16 July 2019 Theoretical and experimental characterization of emission and transmission spectra of Cerenkov radiation generated by 177Lu in tissue
Nallely P. Jiménez-Mancilla, Keila Isaac-Olivé, Eugenio Torres-García, Miguel A. Camacho-López, Gerardo J. Ramírez-Nava, Héctor J. Mendoza-Nava
Author Affiliations +
Abstract

Cerenkov radiation (CR) is the emission of UV-vis light generated by the de-excitation of the molecules in the medium, after being polarized by an excited particle traveling faster than the speed of light. When β particles travel through tissue with energies greater than 219 keV, CR occurs. Tissues possess a spectral optical window of 600 to 1100 nm. The CR within this range can be useful for quantitative preclinical studies using optical imaging and for the in-vivo evaluation of Lu177-radiopharmaceuticals (β-particle emitters). The objective of our research was to determine the experimental emission light spectrum of Lu177-CR and evaluate its transmission properties in tissue as well as the feasibility to applying CR imaging in the preclinical studies of Lu177-radiopharmaceuticals. The theoretical and experimental characterizations of the emission and transmission spectra of Lu177-CR in tissue, in the vis-NIR region (350 to 900 nm), were performed using Monte Carlo simulation and UV-vis spectroscopy. Mice Lu177-CR images were acquired using a charge-coupled detector camera and were quantitatively analyzed. The results demonstrated good agreement between the theoretical and the experimental Lu177-CR emission spectra. Preclinical CR imaging demonstrated that the biokinetics of Lu177-radiopharmaceuticals in the main organs of mice can be acquired.

1.

Introduction

Cerenkov radiation (CR) is the energy produced in the ultraviolet (UV)-visible region of the electromagnetic spectrum, when the charged particles in a medium travel faster than the speed of light in the same medium. Because of the high particle velocity, the molecules of the medium are polarized and are immediately de-excited by releasing energy in the form of UV and visible radiation. CR generation is related to the velocity and energy of the particles as well as the refractive index of the environment. The threshold condition for CR generation is vp>c/n, where n is the refractive index of the medium, and vp and c are the speed of the particle and the speed of light in vacuum, respectively.1,2 For β particles traveling in water (n=1.33), the energy threshold for CR generation is 261 keV, and it changes in muscle tissue (n=1.4) to 219 keV.3

The variability of the tissue refractive index plays an important role in the production of photons through CR. The shape of the CR spectrum is independent of the radionuclide, but the intensity is regulated by the β particle source. Therefore, the CR intensity is determined by the radionuclide radioactive decay scheme and the number of transitions per unit time (activity). For example, β particles with higher end-point energy (Emax) will result in the emission of more Cerenkov photons.4

The optical imaging of radiopharmaceuticals using CR is a preclinical molecular imaging modality that describes the distribution, in time and space, of the charged-particle emitters through the tissue, using highly sensitive charge-coupled detectors (CCDs).5,6 This imaging modality has advantages, such as lower infrastructure cost, higher resolution, and lower image acquisition time, compared to micropositron-emission tomography (PET),7,8 a conventional nuclear imaging technique widely used.

Lu177 is a radionuclide β emitter, with a half-life of 6.71 days and Emax emission of 0.497 MeV (78%). It has been successfully used for radiopeptide therapy with an efficient crossfire effect in cancer cells.9,10 Hence the development of new Lu177-radiopharmaceuticals is of immense interest in the global medical field.11

Some of the β particles emitted by Lu177 reach the relativistic limit because they travel faster than the speed of light in different media. Nevertheless, to successfully use Lu177 for applications in preclinical CR imaging, it is necessary to establish several physical parameters, such as the Lu177-CR emission spectrum in the tissue optical window (600 to 1100 nm),12 number of photons produced per unit of Lu177 activity considering the loss of energy in its path through the tissue, Lu177-CR transmission spectrum in the tissue, and the maximum depth at which the Lu177 source should be located in the tissue to obtain CR signals suitable for imaging. These parameters, so far unknown, determine the applicability of Lu177-CR in the preclinical assessment (biokinetic models) of Lu177-radiopharmaceuticals, based on the optical images obtained in murine models.1,2,13,14

The objective of this work was to determine the experimental emission light spectrum of Lu177-CR, and evaluate its transmission properties in tissue as well as the feasibility of using CR imaging in the preclinical studies of Lu177-radiopharmaceuticals. For this purpose, the theoretical and experimental characterizations of the emission and transmission of CR generated by Lu177 in tissue were performed using the Frank–Tamm and relativistic theory, Monte Carlo simulation (MCLTmx code), and UV-vis spectroscopy techniques. The Lu177-CR images of mice were acquired using a CCD camera. The total number of photons per unit time per unit area (p/s/mm2) was quantitatively analyzed in each region of interest to obtain the total number of disintegrations, N ([MBq*h/MBq]), and the biokinetic models in the kidneys and tumor regions.

2.

Materials and Methods

2.1.

Theoretical Characterization of 177Lu-CR

From the emission spectrum of Lu177 β particles (Emax=0.497  MeV), the threshold energies required for CR generation in water and tissue were determined using Eq. (1):4

Eq. (1)

vp>cn,
where c is the speed of light in vacuum, n is the refractive index of the medium, and υp is the speed of the emitted β particle, which can be calculated using Eq. (2). Here, E0 is the energy of the β particle at rest and E is the energy of the β particle:

Eq. (2)

vp=c[1E02(E+E0)2]1/2.

