Minimally invasive robotic surgery using fluorescence-guided images with a video laparoscope has been widely used because of its advantages of small incision, fast recovery time, and efficiency. However, the penetration depth limitation of fluorescence is a disadvantage caused by the absorption and scattering in tissues and blood cells. If this limitation can be overcome by additional imaging modalities, the surgical procedure can be quite efficient and precise. High-energy annihilation-gamma photons have a stronger penetration capability than visible and fluorescence photons. To characterize and validate a multimodal annihilation-gamma/near-infrared (NIR)/visible laparoscopic imaging system, an internal detector composed of an annihilation-gamma detector and an optical system was assembled inside a surgical stainless pipe with an outer diameter of 15.8 mm and an external detector with a dimension of 100 × 100 mm2 placed at the opposite side of the internal detector. Integrated images of 511-keV gamma rays, NIR fluorescence, and visible light were obtained simultaneously. The 511-keV gamma image could be clearly seen with the acquisition of 5 s, while NIR and visible images could be presented in real time. This multimodal system has the potential for improving the surgery time and the quality of patient care. |
1.IntroductionCompared to conventional open surgery, which utilizes images acquired before the surgery from computed tomography (CT), magnetic resonance imaging (MRI), positron emission tomography (PET), and other diagnostic modalities, minimally invasive robotic surgery (MIRS) has been widely used because of advantages such as less incision, fast recovery time, and the quality of patient care.1–3 Recently, near-infrared (NIR) image-guided surgeries have been used for the detection of sentinel lymph nodes (SLNs) and micro tumors during surgery, because these fluorescence systems provide relatively high sensitivity and superior resolution of .4–7 They have also been used for colorectal, urological, and gynecological diseases to define the margin of the target to achieve minimal damage of the normal tissue and sensitive nerves. Among various fluorescence dyes for MIRS, a U.S. Food and Drug Administration-approved indocyanine green (ICG) dye with a peak emission wavelength of 820 nm has been used for SLN mapping and medical imaging for other lesions.8–12 However, the fluorescence has limited penetration depth (below 10 mm) because of its absorption and scattering in tissue such as skin, fat, and blood.13–15 For this reason, it is difficult to find SLNs or other target tissues covered with normal tissue. To overcome the depth limitation, multimodal imaging and hybrid methods have been used with radiation, ultrasound, and other compatible modalities. High-energy photons, such as x-rays and gamma rays, have relatively longer penetration depths despite radiation exposure, relatively low sensitivity, and poor spatial resolution. Recent works validated the feasibility of intraoperative fluorescence-guided surgery with an image of preoperative single-photon emission computed tomography/CT, although they did not provide real-time integrated images showing the target margin clearly.16–18 Another study used trimodal imaging probes of ultrasound, photoacoustic, and fluorescence to capture sequential SLN images with a preclinical animal model.19 Kang et al.20 studied an endoscopic NIR/gamma/visible imaging system that integrated the images of an SLN in a mouse model using a –nanocolloid and an ICG fluorescence dye with a dichroic beam splitter module and respective charge-coupled diode (CCD) cameras. Recently, the feasibility of PET/NIR hybrid tracers that emit annihilation-gamma rays with higher sensitivity and spatial resolution than single gamma rays has been reported.21–26 With these dual-labeled hybrid tracers, fluorescence-guided surgery can be more convenient and expedited without unexpected excessive bleeding and damage of the normal tissue. To our knowledge, a multimodal laparoscopic system that detects the annihilation-gamma rays has not been developed for fluorescence-guided surgery. If a multimodal system can display the simultaneous images of gamma rays and optical fluorescence for intraoperative surgery, the surgical procedure will be faster and more precise. In this study, we present a multimodal imaging system that shows real-time images of positron annihilation-gamma rays, NIR, and visible light with higher spatial resolution and sensitivity than conventional intraoperative gamma probes. 