19 March 2024 Video-rate endocavity photoacoustic/harmonic ultrasound imaging with miniaturized light delivery
Donghyeon Oh, Hyunhee Kim, Minsik Sung, Chulhong Kim
Author Affiliations +
Abstract

Significance

Endocavity ultrasound (US) imaging is a frequently employed diagnostic technique in gynecology and urology for the assessment of male and female genital diseases that present challenges for conventional transabdominal imaging. The integration of photoacoustic (PA) imaging with clinical US imaging has displayed promising outcomes in clinical research. Nonetheless, its application has been constrained due to size limitations, restricting it to spatially confined locations such as vaginal or rectal canals.

Aim

This study presents the development of a video-rate (20 Hz) endocavity PA/harmonic US imaging (EPAUSI) system.

Approach

The approach incorporates a commercially available endocavity US probe with a miniaturized laser delivery unit, comprised of a single large-core fiber and a line beamshaping engineered diffuser. The system facilitates real-time image display and subsequent processing, including angular energy density correction and spectral unmixing, in offline mode.

Results

The spatial resolutions of the concurrently acquired PA and harmonic US images were measured at 318 μm and 291 μm in the radial direction, respectively, and 1.22 deg and 1.50 deg in the angular direction, respectively. Furthermore, the system demonstrated its capability in multispectral PA imaging by successfully distinguishing two clinical dyes in a tissue-mimicking phantom. Its rapid temporal resolution enabled the capture of kinetic dye perfusion into an ex vivo porcine ovary through the depth of porcine uterine tissue. EPAUSI proved its clinical viability by detecting pulsating hemodynamics in the male rat’s prostate in vivo and accurately classifying human blood vessels into arteries and veins based on sO2 measurements.

Conclusions

Our proposed EPAUSI system holds the potential to unveil previously overlooked indicators of vascular alterations in genital cancers or endometriosis, addressing pressing requirements in the fields of gynecology and urology.

1.

Introduction

Genital cancer is a serious global health concern, comprising 14.8% of all reported global cancer cases in 2020.1 Specifically, prostate cancer stands as the most prevalent malignancy among the male population, with an estimated 1,414,259 cases in 2020, leading to 375,304 fatalities.1 Simultaneously, ovarian cancer ranks as the seventh most prevalent cancer and the fifth leading cause of death among women worldwide, accounting for 314,000 new cases and 207,000 deaths in 2020.1 Transrectal or transvaginal (together noted as “endocavity”) ultrasound (US) imaging is the most representative diagnostic imaging technique for visualizing deeply situated reproductive organs. By accessing these organs through the rectal or vaginal canal, endocavity US enables close examination, revealing clinical symptoms that may remain undetectable using standard transabdominal US imaging.2,3 This approach is often employed in the initial evaluation of malignant lesions in the ovaries or prostate, including tumor masses and cysts.4,5 Additionally, it aids in guiding precise needle placement during needle aspiration biopsy, a confirmatory step in cancer diagnosis.68 For its specialized application, the endocavity US probe is characterized by a long neck for insertion and features a microconvex transducer array at its distal end, providing a wide-angle fan-shaped view.

Although endocavity US is a valuable imaging tool for assessing genital abnormalities, it is not typically employed as the sole diagnostic test owing to its relatively modest diagnostic accuracy. Transrectal US exhibits a sensitivity of 53.3% and a specificity of 75% for prostate cancer,9 whereas transvaginal US demonstrates a sensitivity of 88% and a specificity of 79% for deep endometriosis.10 Thus, definitive diagnoses usually rely on a combination of diagnostic methods, encompassing physical examinations (vaginal/rectal palpation), blood tests, advanced medical imaging (e.g., computed tomography or magnetic resonance imaging), and when necessary, histological analysis following surgical removal.

Moreover, the integration of photoacoustic (PA) imaging with clinical US imaging is actively progressing in clinical studies.1120 This integration enhances the ability to capture subtle pathological signs and improves diagnostic accuracy.2123 As a hybrid optical and acoustical imaging modality, PA imaging provides unique photochemical contrast capabilities at previously uncharted depths within biological tissue,2429 surpassing the capabilities of other optical imaging methods.3037 It mirrors the US imaging process3842 but utilizes pulsed laser excitation to induce the PA effect in tissue chromophores (e.g., hemoglobin, melanin, lipid, and collagen).4346 The resultant thermal expansion generates an acoustic signal,4749 which can be reconstructed into cross-sectional distributions of chromophores akin to US B-mode images.5052 The technical similarities between PA and US imaging enable simultaneous acquisition of both types of images, with PA images offering valuable physiological information about the lesion’s photochemical characteristics, such as total hemoglobin and oxygen saturation.5257 This ability to detect changes in oxygen saturation proves beneficial in diagnosing urological and gynecological conditions, identifying peritumoral hypoxia in genital cancer, and detecting menstrual cycle-related lesions such as endometrioma or uterine fibroids.

The main challenge in integrating PA imaging with endocavity US imaging is the size of the imaging probe, given the necessity of probe insertion into confined application sites. Most current PA imaging probes are primarily designed for on-skin use with direct cutaneous contact, overlooking size considerations. Of all components, the optical delivery module occupies the largest volume, necessitating miniaturization for endocavity use. Most PA imaging probes employ fiber bundles for high-energy laser pulse delivery, which are bulky and result in low optical coupling efficiency due to the dead spaces created when arranging thin fibers consecutively.11,12,58,59 This requirement for customized probe fabrication for embedding impacts the compatibility with clinical US imaging systems.6064 Although some optical fiber-based endocavity PA imaging studies have been reported,6568 they require customized fiber modifications or arrangements of multiple optical elements, which are less practical in clinical settings. Additionally, most previous probe designs adopt a bright-field illumination scheme, which necessitates a void space (referred to as “stand-off”) to position the optical beam in the US imaging plane. This configuration hinders the effective field of view (FOV) and weakens the acoustic signal’s strength. Direct laser exposure to the probe surface also results in notable photo-induced artifacts from the transducer elements, potentially obscuring PA contrasts originating from tissue chromophores.

In this study, we present an endocavity PA and harmonic US imaging (EPAUSI) system equipped with a miniaturized light delivery module composed of a single large-core optical fiber and line beamshaping engineered diffuser. By integrating this system with a commercially available endocavity US probe, we achieve a probe thickness of 25 mm. Our system, combined with a portable optical parametric oscillator (OPO) laser, provides real-time display of both PA and harmonic US at a video rate of 20 Hz, with a spatial resolution benchmarked at 318/291  μm in the radial direction and 1.22/1.50° in the angular direction. Multispectral PA imaging further enables spectral multiplexing of various chromophores as demonstrated through clinical dye separation in vitro. The system’s video-rate temporal resolution permits the observation of real-time dynamic changes in PA contrasts in clinically relevant scenarios such as clinical dye perfusion and pulsatile hemodynamic motions, as evidenced in experiments involving an ex vivo porcine ovary and an in vivo rat prostate. Finally, functional EPAUSI effectively measures oxygen saturation (sO2) levels in the human arteries and veins of a healthy volunteer, highlighting its capability to detect pathological hypoxia in reproductive organs.

2.

Materials and Methods

2.1.