The number of photons produced (N) over the distance traveled (x) by the β particle was obtained using the Frank–Tamm formula depicted in Eq. (3),15 which mathematically describes the light dependence with respect to the radionuclide energy and refractive index of the medium

Eq. (3)

Nx=2πα(11B2n2)λ1λ21λ2dλ,
where α is the fine-structure dimensionless constant (1/137), B is the ratio of the particle velocity to the speed of light, n is the refractive index of the medium, and λ1 and λ2 are the inferior and superior wavelengths of the emission spectrum, respectively.

The mean range (R) of the charged particles is an extensively used parameter in nuclear medicine.16 Here, R is defined as the average linear scope of the charged particle in a material medium; for β particles, it can be determined using Eqs. (4) and (5):

Eq. (4)

Rm=412Emax1.2650.0954lnEmax0.01<Emax<2.5  MeV,

Eq. (5)

Rm=530Emax106Emax>2.5  MeV,
where Emax is the maximum energy of the emitted particle and Rm is the mass range in milligram per square centimeter. To obtain the mean range in a medium or material, Rm is divided by the density of the material (ρwater=1.00  g/cm3 and ρtissue=1.07  g/cm3).

2.2.

Experimental 177Lu-CR Emission Spectrum

Lu177 was procured from ITG Isotope Technologies, with a radionuclidic purity of 99.9% and activity of 37 GBq in 2 mL of 0.04 of hydrochloric acid (HCl) within a glass ampoule vial. To characterize the CR emission spectrum, a BLUE-wave optical fiber spectrometer (StellarNet Inc.) was used; the spectrum was obtained within the 350- to 900-nm range.

An experimental setup was designed, including the construction of a lead-shielded insulated dark chamber with a circular aperture for the optical fiber. A glass ampoule vial was placed within the chamber and an optical fiber was placed in contact with the vial. This arrangement isolated the system from the background light and protected the involved personnel from the Lu177 ionizing radiation.

The Lu177-CR(λ) emission spectrum [gross spectrum, Lu177-CRgross(λ)] was registered continuously in the 350- to 900-nm range. In addition, the background spectrum was monitored similarly. The transmission spectrum (T) of the empty glass ampoule vial was also recorded in a Perkin Elmer spectrophotometer (model 551S) in the same wavelength range (350 to 900 nm). The absorption spectrum of only one wall of the vial (Abswall vial) was obtained as follows:

Eq. (6)

Abswall vial(λ)=log[T(λ)].

Finally, the absorption of the optical fiber (Absfiber) was also determined. In this case, the StellarNet spectrophotometer was illuminated with an aligned white light source and the transmission spectra in the presence (I) and the absence (I0) of the optical fiber were collected. From these spectra, the absorption spectrum in the range from 350 to 900 nm was determined as follows:

Eq. (7)

Absfiber(λ)=logI(λ)I0(λ).

All obtained spectra were normalized. The final Lu177-CR emission spectrum [Lu177-CR(λ)] corrected by background, glass vial, and optical fiber absorption was obtained as follows:

Eq. (8)

Lu177-CR(λ)=Lu177-CRgross(λ)Background(λ)Abswall vial(λ)*Absfiber(λ).

2.2.1.

Validation of the experimental 177Lu-CR emission spectrum

The experimental Lu177-CR(λ) emission spectrum was validated using the CR spectrum of Ga68 reported by Thorek et al.,2 because CR generation phenomenon does not depend on the type of radionuclide. Gallium-68 is a positron emitter that has a half-life of 68 min with a β+ particle (Emax=1.9  MeV) abundance of 89%. The validation was performed by a point-by-point comparison of the normalized CR values reported by Thorek et al.2 and the normalized values obtained from the experimental Lu177-CR(λ) spectrum in the spectral range of 380 to 600 nm. It is to be noted that Thorek et al. only reported the CR spectrum in the 300- to 600-nm range; hence the Lu177-CR(λ) emission spectrum emitted in the tissue optical window is unknown.

2.3.

Determination of the Optical 177Lu-CR Transmission in Tissue

Initially, the light transmission profile through tissue was obtained by visible photon transport simulation using the MCLTmx code.17 The Lu177-CR light transmission in tissue (transmission efficiency) was then obtained by the product of the experimental Lu177-CR(λ) emission spectrum and the light transmission profile simulated by the Monte Carlo MCLTmx code at different wavelengths and tissue thicknesses. Finally, the percentage of total emergent photons from the tissue optical window was obtained as a sum of the integrals of each function of the Lu177-CR(λ) transmission efficiency in the 600- to 900-nm range.

2.3.1.

Light-transmission profile in tissue using the Monte Carlo MCLTmx code

The Monte Carlo MCLTmx code simulates a monoenergetic and isotropic point source, because the beta emitter (Lu177) was considered homogenously distributed within the interest region (tumor and kidneys). Since the beta emission is isotropic, CR emission is also isotropic.

The source was located at the center of a biological tissue plate of finite thickness and sides with the following girth at each side of the source: 0.05, 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, 0.9, 1, 1.2, 1.4, 1.5, and 2 cm. For each girth, photons with the following wavelengths were, respectively, generated: 350, 400, 450, 500, 550, 600, 650, 700, 750, 800, 850, and 900 nm.