2.Methods and Materials2.1.Schematic of the Annihilation-Gamma/Near-Infrared/Visible Multimodal SystemFigure 1 shows a schematic diagram of an annihilation-gamma/NIR/visible multimodal imaging system. When hybrid tracers are accumulated in patients, they emit annihilation-gamma rays in the opposite direction and NIR after absorbing external excitation photons with a range of 760 to 800 nm. If the wavelength band of excitation light is too wide, reabsorption occurs in the higher wavelength band of excitation light.27 To limit the wavelength band of excitation light to the range of 760 to 800 nm, a commercial ICG filter kit that includes excitation, emission, and dichroic imaging filters was used. The proposed laparoscopic module consisted of an internal detector for gamma ray detection and objective lens gathering optical and NIR photons. The external detector was placed outside patients for coincidence detection of annihilation photons with the internal detector. Scintillation light derived from the annihilation-gamma rays was directly detected by an array of multipixel photon counters (MPPCs). Visible light (500 to 700 nm) and emission NIR (800 to 850 nm) were transmitted through a gradient index (GRIN) relay lens and separated with a dichroic mirror module. The visible and NIR images from CCD cameras and the gamma images from MPPCs were merged and displayed on the monitor. The multimodal imaging system provides the simultaneous images of NIR and visible light in real time and the image of gamma rays within 1 s. 2.2.Annihilation-Gamma Detection for 511-keV Gamma RayThe annihilation-gamma detector consisted of two types of detectors. The internal (laparoscopic) detector was inserted into a surgical stainless pipe with an outer diameter of 15.8 mm and a thickness of 0.5 mm. A crystal array [Fig. 2(a)] of cerium-doped LYSO (EPIC Crystal, China) with the total dimension of (crystal size of and pitch of 1.4 mm) was directly coupled to a custom-assembled MPPC array [Fig. 2(b)]: One type of eight central MPPCs (S13360-2050VE, Hamamatsu, Japan) which has a photosensitive area of and a microcell size of , and the other type of peripheral 12 MPPCs (S13190-1015HDR, Hamamatsu, Japan) that has a photosensitive area of and a microcell size of 15 μm. Figure 2(b) shows the MPPC array attached to a custom-designed front-end board. The central blank area of the internal detector was reserved for a GRIN lens for the optical path of the visible light and NIR, as shown in Figs. 2(a) and 2(b). The crystal array consisted of 32 LYSO crystals because of the central blank area. The external detector was made up of four blocks of LYSO (EPIC Crystal, China) crystal array (crystal size of and pitch of ) dimension of [Fig. 2(c)]. All the four crystal blocks were coupled one to one with four arrays of MPPCs (S13361-3050-NE04, Hamamatsu, Japan). Figure 2(d) shows the front-end board containing the array of MPPCs. The surfaces between crystal array and MPPCs of both detectors were coupled with optical grease (BC-630, Saint-Gobain, France). All sides of the scintillation crystals in the internal detector, except for the surface facing the MPPCs, were wrapped with Teflon tape with a reflectivity of at 550 nm.28 The crystal surfaces in the external detector, except for the surface facing the MPPCs, were painted with barium sulfate () mixed powder ( at 550 nm).29 Each crystal surface of both detectors was chemically polished. To provide an external power source with each MPPC, a power supply board (BASP-10002, Brightonix, Korea), which has eight output channels, was used, as shown in Fig. 2(e). 2.3.Signal Processing of the Annihilation-Gamma DetectorThe analog signals from the internal detector were reduced to four position signals (A, B, C, and D) by a discretized positioning circuit (DPC). All the MPPCs in the internal detector were connected to a resistive chain, which generates the four position signals. The analog signals from the external detector were transformed into position signals (, , , and ) by a symmetric charge division (SCD) circuit. In general, the SCD circuit generates more position-sensitive signals than the DPC. However, the SCD circuit is not proper for the internal detector, which requires compact size since the SCD circuit requires more resistors than the DPC. The horizontal () and vertical () position signals were calculated by the equations, as shown in Fig. 3. The position signals from both detectors were discriminated by an on-board leading-edge discriminator with a 0.5-V threshold. The discriminated signals were fed into a coincidence logic unit (BASP-10003, Brightonix, Korea) which has 16 trigger input channels, 2 AND output channels, and 8 OR output channels [Fig. 2(f)]. The four position signals of both detectors were converted to analog-to-digital converter (ADC) values by a 32-channel waveform desktop digitizer (DT5740, CAEN, Italy) with a sampling frequency of 62.5 MHz [Fig. 2(g)]. The ADC values were used to calculate the and positions of gamma interaction points within the LYSO crystal array and to obtain the energy information by summing the four position signals. Figure 3 shows the block diagram to process the detected signal from the annihilation-gamma detector. 