Endocavity Photoacoustic and Harmonic Ultrasound Imaging System Configuration

Our EPAUSI system comprises a clinical endocavity probe for light delivery [as depicted in Fig. 1(a)], which is connected to a combination of a research US system (Vantage 256, Verasonics, Washington, United States) and a cart-mounted tunable pulse laser (PhotoSonus, Ekspla, Lithuania) [as shown in Fig. 1(b)]. For both PA and US acquisitions, we employed a commercially available endocavity US probe (6EIX, Humanscan, Republic of Korea) with a center frequency of 6 MHz and a 95% bandwidth. This probe features a microconvex transducer array positioned along the periphery of a 10 mm-radius circular edge, offering a wide-angle 160 deg fan-shaped FOV beneath the application site. In light of the practical scenario involving direct contact with a canal wall, we implemented a dark-illumination scheme, ensuring that the transducer surface is not directly exposed to a laser beam reflected from the tissue surface. This dark-field illumination scheme presents several practical advantages. First, it eliminates the need for a transparent standoff layer filled with water or gel to clear the light path, resulting in the efficient reduction of the overall probe size by economizing the void volume. Moreover, the removal of the standoff layer helps maintain the signal intensity by reducing subsidiary attenuation and preserving the original FOV. Finally, this scheme prevents the direct exposure of the laser beam to the probe elements, mitigating strong photoinduced artifacts generated from the probe lens surface, which often interfere with PA contrast from microvessels at shallow depths. To further minimize PA artifacts arising from the transducer surface, we replaced the probe’s acoustic lens with a white one. To preserve the identical physical properties and geometrical form factors of the transducer, the replaced lens was crafted using the same elastomer and lens mold as employed in the original lens fabrication process, with the only variation being in color through adjustments in the pigment ratio.

Fig. 1

EPAUSI system configuration: (a) photograph and 3D rendering of EPAUSI probe assembled with optical casing. Optical and acoustic paths are depicted in red and blue, respectively. (b) System configuration of EPAUSI system on a cart. (c) Magnified rendering of EPAUSI probe head describing optical assemblies for line-beam shaping. (d) Photograph of a laser beam trajectory on the x-z and y-z planes imaged under a scattering suspension medium. PC, personal computer; OPO, optical parametric oscillator; DAQ, data acquisition unit; ED, engineered diffuser; LCF, large core optical fiber; ECP, endocavity ultrasound probe; and PC, probe casing.

JBO_29_S1_S11528_f001.png

For laser transport, we employed a large-core optical fiber, significantly reducing the size of the laser delivery module while achieving superior optical coupling efficiency (approximately 90%), in contrast to using a fiber bundle (approximately 50%). Specifically, a 1 mm core size multimodal fiber (Optibase, South Korea), equipped with air-gapped SMA-905 fiber connectors and bare FC/PC ferrules on each end, was coupled to a customized fiber coupling system within the portable OPO laser system (operating at a wavelength of 660 to 1064 nm, a pulse width of 3 to 5 ns, and a pulse repetition rate of 20 Hz). This fiber delivered the laser beam to the distal end of the endocavity probe, and the beam was directed into a line beam-shaping engineered diffuser (EDL-150, VIAVI Solutions, Arizona, United States) installed in front of the fiber outlet [as depicted in Fig. 1(c)]. The initially circular-shaped beam diverged into a single direction as it passed through the engineered diffuser, eventually transforming its pattern into a line shape [Fig. 1(d)]. These optical components were integrated within a biodegradable 3D printed casing, which was attached to the side of the probe head. To maximize the PA signal, the fiber channel within the casing was slightly tilted toward the imaging plane. Consequently, the total thickness of the probe head was 25 mm.

2.2.

Real-time, Video-rate PA and Harmonic US Image Processing

The signal acquisition, image reconstruction, and real-time display followed the event sequence timing outlined in Fig. 2. The acquisition of PA signals commenced with a laser trigger synchronized to a laser shot. This was swiftly followed by the acquisition of US signals, in which we implemented a pulse inversion harmonic US imaging technique to enhance the resolution and tissue contrast (Fig. S1 in the Supplementary Material). This method involves transmitting a pair of normal and inverted US pulses at each location, subsequently isolating the harmonic signal by canceling out the fundamental frequency signal. To match the probe’s bandwidth, we designed a pair of normal and inverted bipolar 4 MHz US transmit pulses, which ultimately resulted in an 8 MHz harmonic signal after the successful cancellation of the fundamental frequency signal by adding the radiofrequency (RF) signal pairs.

Fig. 2

The 20 Hz video-rate timing sequence of multispectral PA and harmonic US imaging. FWL, first wavelength laser trigger; PRF, pulse repetition frequency; DAQ, data acquisition unit; and RF, radiofrequency data.

JBO_29_S1_S11528_f002.png

For each US frame, 96 line-focused US beams swept across the entire FOV, with the transmit pairs being sequentially positioned to minimize tissue motion between the two acquisitions. In total, 1 PA receive event and 192 harmonic US receive events (comprising 96 pairs) were accommodated within the elapsed time between two laser emissions, amounting to 50 ms. Unlike US signal reception involving active pulse transmission, PA signal reception experiences greater attenuation in both optical and acoustical domains. Accordingly, an additional gain was allocated for the PA acquisition event, utilizing a programmable variable-gain preamplifier consisting of a time-gain compensator (TGC), programmable gain amplifier (PGA), and low noise amplifier (LNA) relay. Although the TGC was adjustable, it was maintained at the maximum value across all depths to ensure consistency and maximize the preservation of deep PA signals. Additionally, an extra 12 dB gain (PGA: 30 dB; LNA: 24 dB) was selectively applied in PA acquisition compared with harmonic US acquisition (PGA: 24 dB; LNA: 18 dB), with the preamplifier settings remaining constant throughout the experiments. The acquired PA and US RF data were transferred twice during a single frame acquisition to the host PC for online image reconstruction, utilizing a real-time Vantage reconstruction algorithm (Verasonics, WA, USA). Consequently, a pair of PA and harmonic US images was displayed on the screen at a real-time frame rate of 20 Hz. For data export and subsequent offline processing, such as energy density compensation and spectral unmixing (as detailed in Sec. 2.3), a conditional loop was activated. This loop involved an additional trigger event and a data save event. Operating in a multiwavelength mode, the laser system emitted an extra trigger of 8 microseconds in advance of the emission of the first wavelength of the laser pulse, allowing the US system to organize the data in the order of the laser pulse relay. After repeating the acquisition event as commanded by the user, the multi-frame PA RF, PA image, and harmonic US image data were saved, and the system returned to the real-time mode.

2.3.