The reduced dispersion coefficient [μs(λ)] was obtained using Eq. (9), and the anisotropy [g(λ)] was calculated using18

Eq. (9)

μs(λ)=46[λ500  nm]1.421,

Eq. (10)

g(λ)=0.9435.6e0.0084λ.

Using Eqs. (9) and (10), the dispersion coefficient was obtained through Eq. (11). The absorption coefficient, as a function of the wavelength [μa(λ)], was obtained from the Svaasand-1995 data reported by Lister et al.,19 which can be considered an average of all biological tissue containing blood, and Eqs. (12) and (13):

Eq. (11)

μs(λ)=μs1g(λ),

Eq. (12)

μa(λ)=5.54+2939.68[32.74(λ414.8)2+32.72]  if  350λ<510(nm),

Eq. (13)

2.30+1073.01[81.24(λ550)2+81.22]  if  510λ900(nm).

The Svaasand-1995 data were reported for μa(λ), ranging from 350 to 750 nm, which was extrapolated to obtain the μa(λ) values for wavelengths between 750 and 900 nm using Eq. (13).

Finally, the primary and secondary transmission data were collected through a tissue plate. A total of 107 histories were run; three runs were conducted for each wavelength.

2.4.

177Lu-CR Image Acquisition and Quantitative Analysis

T47D human breast cancer tumors were induced subcutaneously in 8-week-old athymic mice weighing 22 g. All the studies on athymic mice were conducted according to the official Mexican norm, 062-ZOO-1999.

Lu177-radiopharmaceutical with a radiochemical purity of >98% was administered. The preparation of the radiopharmaceutical was previously reported by Aranda-Lara.20

Lu177-CR and x-ray images were acquired using Bruker’s in-vivo xtreme (BVX) system which is a preclinical multimodal imaging system that records images using different modalities (radiographic, radioisotopic, fluorescence, and Cerenkov luminescence) through an ultrathin phosphor screen and CCD camera. All the images were acquired and reconstructed following the methodology reported by Ramírez-Nava.21 Briefly, CR and x-ray images of the mice were acquired, as represented in Fig. 1, with BVX and its multimodal animal rotation system (MARS) after the administration of Lu177 (2, 4, 24, 48, and 120 h). Each Lu177-CR acquisition gathered a set of 13 projections, covering the 360 deg of rotation (30-deg steps). The acquisition time of each projection was 5 min, the binning was 4×4, in a field of view (FOV) of 12.5 cm.

Fig. 1

Schematic representation of x-ray and Cerenkov images acquisition using BVX system and projection images acquired with MARS.

JBO_24_7_076002_f001.png

The x-ray images were used to obtain the anatomical tomographic reconstruction of the rodents. These images were registered with a tube voltage of 45 kVp, a current of 497  μA, and an aluminum filter of 0.8 mm. In this modality, a batch of 181 projections (360-deg coverage and 2-deg steps) was acquired. The acquisition time of each projection was 10 s, using a 1×1 binning, and a 12.5-cm FOV. Both CR and x-ray images were reconstructed with the filtered backprojection algorithm, which was implemented in MATLAB R2108b (MathWorks, 2018).

The regions of interest (kidneys and the tumor) were located in the planar images of mice. From the 13 projection images acquired with the MARS, the mean depth of the tumor and kidneys were determined. The Lu177-CR images were quantitatively analyzed by counting the photons per unit time per unit area (p/s/mm2) in each region of interest.

2.4.1.

Determination of the total number of p/s/mm2 as a function of the activity

The total activity in the regions of interest was determined by calibrating the CCD camera, by counting the total number of p/s/mm2 detected by it from different known radiation activities (0.5, 1, 2, 4, 8, and 16 MBq). The mathematical function describing p/s/mm2 as a function of the activity was then obtained.

Biokinetic models [Ah(t)] for the tumor and kidneys were constructed using the activity registered at 2, 4, 24, 48, and 120 h postinjection in the regions of interest, and the number of total disintegrations (N) normalized to the unit-administered activity (MBq) was obtained as follows:

Eq. (14)

Nregion=t=0t=Ah(t)dt.

The total number of disintegrations, Nregion, obtained was compared with that obtained through a method involving the counting of the activity in a well-type scintillation detector from ex-vivo organs, reported by Aranda-Lara et al.20

3.

Results and Discussion

3.1.

Theoretical Characterization of 177Lu-CR

Figure 2 shows the theoretical characterization of the physical properties of Lu177. Figure 2(a) presents the energy spectrum of the β particles emitted by Lu177; Fig. 2(b) depicts the relationship between the energy and velocity in water (n=1.33) and the tissue (n=1.34 to 1.36); Fig. 2(c) displays the number of photons produced by the traveling of the emitted β particles for unit distance. This number was obtained using the Frank–Tamm equation in different media (water and tissue); Fig. 2(d) depicts the linear range (R) of the β particles in water and tissue. It can be observed that the threshold energy necessary for CR generation in water and tissue are 246.06 keV and 219.19 to 242.97 keV, respectively, depending on the tissue refractive index. Around 25% of the β particles emitted by Lu177 presented an energy greater than the threshold energy required for CR generation. The R, where the β particles can produce CR, was 1  cm in water and 0.9 cm in tissue. This characteristic is important in image reconstruction because it enables the improvement of the spatial and temporal resolutions of the image by differentiating between the material interfaces.22,23 Moreover, the number of the total photons produced per unit of activity can be calculated by the functions shown in Fig. 2(c), corrected by the probability of emission of the β particle [Fig. 2(a)]; in the case of water, the total numbers of photons by a unit of activity was 556,089.

