14 December 2022 Tissue harvest with a laser microbiopsy
Jason B. King, Nitesh Katta, Sapun H. Parekh, Thomas E. Milner, James W. Tunnell
Author Affiliations +
Abstract

Significance

Traditional pathology workflow suffers from limitations including biopsy invasiveness, small fraction of large tissue samples being analyzed, and complex and time-consuming processing.

Aim

We address limitations of conventional pathology workflow through development of a laser microbiopsy device for minimally invasive harvest of sub-microliter tissue volumes. Laser microbiopsy combined with rapid diagnostic methods, such as virtual hematoxylin and eosin (H&E) imaging has potential to provide rapid minimally invasive tissue diagnosis.

Approach

Laser microbiopsies were harvested using an annular shaped Ho:YAG laser beam focused onto the tissue surface. As the annulus was ablated, the tissue section in the center of the annulus was ejected and collected directly onto a glass slide for analysis. Cryogen spray cooling was used before and after laser harvest to limit thermal damage. Microbiopsies were collected from porcine skin and kidney. Harvested microbiopsies were imaged with confocal microscopy and digitally false colored to provide virtual H&E images.

Results

Microbiopsies were successfully harvested from porcine skin and kidney. Computational and experimental results show the benefit of cryogen pre- and post-cooling to limit thermal damage. Virtual H&E images of microbiopsies retained observable cellular features including cell nuclei.

Conclusions

Laser microbiopsy with virtual H&E imaging shows promise as a potential rapid and minimally invasive tool for biopsy and diagnosis.

1.

Introduction

Tissue harvest and examination is widely used for diagnosis and intraoperative margin assessment of disease including many types of cancer. Examination of thinly sliced and hematoxylin and eosin (H&E) stained tissue sections is the gold standard despite numerous well-known limitations. The traditional workflow involves: (1) harvest of tissue sections on the order of a mm or larger using mechanical tools such as punches, needles, scalpels, or forceps, (2) preparing the tissue for sectioning by either fixing and embedding in paraffin (>1  day) or by freezing, (3) slicing the tissue into thin sections (3 to 10  μm), (4) staining with H&E (>20  min),1 and (5) microscopic examination by a pathologist. Limitations of this workflow include biopsy invasiveness, small fraction of harvested tissue analyzed, relatively long time required for processing, human factors including need for highly trained technicians, and artifacts from processing and sectioning.2

Limitations of the traditional H&E workflow have been well recognized leading to significant research and development of alternative techniques. Non-invasive or minimally invasive sampling methods have been proposed to replace traditional biopsies. Non-invasive methods include probe devices such as mass spectrometry3 or optical spectroscopy4 probes as well as tape stripping.5 Minimally invasive microbiopsy tools have been designed to mechanically cut and harvest sub-microliter (<1  mm3) tissue volumes.69 These microbiopsies are hundreds of microns in diameter, making them compatible with the traditional H&E workflow in addition to compatibility with alternative methods, such as mass spectrometry, optical spectroscopy, and genetic testing. We present an alternative approach using a laser for minimally invasive microbiopsy. While lasers have been utilized previously for laser microdissection to harvest cells from thin histologic sections or cell cultures,1014 the laser microbiopsy approach described here harvests three-dimensional tissue volumes from thick tissues. In addition, we describe a photothermal cutting mechanism as opposed to plasma mediated ablation used for laser microdissection.13,14 Advantages of laser vs. mechanical tissue harvest include non-contact cutting, precise adjustable laser beam shaping, and fiber optic delivery.

Additionally, progress has been achieved on creating H&E looking images without sectioning and with minimal to no tissue processing. These “virtual H&E” images can be generated using microscopy techniques including confocal,15,16 two-photon,1719 light-sheet,20 ultraviolet surface excitation,21 structured illumination,22 optical coherence tomography (OCT),23 and stimulated Raman scattering.24,25 Virtual H&E images can be generated within minutes of tissue harvest, do not require a histology technician to process the tissue, and do not have artifacts that result from traditional processing and sectioning.

We propose a workflow for rapid, minimally invasive tissue harvest and analysis. An annular shaped laser beam is used to cut sub-microliter tissue volumes. The tissue volumes are ejected from the bulk tissue and are collected directly onto a glass slide for analysis. Cryogen spray cooling is used before and after tissue harvest to reduce tissue damage. Harvested tissue sections are rapidly stained, imaged with confocal microscopy, and false colored to create virtual H&E images within minutes of harvest.

2.

Materials and Methods

2.1.

Laser Microbiopsy Overview

An overview of the laser microbiopsy approach is shown in Fig. 1. A pulsed laser beam is shaped through a lens system to generate a converging annular beam profile at the tissue surface. The annular beam shape ablates the tissue, and the tissue section in the center of the annulus is ejected. Cryogen spray cooling is used before and after the laser pulse to limit tissue damage and patient discomfort. These tissue sections, with volumes <1  μl (=1  mm3), are collected onto a glass slide where they can be analyzed using desired methods including, but not limited to, conventional H&E staining, virtual H&E or other microscopy methods, and molecular profiling.

Fig. 1

Overview of laser microbiopsy workflow. An annular beam is used to cut a sub-microliter tissue section. Cryogen spray cooling is used pre- and post-ablation to limit thermal damage. The tissue is ejected and collected directly onto a glass slide for imaging/analysis. Figure and portions of methods adapted from Refs. 26 and 27.

JBO_27_12_125001_f001.png

2.2.