2.4.Analysis of Digitized List-Mode DataCrystal maps containing the gamma-struck positions were segmented using a Voronoi diagram algorithm implemented in MATLAB 2018b. The two-dimensional peak position of the crystal map was automatically picked and segmented into individual crystal addresses corresponding to the number of crystals. Each segmented area in the Voronoi diagram was assigned to an individual crystal, for which the energy spectrum was obtained. The energy spectrum in each crystal was normalized to compensate the different gain of the photodetector and light yield of each crystal using the following equation: where is the normalizing factor defined by mean ADC value of the spectrum with the highest energy resolution, is the weight factor of temperature change, and is the normalized energy spectrum from the individual energy spectrum (). After obtaining a lookup table (LUT), which contained the mean photopeak ADC value versus the temperature, at the respective temperature was calculated by dividing the mean photopeak ADC at 20°C.2.5.Evaluation of the Annihilation-Gamma DetectorBecause the gain of each MPPC channel varies with the bias voltage and temperature, it is necessary to measure the ADC values by changing these external conditions. To evaluate the intrinsic performance of the external detector, a reference detector was used. The reference detector consisted of a 16-channel MPPCs coupled with a gadolinium aluminum gallium garnet (C&A Corporation, Japan) crystal with a size of . A calibration disk source (Eckert & Ziegler, Germany) with an activity of 50 kBq was placed vertically between the reference detector and the external detector. The distances from the source to the reference detector and to external detector were 10 and 300 mm, respectively. All the detectors were placed inside a temperature-controlled chamber at 20°C. A total of 5.1 million prompt (coincidence) events were collected. 2.6.Design of Optical Lens, Parts and a Dichroic Mirror ModuleTo aid the assembly of the optical components such as optical lens and a dichroic mirror, the distance between the optical parts was optimized by using optical simulation software (ZEMAX) [Fig. 4(a)]. The optical imaging system was made up of a GRIN objective lens (GRINTECH, Germany) with an numerical aperture (NA) of 0.5 and a GRIN relay lens (GRINTECH, Germany) with an NA of 0.1 to transmit the properly focused optical light, because the amount of escaped light was smaller than with a higher NA value. Between the GRIN relay lenses, achromatic doublet lenses (Edmund Optics), which has a diameter of 6.25 mm and effective focal length of 7.5 mm, were placed at the center of optical axis with magnification of 1. A dichroic mirror module consisted of an aspherized achromatic lens with a diameter of 12.5 mm (Edmund Optics), a dichroic mirror (Edmund Optics), and CCD cameras (Ace 800-510, Basler, Germany). The dichroic mirror with a cut-off wavelength of 801 nm ( reflection) was placed at the center of a beam splitter box [Fig. 4(b)]. The CCD cameras with the same pixel size of were used for the acquisition of the reflected visible light and the passed NIR images. Figure 4(c) shows the transmission curves of the optical filters and the emission spectrum of the ICG fluorescence dye. The transmission curve of the dichroic mirror was in agreement with the emission spectrum of the ICG dye. 2.7.Integration of Annihilation-Gamma, Near-Infrared, and Visible ImagesThe calibration disk source with a 1.5-mm diameter and a 3D-printed plastic phantom containing an ICG dissolved in dimethyl sulfoxide (DMSO) were used to test the integration of sequentially acquired images. The hole at the center of the ICG plastic phantom has the diameter of 2 mm and height of 5 mm. Figure 5 shows the procedure of merging optical and gamma images. The NIR image rescaled to the gray level in Fig. 5(a) was changed to an RGB true color image with green-fluorescence pseudocolor for clear identification [Fig. 5(b)]. Because the CCD cameras have the same pixel size and the identification number, only off-axis calibration was carried out to merge the two images [Figs. 5(b) and 5(c)]. The reconstructed gamma ray image having a field of view (FOV) of was cropped to a rectangle (red box) corresponding to the FOV of the respective CCD cameras and resized to by using a two-dimensional cubic spline interpolation algorithm provided with MATALB software [Figs. 5(d) and 5(e)]. 