Offline Angular Energy Density Correction and Spectral Unmixing

We observed an angular deviation in laser intensity between the center and the sides of the wide-angle FOV. Consequently, we conducted post-processing on the PA images to achieve uniform PA intensity along the angular direction. Utilizing the original PA image of a tissue-mimicking phantom (as described in Sec. 3.1), we estimated the angular energy density distribution by analyzing the PA intensity of radial-patterned string targets. Initially, the PA image was transformed from Cartesian coordinates to polar coordinates. Subsequently, point targets spanning between ±70  deg and 15 to 25 mm in depth, along with their neighboring pixels, were windowed [as illustrated in Fig. 3(a)]. We calculated to the average amplitudes of the top 10% of pixels from each window, sorted them by depth, and generated an angular distribution of PA amplitude, as shown in Fig. 3(b). We fitted Gaussian curves to the measured data points, resulting in a consistent standard deviation of approximately 60 at all depths. Based on this observation, we created an estimated angular energy density map that covered the entire FOV. This map was generated by transforming the rectangular Gaussian distribution block into Cartesian coordinates, as depicted in Fig. 3(c). Subsequently, all experimental PA images were homogenized using inverse compensatory weighting based on the estimated angular energy distribution map.

Fig. 3

Estimation of angular energy density distribution using a tissue-mimicking phantom: (a) PA imaging result of TiO2-gelatin phantom (45 dB). To build a fluence distribution map based on angles, the image was changed from Cartesian coordinates to polar coordinates. (b) PA amplitude distribution by angles. (c) Estimated angular fluence distribution map.

JBO_29_S1_S11528_f003.png

We conducted linear spectral unmixing on the multispectral PA images to distinguish between multiple contrasts originating from distinct chromophores, such as clinical dyes and oxy and deoxyhemoglobin (as detailed in Secs. 3.2 and 3.5). Multispectral PA amplitudes at a specific location represent a combination of PA contrasts from various chromophores with distinct spectral properties, which are expressed using the following equation:

Eq. (1)

P=[p(λ1)p(λ2)  p(λn)]=[μ1(λ1)μ2(λ1)μm(λ1)μ1(λ2)μ2(λ2)μm(λ2)μ1(λn)μ2(λn)μm(λn)][c1c2cm]=MC,
where P represents a single-column PA amplitude vector composed of p(λi); M is a [n×m] matrix representing the molecular extinction coefficients and is composed of μj(λi); and C is a single-column vector representing relative concentrations and is composed of cj, where i and j denote the arbitrary indices of wavelengths and chromophores, both ranging from 1 to n and 1 to m, respectively. Referring to their measured molecular optical extinction coefficient available online (Ref. 69; Oregon Medical Laser Center, Oregon, United States), we inversely calculated the local concentration of the chromophores by implementing a pseudo-inverse matrix scheme:18

Eq. (2)

C=(M+)P=(MTM)1MTP.

Considering the clinical scenario in which the probe is manually maneuvered by hand, we reduced the number of wavelengths (n) to maximize the spatial correlation among the PA images. To enhance the accuracy of sO2 measurements, we estimated an optical fluence for each original spectral PA image, and we performed preliminary compensation prior to its utilization as input for spectral unmixing. We divided the provided PA image into 1 cm depth increments, and within each step, we determined the mean value of the upper 50% of PA signals. These values were fitted with an exponential curve, assuming that optical fluence attenuation follows Beer–Lambert’s law. Distinct attenuation coefficients were obtained for each wavelength, and the respective PA images were inversely compensated by their corresponding exponential curves. Data post-processing was conducted using a custom script developed with MATLAB (MATLAB R2021a, MathWorks, Massachusetts, United States).

3.

Results and Discussion

3.1.

Spatial Resolution Benchmarking in TiO2-gelatin Phantom

The measurement of the image resolution and angular energy density distribution was conducted using an in vitro experiment with a tissue-mimicking phantom. To create the phantom mold, transparent acrylic boards were designed, and holes were evenly punched on both sides of the side face in a radial pattern, spaced at 10 deg intervals in the angular direction and 5 mm intervals in the depth direction. A 100  μm black nylon string was threaded through these holes to serve as the imaging target. For the medium, a gelatin solution was prepared by adding 16.7 w% gelatin to water, which was thoroughly dissolved through heating at 60°C. The medium’s optical turbidity was achieved by adding 1  g/L of TiO2 while constantly stirring.70 This prepared medium was subsequently poured into the threaded mold and allowed to cool to room temperature, solidifying into a gel phantom. To ensure complete contact with the cylindrical surface of the probe, a groove with a 12.5 mm radius was created by dipping a cylindrical mold during fixation. For imaging, the assembled endocavity probe was acoustically matched with the phantom using a droplet of water and securely fixed upright onto the groove, with adjustments guided by a real-time harmonic US view. PA images were acquired at 700 nm with an energy density of 20  mJ/cm2, adhering to the American National Standards Institute (ANSI) criteria for maximum permissible exposure (MPE) at NIR wavelengths. As a result of this experiment, the simultaneous acquisition of PA and harmonic US B-mode images with a strong spatial correlation of the point targets was achieved, as illustrated in Fig. 4(a) and Fig. S2 in the Supplementary Material. The PA contrast appeared to be evenly distributed across the full 160 deg fan-shaped FOV scan range, similar to the harmonic US images, which were reconstructed with geometric homogeneity. However, contrary to the US image contrast, which extended down to 50 mm, the PA contrast logarithmically decreased with depth and reached the noise-equivalent depth at approximately 35 mm due to optical attenuation in the phantom media.

Fig. 4

EPAUSI spatial resolution measured from a tissue-mimicking phantom. (a) overlaid PA and harmonic US images highlight the spatially correlated contrasts from points in a radial pattern. Quantification of (b) radial and (c) angular resolution from line profiles of five 20-mm depth point targets, marked with dashed white box in panel (a). Mean ± standard error.

JBO_29_S1_S11528_f004.png

In terms of spatial resolution, we observed that the line profile of the point targets broadened with increasing the depth. The distinctive shape of the array dimensions, being convex, led us to measure the radial and angular resolutions in polar coordinates, in contrast to the lateral and axial resolution measurements in Cartesian coordinates. We obtained line profiles of the five central point targets at a depth of 20 mm, considering the elevational focal length of the probe. The maximum amplitude projections in the radial and angular directions were normalized, and the peak widths were measured following the full-width–half-maximum criteria [as depicted in Figs. 4(b) and 4(c)]. The radial resolution, which was narrower in harmonic US (291±22  μm) compared with PA (318±16  μm), can be attributed to the inherent disparity in carrier frequency between PA and US signals. PA signals are modulated by the fundamental frequency of the transducer (6.25 MHz), whereas harmonic US signals are modulated at the second harmonic frequency of the transmit pulse (8 MHz). The angular resolution of PA (1.22±0.02deg and 639±11  μm in lateral) was finer than that of harmonic US (1.50±0.09  deg and 785±49  μm in lateral), as shown in Fig. 4(c). This observation can be explained by the greater limited view effect induced by the transducer’s geometry. Unlike the round trip travel of US signals, PA signals travel in one direction, and PA images are affected by the limited acceptance angle of the piezo elements. Additionally, the curved arrangement of piezoelements in microconvex probes results in a sparser acceptance angle overlay, contributing to a more pronounced limited view effect. Consequently, PA images may provide a better resolution in depicting microvessels, which can be particularly useful in identifying cancerous angiogenesis.

3.2.