Benchtop Laser Microbiopsy Implementation

We have designed and built a benchtop laser microbiopsy system consisting of the following components (Fig. 2): (1) A Ho:YAG laser (MOSES Pulse 120H, Lumenis, Yokneam, Israel) delivered through a 230-μm core diameter multimode fiber (Moses 200 D/F/L, Lumenis, Yokneam, Israel). An SMA connector was added to the distal end of the fiber to allow connection to a fiber mount with x-y-z positioning and tilt control for precise optical alignment. (2) A ZnSe aspheric collimation lens (ASPH-ZC-25-50, ISP Optics, Orlando, Florida) to collimate light from the fiber. (3) Two fused silica axicons (α=20  deg, Thorlabs, Newton, New Jersey) to shape the collimated circular beam into an annular beam. The axicons are oriented with the points of the conical surface facing in the direction of light propagation as opposed to the typical orientation with the axicon points facing each other. This axicon orientation provides an annular beam diverging at an angle of 0.6 deg rather than perfect collimation, which results in the beam retaining an annular shape at the focal plane. (4) A ZnSe aspheric focusing lens (ASPH-ZC-25-25, ISP Optics, Orlando, Florida) is utilized to focus the annular shape. The optical components produce a converging annular shape with decreasing diameter towards the beam focus. A similar beam shaping approach has been reported for material processing using an annular beam.28 The converging annular shape conveniently allows a variable annulus diameter at the tissue surface by adjusting the distance between the focusing lens and the tissue surface. Additionally, the beam converges into the tissue, generating a pressure gradient necessary to eject the harvested tissue upward where it is collected directly onto a glass slide. Cryogen spray cooling (R134A) is used before and after pulsed laser emission to limit thermal damage. A solenoid fuel injector valve (800-1257N, GP Sorensen, Long Island City, New York) is used to deliver the cryogen spray pre-cooling directed at the tissue surface from 4 cm away and a second valve is used to deliver post-cooling directed at the biopsy collected on the glass slide from 4 cm away.

Fig. 2

Diagram of optics used in this study. A Ho:YAG laser is shaped into an annular beam using a pair of axicons. The annular beam is focused on the tissue surface through an aspheric focusing lens. Cooling and harvesting slide not shown.

JBO_27_12_125001_f002.png

2.3.

Transmission Measurement

Transmission through the beam-shaping optics was measured. A power meter (PM10, Coherent, Santa Clara, California) was used to record average power before and after the beam-shaping optics, and percent transmission was calculated.

2.4.

Beam Profile Measurement

The annular beam profile produced by our benchtop system was explored computationally and experimentally. The optical components were imported into raytracing software (Zemax), and the resulting beam profile was measured in software with a virtual detector. The beam profile was measured experimentally by irradiating a 100-μm layer of water in a rectangular capillary tube and imaging the emitted radiation using an IR camera (420-0044-04-00, FLIR, Wilsonville, Oregon) [Fig. 5(a)]. For these measurements, the laser pulse energy was set to the minimum level (0.2 J). Pulse energy was further reduced by coupling light between two optical fibers with intentionally poor coupling efficiency. Beam-shaping optics were oriented horizontally, and the rectangular capillary was placed a variable distance, Z, from the back surface of the focusing lens. The capillary was initially placed at Z=20  mm and increased in increments of 0.25 mm to measure the beam profile at varying focal locations. A ZnSe lens (ASPH-ZC-25-50, ISP Optics, Orlando, Florida) was placed between the capillary and the IR camera to magnify the beam profile image. A video was recorded by the IR camera at a frame rate of 432.5 Hz and an integration time of 0.50 ms. The first frame after laser emission was recorded and used for beam profile analysis. Computational and experimental beam profiles were compared.

2.5.

Computational Modeling

The laser microbiopsy processes were modeled computationally to gain insight in beam-shaping optimization, interaction with annular beam and tissue, tissue ablation, and thermal damage. The simulation is based off of a blow-off model as described by Vogel and Venugopalan.29 This model is conditional on the laser pulse being shorter than the thermal diffusion time providing thermal confinement. In the case of annular ablation, we calculate the thermal diffusion time using Eq. (1), where l is the width of the annular beam from inner to outer diameter, α is the thermal diffusivity, k is the thermal conductivity, ρ is the tissue density, and Cp is the tissue specific heat capacity (Table 1)

Eq. (1)

td=l24α,α=kρCp.

Table 1

Parameters used in computational modeling.

ParameterValue
Pulse energy1 J
Absorption coefficient, μa30  cm1
Reduced scattering coefficient, μs6  cm1, 3  cm1, 0
Scattering phase functionHenyey–Greenstein30
Anisotropy factor, g0.9
Tissue density, ρ1109  kg/m3
Specific heat capacity, Cp3391 J/kgK
Thermal conductivity, k0.4 W/mK
Convective heat transfer coefficient, h10  W/m2K, 10,000  W/m2K
External convective medium temperature, Text23°C, −40°C31
Arrhenius frequency factor, A4.585e72s132
Arrhenius activation energy, Ea4.710e5 J/mol32

Using an annular beam width of 100  μm gives a thermal diffusion time of 24 ms. The pulse duration of the laser is much shorter with a FWHM duration of 71  μs.33 Therefore, thermal confinement during the laser pulse is a justified assumption, and the blow-off model can be used.

The computational model consisted of the following steps. (1) The optical fiber and beam-shaping optics were modeled in raytracing software (Zemax), and the beam profile at varying distance from the beam focus was simulated. (2) A volume detector with the absorption and scattering properties of soft tissue (Table 1) was inserted at a distance such that the outer diameter of the annular beam was 800  μm. (3) The simulated fluence in the tissue was computed with raytracing and exported to MATLAB [Fig. 3(a)]. 4) Regions with a radiant energy density greater than a threshold of 2.5  J/mm3 were assumed to be ablated, and the tissue inside the ablation annulus was assumed to be ejected as a microbiopsy at the conclusion of the laser pulse [Fig. 3(b) and 3(c)]. The ablation threshold is derived from the energy required to vaporize water within the tissue.34 Energy absorbed in the ablation region was removed from the model as this energy is assumed to be spent on ablation and kinetic energy of tissue expulsion. The energy absorbed in the regions of tissue that are not ablated contributes to tissue heating. The temperature at the end of the laser pulse at each depth and radius, is calculated using Eq. (2), where 37°C is the initial body temperature, F(r,z) is the fluence at each location, μa is the absorption coefficient, ρ is the tissue density, and Cp is the tissue specific heat capacity (Table 1):29

Eq. (2)

T(r,z)=37°C+F(r,z)μaρCp.