2.8.Tissue-Like Phantom Test for Penetration Depth MeasurementAs shown in Fig. 6(a), the integrated images of the ICG-contained plastic phantom and the disk source covered with optical scattering material were evaluated to measure the penetration depth of NIR optical photons and to validate the annihilation-gamma rays independent on the optical scattering and absorption. The phantom and the source were located on a support jack, which is 10 mm apart from the center of origin. The acquisition time of the NIR CCD and the digitizer were 120 ms and 30 s, respectively. A tissue-like phantom was fabricated with 89% of water, 10% of gelatin powder from porcine skin (Sigma Aldrich), and 1% by weight of intralipid 20% (JW pharmaceutical, Korea). The mixture was heated to 37°C and stirred for 30 min until completely dissolved. After pouring into a silicon mold, the mixture was cooled for 20 h at room temperature. Figure 6(b) shows the tissue-like phantoms with different thicknesses of 5, 10, 15, and 20 mm, respectively. Only an area of of phantoms was used because the hole of the plastic phantom had a diameter of 2 mm. To measure the signal loss of NIR and the annihilation-gamma ray by the thickness of the tissue-like phantom, the total pixel value at region of interest (ROI) of the signal (red box) and the background (white box) in the obtained images was calculated [Fig. 6(c)]. 2.9.Phantom Image Test with and ICGTo validate the feasibility of the proposed multimodal imaging system, an image of 3D-printed acrylonitrile butadiene styrene-like plastic phantom that contained radionuclides and an ICG-DMSO mixture was obtained. Figure 7(a) shows the experimental setup of the phantom test. A tunable white light source (White 6500K, ScopeLED) and an excitation source (ML780L3, Thorlabs) with the peak wavelength of 780 nm were connected to a dichroic mirror module that transmits light from these light sources. The 0.25-mm-diameter excitation fibers (Eska-CK10, Mitsubishi, Japan) were assembled into four-layered ring shape (total 1-mm width) and were inserted at the edge of the laparoscopic (internal) detector. As shown in Fig. 7(b), the distance of the source to the internal detector and the external detector was 30 and 90 mm, respectively, to cover the full area of the plastic phantom. Half of the volumes of the holes with diameters of 2.6, 3.5, and 4.6 mm were filled with a mixture of and ICG. The activity of each hole was 0.78 (4.6 mm hole), 0.45 (3.5 mm hole), and 0.25 MBq (2.6 mm hole), respectively, and each hole has the same activity per volume (). Events were collected for 30 s to obtain the gamma image. Acquisition time and other parameters of the digitizer and respective CCD cameras were the same as in Secs. 2.4 and 2.6. 3.Results3.1.Gain Normalization of the Annihilation-Gamma DetectorFigure 8 shows the segmented crystal map, the energy resolution map, and the entire energy spectra from the crystals without and with gain normalization (photopeak alignment). The normalized photopeak ADC values were processed by the algorithm described in Sec. 2.4. The individual crystal segmentation of the internal detector and bottom-left blocks of the external detector performed well, as shown in Figs. 8(a) and 8(e). The other three blocks of the external detector showed similar crystal maps. The events in the central crystals of the segmented map of the internal detector were excluded from the true coincidence events because these events are dominantly derived from intercrystal scattering [Fig. 8(b)]. At the bias voltage of 58 V, without gain normalization, the mean photopeak ADC value and the energy resolution of the internal detector were ADC values and 52%, respectively [Fig. 8(c)]. After gain normalization, the energy resolution of the internal detector was 21%, as shown in Fig. 8(d). Similarly, the ADC values of the external detector were also normalized. Figure 8(f) represents the energy resolution maps of the external detector. The energy resolution of top-left and top-right crystal blocks were slightly worse than that of other crystal blocks. This degradation was mainly caused by light output variation between the crystals, and slightly nonuniform optical coupling condition between the crystal and the MPPCs, and the gain difference of the MPPC channels. For these reasons, the energy spectra of the four crystal blocks of the external detector were widely different from each other so that the mean energy resolution could not be calculated, as shown in Fig. 8(g). With gain normalization, the energy resolution of the external detector was 18% at the bias voltage of 58 V [Fig. 8(h)]. 3.2.Performance of the Annihilation-Gamma DetectorFigure 9 shows the normalized mean photopeak ADC values and the energy resolution obtained from the annihilation-gamma detector depending on the bias voltage and temperature. The mean photopeak ADC values and the energy resolution of top-left (dotted-line) and top-right (one-dot chain line) crystal blocks were slightly worse than the other crystal blocks, as described in Sec. 3.1. Because the maximum overvoltage of the MPPCs is 61 V (kapd1053e.pdf, Spec sheet, Hamamatsu, Japan), the performance of all the MPPCs was evaluated from 56 to 61 V. The best (lowest) energy resolution of the internal detector was 16% at 58.5 V, and the normalized energy resolution of the four crystal blocks of the external detector was 19% at 59 V [Fig. 9(b)]. The bias of 72 V for the 12-edge MPPCs with a photosensitive area of was applied. As shown in Figs. 9(c) and 9(d), the ADC values of the annihilation gamma detector were estimated depending on the external temperature. For simplicity, all crystal blocks in the external detector were applied at bias voltage of 59 V. The applied bias voltage to the internal detector was 58.5 V. The mean photopeak ADC value of both detectors shows the linear dependence on external temperature. Because the data acquired at real-time basis are insufficient to generate true coincidence event, all the mean photopeak ADC values and , as described in Sec. 2.4, were automatically saved as a calibration LUT for continuous data acquisition. 