Fig. 2

(a) Energy spectrum of the β particles emitted by Lu177, reported by radiation dose assessment resource24 and (b) relationship between β energy and their velocity in water (n=1.33) or tissue (n=1.341.36); the star represents the energy at which CR can be produced. (c) Number of photons produced by unit of β particle traveled path obtained with the Frank–Tamm theory in water and tissue, and (d) mean ranges of β particles emitted by Lu177 in water and tissue determined by nuclear medicine technique.16

JBO_24_7_076002_f002.png

3.2.

Experimental 177Lu-CR Emission Spectrum

Thorek et al.2 previously reported the Cerenkov spectrum of Ga68, and since this spectrum does not depend on the type of radionuclide used, it is expected that the normalized CR spectrum of Lu177 is equal to that of Ga68. In a previous work, CR spectrum was recorded from Ga68 but for wavelengths lower than 600 nm. In this work, the area of interest is the optical window of the tissue (λ>600  nm), therefore CR spectrum was recorded from Lu177 reaching the wavelength of 900 nm.

Figure 3 shows the experimental CR emission spectrum, corrected for the background, glass vial, and optical fiber absorption. The spectral range of the tissue optical window (600 to 900 nm) is an important contributor to the total intensity of the spectrum.

Fig. 3

Experimental CR spectrum, corrected for background, glass vial, and optical fiber absorption, obtained in the wavelength range from 300 to 900 nm.

JBO_24_7_076002_f003.png

Figure 4(a) shows the normalized emission spectrum of Ga68 reported by Thorek et al.,2 while Fig. 4(b) presents the normalized experimental Lu177-CR spectrum of Lu177 obtained in this study. Figure 4(c) depicts the comparison between the experimental Lu177-CR spectrum and that reported by Thorek et al.2 The comparison does not show significant differences in the 380 to 600 nm spectral window (<2%) because CR generation does not depend upon the radionuclide and only depends on the emission energy. Thus, this result validates the complete Lu177-CR spectrum (380 to 900 nm). The results of the normalized experimental characterization demonstrated that 45% of the total intensity of the Lu177-CR emission spectrum corresponds to the photons within the tissue optical window, supporting possible applications for the acquisition of optical images similar to those of positron emitters (e.g., F18, Cu64, and Zr89).8,25,26

Fig. 4

(a) Normalized emission spectrum of Ga68 reported by Thorek et al.2 (b) Normalized emission spectrum of Ga68 obtained from the original spectrum, which was reported by Thorek et al.2 (c) Comparison between both CR spectra obtained in the spectral range from 380 to 600 nm.

JBO_24_7_076002_f004.png

Table 1 shows the mathematical models that describe the normalized Lu177-CR experimental emission spectrum; the mathematical models for the two subregions (380 to 600 nm and 600 to 900 nm) are indicated.

Table 1

Normalized mathematical model of the experimental Lu177-CR emission spectrum in the spectral ranges of 350 to 600 nm and 600 to 900 nm.

Mathematical model I(λ)
I(λ)=A+B−C1+10(a−λ)(−b)
Wavelength (nm)Constant valuesR-square
380 to 600A=0.1876
B=1.0828
C=0.1876
0.9998
a=423.3840
b=0.0216
600 to 900A=0.4292
B=1.0004
C=0.4292
1
a=707.7825
b=0.01768

3.3.

Determination of the Optical Transmission 177Lu-CR in Tissue

In general, light transport by Monte Carlo method is performed in models that consider turbid homogeneous media or multilayer.27,28 But if it is considered that the refraction index is similar for different biological tissues (skin, brain, liver, kidneys, etc.), then a homogeneous media can be simulated, using absorption and scattering coefficients that represent various tissues and blood as a “generic tissue.”12,27,29 In this work, one layer of such tissue was considered in the simulation.

In Fig. 5, the normalized values of the light transmission for different tissue thicknesses obtained by Monte Carlo simulation are presented. As expected, greater light transmission was obtained in the 600- to 900-nm wavelength interval, which falls within the interval of the tissue optical window (600 to 1100 nm), where the tissues become optically transparent.12

Fig. 5

Light transmission profile for different thicknesses of tissue as a function of the wavelength obtained by Monte Carlo simulation (MCLTmx code).

JBO_24_7_076002_f005.png

Considering the general light transmission profile as a function of the wavelength at different tissue thicknesses (Fig. 5), as well as the experimental Lu177-CR emission spectrum, the efficiency of the Cerenkov light transmission generated by Lu177 at different tissue thicknesses and wavelengths was obtained (Fig. 6).

Fig. 6

Cerenkov light transmission profile generated by Lu177 at different tissue thicknesses (efficiency of transmission) obtained by the product of the experimental Lu177-CR spectrum [Fig. 4(a)] and the light transmission profile (Fig. 5).