Fig. 3

Computational model steps. (a) Fluence through the tissue was modeled using Zemax. (b) Regions with fluence above the ablation threshold were removed [displayed as dark blue in (b) and (c)]. (c) The microbiopsy separated from the bulk tissue at the conclusion of the laser pulse and the post ablation temperature evolution was calculated. (d) The microbiopsy volume and initial temperature were exported to COMSOL for modeling of bioheat transfer and thermal damage.

JBO_27_12_125001_f003.png

(5) The simulated biopsy was severed at half of the maximum ablation depth to give a simulated microbiopsy volume within the range of experimental microbiopsy volumes. The volume and geometry of the tissue in the middle of the ablated annulus and above the severed depth were saved to an STL file and imported into COMSOL multiphysics (COMSOL) as the volume of the ejected microbiopsy [Fig. 3(d)]. The temperature at the end of the laser pulse [Eq. (2)] within the ejected microbiopsy volume was imported into COMSOL [Figs 3(c) and 3(d)]. (6) The heat transfer through the tissue as it cools was modeled in COMSOL. (7) The Arrhenius damage integral [Eq. (3)] and the percentage of cell death [Eq. (4)] were calculated in COMSOL for each location in the microbiopsy.29,35,36

In these equations, Ω is the Arrhenius thermal damage parameter, A is the frequency factor, Ea is the activation energy per mole, R is the universal gas constant, and T is the tissue temperature in Kelvin. Values used for these parameters are given in Table 1. Importantly, the Arrhenius damage integral is used here as an overall measure of thermal damage to make comparisons for optimization but does not necessarily indicate whether the tissue will retain integrity for analysis by virtual H&E or other analysis methods. An Arrhenius thermal damage parameter of Ω=1 indicates 63% cell death and is used as a threshold for thermal damage.

Eq. (3)

Ω(τ)=0τA*e(EaR*T(t))dt,

Eq. (4)

Damage(%)=100*(1eΩ).

This model allows for various parameters to be varied to better understand the laser microbiopsy process and aid in optimization to limit microbiopsy thermal damage. Important parameters that were explored included tissue scattering coefficient, convective heat transfer coefficient, and external temperature. The effects of cryogen spray cooling before (pre-cooling) and/or after (post-cooling) the laser pulse were explored. To model pre-cooling, the heat transfer through tissue was calculated with a surface temperature of 40°C and a convective heat transfer coefficient at the surface of 10,000  W/m2K for a duration of 200 ms. Detailed descriptions of cryogen spray cooling including computational models and experimental results can be found elsewhere.31,3739 To model post-cooling the convective heat transfer coefficient and external temperature were set to 10,000  W/m2K and 40°C. To model cooling in room air, the convective heat transfer coefficient and external temperature were set to 10  W/m2K and 23°C.

The effects of tissue scattering were also explored. The model was run with normal tissue reduced scattering coefficient (μs=6  cm1),40 scattering reduced by half (μs=3  cm1), and without scattering (μs=0). Experimental results with reduced tissue scattering are not available for microbiopsy volume comparison. Therefore, biopsy volumes were assumed to separate from the bulk tissue at the maximum ablation depth for these models.

The full set of parameters used in the computational model is summarized in Table 1. The key assumptions made in this model are summarized in Fig. 4.

Fig. 4

Key assumptions made in the thermal damage model for each stage: pre-cooling, ablation, and post-cooling.

JBO_27_12_125001_f004.png

2.6.

Microbiopsy Harvest Experiments

Microbiopsy harvest experiments were completed on ex vivo porcine tissues (Animal Technologies, Tyler, Texas). The protocol involved the following: (1) the tissue surface was aligned to the desired position relative to the beam focus using an aiming beam. This position determines the diameter of the annular beam at the tissue surface. (2) A 100-μm thick glass coverslip was positioned 5  mm above the tissue surface to collect the ejected microbiopsy. (3) A custom shutter consisting of a foil paddle connected to a rotational stepper motor (17md302s-00, Anaheim Automation, Anaheim, California) was placed with the foil paddle initially directly below the focusing lens, preventing initial pulses emitted by the laser from reaching the tissue. (4) Pre- and post-cooling solenoid valves were directed to the tissue surface and glass coverslip, respectively. (5) The laser pulse energy was set. (6) A train of pulses was manually initiated from the laser. The first pulse emitted by the laser was detected by an InGaAs photodetector (2034, New Focus, San Jose, California) connected to a digital delay generator. This event was used to control the shutter and spray cooling timing. (7) Pre-cooling began 600 ms before microbiopsy collection for a duration of 400 ms. (8) The shutter rotated open to allow a single laser pulse to perform the microbiopsy before returning to the closed position to block the following pulse. (9) Post-cooling began immediately after microbiopsy collection for a duration of 200 ms.

This procedure was tested on ex vivo porcine skin and kidney tissues. Optimal pulse energy and tissue location in the beam (beam diameter) were experimentally determined for each tissue type. A fast video camera (Fastcam Mini UX100, Photron, Tokyo, Japan) was used to record the skin ablation process at 50,000 fps.

2.7.

Microbiopsy Volume Measurements

Microbiopsies were collected from ex vivo porcine skin at two distances from the focusing lens (separated by 0.5 mm) providing a larger and smaller beam with estimated inner annulus diameters of 530 and 420  μm, respectively. Beam diameters outside this range were either too large or small to reliably harvest microbiopsies. Pulse energies of 1 and 1.3 J incident on the tissue surface were used. Pulse energies <1  J were insufficient to reliably harvest microbiopsies from porcine skin. Five microbiopsies were collected for each combination of beam diameter and pulse energy. Harvested microbiopsies were imaged using a custom-built OCT system.36 The OCT system used a single mode optical fiber with a swept source laser (1310 nm, Axsun, Billerica, Massachusetts). The OCT system used a Mach–Zehnder fiber interferometer configuration collecting field of views up to 6.5  mm(X)×6.5  mm(Y)×3  mm (depth). A custom MATLAB program was used to detect the microbiopsy surface from recorded OCT data, display a 3D reconstruction of each microbiopsy, and calculate microbiopsy volumes.