Spectral Unmixing of Multiplexed Clinical Dyes In Vitro

To validate the molecular imaging capabilities of our EPAUSI system, we conducted a study in which we analyzed the constituent ratio of two Food and Drug Administration (FDA)-approved contrast dyes using multispectral PA imaging under in vitro conditions. Multispectral PA imaging, which leverages the distinctive optical spectral characteristics of chromophores in biological tissues, is a valuable technique for distinguishing the molecular-specific distribution of substances. This technique is widely employed in the current clinical translation of PA imaging. We prepared aqueous solutions of 65  μM methylene blue (MB) and 100  μM indocyanine green (ICG) and adjusted their concentrations to achieve absorbance levels comparable to that of hemoglobin in the NIR range, as documented on the Oregon Medical Laser Center’s spectral data repository.69 To mimic the quantification of oxygen saturation, we mixed the MB and ICG solutions in volume ratios of 3:1, 2:2, and 1:3 [Fig. 5(a)]. Using a 27G syringe, these mixtures were subsequently injected into silicone microtubing threaded into the TiO2-gelatin phantom, as described in Sec. 2.1, with the MB content decreasing from left to right. To minimize variations in optical attenuation, all tubings were placed at the same depth (25 mm). For multispectral PA imaging, we selected 20 wavelengths at 2 nm intervals, ranging from 660 nm to 718 nm, as this range exhibited the most significant spectral absorbance variations for the two dyes.

Fig. 5

Differentiation of multiplexed clinical dye contrast with multispectral EPAUSI: (a) a photograph of five MB – ICG cocktail samples arranged in order of ICG constituent ratio. (b) An overlaid PA and US image of the microtubing phantom filled with five samples. Scale bar = 10 mm. (c) PA spectroscopy of five samples in the NIR range. (d) A series of PA images corresponding to four wavelengths (660, 680, 700, and 718 nm) involved in spectral unmixing. Multiplexed mapping of (e) a normalized MB concentration and (f) a normalized ICG concentration. Scale bar = 5 mm. (g) Quantified MB and ICG concentration from each corresponding tubing location. (h) Measured ICG constituent ratio and ground truth.

JBO_29_S1_S11528_f005.png

In the resulting PA and harmonic US images obtained at each wavelength, the upper and lower boundaries of the tubing were clearly visible in the harmonic US images, whereas the PA contrast of the dye appeared as small dots in the center of the lumen in the PA images [see Fig. 5(b)]. As the wavelength changed from 660 to 720 nm, distinct spectral trends in the PA contrast were observed within the tubing samples containing MB and ICG. Specifically, MB exhibited a consistent diminishing pattern, whereas ICG displayed a consistent increase [see Fig. 5(c)]. For practicality, we performed spectral unmixing using only four out of the total 20 multispectral image ensembles. These four images were selected from the quartile range at wavelengths 660, 680, 700, and 718 nm [see Fig. 5(d)]. The resultant unmixed distribution images of MB and ICG demonstrated a consistent rise and decline, respectively, which strongly correlated with the actual fractional composition of each dye [see Figs. 5(e) and 5(f)]. We quantified the upper 10% pixel intensity within the designated region of interest (ROI) within each tube. This analysis allowed for clear discrimination between MB and ICG in the pure tubing samples, whereas an equal constituent ratio was observed in the 2:2 cocktail tubing [see Fig. 5(g)]. Furthermore, when we calculated the fractional ICG concentration from the unmixed MB and ICG images, a monotonic increase was highlighted; it closely correlating with the actual fractional ICG concentration [see Fig. 5(h)]. However, the values for B and D, which should exhibit a volumetric ratio of 25% and 75%, respectively, appeared to deviate from the most representative reference ratio among the measured values. This discrepancy may stem from variances between the reference molar absorption coefficient used in the pseudoinverse calculations and the actual absorption coefficients of ICG and MB.

3.3.

ICG Infusion Kinetics into Ex Vivo Porcine Ovary

To assess the feasibility of using EPAUSI for transvaginal applications, we conducted real-time imaging of the dynamic ICG infusion into ex vivo female porcine genital tissue. In gynecological conditions necessitating surgical intervention, such as genital cancer or endometriosis, ICG is frequently employed for intraoperative fluorescent staining of malignant lesions, such as tumor masses or sentinel lymph nodes. For this experiment, we utilized a freshly excised porcine uterus, which included the ovary, fallopian tube (FT), uterus, uterine cervix, and vaginal canal [refer to Fig. 6(a)]. To replicate the clinical scenario, we enveloped the lower and surrounding regions of the excised ovary with FT tissue and subsequently covered them with an unfolded vaginal tissue flap. The EPAUSI probe was positioned over the vaginal tissue, and PAUS imaging was performed to penetrate through these tissue layers. Following probe positioning, harmonic US images clearly visualized the ovary at a depth of 23 mm, along with the adjacent FT tissues and the upper vaginal tissue layer. Subsequently, a 27G syringe needle was guided into the center of the ovary using real-time US imaging. Both 700 nm PA and harmonic US images were acquired at a rate of 20 Hz for a duration of 4 s, starting simultaneously with the injection of 3 mL of a 200  μM ICG solution through the needle.

Fig. 6

Ex vivo imaging of a porcine ovary with ICG infusion: (a) a photograph of a porcine female genital. A red inset image shows a magnified appearance of a porcine ovary in the end of a long FT. Comparative overlaid PA and harmonic US images of a tissue-surrounded ovary (b) in the preinjection state and (c) after ICG injection highlights the PA contrast of infused ICG (Video 1). (d) ICG infusion dynamics under the conditions of increasing PA amplitude inside the ovary. (e) A photograph of an ovarian hemisection after ICG infusion with topical ICG stain inside the ovary. UT, uterus; UC, uterus cervix; and VC, vaginal canal. Scale bar = 10 mm.

JBO_29_S1_S11528_f006.png

Before the injection, the syringe needle piercing through the ovarian membrane and into the ovarian center was clearly visualized in both the 700 nm PA and US images [see Fig. 6(b)]. However, upon the initiation of ICG perfusion, the PA image exclusively displayed the formation of a strong PA contrast cloud at the distal tip of the needle. This PA contrast area continued to expand over a span of 2 s, whereas such an observation was not evident in the US image [see Fig. 6(c) and Video 1]. We extended our analysis to compare the temporal changes in PA and US amplitudes. This involved selecting an ROI along the ovarian boundary and calculating the upper 10% average of pixel amplitudes from both images. As a result, the PA signal exhibited a remarkable increase of up to 10-fold within 2 s from the point of injection, compared with the pre-injection levels [see Fig. 6(d)]. After acquiring the images, the ovary used in the experiment was dissected into halves. The localized ICG stain was distinctly observable in a location corresponding to where the ICG cloud was evident in the PA image [see Fig. 6(e)].

3.4.