3.3.Performance of Acquired Optical ImagesFigures 10(a) and 10(b) show the USAF 1951 resolution target images and the smoothed line profiles of selected ROI (red box) obtained by the optical system inside the laparoscopic module. The bar phantom images acquired by two CCD cameras were clearly visible. Figure 10(c) shows the horizontal and vertical contrast transfer functions (CTFs) at the square ROI corresponding to each size of pattern. Even though the base noise signals were caused by the fiber-optic light sources, the differences in the peak and the valley of the images were clearly shown in all line profiles. There was some mismatched contrast region (vertical redline) because of unproperly calibrated flat-field mask images. The minimum distinguishable bar pattern of the USAF target corresponds to 4 line-pair per millimeter. 3.4.Penetration Depth Measurements of Near-Infrared and Annihilation-Gamma RaysFigure 11 shows the integrated images covered with the tissue-like phantoms having different thicknesses ranging from 0 (left) to 15 mm (right). At the phantom thickness of 10 mm, the pseudo-fluorescence-green-colored NIR image was not clearly seen because of severe scattering by intralipid compounds [Fig. 11(a)]. However, the images of annihilation-gamma rays were not influenced by the thickness of the scattering material even at the thickness of 15 mm [Fig. 11(b)]. Figure 12 shows the relative and total pixel intensity curves versus the thickness of the tissue-like phantom. To calculate the total pixel intensity, pixel values in the signal ROI was summed and subtracted by pixel values in the background ROI. The signal and background ROIs have the same size for NIR () and gamma rays (). In case of gamma rays, there was no signal in the background ROI. The total pixel intensity of the NIR image significantly decreased by the thickness of the scattering material. Because the value at the thickness of 20 mm was almost the same as the thickness of 15 mm, the result at 20-mm thickness was excluded. The total pixel intensity of the image of annihilation-gamma rays was almost the same over the thickness and not affected by the thickness of the tissue-like phantom. 3.5.Simultaneously Acquired 511-keV Gamma Ray, Near-Infrared, and Visible Phantom ImageFigure 13(a) shows the reconstructed 511-keV gamma images using simple backprojection algorithm. The image was clearly seen at the acquisition time of 5 s, and the true coincidence count rate at the distance of 30 mm from the internal detector was . As shown in Fig. 13(e), the acquired gamma image was resized by a factor of 10 and the average pixel intensity values on the axis and the full width at half maximum (FWHM) of the rectangular ROI (white line) were calculated. Figure 13(c) shows the intensity profile of each hot phantom hole. After the quadratic subtraction of the size of phantom holes, the FWHM of the gamma image in the hole of 2.6, 3.5, and 4.6 mm was, respectively, 2.8, 3.0, and 3.8 mm at the acquisition time of 30 s. Figure 14 shows the integrated image of gamma ray, NIR, and visible light of the plastic phantom-contained and the ICG-DMSO compound solution. The pixel size of the reconstructed image was 1.4 mm, corresponding to a real crystal pitch of the internal detector, and the reconstructed image was resized to the same size of the CCD cameras, as shown Fig. 14(c). After the measurement for 30 s, the detected true coincidence events were reconstructed and were updated to the image generated from the measurement for 1 s. As shown in Fig. 14(f), the image overlaid with the pseudocolored NIR and gamma image was well matched. 4.DiscussionRecently, optical imaging with fluorescence has become a new paradigm for intraoperative surgery because of its superior sensitivity and spatial resolution. Particularly, NIR has a relatively long penetration depth (), compared to other fluorescence types. Because NIR optical images provide information on the clear margins of suspicious lesions, they have been widely used for the detection of various diseases. However, fluorescence has limited penetration depth because of the absorption and scattering by tissues such as lipid, blood, and skin. The multimodal system with radiation would be effective to overcome the limited depth problem of fluorescence, because it has higher penetration depth and the additional possibility to display a molecular image. Latest research shows that the spatial