2.8.

Virtual H&E

To generate H&E images, acridine orange (AO) was chosen as the nuclear stain with sulforhodamine 101 (SR101) as a counter stain as described by Cahill et al.17 The stain was prepared by mixing 8 mg AO (A6014, Sigma-Aldrich, St. Louis, Missouri) and 5 mg SR101 (HB0838, Hello Bio, Princeton, New Jersey) in 50 ml 1:1 water:ethanol. Tissues were stained for 2 min followed by a 20 s rinse in 1:1 water:ethanol. Tissues were then imaged with a confocal microscope (Ultima IV, Bruker, Billerica, Massachusetts) with a 20× objective. Two excitation wavelengths were used: 488 nm for AO excitation and 561 nm for SR101 excitation. Two emission channels were used: 525±25  nm for AO emission and 605±35  nm for SR101 emission. The two channels were false colored and merged as described by Gareau15 with the SR101 channel taking the place of the reflectance channel. The R and G values for the AO or H channel were increased to reduce blueness after feedback from a pathologist. The RGB values used for the H&E digital staining were H=[0.5,0.4,1] and E=[1,0.55,0.88]. Virtual H&E imaging was first done on bulk tissues as a control. Microbiopsies and ablation craters were then imaged and compared to bulk tissues to evaluate damage.

3.

Results

3.1.

Transmission Measurements

Transmission through the beam-shaping optics was measured to be 66% of pulsed radiant energy released from the fiber tip. Therefore, a pulse energy setting at the laser of 2 J results in a pulse energy at the tissue surface of 1.3 J while a pulse energy setting of 1.5 J results in a pulse energy at the tissue surface of 1 J. All values reported in this study are the Ho:YAG radiant pulse energy incident on the tissue surface.

3.2.

Simulated and Experimental Beam Profiles

Simulated and experimental beam profile measurements show an annular shape with good intensity uniformity around the circumference of the annulus [Fig. 5(b)]. The full width at half maximum (FWHM) width from outer to inner radius was larger for experimental vs. simulated beams [Fig. 5(c)]. The experimental FWHM outer beam diameter decreased from 874  μm (at position Zo=20  mm from the back side of focusing lens) to 795  μm (Z=20.25  mm) to 721  μm (Z=20.50  mm) to 643  μm (Z=20.75  mm) as the beam was measured in increments of 0.25 mm away from the focusing lens.

Fig. 5

(a) Beam profiles were measured by irradiating a thin layer of water and imaging with an infrared camera. (b) and (c) Simulated and measured beam profiles show a uniform annular shape that decreases in diameter with increased distance from the beam-shaping optics.

JBO_27_12_125001_f005.png

3.3.

Computational Thermal Damage

Computational results displaying biopsy volume geometry, post-ablation temperature, cooling profile, and resulting tissue damage are compared for biopsies with and without pre- and post-cooling are shown in Fig. 6. Computational results indicate that cooling does not significantly affect ablation depth. Results indicate that pre- and post-cooling works best in combination to preserve the largest volume of tissue. The model predicts that without post-cooling the entire biopsy would be thermally damaged at a level Ω>1 (Fig. 6, left two columns). With post-cooling alone a small fraction of tissue at the surface up to 55  μm into the tissue remains viable. With a combination of pre- and post-cooling a larger volume up to 140  μm into the tissue remains viable (Ω<1). The mean temperature in the simulated harvested microbiopsies over time, including before, during, and after the laser pulse is shown in Fig. 7. The time axis is shown on a log scale to display the orders of magnitude difference in cooling time for cases with and without post-cooling. From Figs. 6 and 7, we observe that pre-cooling limits the maximum temperature reached within the microbiopsy, while post-cooling with cryogen spray leads to much a much faster cooling time on the order of 50 ms. When used together, pre- and post-cooling minimize the extent of thermal damage.

Fig. 6

Temperature and thermal damage calculations for biopsies under different cooling schemes. Temperature profiles are shown immediately post ablation and after 50 ms of cooling time. Thermal damage profiles are shown as percentage of cell death.

JBO_27_12_125001_f006.png

Fig. 7

Average temperature within biopsy volume over time for each cooling scheme. The time axis is shown in log scale.

JBO_27_12_125001_f007.png

The model also indicates that laser biopsy thermal damage is primarily due to light scattered from the annular beam into the center causing increased absorption, heating, and damage of the tissue. Results for normal scattering (μs=6  cm1), half scattering (μs=3  cm1), and zero scattering are compared with all cases including pre- and post-cooling (Fig. 8). By reducing scattering strength, depth of viable tissue (Ω<1) increased from 140  μm for normal scattering to 260  μm for half scattering, and 620  μm for zero scattering, suggesting that optical clearing methods that reduce tissue scattering could reduce microbiopsy damage. Optical clearing methods that could be used in vivo include mechanical optical clearing devices41 or chemicals such as glycerol which provide reversable optical clearing.42

Fig. 8

Computed fluence (top row) and tissue damage percentage (bottom row) of microbiopsies collected with pre- and post-cooling with different tissue scattering properties. (a) Normal scattering, μs=6  cm1, (b) half scattering, μs=3  cm1, and (c) no scattering.

JBO_27_12_125001_f008.png

3.4.

Microbiopsy Harvest and Ablation Volume Measurements

A pulse energy of 1.3 J was required to provide consistent and repeatable biopsy harvest when cooling was used. Fast camera videos show the microbiopsies ejecting from the center of the annular ablation region and reaching the coverslip after 100 to 150  μs [Fig. 9(a)]. OCT images show that the dimensions of ablation craters and harvested biopsies are on the order of 200 to 500  μm [Figs. 9(b) and 9(d)]. Ejected microbiopsy volumes collected on the glass slides increased with increasing beam diameter and volumes ranged from 0.02 to 0.08  mm3 [Fig. 9(e)].