In Vivo Monitoring of Rat Prostate Vessel Pulsation with EPAUSI

We evaluated the in vivo imaging capabilities of EPAUSI by imaging the hemodynamics in the male reproductive organ of a rat. All experimental procedures involving animals were conducted in compliance with the protocol approved by the Institutional Animal Care and Use Committee of Pohang University of Science and Technology (Protocol POSTECH-2021-0052-C2, approved on February 17, 2022). A healthy 12-week-old male Sprague Dawley rat with a body weight of 347 g was initially anesthetized in an induction chamber with 3% to 5% isoflurane/O2 at a flow rate of 1.5  L/min, using a preclinical anesthetic gas system (VIP 3000 Veterinary Vaporizer, Midmark, Ohio, United States). The rat was supplied with a continuous stream of anesthetic gas through a customized nasal cap. About 75% of its abdominal hair was removed using clippers and depilatory cream, and the rat was placed on a heating pad in a supine position to expose the lower abdomen. An EPAUSI probe was placed over the lower abdomen of the rat after applying ultrasonic gel for PAUS imaging. The probe was aligned on the sagittal plane with the left side directed toward the cranial direction [refer to Fig. 7(a)]. Below the probe, the male murine genitalia were located, as evidenced by the post-experiment autopsy image of the organ’s arrangement [refer to Fig. 7(b)]. The imaging plane encompassed the urinary bladder (UB), ventral prostate (VP), and urethra, with each organ being distinctly visible in the harmonic US image, showing unique echogenicity and structure [refer to Fig. 7(c)].

Fig. 7

Arterial pulsation in a murine prostate monitored in vivo with EPAUSI: (a) illustration of a transabdominal imaging setting of a male rat in a supine position. (b) Anatomical photograph of a murine male reproductive organ. The dotted line indicates the imaging plane of the probe. (c) Harmonic US B-mode image in the sagittal plane. (d) Overlaid PA and harmonic US image corresponding to (c). (e) Time-resolved normalized US and PA amplitude change in prostatic junction (yellow box in panels (c) and (d); Video 2). (f) US and PA temporal frequency spectra corresponding to (e). UB, urinary bladder; SV, seminal vesicle; VP, ventral prostate; DP, dorsal prostate; and USp, urethral sphincter. Scale bar = 10 mm.

JBO_29_S1_S11528_f007.png

Upon laser excitation, 700 nm PA and harmonic US images of the site were simultaneously displayed at a real-time frame rate of 20 Hz [refer to Fig. 7(d)]. In the center of both real-time views, we observed a repetitive pulsating motion at the boundary of the ventral and dorsal prostate junction at a depth of 8 mm (see Video 2). The prostate is a highly perfused organ with numerous interconnected blood vessels. Near its anatomical correspondence with the urethral axis, the identified vasculature appears to be a major branch from the central prostatic artery. Scattered PA contrast was also observed around the prostatic region, contributing to the subsidiary prostatic microvascular network. As we quantified the temporal amplitude changes in both PA and US images by comparing small regions of interest (ROI) at the pulsating site, we noticed synchronization between PA and US signals, evidenced by identical peaks [refer to Fig. 7(e)]. The signal changes were further analyzed by transforming them into temporal frequency spectra using fast Fourier transform (MATLAB R2021, Mathworks, Massachusetts, United States).71 These analyses revealed peaks at 4.7 and 9.4 Hz in both the US and PA spectra, representing the fundamental heartbeat frequency and its second harmonic [refer to Fig. 7(f)]. In summary, these findings confirm the practicality of our video-rate EPAUSI probe for capturing dynamic hemodynamics in live subjects.

3.5.

Hemoglobin Oxygen Saturation (sO2) Mapping of a Human Blood Vessel In Vivo

We further demonstrated the feasibility of EPAUSI by imaging representative blood vessels from various body parts of a healthy volunteer. These experiments were conducted in strict accordance with protocols approved by the Institutional Review Board of Pohang University of Science and Technology (Protocol POSTECH-PRIB-2023-R009, approved on April 3, 2023). Safety goggles were worn by both the examinee and the examiner to ensure laser safety, and the laser density was controlled to not exceed 25  mJ/cm2 at 700 nm, complying with the ANSI MPE criterion. Various body parts of the examinee, including the neck, upper limb, and lower limb, were imaged by placing the EPAUSI probe in cutaneous contact with US gel. This allowed for the visualization of subcutaneous major vessels such as the carotid artery, jugular vein, radial artery, and popliteal vein. We conducted 20 Hz PAUS bimodal imaging, with PA images acquired at different laser wavelengths (690, 730, 756, and 796 nm) in series. Out of the acquired multispectral PA images, we decoupled oxyhemoglobin and deoxyhemoglobin via spectral unmixing, referring to the molecular extinction coefficient of hemoglobin (source: Oregon Medical Laser Center, Oregon, United States). We calculated pixelwise oxygen saturation (sO2) by dividing oxyhemoglobin by the total hemoglobin, and the sO2 map was thresholded based on the spatially coherent 866 nm PA pixel intensity. This thresholded sO2 map was overlaid onto the ultrasound image, providing both structural and photochemical features of the vessels. To quantify sO2, we manually labeled sO2 pixels neighboring the vascular contrasts, and their upper 75% means were measured as nominal sO2 values. For each blood vessel, five independent measurements of sO2 were taken from consecutive sO2 frames, and the statistical presentation included calculating their mean values and standard errors.

Among the numerous PA images, we identified the deepest vascular contrast at a depth of 17.2 mm, originating from the lower boundary of a popliteal vein [Fig. 8(a)]. Both the upper and lower boundaries were confirmed in the PA image, and their appearances were also presented in the subsequent sO2 images at the same depth [Fig. 8(b)]. The nominal sO2 was measured to be 76.7%±1.4%, a value well within the known range of oxygen saturation in venous flow. Similar strong PA signals and highly correlating sO2 values were observed for the rest of the vessels [Fig. 8(c)]. In the case of larger vessels, such as the jugular vein or carotid artery, PA contrasts were only visible at the upper boundary of the vessels, indicating both strong optical attenuation from the blood within the lumen and an enhanced limited view effect due to the large vessel diameter. Interestingly, it was possible to classify the imaged vessels as arteries or veins based solely on their nominal sO2 values [see Fig. 8(d)]. The sO2 levels were measured to be high at 89.9% and 88.7% for the carotid artery and radial artery, respectively. By contrast, the veins exhibited comparatively lower sO2 levels, with measurements of 74.9%, 75.1%, 76.7%, and 77.2% for the jugular vein, radial vein, and two popliteal veins, respectively. The high consistency in sO2 levels within the same vascular group, both across measurements and between different vessels, underscores the robustness of multispectral EPAUSI. This suggests its potential applicability in measuring sO2 in deep reproductive organs, which could efficiently capture hypoxic lesions commonly found in reproductive diseases, such as cancer or endometrioma.

Fig. 8

Human in vivo transcutaneous multispectral EPAUSI: (a) overlaid harmonic US B-mode and 796 nm PA images of a popliteal vein at a depth of 17.2 mm. (b) A corresponding oxygen saturation (sO2) map overlaid on the harmonic US B-mode image. (c) Series of sO2 maps identified from major vessels in neck, upper limb, and lower limb. (d) Quantitative statistics of nominal sO2 classified into arterial and venous group (two arteries, four veins, n=5). CA, carotid artery; RA, radial artery; JV, jugular vein; RV, radial vein; PV1, popliteal vein 1; and PV2, popliteal vein 2. Scale bar = 10 mm.

JBO_29_S1_S11528_f008.png

4.