JBO_24_7_076002_f006.png

Figure 7 shows the percentage of the total emergent photons in the tissue optical window. It was calculated as the sum of the integral of each Lu177-CR transmission efficiency function obtained at different tissue thickness in the 600- to 900-nm range. Given the fact that the main organs of mice have an average depth of 1.25 cm (liver, kidney, spleen, intestine, lungs, heart, and induced tumors)30 it can be seen from Fig. 7 that about 20% of the photons produced emerged from tissue. This percentage sustains the potential acquisition of images from the CR, which supports the biokinetics of Lu177-radiopharmaceuticals in the main organs of mice. Nevertheless, the imaging correction factors due to the overlapping of organs must be determined for improving the quantification by the optical imaging of the parameters related to the Lu177-radiopharmaceutical biokinetics.

Fig. 7

Percentage of transmission of total photons in the optical spectral window with respect to depth and its function fitting.

JBO_24_7_076002_f007.png

The function fitting of the percentage of total emergent photons from the tissue optical window with respect to the depth is represented in Eq. (15), where A=33.51, B=41.60, C=42.01, D=1.3187, a=0.12, b=0.95, and c=0.95. The inverse of this function can be considered as the tissue attenuation function; knowing the percentage of the total emergent photons at the tissue depth to be studied, the optical image can be corrected by the attenuation factor

Eq. (15)

%total emergent photons(depth)=Ae(depth/a)+Be(depth/b)+Ce(depth/c)D.

3.4.

177Lu-CR Image Acquisition and Quantitative Analysis

The Lu177-CR images of mice acquired using the BVX, and reconstructed as per the methodology reported by Ramírez-Nava et al.21 at 2 and 24 h, are depicted in Fig. 8; images were also acquired at 4, 48, and 120 h. The quantitative analysis of the Lu177-CR images shows that the mean depths of the tumor and kidneys are 0.4 and 0.9 cm, respectively. The number of photons per unit time per unit area (p/s/mm2) in the kidneys and tumor at different time durations and their correction by the attenuation factor are listed in Table 2.

Fig. 8

Tridimensional fusion of Lu177-CR (green color) and x-ray images of mice acquired in a Bruker’s in-vivo xtreme system at times 2 and 24 h.

JBO_24_7_076002_f008.png

Table 2

Number of photons per unit time per unit area (p/s/mm2) for tumor and kidneys at different times obtained from the optical images and their correction by the attenuation factor.

Photons/s/mm2
TissueDepth (cm)Time (h)
242448120
Kidneys0.91950.85±35.611840.15±33.59649.46±11.85418.21±07.63324.73±05.93
Tumor0.42361.25±43.352150.36±39.47986.68±18.11643.98±11.82591.25±10.85
Experimental data corrected by the attenuation factor [Photons/s/mm2*100/% Transmission of total photons (depth)]
Kidneys0.96211.02±113.375858.56±106.932067.73±37.741331.49±24.301033.86±18.87
Tumor0.44251.77±78.053872.02±71.081776.66±32.611159.57±21.291064.64±19.54

The total activity in MBq found in the kidneys and tumor at different time durations (Table 3) was obtained from Eq. (16), which relates the photons per unit time per unit area as a function of the activity. Equation 16 was obtained from the calibration curve of the CCD in-vivo xtreme system as the function of different activities (MBq) (Fig. 9). In Table 3, the biokinetic models in both regions of interest are also presented. It is to be noted that for obtaining high-quality optical images, it is necessary to characterize the CCD, including its limit of detection, dependence of the limit of detection on the production of photons, and its dependence on the initial activity because low signal intensity relative to the CR is a major limitation. Tamura et al.7 argue that preclinical instrumentation should focus on maximizing the light generated by a radionuclide source, while minimizing the light loss (by absorbance in the optical path) related to the detector wavelength sensitivity. Solving these shortcomings, the clinical translation to in-vivo models can be accelerated

Eq. (16)

p/s/mm2(Activity)=ABe(Activity/a),
where A=13321.9803, B=13302.702, and a=18.980.

Table 3

Total activity in MBq in the tumor and kidneys and their biokinetic models.

Activity (MBq)
TissueTime (h)
242448120
Kidneys11.89±0.3110.97±0.273.18±0.061.97±0.041.51±0.03
Tumor7.27±0.166.49±0.142.69±0.051.70±0.031.55±0.02
Biokinetic models
Ah(t)=Aeat±Bebt
Constant values
KidneysA=12.40B=1.62
a=8.07E2b=5.70E4
TumorA=6.65B=1.539
a=7.31E2b=5.00E

Fig. 9

Calibration of the CCD in-vivo xtreme system and function fitting.

JBO_24_7_076002_f009.png

Finally, the total number of disintegrations, N (MBq*h/MBq), in the kidneys and tumor obtained by the integration over time of the biokinetic models are presented in Table 4. The results obtained in this study using Lu177-CR imaging and Monte Carlo simulation do not significantly differ (<3%) with those obtained from an ex-vivo model that used the same Lu177-radiopharmaceutical reported by Aranda-Lara et al.20

Table 4

Total number of disintegrations N (MBq*h/MBq) obtained in this research by optical imaging and the comparison with that reported by Aranda-Lara et al.20 from an ex-vivo method.