resolutions of single gamma ray imaging systems employing multipinholes exceed those of positron annihilation-gamma ray images.30 On the other hand, it is challenging to incorporate the multipinhole single gamma imaging system into the image-guided laparoscopy because of spatial constraints. In general, positron annihilation-gamma ray images have an advantage of better image quality than other single gamma ray imagers, except a multipinhole imager, because of their higher spatial resolutions. The most widely used positron-emitting nuclides have relatively short half-lives of 110 min () and 68 min (). For this reason, molecular-guided surgery and biopsy using PET radiotracers have recently been used for various diseases, such as brain, lymph node, and prostate (prostate-specific antigen negative) cancers, in spite of radiation exposure during radio-guided surgery using high-energy gamma rays.31 Furthermore, according to the research by Povoski et al.,32 occupational radiation exposure is relatively small during intraoperative or perioperative surgical procedures because of the short half-lives of radionuclides. With the mean -FDG activity of injected to the patients during radio-guided surgeries, the mean deep-dose equivalents per hour of exposure were and for preoperative nurse and surgeon, respectively. The annual occupational exposure limit for adults defined by the International Commission on Radiological Protection (ICRP) is of total effective dose equivalent.33 According to another report by Heckathorne et al.,34 all the surgical staff including surgeons and scrub nurses received per procedure. Recently, laparoscopes and endoscopes assembled to a fiber-optic image guide bundle have been developed because of their convenience to be coupled with other optical parts, including optical lenses or laser sources. In our unpublished study, integrated phantom images were obtained by using the first prototype multimodal system based on the glass optical fiber bundle. However, there was significant loss of the scintillation lights of (peak emission wavelength of scintillation light was 420 nm for LYSO crystals) through the bundle, which was derived for light absorption by glass materials and the reflection of cladding materials. To overcome the loss of transmitted scintillation lights, the internal detector (MPPCs) was directly coupled with scintillation crystals without the optical fiber bundle. The energy resolution of the internal detector was 15% better than the previous optical fiber-based internal detector. Because the performance of the coincidence detectors was dominantly affected by light collection efficiency, the internal detector in this study had a better true coincidence count rate. The FWHM of the gamma ray image was found to be 3.8 mm at 4.6-mm-diameter hole using the proposed annihilation-gamma/NIR/visible imaging system, and the spatial resolution of the gamma ray images of 5-mm-diameter-cylinder sources acquired by previous endoscopic gamma/NIR/visible module was about 10 mm with the pinhole diameter of 1.5 mm.35 Moreover, the minimum acquisition time to acquire the image of radioactive source was six times faster than that of previously developed endoscopic module. Because of high energy (511 keV) and coincidence detection, the proposed annihilation-gamma-ray/NIR/visible multimodal system has the higher sensitivity and the spatial resolution than the single gamma multimodal system not employing the multipinholes. However, incorporating the multipinhole single gamma imaging system into the image-guided laparoscopy seems to be difficult because of spatial constraints. In this study, we acquired the transaxial image of the reconstructed image matrix that has a slice thickness of 10 mm at the depth of an optical image, because we already knew the position of the phantom. However, it was necessary to obtain the relative position information between the laparoscopic module and the tracers, because the position of the hybrid tracers will be unknown during the surgical procedure. The four-dimensional optical tracker will be adopted for future work, because the position of the laparoscope can be changed during intraoperative surgery. The optical tracker provides the position, distance, and the angular information in near real time, which is suitable for laparoscopic surgery. It is also important to provide a continuous gamma ray image without delay, as well as the sound information that more clearly provides the distribution of radio pharmaceuticals. For three-dimensional image reconstruction, the lack of angular samples compared to the conventional ring PET scanner can cause distortion in the reconstructed image matrix.36 It is important to know the actual geometric position reflected by the movement and the rotation of the coincidence detectors. Therefore, if there are limited angular samples, additional methods to compensate for these missing angular samples should be implemented to display the real-time images. 