Conclusion

In this study, we have developed a video-rate simultaneous endocavity PA and harmonic US imaging system designed for the diagnosis of gynecologic or urologic pathology via physically confined vaginal or rectal canals. This innovative system integrates a commercially available endocavity US probe with a miniaturized light delivery system consisting of a single large core fiber and a line beam engineered diffuser, resulting in a 25 mm thickness and ensuring high laser coupling efficiency. The system offers fine spatial resolution in both the radial (341  μm) and angular (1.23 deg) dimensions, combined with a temporal resolution of 20 Hz. These features prove highly advantageous for conducting multispectral PA imaging and capturing dynamic changes. The system’s capabilities were clearly demonstrated through various applications, including spectral unmixing of MB and ICG in vitro, ICG perfusion studies in an ex vivo porcine ovary, and monitoring of prostatic hemodynamics in an in vivo male rat. Furthermore, our functional EPAUSI successfully assessed the levels of sO2 in the arteries and veins of a healthy volunteer, showcasing its potential for identifying pathological hypoxia in reproductive organs. We believe that EPAUSI holds promise in enhancing diagnostic accuracy in cases of genital pathology, particularly in the diagnosis of conditions such as cancers or endometriosis, by revealing previously undetected signs of vascular changes.

5.

Appendix: Video captions

The following videos are mentioned in the text:

Disclosures

D. Oh and H. Kim have no competing interests to declare. C. Kim has a financial interest in OPTICHO, which did not support this work.

Code and Data Availability

Data will be made available on request.

Acknowledgments

This work was supported by the National Research Foundation (NRF) (Grant Nos. 2021M3C1C3097624, 2020R1A6A1A03047902, and 2023R1A2C3004880), the Korea Medical Device Development Fund (Grant Nos. 1711195277 and RS-2020-KD000008), and BK21 FOUR projects (POSTECH) funded by the Korean government (Ministry of Science and ICT, Ministry of Education, Ministry of Trade, Industry and Energy, Ministry of Health and Welfare, and Ministry of Food and Drug Safety).

References

1. 

H. Sung et al., “Global cancer statistics 2020: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries,” Cancer J. Clin., 71 (3), 209 –249 https://doi.org/10.3322/caac.21660 (2021). Google Scholar

2. 

W. Dähnert et al., “Prostatic evaluation by transrectal sonography with histopathologic correlation: the echopenic appearance of early carcinoma,” Radiology, 158 (1), 97 –102 https://doi.org/10.1148/radiology.158.1.3510032 RADLAX 0033-8419 (1986). Google Scholar

3. 

A. M. Sassone et al., “Transvaginal sonographic characterization of ovarian disease: evaluation of a new scoring system to predict ovarian malignancy,” Obstet. Gynecol., 78 (1), 70 –76 (1991). Google Scholar

4. 

K. Shinohara, T. M. Wheeler and P. T. Scardino, “The appearance of prostate cancer on transrectal ultrasonography: correlation of imaging and pathological examinations,” J. Urol., 142 (1), 76 –82 https://doi.org/10.1016/S0022-5347(17)38666-4 (1989). Google Scholar

5. 

S. C. Modesitt et al., “Risk of malignancy in unilocular ovarian cystic tumors less than 10 centimeters in diameter,” Obstet. Gynecol., 102 (3), 594 –599 https://doi.org/10.1016/S0029-7844(03)00670-7 (2003). Google Scholar

6. 

L. V. Rodriguez and M. K. Terris, “Risks and complications of transrectal ultrasound guided prostate needle biopsy: a prospective study and review of the literature,” J. Urol., 160 (6 Part 1), 2115 –2120 https://doi.org/10.1016/S0022-5347(01)62255-9 (1998). Google Scholar

7. 

C. Harvey et al., “Applications of transrectal ultrasound in prostate cancer,” Br. J. Radiol., 85 (special_issue_1), S3 –S17 https://doi.org/10.1259/bjr/56357549 BJRAAP 0007-1285 (2012). Google Scholar

8. 

J. C. Applewhite et al., “Transrectal ultrasound and biopsy in the early diagnosis of prostate cancer,” Cancer Control, 8 (2), 141 –150 https://doi.org/10.1177/107327480100800204 (2001). Google Scholar

9. 

M. K. Terris et al., “Efficacy of transrectal ultrasound for identification of clinically undetected prostate cancer,” J. Urol., 146 (1), 78 –84 https://doi.org/10.1016/S0022-5347(17)37718-2 (1991). Google Scholar

10. 

K. Chen-Dixon et al., “Effectiveness of ultrasound for endometriosis diagnosis,” Curr. Opin. Obstet. Gynecol., 34 (5), 324 –331 https://doi.org/10.1097/GCO.0000000000000812 COOGEA 1040-872X (2022). Google Scholar

11. 

B. Park et al., “3D wide-field multispectral photoacoustic imaging of human melanomas in vivo: a pilot study,” J. Eur. Acad. Dermatol. Venereol., 35 (3), 669 –676 https://doi.org/10.1111/jdv.16985 JEAVEQ 0926-9959 (2021). Google Scholar

12. 

J. Kim et al., “Multiparametric photoacoustic analysis of human thyroid cancers in vivo,” Cancer Res., 81 (18), 4849 –4860 https://doi.org/10.1158/0008-5472.CAN-20-3334 CNREA8 0008-5472 (2021). Google Scholar

13. 

A. Karlas et al., “Optoacoustic imaging in endocrinology and metabolism,” Nat. Rev. Endocrinol., 17 (6), 323 –335 https://doi.org/10.1038/s41574-021-00482-5 (2021). Google Scholar

14. 

V. G. Andreev et al., “Optoacoustic tomography of breast cancer with arc-array transducer,” Proc. SPIE, 3916 36 –47 https://doi.org/10.1117/12.386339 (2000). Google Scholar

15. 

N. Nyayapathi and J. Xia, “Photoacoustic imaging of breast cancer: a mini review of system design and image features,” J. Biomed. Opt., 24 (12), 121911 https://doi.org/10.1117/1.JBO.24.12.121911 JBOPFO 1083-3668 (2019). Google Scholar

16. 

S. M. Schoustra et al., “Twente Photoacoustic Mammoscope 2: system overview and three-dimensional vascular network images in healthy breasts,” J. Biomed. Opt., 24 (12), 121909 https://doi.org/10.1117/1.JBO.24.12.121909 JBOPFO 1083-3668 (2019). Google Scholar

17. 

S. Manohar and M. Dantuma, “Current and future trends in photoacoustic breast imaging,” Photoacoustics, 16 100134 https://doi.org/10.1016/j.pacs.2019.04.004 (2019). Google Scholar

18. 

W. Choi et al., “Three-dimensional multistructural quantitative photoacoustic and US imaging of human feet in vivo,” Radiology, 303 (2), 467 –473 https://doi.org/10.1148/radiol.211029 RADLAX 0033-8419 (2022). Google Scholar

19. 

N.-A. Fasoula et al., “Non-invasive multispectral optoacoustic tomography resolves intrahepatic lipids in patients with hepatic steatosis,” Photoacoustics, 29 100454 https://doi.org/10.1016/j.pacs.2023.100454 (2023). Google Scholar

20. 

A. P. Regensburger et al., “Detection of collagens by multispectral optoacoustic tomography as an imaging biomarker for Duchenne muscular dystrophy,” Nat. Med., 25 (12), 1905 –1915 https://doi.org/10.1038/s41591-019-0669-y 1078-8956 (2019). Google Scholar

21. 