TissueN=∫t=0t=∞Ah(t)dt Total disintegrations  (MBq*h/MBq)*N (MBq*h/MBq) Reported by Aranda-Lara et al.20
Kidneys4.88±0.194.75±0.67
Tumor4.40±0.094.33±0.31

Various biomolecules, peptides, and proteins labelled as F18, Y90, Cu64, Ga68, Zr89, Lu177, etc., are administered for specific diseases. These radiopharmaceuticals, which have demonstrated significant improvement in the condition and life quality of patients,31,32 can also generate CR. Several authors have mentioned that in addition to their use in PET or SPECT, they have significant potential for employment in preclinical studies using optical image evaluation techniques. Owing to the absorption of the CR generated in human tissue, optical imaging by CR cannot be used in clinical studies. Nevertheless, optimal characterization of the CR images can improve the image resolution because CR generation takes place close to the point where radioactive disintegration occurs and therefore, less image reconstruction would be required in comparison with the PET reconstruction techniques.7,8,33

It has been reported that the CR optical image technique is better characterized for evaluating tissue under hypoxia.34 This technique is the only one that can be corroborated by other imaging techniques, such as the PET or SPECT (different imaging principle). Moreover, it has additional advantages compared to the traditional nuclear medicine imaging techniques in terms of the cost, rapidity of image acquisition, and image evaluation.7,8

4.

Conclusions

This study presented a methodology based on the theoretical and experimental characterizations of the generation and range of Lu177-CR. By Monte Carlo simulation, the absorption and/or attenuation parameters in tissue were established and subsequently, the biokinetic model of the Lu177-radiopharmaceutical was successfully obtained.

The results of experimental characterization showed that 45% of the total intensity of the Lu177-CR emission spectrum corresponds to photons with energies within the tissue optical window. The total percentage of emerging photons as a function of the depth of Lu177-CR, obtained by Monte Carlo simulation and the experimental spectrum of Lu177-CR, demonstrated that emergent photons existed even at a tissue depth of 1.25 cm (average depth of main organs of mice). The results obtained from the preclinical Lu177-CR images agreed with the theoretical transmission spectrum and demonstrated that the biokinetics of Lu177-radiopharmaceuticals in the main organs of mice can be acquired from the CR optical images. The results indicate that Lu177 can be applied for the acquisition of optical images by CR generation in tissue and can be implemented in routine preclinical studies. Lu177-CR optical images are an alternative for the preclinical evaluation of new Lu177-radiopharmaceuticals.

Disclosures

The authors have no relevant financial interests in the manuscript and no other potential conflicts of interest to disclose.

Acknowledgments

This study was supported by the Mexican National Council of Science and Technology (CONACYT) through the CATEDRAS-CONACYT-ININ-337 and CONACYT-SEP-CB-2016-286753 projects. It was carried out as part of the activities of the “Laboratorio Nacional de Investigación y Desarrollo de Radiofármacos, CONACyT (LANIDER-CONACYT)” and the Red-Biofotónica, CONACYT. The financial support of the SIEA-UAEMex through the Grant No 4348/2017/CI is also acknowledged. We would like to thank Editage ( www.editage.com) for English language editing.

References

1. 

K. Tanha, A. M. Pashazadeh and B. W. Pogue, “Review of biomedical Čerenkov luminescence imaging applications,” Biomed. Opt. Express, 6 3053 –3065 (2015). https://doi.org/10.1364/BOE.6.003053 BOEICL 2156-7085 Google Scholar

2. 

D. L. Thorek et al., “Cerenkov imaging—a new modality for molecular imaging,” Am. J. Nucl. Med. Mol. Imaging, 2 163 –173 (2012). Google Scholar

3. 

A. E. Spinelli and F. Boschi, “Novel biomedical applications of Cerenkov radiation and radioluminescence imaging,” Phys. Med., 31 (2), 120 –129 (2015). https://doi.org/10.1016/j.ejmp.2014.12.003 Google Scholar

4. 

A. E. Spinelli et al., “Cerenkov radiation allows in vivo optical imaging of positron emitting radiotracers,” Phys. Med. Biol., 55 483 –495 (2010). https://doi.org/10.1088/0031-9155/55/2/010 PHMBA7 0031-9155 Google Scholar

5. 

X. Ma, J. Wang and Z. Cheng, “Cerenkov radiation: a multi-functional approach for biological sciences,” Front. Phys., 2 4 (2014). https://doi.org/10.3389/fphy.2014.00004 Google Scholar

6. 

B. J. Beattie et al., “Quantitative modeling of Cerenkov light production efficiency from medical radionuclides,” PLoS One, 7 e31402 (2012). https://doi.org/10.1371/journal.pone.0031402 POLNCL 1932-6203 Google Scholar

7. 

R. Tamura, E. C. Pratt and J. Grimm, “Innovations in nuclear imaging instrumentation: Cerenkov imaging,” Semin. Nucl. Med., 48 (4), 359 –366 (2018). https://doi.org/10.1053/j.semnuclmed.2018.02.007 SMNMAB 0001-2998 Google Scholar

8. 

R. Robertson et al., “Optical imaging of Cerenkov light generation from positron-emitting radiotracers,” Phys. Med. Biol., 54 (16), N355 –N365 (2009). https://doi.org/10.1088/0031-9155/54/16/N01 Google Scholar

9. 