Fig. 9

Biopsy harvest results: (a) select frames recorded by a fast camera (50,000  frames/s) of microbiopsy harvest. (b) OCT image of a harvested microbiopsy. (c) OCT image of a bulk tissue crater at the microbiopsy site. (d) 3D reconstruction of a microbiopsy volume. E) Mean microbiopsy volumes for two pulse energies and two beam diameters (estimated inner annulus diameters of 420 and 530  μm for small and large, respectively).

JBO_27_12_125001_f009.png

3.5.

Virtual H&E Images

An example of the H&E virtual staining process for bulk porcine skin is shown in Fig. 10. AO fluorescence highlights the cell nuclei [Fig. 10(a)] while SR101 fluorescence highlights the cytosol and collagen [Fig. 10(b)]. The two channels were merged [Fig. 10(c)] and false colored to the purple and pink colors of H&E stains [Fig. 10(d)]. Example virtual H&E images for bulk porcine skin and kidney tissue, ablation craters, and harvested biopsies, are shown in Figs. 11 and 12, respectively.

Fig. 10

Virtual H&E imaging of bulk skin sample. (a) AO fluorescence; (b) SR101 fluorescence; (c) merged image; and (d) false colored virtual H&E image.

JBO_27_12_125001_f010.png

Fig. 11

Examples of skin virtual H&E images, including (a) undamaged bulk skin, (b) microbiopsy crater in bulk skin, and (c) and (d) representative microbiopsies. Scale bars=100  μm.

JBO_27_12_125001_f011.png

Fig. 12

Example kidney virtual H&E images, including (a) undamaged bulk kidney, (b) microbiopsy crater in bulk kidney, and (c) and (d) representative microbiopsies. Scale bars=100  μm.

JBO_27_12_125001_f012.png

The virtual H&E image of the skin ablation crater [Fig. 11(b)] suggests that the entire epidermis and a portion of the dermis was removed. The crater shows increased signal in the SR101 channel at the left and right edges of the crater in the epidermis, indicating thermal damage. No clear changes were observed in the image along the base of the crater in the dermis. While our model did not focus on predicting or preventing thermal damage to the bulk tissue near the microbiopsy crater, one might expect to see significant thermal damage along the full length of the crater due to high predicted temperature increase. The absence of observable damage along the base of the crater indicates less light scattering than predicted into the center of the annulus and/or limited thermal exposure due to pre- and post-cooling. Some skin microbiopsies appear to be entirely epidermis [Fig. 11(c)], while others appear to contain dermis and epidermis [Fig. 11(d)]. The biopsy edge in the epidermis shows increased signal in the SR101 channel [Fig. 11(c)].

The virtual H&E image of the kidney crater [Fig. 12(b)] shows increased signal in the SR101 channel along the full length of the ablation crater. Due to the weak mechanical properties of kidney, many smaller tissue fragments were ejected in addition to the larger microbiopsy. Some kidney microbiopsies largely retained their shape [Fig. 12(c)], while others were fragmented [Fig. 12(d)]. Portions of the biopsies appeared to have an increase in signal in the SR101 channel such as at the edges [Fig. 12(c)].

4.

Discussion

Results of this study show that laser microbiopsy combined with virtual H&E imaging has potential to address several of the primary limitations of the traditional histopathology workflow. In contrast to traditional invasive biopsy tools which harvest tissues millimeters in size, laser microbiopsy harvests tissues that are several hundred microns in diameter with volumes <0.1  mm3. We expect that this approach may increase the number of biopsies that can be tolerated by a patient in a single visit. One pathology where laser microbiopsy may provide value is for early detection of skin cancer. Clinicians are sometimes challenged to determine if a skin lesion is cancerous without biopsy. Twenty biopsies are taken per melanoma diagnosis and five biopsies per non-melanoma skin cancer diagnosis.43 Laser microbiopsy might be used, for example, as a first pass screening method to determine if a full-scale biopsy is needed. This could increase the number of lesions that could be tested and reduce the number of larger volume invasive biopsies required. The small scale of laser microbiopsy may also provide value intraoperatively to sample tissue near delicate tissue structures and preserve healthy tissue. Combination with a rapid analysis technique would provide feedback to the surgeon to aid in complete tumor resection.

In traditional invasive biopsy only a few 3- to 10-μm sections are analyzed out of a biopsy several millimeters in diameter. This equates to analyzing <1% of the harvested tissue volume. In contrast, the small scale of microbiopsy allows rapid imaging and viewing of the entire tissue volume.

Traditional tissue processing, sectioning, and staining require more than 20 min to days to complete and require a highly trained technician. In contrast, virtual H&E staining methods are very simple and require only 2  min to complete. No tissue pre-processing or sectioning is required. The tissue is placed into a single solution containing two stains for 2 min followed by a 20 s rinse step. Little time, expense, or training is required.

Microbiopsies were successfully harvested from both skin, which has a relatively high ultimate tensile strength and is considered a strong tissue,44 and kidney which has a relatively low ultimate tensile strength and is considered a mechanically weak tissue. Therefore, we expect laser microbiopsy to be achievable from various tissues with a range of mechanical properties.

The results of this study show compatibility of laser microbiopsy with virtual H&E imaging using confocal microscopy. Confocal microscopy provides similar H&E image quality to multiphoton microscopy at a much lower cost.45 Histological features including cell nuclei and collogen were observed in virtual H&E stained microbiopsies. A combination of some extent of thermal and mechanical damage was observed in harvested microbiopsies and bulk tissue harvest sites. Future work will continue to optimize the approach to further reduce this damage. An increase in SR101 signal was observed in areas of suspected thermal damage in the skin epidermis and kidney, suggesting that increase in SR101 accumulation is a good indicator of thermal damage in these structures. This increase in SR101 signal was not observed in the dermis. The observation suggests that either collagen in the epidermis was not thermally damaged, or that thermal damage occurred and did not lead to an observable change in SR101 accumulation.

