXFF 1094-4087 Google Scholar

14. 

B. Bhaduri, K. Tangella and G. Popescu, “Fourier phase microscopy with white light,” Biomed. Opt. Express, 4 (8), 1434 –1441 (2013). http://dx.doi.org/10.1364/BOE.4.001434 BOEICL 2156-7085 Google Scholar

15. 

B. Kemper et al., “Characterisation of light emitting diodes (LEDs) for application in digital holographic microscopy for inspection of micro and nanostructured surfaces,” Opt. Lasers Eng., 46 (7), 499 –507 (2008). http://dx.doi.org/10.1016/j.optlaseng.2008.03.007 Google Scholar

16. 

J. Rosen and M. Takeda, “Longitudinal spatial coherence applied for surface profilometry,” Appl. Opt., 39 (23), 4107 –4111 (2000). http://dx.doi.org/10.1364/AO.39.004107 APOPAI 0003-6935 Google Scholar

17. 

A. Nava-Vega et al., “Phase map by fringe projection with dammann gratings: an application to measure small objects,” Rev. Mex. Fis., 61 (5), 330 –337 (2015). RMXFAT 0035-001X Google Scholar

18. 

S. Zhang, “Recent progresses on real-time 3D shape measurement using digital fringe projection techniques,” Opt. Laser Eng., 48 (2), 149 –158 (2010). http://dx.doi.org/10.1016/j.optlaseng.2009.03.008 Google Scholar

19. 

M. Thakur, C. Quan and C. J. Tay, “Surface profiling using fringe projection technique based on Lau effect,” Opt. Laser Technol., 39 (3), 453 –459 (2007). http://dx.doi.org/10.1016/j.optlastec.2005.12.002 OLTCAS 0030-3992 Google Scholar

20. 

K. Creath, J. C. Wyant, “Moiré and fringe projection techniques,” Optical Shop Testing, 653 –685 2nd ed.Wiley, Hoboken, New Jersey (1992). Google Scholar

21. 

R. Rodriguez-Vera, D. Kerr and F. Mendoza-Santoyo, “3-D contouring of diffuse objects by Talbot-projected fringes,” J. Mod. Opt., 38 (10), 1935 –1945 (1991). http://dx.doi.org/10.1080/09500349114552051 JMOPEW 0950-0340 Google Scholar

22. 

R. Rodriguez-Vera et al., “Calibration of a micro-topographic measurement system based on Talbot fringe projection,” in Proc. 2007 SEM Annual Conf. and Exposition on Experimental and Applied Mechanics, (2007). Google Scholar

23. 

R. Rodriguez-Vera et al., “Vibration analysis at microscale by Talbot fringe projection method,” Strain, 45 (3), 249 –258 (2009). http://dx.doi.org/10.1111/str.2009.45.issue-3 Google Scholar

24. 

R. Onodera, H. Watanabe and Y. Ishii, “Interferometric phase-measurement using a one-dimensional discrete Hilbert transform,” Opt. Rev., 12 (1), 29 –36 (2005). http://dx.doi.org/10.1007/s10043-005-0029-7 1340-6000 Google Scholar

25. 

H. F. Talbot, “Facts relating to optical science. No. IV,” J. Philos. Mag. Ser. 3, 9 401 –407 (1836). http://dx.doi.org/10.1080/14786443608649032 Google Scholar

26. 

O. Bryngdahl, “Image formation using self-imaging techniques,” J. Opt. Soc. Am., 63 (4), 416 –419 (1973). http://dx.doi.org/10.1364/JOSA.63.000416 JOSAAH 0030-3941 Google Scholar

27. 

K. Patorski, “The self-imaging phenomenon and its applications,” Progress in Optics, 27 1 –108 Elsevier Science Publishers B.V., North Holland, Amsterdam (1989). Google Scholar

28. 

J. M. Cowley and A. F. Moodie, “Fourier images: I-The point source,” Proc. Phys. Soc. London Sec. B, 70 (5), 486 –496 (1957). Google Scholar

29. 

J. T. Winthrop and C. R. Worthington, “Theory of Fresnel images. I. Plane periodic objects in monochromatic light,” J. Opt. Soc. Am., 55 (4), 373 –381 (1965). http://dx.doi.org/10.1364/JOSA.55.000373 JOSAAH 0030-3941 Google Scholar

30. 

S. Agarwal, V. Kumar and C. Shakher, “Cell imaging by Talbot projected fringes,” in Imaging Systems and Applications, (2017). Google Scholar

31. 

Y. Takeda et al., “X-ray phase imaging with single phase grating,” Jpn. J. Appl. Phys., 46 L89 –L91 (2007). http://dx.doi.org/10.1143/JJAP.46.L89 Google Scholar

32. 

N. Silin, S. Shin and Y. Yu, “Three dimensional measurements by deflectometry and double Hilbert transform,” J. Korean Phys. Soc., 69 (3), 286 –290 (2016). http://dx.doi.org/10.3938/jkps.69.286 KPSJAS 0374-4884 Google Scholar

33. 