M. Luna-Gutiérrez et al., “177Lu-labeled monomeric, dimeric and multimeric RGD peptides for the therapy of tumors expressing α(ν)β(3) integrins,” J. Label. Compd. Radiopharm., 55 140 –148 (2012). https://doi.org/10.1002/jlcr.v55.4 Google Scholar

10. 

N. Jimenez-Mancilla et al., “Multifunctional targeted radiotherapy system for induced tumours expressing gastrin-releasing peptide receptors,” Curr. Nanosci., 8 193 –201 (2012). https://doi.org/10.2174/157341312800167678 Google Scholar

11. 

M. Cremonesi et al., “Correlation of dose with toxicity and tumour response to 90Y- and 177Lu-PRRT provides the basis for optimization through individualized treatment planning,” Eur. J. Nucl. Med. Mol. Imaging, 45 (13), 2426 –2441 (2018). https://doi.org/10.1007/s00259-018-4044-x Google Scholar

12. 

S. L. Jacques, “Optical properties of biological tissues: a review,” Phys. Med. Biol., 58 R37 (2013). https://doi.org/10.1088/0031-9155/58/11/R37 PHMBA7 0031-9155 Google Scholar

13. 

G. S. Mitchell et al., “In vivo Cerenkov luminescence imaging: a new tool for molecular imaging,” Philos. Trans. R. Soc. A Math. Phys. Eng. Sci., 369 (1955), 4605 –4619 (2011). https://doi.org/10.1098/rsta.2011.0271 Google Scholar

14. 

Z. Hu et al., “Experimental Cerenkov luminescence tomography of the mouse model with SPECT imaging validation,” Opt. Express, 18 (24), 24441 –24450 (2010). https://doi.org/10.1364/OE.18.024441 OPEXFF 1094-4087 Google Scholar

15. 

H. Liu et al., “Molecular optical imaging with radioactive probes,” PLoS One, 5 (3), e9470 (2010). https://doi.org/10.1371/journal.pone.0009470 POLNCL 1932-6203 Google Scholar

16. 

H. Cember and T. E. Johnson, Introduction to Health Physics, 4 ed.McGraw-Hill Education, New York (1969). Google Scholar

17. 

E. Torres-García et al., “A new Monte Carlo code for light transport in biological tissue,” Med. Biol. Eng. Comput., 56 (4), 649 –655 (2018). https://doi.org/10.1007/s11517-017-1713-z MBECDY 0140-0118 Google Scholar

18. 

X. Ma et al., “Bulk optical parameters of porcine skin dermis at eight wavelengths from 325 to 1557 nm,” Opt. Lett., 30 (4), 412 –414 (2004). https://doi.org/10.1364/OL.30.000412 OPLEDP 0146-9592 Google Scholar

19. 

T. Lister, P. A. Wright and P. H. Chappell, “Optical properties of human skin,” J. Biomed. Opt., 17 (9), 090901 (2012). https://doi.org/10.1117/1.JBO.17.9.090901 JBOPFO 1083-3668 Google Scholar

20. 

L. Aranda-Lara et al., “Synthesis and evaluation of Lys1(α, γ-Folate)Lys3(177Lu-DOTA)-Bombesin(1-14) as a potential theranostic radiopharmaceutical for breast cancer,” Appl. Radiat. Isot., 107 214 –219 (2016). https://doi.org/10.1016/j.apradiso.2015.10.030 ARISEF 0969-8043 Google Scholar

21. 

G. J. Ramírez-Nava et al., “Multimodal molecular 3D imaging for the tumoral volumetric distribution assessment of folate-based biosensors,” Med. Biol. Eng. Comput., 56 (7), 1135 –1148 (2018). https://doi.org/10.1007/s11517-017-1755-2 MBECDY 0140-0118 Google Scholar

22. 

C. Li, G. S. Mitchell and S. R. Cherry, “Cerenkov luminescence tomography for small-animal imaging,” Opt. Lett., 35 (7), 1109 –1111 (2010). https://doi.org/10.1364/OL.35.001109 OPLEDP 0146-9592 Google Scholar

23. 

E. Brayfindley et al., “Automated defect detection in spent nuclear fuel using combined Cerenkov radiation and gamma emission tomography data,” Nucl. Technol., 204 343 –353 (2018). https://doi.org/10.1080/00295450.2018.1490123 Google Scholar

24. 

“RADAR (RAdiation Dose Assessment Resource),” http://www.doseinfo-radar.com/ Google Scholar

25. 

E. C. Pratt, T. M. Shaffer and J. Grimm, “Nanoparticles and radiotracers: advances toward radionanomedicine,” Wiley Interdiscip. Rev. Nanomed. Nanobiotechnol., 8 (6), 872 –890 (2016). https://doi.org/10.1002/wnan.1402 Google Scholar

26. 

A. Ruggiero et al., “Cerenkov luminescence imaging of medical isotopes,” J. Nucl. Med., 51 (7), 1123 –1130 (2010). https://doi.org/10.2967/jnumed.110.076521 JNMEAQ 0161-5505 Google Scholar

27. 

S. A. Prahl et al., “A Monte Carlo model of light propagation in tissue,” Proc. SPIE, 10305 1030509 (1989). https://doi.org/10.1117/12.2283590 Google Scholar

28. 

Y. Wang, “Accurate Monte Carlo simulation of frequency-domain optical coherence tomography,” 35 (4), e3177 (2019). https://doi.org/10.1002/cnm.3177 Google Scholar

29. 