While this study focused on virtual H&E as a diagnostic method, laser microbiopsy could potentially be used with a range of analytical techniques. For example, laser microbiopsy could be used with genetic or protein testing methods for analysis of cancer biomarkers. For example, laser microbiopsy may be valuable in applications such as determining BRAF mutation status for melanoma.46 DNA and RNA are expected to remain stable even at high temperatures for a short time.47,48 Therefore, we expect DNA/RNA testing to be compatible even with thermally damaged microbiopsies possibly including those harvested without spray cooling.

Harvested microbiopsies showed signs of mechanical damage in addition to the minor thermal injury. The ablation process that causes microbiopsy is explosive causing rapid ejection of the microbiopsy. The microbiopsy is ejected from the bulk tissue 60  μs after the beginning of the laser pulse and reaches the coverslip 80  μs after ejection. The 1 J short pulse setting used for these experiments has a mean full width half max pulse duration of 71  μs with a tail in the temporal profile that extends to 250  μs. Temporal profile measurements show that 56% of the pulse energy is delivered in the first 60  μs on average. Therefore, up to 44% of the pulse energy may be delivered after tissue ejection leading to unnecessary tissue heating. The microbiopsies can spin and fold as they travel in air to the coverslip. The collision with the coverslip can cause further mechanical damage. Therefore, future work will include developing a more controlled microbiopsy harvest to limit mechanical damage.

The current optical design of the laser microbiopsy device utilizes many free space optics, e.g., lenses that are 25 mm in diameter, which limits microbiopsy use to applications where sufficient space is available, such as for the skin or during open surgery. We aim to develop a miniaturized device with fiber optic delivery to allow implementation with endoscopic systems.

The Ho:YAG laser source used in this study is a commercial laser system that is cleared by the FDA for clinical use for indications including biopsy. The source is a class 4 laser that has potential to cause damage to the eye or skin. This risk is mitigated through precautions including the use of laser safety glasses during experimental procedures. Therefore, we expect laser microbiopsy can be implemented safely in a clinical setting when proper safety precautions are followed.

5.

Conclusions

Laser microbiopsy with virtual H&E imaging shows promise as a potential rapid-minimally invasive tool for biopsy and intraoperative diagnosis. Microbiopsies were successfully harvested from both mechanically strong and weak tissues. A combination of pre- and post-cooling was required to preserve tissue integrity. Histological features including cell nuclei and collagen were observed in harvested microbiopsies.

Disclosures

The authors have no relevant financial interests or conflicts of interest to disclose.

Acknowledgments

This work was partially funded by a Texas 4000 cancer research grant to J.W.T and S.H.P. J.B.K. received funding from the National Institutes of Health (Grant No. T32EB007507).

Code, Data, and Materials Availability

Code is available upon reasonable request.

References

1. 

D. A. Novis and R. J. Zarbo, “Interinstitutional comparison of frozen section turnaround time. A College of American Pathologists Q-Probes study of 32868 frozen sections in 700 hospitals,” Arch. Pathol. Lab. Med., 121 (6), 559 –567 APLMAS 0003-9985 (1997). Google Scholar

2. 

“Histopathology is ripe for automation,” Nat. Biomed. Eng., 1 925 https://doi.org/10.1038/s41551-017-0179-5 (2017). Google Scholar

3. 

J. Zhang et al., “Nondestructive tissue analysis for ex vivo and in vivo cancer diagnosis using a handheld mass spectrometry system,” Sci. Transl. Med., 9 (406), eaan3968 https://doi.org/10.1126/scitranslmed.aan3968 STMCBQ 1946-6234 (2017). Google Scholar

4. 

M. Sharma et al., “Design and characterization of a novel multimodal fiber-optic probe and spectroscopy system for skin cancer applications,” Rev. Sci. Instrum., 85 (8), 1 –10 https://doi.org/10.1063/1.4890199 RSINAK 0034-6748 (2014). Google Scholar

5. 

P. Gerami et al., “Development of a novel noninvasive adhesive patch test for the evaluation of pigmented lesions of the skin,” J. Am. Acad. Dermatol., 71 (2), 237 –244 https://doi.org/10.1016/j.jaad.2014.04.042 JAADDB 0190-9622 (2014). Google Scholar

6. 

S. Park et al., “A novel microactuator for microbiopsy in capsular endoscopes,” J. Micromech. Microeng., 18 (2), 025032 https://doi.org/10.1088/0960-1317/18/2/025032 (2008). Google Scholar

7. 

S. Byun et al., “Barbed micro-spikes for micro-scale biopsy,” J. Micromech. Microeng., 15 (6), 1279 –1284 https://doi.org/10.1088/0960-1317/15/6/020 (2005). Google Scholar

8. 

E. Gultepe et al., “Biopsy with thermally-responsive untethered microtools,” Adv. Mater., 25 (4), 514 –519 https://doi.org/10.1002/adma.201203348 ADVMEW 0935-9648 (2013). Google Scholar

9. 

L. L. Lin et al., “Microbiopsy engineered for minimally invasive and suture-free sub-millimetre skin sampling,” F1000Research, 2 120 https://doi.org/10.12688/f1000research.2-120.v2 (2013). Google Scholar

10. 

K. Schütze and G. Lahr, “Identification of expressed genes by laser-mediated manipulation of single cells,” Nat. Biotechnol., 16 (8), 737 –742 https://doi.org/10.1038/nbt0898-737 (1998). Google Scholar

11. 

S. Thalhammer et al., “Laser microtools in cell biology and molecular medicine,” Laser Phys., 13 (5), 681 –691 (2003). Google Scholar

12. 

D. Elvers et al., “Laser microdissection of biological tissues: Process optimization,” Appl. Phys. A Mater. Sci. Process., 80 (1), 55 –59 https://doi.org/10.1007/s00339-004-3017-z (2005). Google Scholar

13. 