5.ConclusionBy using the multimodal imaging system, simultaneously integrated annihilation-gamma rays, NIR, and visible images can be obtained. In this study, we proved the feasibility of a multimodal annihilation-gamma/NIR/visible imaging system for intraoperative laparoscopic surgery. In future work, an additional backup study and preclinical test with mouse will be conducted with a completed real-time multimodal imaging system. AcknowledgmentsThis work was supported by grants from the National Research Foundation (NRF) of Korea of the Ministry of Science, ICT, and Future Planning Nuclear R&D Program (No. NRF-2016M2A2A4A03913619) and NRF of Korea of the Ministry of Science, ICT and Future Planning, Nuclear R&D Program (No. NRF-2017M2A2A4A01071175). ReferencesB. J. Ramshaw,
“Laparoscopic surgery for cancer patients,”
CA: Cancer J. Clin., 47
(6), 327
–350
(1997). https://doi.org/10.3322/canjclin.47.6.327 Google Scholar
R. Veldkamp et al.,
“Laparoscopic surgery versus open surgery for colon cancer: short-term outcomes of a randomised trial,”
Lancet Oncol., 6
(7), 477
–484
(2005). https://doi.org/10.1016/S1470-2045(05)70221-7 LOANBN 1470-2045 Google Scholar
M. A. Carbajo et al.,
“Laparoscopic treatment vs open surgery in the solution of major incisional and abdominal wall hernias with mesh,”
Surg. Endosc., 13
(3), 250
–252
(1999). https://doi.org/10.1007/s004649900956 Google Scholar
S. Luo et al.,
“A review of NIR dyes in cancer targeting and imaging,”
Biomaterials, 32
(29), 7127
–7138
(2011). https://doi.org/10.1016/j.biomaterials.2011.06.024 BIMADU 0142-9612 Google Scholar
R. A. Cahill et al.,
“Near-infrared (NIR) laparoscopy for intraoperative lymphatic road-mapping and sentinel node identification during definitive surgical resection of early-stage colorectal neoplasia,”
Surg. Endosc., 26
(1), 197
–204
(2012). https://doi.org/10.1007/s00464-011-1854-3 Google Scholar
T. B. Manny, M. Patel and A. K. Hemal,
“Fluorescence-enhanced robotic radical prostatectomy using real-time lymphangiography and tissue marking with percutaneous injection of unconjugated indocyanine green: the initial clinical experience in 50 patients,”
Eur. Urol., 65
(6), 1162
–1168
(2014). https://doi.org/10.1016/j.eururo.2013.11.017 EUURAV 0302-2838 Google Scholar
G. Spinoglio et al.,
“Real-time near-infrared (NIR) fluorescent cholangiography in single-site robotic cholecystectomy (SSRC): a single-institutional prospective study,”
Surg. Endosc., 27
(6), 2156
–2162
(2013). https://doi.org/10.1007/s00464-012-2733-2 Google Scholar
B. E. Schaafsma et al.,
“The clinical use of indocyanine green as a near-infrared fluorescent contrast agent for image-guided oncologic surgery,”
J. Surg. Oncol., 104
(3), 323
–332
(2011). https://doi.org/10.1002/jso.v104.3 JSONAU 0022-4790 Google Scholar
M. D. Jafari et al.,
“The use of indocyanine green fluorescence to assess anastomotic perfusion during robotic assisted laparoscopic rectal surgery,”
Surg. Endosc., 27
(8), 3003
–3008
(2013). https://doi.org/10.1007/s00464-013-2832-8 Google Scholar
S. L. Troyan et al.,
“The FLARE intraoperative near-infrared fluorescence imaging system: a first-in-human clinical trial in breast cancer sentinel lymph node mapping,”
Ann. Surg. Oncol., 16
(10), 2943
–2952
(2009). https://doi.org/10.1245/s10434-009-0594-2 Google Scholar
F. P. Verbeek et al.,
“Optimization of near-infrared fluorescence cholangiography for open and laparoscopic surgery,”
Surg. Endosc., 28
(4), 1076
–1082
(2014). https://doi.org/10.1007/s00464-013-3305-9 Google Scholar
E. C. Rossi, A. Ivanova and J. F. Boggess,
“Robotically assisted fluorescence-guided lymph node mapping with ICG for gynecologic malignancies: a feasibility study,”
Gynecol. Oncol., 124
(1), 78
–82
(2012). https://doi.org/10.1016/j.ygyno.2011.09.025 GYNOA3 Google Scholar
C. Chi et al.,
“Intraoperative imaging-guided cancer surgery: from current fluorescence molecular imaging methods to future multi-modality imaging technology,”
Theranostics, 4
(11), 1072
–1084
(2014). https://doi.org/10.7150/thno.9899 Google Scholar
A. L. Vahrmeijer et al.,
“Image-guided cancer surgery using near-infrared fluorescence,”
Nat. Rev. Clin. Oncol., 10
(9), 507
–518
(2013). https://doi.org/10.1038/nrclinonc.2013.123 Google Scholar
J. R. van der Vorst et al.,
“Near-infrared fluorescence-guided resection of colorectal liver metastases,”
Cancer, 119
(18), 3411
–3418
(2013). https://doi.org/10.1002/cncr.28203 CANCAR 0008-543X Google Scholar
H. G. van der Poel et al.,
“Intraoperative laparoscopic fluorescence guidance to the sentinel lymph node in prostate cancer patients: clinical proof of concept of an integrated functional imaging approach using a multimodal tracer,”
Eur. Urol., 60
(4), 826
–833
(2011). https://doi.org/10.1016/j.eururo.2011.03.024 EUURAV 0302-2838 Google Scholar
N. S. van den Berg et al.,
“Sentinel lymph node biopsy for prostate cancer: a hybrid approach,”
J. Nucl. Med., 54
(4), 493
–496