W. Choi et al., “Clinical photoacoustic imaging platforms,” Biomed. Eng. Lett., 8 139 –155 https://doi.org/10.1007/s13534-018-0062-7 (2018). Google Scholar

22. 

A. B. E. Attia et al., “A review of clinical photoacoustic imaging: current and future trends,” Photoacoustics, 16 100144 https://doi.org/10.1016/j.pacs.2019.100144 (2019). Google Scholar

23. 

B. Park, C. Kim and J. Kim, “Recent advances in ultrasound and photoacoustic analysis for thyroid cancer diagnosis,” Adv. Phys. Res., 2 (4), 2200070 https://doi.org/10.1002/apxr.202200070 (2023). Google Scholar

24. 

D. Kim et al., “An ultraviolet-transparent ultrasound transducer enables high-resolution label-free photoacoustic histopathology,” Laser Photonics Rev., 18 (2), 2300652 https://doi.org/10.1002/lpor.202300652 (2024). Google Scholar

25. 

E. Park et al., “Azimuth mapping of fibrous tissue in linear dichroism-sensitive photoacoustic microscopy,” Photoacoustics, 31 100510 https://doi.org/10.1016/j.pacs.2023.100510 (2023). Google Scholar

26. 

J. Yao et al., “Breaking the speed limits in photoacoustic microscopy,” Photoacoustics, 32 100541 https://doi.org/10.1016/j.pacs.2023.100541 (2023). Google Scholar

27. 

S. H. Hwang et al., “3D printed multi-growth factor delivery patches fabricated using dual-crosslinked decellularized extracellular matrix-based hybrid inks to promote cerebral angiogenesis,” Acta Biomater., 157 137 –148 https://doi.org/10.1016/j.actbio.2022.11.050 (2023). Google Scholar

28. 

J. Yoo et al., “Switchable preamplifier for dual modal photoacoustic and ultrasound imaging,” Biomed. Opt. Express, 14 (1), 89 –105 https://doi.org/10.1364/BOE.476453 BOEICL 2156-7085 (2023). Google Scholar

29. 

J. Kim et al., “High-resolution photoacoustic/ultrasound imaging of the porcine stomach wall: an ex vivo feasibility study,” Biomed. Opt. Express, 12 (11), 6717 –6729 https://doi.org/10.1364/BOE.441241 BOEICL 2156-7085 (2021). Google Scholar

30. 

S. Jeon et al., “Review on practical photoacoustic microscopy,” Photoacoustics, 15 100141 https://doi.org/10.1016/j.pacs.2019.100141 (2019). Google Scholar

31. 

J. Kim et al., “Deep learning acceleration of multiscale superresolution localization photoacoustic imaging,” Light: Sci. Appl., 11 (1), 131 https://doi.org/10.1038/s41377-022-00820-w (2022). Google Scholar

32. 

J. Park et al., “Quadruple ultrasound, photoacoustic, optical coherence, and fluorescence fusion imaging with a transparent ultrasound transducer,” Proc. Natl. Acad. Sci., 118 (11), e1920879118 https://doi.org/10.1073/pnas.1920879118 (2021). Google Scholar

33. 

J. W. Baik et al., “Super wide-field photoacoustic microscopy of animals and humans in vivo,” IEEE Trans. Med. Imaging, 39 (4), 975 –984 https://doi.org/10.1109/TMI.2019.2938518 ITMID4 0278-0062 (2019). Google Scholar

34. 

N. Kwon et al., “Hexa-BODIPY-cyclotriphosphazene based nanoparticle for NIR fluorescence/photoacoustic dual-modal imaging and photothermal cancer therapy,” Biosens. Bioelectron., 216 114612 https://doi.org/10.1016/j.bios.2022.114612 BBIOE4 0956-5663 (2022). Google Scholar

35. 

Y. Ding et al., “Surfactant-stripped semiconducting polymer micelles for tumor theranostics and deep tissue imaging in the NIR-II window,” Small, 18 (6), 2104132 https://doi.org/10.1002/smll.202104132 SMALBC 1613-6810 (2022). Google Scholar

36. 

D. Maji et al., “Copper-catalyzed covalent dimerization of near-infrared fluorescent cyanine dyes: synergistic enhancement of photoacoustic signals for molecular imaging of tumors,” Anal. Sens., 2 (1), e202100045 https://doi.org/10.1002/anse.202100045 (2022). Google Scholar

37. 

J. Park et al., “Bi-modal near-infrared fluorescence and ultrasound imaging via a transparent ultrasound transducer for sentinel lymph node localization,” Opt. Lett., 47 (2), 393 –396 https://doi.org/10.1364/OL.446041 OPLEDP 0146-9592 (2022). Google Scholar

38. 

S. Manohar and D. Razansky, “Photoacoustics: a historical review,” Adv. Opt. Photonics, 8 (4), 586 –617 https://doi.org/10.1364/AOP.8.000586 AOPAC7 1943-8206 (2016). Google Scholar

39. 

L. V. Wang, “Prospects of photoacoustic tomography,” Med. Phys., 35 (12), 5758 –5767 https://doi.org/10.1118/1.3013698 MPHYA6 0094-2405 (2008). Google Scholar

40. 

L. V. Wang and J. Yao, “A practical guide to photoacoustic tomography in the life sciences,” Nat. Methods, 13 (8), 627 –638 https://doi.org/10.1038/nmeth.3925 1548-7091 (2016). Google Scholar

41. 

D. Oh et al., “Contrast agent-free 3D renal ultrafast doppler imaging reveals vascular dysfunction in acute and diabetic kidney diseases,” Adv. Sci., 10 (36), 2303966 https://doi.org/10.1002/advs.202303966 (2023). Google Scholar

42. 

W. Kim et al., “Wide-field three-dimensional photoacoustic/ultrasound scanner using a two-dimensional matrix transducer array,” Opt. Lett., 48 (2), 343 –346 https://doi.org/10.1364/OL.475725 OPLEDP 0146-9592 (2023). Google Scholar

43. 

J. Ahn et al., “In vivo photoacoustic monitoring of vasoconstriction induced by acute hyperglycemia,” Photoacoustics, 30 100485 https://doi.org/10.1016/j.pacs.2023.100485 (2023). Google Scholar

44. 

B. Park et al., “Functional photoacoustic imaging: from nano-and micro-to macro-scale,” Nano Converg., 10 (1), 29 https://doi.org/10.1186/s40580-023-00377-3 (2023). Google Scholar

45. 

S. Choi et al., “Deep learning enhances multiparametric dynamic volumetric photoacoustic computed tomography in vivo (DL-PACT),” Adv. Sci., 10 (1), 2202089 https://doi.org/10.1002/advs.202202089 (2023). Google Scholar

46. 

J. Ahn et al., “Fully integrated photoacoustic microscopy and photoplethysmography of human in vivo,” Photoacoustics, 27 100374 https://doi.org/10.1016/j.pacs.2022.100374 (2022). Google Scholar

47. 

W. Choi et al., “Recent advances in contrast-enhanced photoacoustic imaging: overcoming the physical and practical challenges,” Chem. Rev., 123 (11), 7379 –7419 https://doi.org/10.1021/acs.chemrev.2c00627 CHREAY 0009-2665 (2023). Google Scholar

48. 