A. Vogel et al., “Mechanisms of laser-induced dissection and transport of histologic specimens,” Biophys. J., 93 (12), 4481 –4500 https://doi.org/10.1529/biophysj.106.102277 (2007). Google Scholar

14. 

V. Horneffer, N. Linz and A. Vogel, “Principles of laser-induced separation and transport of living cells,” J. Biomed. Opt., 12 (5), 054016 https://doi.org/10.1117/1.2799194 (2007). Google Scholar

15. 

D. S. Gareau, “Feasibility of digitally stained multimodal confocal mosaics to simulate histopathology,” J. Biomed. Opt., 14 (3), 034050 https://doi.org/10.1117/1.3149853 JBOPFO 1083-3668 (2009). Google Scholar

16. 

D. Gareau et al., “Tri-modal confocal mosaics detect residual invasive squamous cell carcinoma in Mohs surgical excisions Tri-modal confocal mosaics detect residual invasive squamous cell carcinoma in Mohs surgical excisions,” J. Biomed. Opt., 17 (6), 066018 https://doi.org/10.1117/1.JBO.17.6.066018 JBOPFO 1083-3668 (2012). Google Scholar

17. 

L. C. Cahill et al., “Rapid virtual hematoxylin and eosin histology of breast tissue specimens using a compact fluorescence nonlinear microscope,” Lab. Investig., 98 (1), 150 –160 https://doi.org/10.1038/labinvest.2017.116 LAINAW 0023-6837 (2018). Google Scholar

18. 

J. P. Kolb et al., “Virtual H & E histology by fiber-based picosecond two-photon microscopy,” Proc. SPIE, 10882 108822F PSISDG 0277-786X (2019). Google Scholar

19. 

M. G. Giacomelli et al., “Virtual hematoxylin and eosin transillumination microscopy using epi-fluorescence imaging,” PLoS One, 11 (8), e0159337 https://doi.org/10.1371/journal.pone.0159337 POLNCL 1932-6203 (2016). Google Scholar

20. 

A. K. Glaser et al., “Light-sheet microscopy for slide-free non-destructive pathology of large clinical specimens,” Nat. Biomed. Eng., 1 (7), 0084 https://doi.org/10.1038/s41551-017-0084 (2017). Google Scholar

21. 

F. Fereidouni et al., “Microscopy with ultraviolet surface excitation for rapid slide-free histology,” Nat. Biomed. Eng., 1 (12), 957 –966 https://doi.org/10.1038/s41551-017-0165-y (2017). Google Scholar

22. 

K. N. Elfer et al., “DRAQ5 and eosin (‘D&E’) as an analog to hematoxylin and eosin for rapid fluorescence histology of fresh tissues,” PLoS One, 11 (10), e0165530 https://doi.org/10.1371/journal.pone.0165530 POLNCL 1932-6203 (2016). Google Scholar

23. 

S.-E. Lin et al., “Rapid pseudo-H&E imaging using a fluorescence-inbuilt optical coherence microscopic imaging system,” Biomed. Opt. Express, 12 (8), 5139 –5158 https://doi.org/10.1364/BOE.431586 BOEICL 2156-7085 (2021). Google Scholar

24. 

D. A. Orringer et al., “Rapid intraoperative histology of unprocessed surgical specimens via fibre-laser-based stimulated Raman scattering microscopy,” Nat. Biomed. Eng., 1 (2), 1 –25 https://doi.org/10.1038/s41551-016-0027 (2017). Google Scholar

25. 

T. C. Hollon et al., “Near real-time intraoperative brain tumor diagnosis using stimulated Raman histology and deep neural networks,” Nat. Med., 26 52 –58 https://doi.org/10.1093/neuros/nyz310_634 (2020). Google Scholar

26. 

J. B. King et al., “Characterization of laser microbiopsy for sub-microliter tissue sampling,” Proc. SPIE, 11631 1163111 https://doi.org/10.1117/12.2577258 PSISDG 0277-786X (2021). Google Scholar

27. 

J. B. King et al., “Laser microbiopsy with virtual H&E imaging for rapid minimally invasive diagnosis,” Proc. SPIE, 11949 1194905 https://doi.org/10.1117/12.2608931 PSISDG 0277-786X (2022). Google Scholar

28. 

W. P. Latham, “Characteristic analysis of a refractive axicon system for optical trepanning,” Opt. Eng., 45 (9), 094302 https://doi.org/10.1117/1.2353119 (2006). Google Scholar

29. 

A. Vogel and V. Venugopalan, “Mechanisms of pulsed laser ablation of biological tissues,” Chem. Rev., 103 (2), 577 –644 https://doi.org/10.1021/cr010379n CHREAY 0009-2665 (2003). Google Scholar

30. 

L. G. Henyey and J. L. Greenstein, “Diffuse radiation in the galaxy,” Astrophys. J., 93 70 –83 https://doi.org/10.1086/144246 ASJOAB 0004-637X (1941). Google Scholar

31. 

T. Dai et al., “Comparative study of cryogen spray cooling with R-134a and R-404a: implications for laser treatment of dark human skin,” J. Biomed. Opt., 11 (4), 041116 https://doi.org/10.1117/1.2338001 JBOPFO 1083-3668 (2006). Google Scholar

32. 

F. Xu, K. A. Seffen and T. J. Lu, “Temperature-dependent mechanical behaviors of skin tissue,” Int. J. Comput. Sci., 35 (1), 92 –101 https://doi.org/10.1002/9780470382776.ch1 (2008). Google Scholar

33. 

J. B. King et al., “Mechanisms of pulse modulated holmium:YAG lithotripsy,” J. Endourol., 35 (S3), S29 –S36 https://doi.org/10.1089/end.2021.0742 (2021). Google Scholar

34. 

A. J. Welch et al., “Laser thermal ablation,” Photochem. Photobiol., 53 (6), 815 –823 https://doi.org/10.1111/j.1751-1097.1991.tb09896.x PHCBAP 0031-8655 (1991). Google Scholar

35. 