(2013). https://doi.org/10.2967/jnumed.112.113746 JNMEAQ 0161-5505 Google Scholar
I. Stoffels et al.,
“Evaluation of a radioactive and fluorescent hybrid tracer for sentinel lymph node biopsy in head and neck malignancies: prospective randomized clinical trial to compare ICG-(99m)Tc-nanocolloid hybrid tracer versus (99m)Tc-nanocolloid,”
Eur. J. Nucl. Med. Mol. Imaging, 42
(11), 1631
–1638
(2015). https://doi.org/10.1007/s00259-015-3093-7 Google Scholar
J. Kang et al.,
“Real-time sentinel lymph node biopsy guidance using combined ultrasound, photoacoustic, fluorescence imaging: in vivo proof-of-principle and validation with nodal obstruction,”
Sci. Rep., 7 45008
(2017). https://doi.org/10.1038/srep45008 SRCEC3 2045-2322 Google Scholar
H. G. Kang et al.,
“A feasibility study of an integrated NIR/gamma/visible imaging system for endoscopic sentinel lymph node mapping,”
Med. Phys., 44
(1), 227
–239
(2017). https://doi.org/10.1002/mp.12029 MPHYA6 0094-2405 Google Scholar
N. S. Van Den Berg et al.,
“Hybrid tracers for sentinel node biopsy,”
Q. J. Nucl. Med. Mol. Imaging, 58
(2), 193
–206
(2014). Google Scholar
F. Chen et al.,
“In vivo tumor vasculature targeted PET/NIRF imaging with TRC105(Fab)-conjugated, dual-labeled mesoporous silica nanoparticles,”
Mol. Pharmaceut., 11
(11), 4007
–4014
(2014). https://doi.org/10.1021/mp500306k MPOHBP 1543-8384 Google Scholar
H. Hu et al.,
“PET and NIR optical imaging using self-illuminating (64)Cu-doped chelator-free gold nanoclusters,”
Biomaterials, 35
(37), 9868
–9876
(2014). https://doi.org/10.1016/j.biomaterials.2014.08.038 BIMADU 0142-9612 Google Scholar
J. Wu, D. Pan and L. W. Chung,
“Near-infrared fluorescence and nuclear imaging and targeting of prostate cancer,”
Transl. Androl. Urol., 2
(3), 254
–264
(2013). https://doi.org/10.3978/j.issn.2223-4683.2013.09.05 Google Scholar
D. A. Heuveling et al.,
“Nanocolloidal albumin-IRDye 800CW: a near-infrared fluorescent tracer with optimal retention in the sentinel lymph node,”
Eur. J. Nucl. Med. Mol. Imaging, 39
(7), 1161
–1168
(2012). https://doi.org/10.1007/s00259-012-2080-5 Google Scholar
S. Lutje et al.,
“Targeted radionuclide and fluorescence dual-modality imaging of cancer: preclinical advances and clinical translation,”
Mol. Imaging Biol., 16
(6), 747
–755
(2014). https://doi.org/10.1007/s11307-014-0747-y Google Scholar
S. Gioux, H. S. Choi and J. V. Frangioni,
“Image-guided surgery using invisible near-infrared light: fundamentals of clinical translation,”
Mol. Imaging, 9
(5), 237
–255
(2010). https://doi.org/10.2310/7290.2010.00034 Google Scholar
M. Janecek,
“Reflectivity spectra for commonly used reflectors,”
IEEE Trans. Nucl. Sci., 59
(3), 490
–497
(2012). https://doi.org/10.1109/TNS.2012.2183385 IETNAE 0018-9499 Google Scholar
E. M. Patterson, C. E. Shelden and B. H. Stockton,
“Kubelka–Munk optical properties of a barium sulfate white reflectance standard,”
Appl. Opt., 16
(3), 729
–32
(1977). https://doi.org/10.1364/AO.16.000729 APOPAI 0003-6935 Google Scholar
N. U. Schramm et al.,
“High-resolution SPECT using multipinhole collimation,”
IEEE Trans. Nucl. Sci., 50
(3), 315
–320
(2003). https://doi.org/10.1109/TNS.2003.812437 IETNAE 0018-9499 Google Scholar
B. Fei and D. M. Schuster,
“PET molecular imaging-directed biopsy: a review,”
Am. J. Roentgenol., 209
(2), 255
–269
(2017). https://doi.org/10.2214/AJR.17.18047 AJROAM 0092-5381 Google Scholar
S. P. Povoski et al.,
“Comprehensive evaluation of occupational radiation exposure to intraoperative and perioperative personnel from -FDG radioguided surgical procedure,”
Eur. J. Nucl. Med. Mol. Imaging, 35
(11), 2026
–2034
(2008). https://doi.org/10.1007/s00259-008-0880-4 Google Scholar
A. D. Wrixon,
“New ICRP recommendations,”
J. Radiol. Prot., 28 161
–168
(2008). https://doi.org/10.1088/0952-4746/28/2/R02 JRPREA 0952-4746 Google Scholar
E. Heckathorne, C. Dimock and M. Dahlbom,
“Radiation dose to surgical staff from positron-emitter-based localization and radiosurgery of tumors,”
Health Phys., 95
(2), 220
–226
(2008). https://doi.org/10.1097/01.HP.0000310962.96089.44 HLTPAO 0017-9078 Google Scholar
H. G. Kang et al.,
“Proof-of-concept of a multimodal laparoscope for simultaneous NIR/gamma/visible imaging using wavelength division multiplxing,”
Opt. Express, 26
(7), 8325
–8339
(2018). https://doi.org/10.1364/OE.26.008325 OPEXFF 1094-4087 Google Scholar
G. Tarantola et al.,
“PET instrumentation and reconstruction algorithm in whole-body applications,”
J. Nucl. Med., 44
(5), 756
–769
(2003). JNMEAQ 0161-5505 Google Scholar
BiographySeong H. Song is a PhD student of senior healthcare at Eulji University. He received his BS degree in radiological science from the Eulji University in 2015 and MS degree in senior healthcare from the Eulji University in 2018. His current research interests include developing a multimodal single-gamma camera with optical imaging for open surgery and a high-energy gamma ray (511 keV)/NIR/visible laparoscopic system for intraoperative surgery. He is a member of SPIE. |