L. Wang, S. L. Jacquesa and L. Zhengb, “MCML—Monte Carlo modeling of light transport in multi-layered tissues,” Comput. Methods Programs Biomed., 47 131 –146 (1995). https://doi.org/10.1016/0169-2607(95)01640-F CMPBEK 0169-2607 Google Scholar

30. 

W. H. Miller et al., “Evaluation of beta-absorbed fractions in a mouse model for 90Y, 188Re, 166Ho, 149Pm, 64Cu, and 177Lu radionuclides,” Cancer Biother. Radiopharm., 20 (4), 436 –449 (2005). https://doi.org/10.1089/cbr.2005.20.436 Google Scholar

31. 

A. O. J. Garcı, L. Dı and Á. L. Rgd, “Preparation and in vitro evaluation of 177 Lu-iPSMA-RGD as a new heterobivalent radiopharmaceutical,” J. Radioanal. Nucl. Chem., 314 (3), 2201 –2207 (2017). https://doi.org/10.1007/s10967-017-5555-9 JRNCDM 0236-5731 Google Scholar

32. 

C. Santos-Cuevas et al., “177Lu-DOTA-HYNIC-Lys(Nal)-Urea-Glu: biokinetics, dosimetry, and evaluation in patients with advanced prostate cancer,” Contrast Media Mol. Imaging, 2018 1 –10 (2018). https://doi.org/10.1155/2018/5247153 Google Scholar

33. 

A. E. Spinelli and F. Boschi, “Optimizing in vivo small animal Cerenkov luminescence imaging,” J. Biomed. Opt., 17 (4), 040506 (2012). https://doi.org/10.1117/1.JBO.17.4.040506 JBOPFO 1083-3668 Google Scholar

34. 

E. Desvaux et al., “Cherenkov luminescence imaging is a fast and relevant preclinical tool to assess tumour hypoxia in vivo,” EJNMMI Res., 8 (1), 111 (2018). https://doi.org/10.1186/s13550-018-0464-7 Google Scholar

Biography

Nallely P. Jiménez-Mancilla received her BSc degree in physics (2007), MSc degree in medical physics (2010), and PhD in health sciences (2014) from the Autonomous University of the State of Mexico. Since 2015, she has worked at the National Council of Science and Technology (Consejo Nacional de Ciencia y Tecnología, CONACYT) and National Institute of Nuclear Research (Instituto Nacional de Investigaciones Nucleares, ININ). She is a researcher in applied physics for therapy and molecular imaging of radiolabeled probes and nanosystems.

Keila Isaac-Olivé received a BSc degree in radiochemistry from the Higher Institute of Nuclear Science and Technology, Havana-Cuba (1998), BSc degree in chemistry from the University of Havana (1999), and MSc (2001) and PhD (2005) from Dalhousie University, Nova Scotia, Canada. She was a postdoctoral fellow from 2007 to 2009 at the Physics Institute, National Autonomous University of Mexico, Mexico. In 2009, she joined the faculty of the Medical Physics Program at the Autonomous University of the State of Mexico (Toluca, Mexico). She is an expert in analytical and radioanalytical techniques, particularly in the use and applications of radionuclides.

Eugenio Torres-García received his BSc degree in physics (2001), and his PhD in medical physics (2007) from the Autonomous University of the State of Mexico. Since 2007, he has worked in the medical physics program of the Autonomous University of the State of Mexico. He is a researcher in medical imaging, ionizing radiation dosimetry, nuclear medicine, and Monte Carlo simulation.

Miguel A. Camacho-López received a BSc degree in physics (1993), MSc in physics (1996), and PhD in physics (Heriot-Watt University, Scotland UK, 2003). He was a postdoctoral fellow at the Liquid Crystal Institute, Kent State University, USA (2000–2002) and a postdoctoral research fellow at the School of Physics and Astronomy, University of St Andrews, Scotland, UK (2002–2004). In 2004, he joined the faculty of the Medical Physics Program at the Autonomous University of the State of Mexico. He is an expert in biophotonics and the optical properties of nanomaterials.

Gerardo J. Ramírez-Nava received his BSc degree in biomedical engineering (2013) and MSc in medical physics (2016). Since 2017, he has been a PhD student of the nanosciences and micro-nanotechnologies program at the National Polytechnic Institute of Mexico. His research focuses on image processing, nuclear imaging and dosimetry.

Héctor J. Mendoza-Nava received his BSc degree in physics (2009), MSc in medical physics (2012), and PhD in health sciences (2017) from the Autonomous University of the State of Mexico (UAEM). He works at the National Institute of Nuclear Research in Mexico (Instituto Nacional de Investigaciones Nucleares, ININ), Metrology Department. He is a researcher in nuclear medicine and metrology of ionizing radiation in medical applications.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Nallely P. Jiménez-Mancilla, Keila Isaac-Olivé, Eugenio Torres-García, Miguel A. Camacho-López, Gerardo J. Ramírez-Nava, and Héctor J. Mendoza-Nava "Theoretical and experimental characterization of emission and transmission spectra of Cerenkov radiation generated by 177Lu in tissue," Journal of Biomedical Optics 24(7), 076002 (16 July 2019). https://doi.org/10.1117/1.JBO.24.7.076002
Received: 27 March 2019; Accepted: 20 June 2019; Published: 16 July 2019