J. Pearce, “Mathematical models of laser-induced tissue thermal damage,” Int. J. Hyperth., 27 (8), 741 –750 https://doi.org/10.3109/02656736.2011.580822 IJHYEQ 0265-6736 (2011). Google Scholar

36. 

N. Katta et al., “Optical coherence tomography image-guided smart laser knife for surgery,” Lasers Surg. Med., 50 202 –212 https://doi.org/10.1002/lsm.22705 LSMEDI 0196-8092 (2018). Google Scholar

37. 

B. Anvari et al., “Selective cooling of biological tissues: application for thermally mediated therapeutic procedures,” Phys. Med. Biol., 40 (2), 241 –252 https://doi.org/10.1088/0031-9155/40/2/003 PHMBA7 0031-9155 (1995). Google Scholar

38. 

B. Chen et al., “Theoretical study of cryogen spray cooling with R134a, R404A and R1234yf: comparison and clinical potential application,” Appl. Therm. Eng., 148 (Aug. 2018), 1058 –1067 https://doi.org/10.1016/j.applthermaleng.2018.11.117 ATENFT 1359-4311 (2019). Google Scholar

39. 

H. Xin et al., “Histopathological evaluation of the R134a multipulsed spray cooling assisted 1210 nm laser lipolysis by the murine model in vivo,” Lasers Surg. Med., 1 –11 https://doi.org/10.1002/lsm.23607 LSMEDI 0196-8092 (2022). Google Scholar

40. 

S. L. Jacques, “Optical properties of biological tissues: a review,” Phys. Med. Biol., 58 (14), 5007 –5008 https://doi.org/10.1088/0031-9155/58/14/5007 PHMBA7 0031-9155 (2013). Google Scholar

41. 

C. G. Rylander et al., “Mechanical tissue optical clearing devices: enhancement of light penetration in ex vivo porcine skin and adipose tissue,” Lasers Surg. Med., 40 (10), 688 –694 https://doi.org/10.1002/lsm.20718 LSMEDI 0196-8092 (2008). Google Scholar

42. 

X. Weny et al., “In vivo skin optical clearing by glycerol solutions: mechanism,” J. Biophotonics, 3 (1–2), 44 –52 https://doi.org/10.1002/jbio.200910080 (2010). Google Scholar

43. 

A. Privalle et al., “Number of skin biopsies needed per malignancy: comparing the use of skin biopsies among dermatologists and nondermatologist clinicians,” J. Am. Acad. Dermatol., 82 (1), 110 –116 https://doi.org/10.1016/j.jaad.2019.08.012 JAADDB 0190-9622 (2020). Google Scholar

44. 

J. T. Walsh, T. F. DeutschJ. T. Walsh, “Pulsed CO2 laser ablation of tissue: effect of mechanical properties,” IEEE Trans. Biomed. Eng., 36 (12), 1195 –1201 https://doi.org/10.1109/10.42114 IEBEAX 0018-9294 (1989). Google Scholar

45. 

T. Yoshitake et al., “Direct comparison between confocal and multiphoton microscopy for rapid histopathological evaluation of unfixed human breast tissue,” J. Biomed. Opt., 21 (12), 126021 https://doi.org/10.1117/1.JBO.21.12.126021 JBOPFO 1083-3668 (2016). Google Scholar

46. 

L. Cheng et al., “Molecular testing for BRAF mutations to inform melanoma treatment decisions: a move toward precision medicine,” Mod. Pathol., 31 (1), 24 –38 https://doi.org/10.1038/modpathol.2017.104 MODPEO 0893-3952 (2018). Google Scholar

47. 

M. Karni et al., “Thermal degradation of DNA,” DNA Cell Biol., 32 (6), 298 –301 https://doi.org/10.1089/dna.2013.2056 (2013). Google Scholar

48. 

A. L. Fabre et al., “An efficient method for long-term room temperature storage of RNA,” Eur. J. Hum. Genet., 22 (3), 379 –385 https://doi.org/10.1038/ejhg.2013.145 EJHGEU 1018-4813 (2014). Google Scholar

Biography

Jason B. King received his BS degree in biomedical engineering from the University of Utah in 2017, and his MS and PhD degrees in biomedical engineering from the University of Texas at Austin in 2019 and 2022, respectively. His research interests include laser ablation of hard and soft tissues, and disease diagnosis using optical imaging and spectroscopy.

Nitesh Katta received his PhD in 2019 from the University of Texas at Austin and is currently a postdoctoral fellow at the Beckman Laser Institute (BLI). At BLI, Nitesh continues to work on conducting basic science and translational research in novel laser energy-based devices and image-guided therapy applications in dermatology and cardiology, among other light–tissue interactions.

Sapun H. Parekh is an assistant professor in the Department of Biomedical Engineering at UT Austin. His lab specializes in molecular microscopy and development of nonlinear fluorescence and vibrational imaging to study disease pathophysiology and molecular biophysics. He is Scialog Fellow for Advanced Bioimaging.

Thomas. E. Milner is the Director of the Beckman Laser Institute and Medical Clinic and a professor in the Departments of Surgery and Biomedical Engineering. He has authored 200 peer-reviewed articles and twelve book chapters. He is a member of the National Academy of Inventors, American Institute for Medical and Biological Engineering, and the American Society for Lasers in Medicine and Surgery. He is an inventor of 70 U.S. patents that have been licensed to six companies.

James W. Tunnell is an associate professor in the Department of Biomedical Engineering at the University of Texas at Austin. He is a fellow of the American Institute of Medical and Biological Engineers (AIMBE) and a member of SPIE. He has served in various leadership roles in SPIE, CLEO, OSA, ASLMS, and BMES. He has a specific research focus on using optical spectroscopy and imaging for disease diagnosis and treatment.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Jason B. King, Nitesh Katta, Sapun H. Parekh, Thomas E. Milner, and James W. Tunnell "Tissue harvest with a laser microbiopsy," Journal of Biomedical Optics 27(12), 125001 (14 December 2022). https://doi.org/10.1117/1.JBO.27.12.125001