X. Liang et al., “Single-shot slightly-off-axis interferometry based Hilbert phase microscopy of red blood cells,” Biomed. Opt. Express, 2 (4), 987 –995 (2011). http://dx.doi.org/10.1364/BOE.2.000987 BOEICL 2156-7085 Google Scholar

34. 

R. M. Goldstein, H. A. Zebker and C. L. Werner, “Satellite radar interferometry: two-dimensional phase unwrapping,” Radio Sci., 23 (4), 713 –720 (1988). http://dx.doi.org/10.1029/RS023i004p00713 RASCAD 0048-6604 Google Scholar

35. 

M. L. Turgeon, Clinical Hematology: Theory and Procedures, Lippincott Williams & Wilkins, Philadelphia, Pennsylvania (2005). Google Scholar

36. 

E. L. Persons, “Studies on red blood cell diameter: III. The relative diameter of immature (reticulocytes) and adult red blood cells in health and anemia, especially in pernicious anemia,” J. Clin. Invest., 7 (4), 615 –629 (1929). http://dx.doi.org/10.1172/JCI100246 Google Scholar

37. 

I. Shock et al., “Optical phase nanoscopy in red blood cells using low-coherence spectroscopy,” J. Biomed. Opt., 17 (10), 101509 (2012). http://dx.doi.org/10.1117/1.JBO.17.10.101509 JBOPFO 1083-3668 Google Scholar

38. 

P. M. S. Roma et al., “Total three-dimensional imaging of phase objects using defocusing microscopy: application to red blood cells,” Appl. Phys. Lett., 104 (25), 251107 (2014). http://dx.doi.org/10.1063/1.4884420 Google Scholar

39. 

W. Jin et al., “Simulation of simultaneous measurement for red blood cell thickness and refractive index,” Opt. Lasers Eng., 50 (2), 154 –158 (2012). http://dx.doi.org/10.1016/j.optlaseng.2011.09.019 Google Scholar

40. 

R. Barer, “Interference microscopy and mass determination,” Nature, 169 366 –367 (1952). http://dx.doi.org/10.1038/169366b0 Google Scholar

41. 

Y. Jang, J. Jang and Y. Park, “Dynamic spectroscopic phase microscopy for quantifying hemoglobin concentration and dynamic membrane fluctuation in red blood cells,” Opt. Express, 20 (9), 9673 –9681 (2012). http://dx.doi.org/10.1364/OE.20.009673 OPEXFF 1094-4087 Google Scholar

42. 

H. Park et al., “3-D refractive index tomograms and deformability of individual human red blood cells from cord blood of newborn infants and maternal blood,” J. Biomed. Opt., 20 (11), 111208 (2015). http://dx.doi.org/10.1117/1.JBO.20.11.111208 Google Scholar

Biography

Shilpi Agarwal received her MSc in physics with a specialization in advanced electronics from A. P. S. University Rewa (M.P.) in 2005. She has worked on a project on the magneto-optical Kerr effect at the Indian Institute of Technology Delhi. She was an assistant professor at an engineering college, Greater Noida, from 2006 to 2012. Currently, she is pursuing her doctorate degree in the area of diffractive optical elements and digital holography.

Varun Kumar received his MTech degree in instrument technology and his PhD in optical engineering from the Indian Institute of Technology, New Delhi, India, in 2010 and 2016, respectively. His research interests include digital holography, digital holographic microscopy and speckle pattern interferometry, and Talbot interferometry.

Chandra Shakher is an emeritus professor with the Instrument Design Development Centre, at the Indian Institute of Technology, New Delhi. He received his MTech degree in applied optics from the Indian Institute of Technology, Delhi in 1976 and his PhD from the Indian Institute of Technology, Madras in 1980. He has handled a large number of research projects and industrial consultancies. His research interest includes holography interferometry, speckle metrology, image processing, fiber optic current sensors, optical coherence tomography, and laser based instrumentation. He has 105 publications in international journals and 73 in conference proceedings. He is a fellow of Indian National Academy of Engineering, a senior member of OSA, and a fellow of SPIE.

© 2017 Society of Photo-Optical Instrumentation Engineers (SPIE) 1083-3668/2017/$25.00 © 2017 SPIE
Shilpi Agarwal, Varun Kumar, and Chandra Shakher "Analysis of red blood cell parameters by Talbot-projected fringes," Journal of Biomedical Optics 22(10), 106009 (13 October 2017). https://doi.org/10.1117/1.JBO.22.10.106009
Received: 16 June 2017; Accepted: 22 September 2017; Published: 13 October 2017
Lens.org Logo
CITATIONS
Cited by 7 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Blood

Refractive index

3D image processing

Microscopy

Phase measurement

CCD cameras

Fringe analysis

Back to Top