B. Park et al., “Shear-force photoacoustic microscopy: toward super-resolution near-field imaging,” Laser Photonics Rev., 16 (12), 2200296 https://doi.org/10.1002/lpor.202200296 (2022). Google Scholar

49. 

B. Park et al., “Listening to drug delivery and responses via photoacoustic imaging,” Adv. Drug Deliv. Rev., 184 114235 https://doi.org/10.1016/j.addr.2022.114235 ADDREP 0169-409X (2022). Google Scholar

50. 

W. Choi, D. Oh and C. Kim, “Practical photoacoustic tomography: realistic limitations and technical solutions,” J. Appl. Phys., 127 (23), 230903 https://doi.org/10.1063/5.0008401 JAPIAU 0021-8979 (2020). Google Scholar

51. 

S. Cho et al., “3D PHOVIS: 3D photoacoustic visualization studio,” Photoacoustics, 18 100168 https://doi.org/10.1016/j.pacs.2020.100168 (2020). Google Scholar

52. 

W. M. MacCuaig et al., “Development of multispectral optoacoustic tomography as a clinically translatable modality for cancer imaging,” Radiol. Imaging Cancer, 2 (6), e200066 https://doi.org/10.1148/rycan.2020200066 (2020). Google Scholar

53. 

A. Taruttis and V. Ntziachristos, “Advances in real-time multispectral optoacoustic imaging and its applications,” Nat. Photonics, 9 (4), 219 –227 https://doi.org/10.1038/nphoton.2015.29 NPAHBY 1749-4885 (2015). Google Scholar

54. 

V. Ntziachristos and D. Razansky, “Molecular imaging by means of multispectral optoacoustic tomography (MSOT),” Chem. Rev., 110 (5), 2783 –2794 https://doi.org/10.1021/cr9002566 CHREAY 0009-2665 (2010). Google Scholar

55. 

V. Gujrati, A. Mishra and V. Ntziachristos, “Molecular imaging probes for multi-spectral optoacoustic tomography,” Chem. Commun., 53 (34), 4653 –4672 https://doi.org/10.1039/C6CC09421J (2017). Google Scholar

56. 

C. Yoon et al., “Motion compensation for 3D multispectral handheld photoacoustic imaging,” Biosensors, 12 (12), 1092 https://doi.org/10.3390/bios12121092 BISSED 0265-928X (2022). Google Scholar

57. 

J. Yang, S. Choi and C. Kim, “Practical review on photoacoustic computed tomography using curved ultrasound array transducer,” Biomed. Eng. Lett., 12 19 –35 https://doi.org/10.1007/s13534-021-00214-8 (2022). Google Scholar

58. 

C. Lee et al., “Panoramic volumetric clinical handheld photoacoustic and ultrasound imaging,” Photoacoustics, 31 100512 https://doi.org/10.1016/j.pacs.2023.100512 (2023). Google Scholar

59. 

C. Lee et al., “Three-dimensional clinical handheld photoacoustic/ultrasound scanner,” Photoacoustics, 18 100173 https://doi.org/10.1016/j.pacs.2020.100173 (2020). Google Scholar

60. 

A. Horiguchi et al., “A pilot study of photoacoustic imaging system for improved real-time visualization of neurovascular bundle during radical prostatectomy,” Prostate, 76 (3), 307 –315 https://doi.org/10.1002/pros.23122 (2016). Google Scholar

61. 

A. Horiguchi et al., “Pilot study of prostate cancer angiogenesis imaging using a photoacoustic imaging system,” Urology, 108 212 –219 https://doi.org/10.1016/j.urology.2017.07.008 (2017). Google Scholar

62. 

S.-R. Kothapalli et al., “Simultaneous transrectal ultrasound and photoacoustic human prostate imaging,” Sci. Transl. Med., 11 (507), eaav2169 https://doi.org/10.1126/scitranslmed.aav2169 STMCBQ 1946-6234 (2019). Google Scholar

63. 

C. Liu et al., “In vivo transrectal imaging of canine prostate with a sensitive and compact handheld transrectal array photoacoustic probe for early diagnosis of prostate cancer,” Biomed. Opt. Express, 10 (4), 1707 –1717 https://doi.org/10.1364/BOE.10.001707 BOEICL 2156-7085 (2019). Google Scholar

64. 

J. Jang et al., “Transrectal ultrasound and photoacoustic imaging probe for diagnosis of prostate cancer,” Sensors, 21 (4), 1217 https://doi.org/10.3390/s21041217 SNSRES 0746-9462 (2021). Google Scholar

65. 

H. S. Salehi et al., “Design of optimal light delivery system for co-registered transvaginal ultrasound and photoacoustic imaging of ovarian tissue,” Photoacoustics, 3 (3), 114 –122 https://doi.org/10.1016/j.pacs.2015.08.003 (2015). Google Scholar

66. 

B. Cong et al., “A low cost sensitive transrectal photoacoustic probe with single-fiber bright-field illumination for in vivo canine prostate imaging and real-time biopsy needle guidance,” IEEE Sens. J., 20 (18), 10974 –10980 https://doi.org/10.1109/JSEN.2020.2993884 ISJEAZ 1530-437X (2020). Google Scholar

67. 

Y. Yan et al., “Endocavity ultrasound and photoacoustic system for fetal and maternal imaging: design, implementation, and ex-vivo validation,” J. Med. Imaging, 8 (6), 066001 https://doi.org/10.1117/1.JMI.8.6.066001 JMEIET 0920-5497 (2021). Google Scholar

68. 

G. Yang et al., “Optimizing light delivery through ball-shaped multimode fiber tips in co-registered photoacoustic and ultrasound endo-cavity imaging: simulation and experimental validation,” Photons Plus Ultrasound Imaging Sens., 2019 465 –471 https://doi.org/10.1117/12.2510914 (2019). Google Scholar

69. 

S. Prahl, “Assorted spectra (Hemoglobin, Indocyanine Green, Methylene Blue),” https://omlc.org/spectra (1999). Google Scholar

70. 

H. G. Akarçay et al., “Determining the optical properties of a gelatin-TiO2 phantom at 780 nm,” Biomed. Opt. Express, 3 (3), 418 –434 https://doi.org/10.1364/BOE.3.000418 BOEICL 2156-7085 (2012). Google Scholar

71. 

L. Li et al., “Single-impulse panoramic photoacoustic computed tomography of small-animal whole-body dynamics at high spatiotemporal resolution,” Nat. Biomed. Eng., 1 (5), 0071 https://doi.org/10.1038/s41551-017-0071 (2017). Google Scholar

Biography

Donghyeon Oh is currently an MS-PhD candidate at Pohang University of Science and Technology in Pohang, Republic of Korea, under the supervision of Prof. Chulhong Kim. His research primarily centers on contrast agent-free microvascular imaging techniques, including photoacoustic imaging or ultrafast ultrasound imaging.

Hyunhee Kim is currently a PhD candidate at POSTECH under the supervision of Prof. Chulhong Kim. Her research is primarily focused on the development of an interventional multimodal imaging system that integrates ultrasound and photoacoustic imaging.

Minsik Sung is working at the Pohang University of Science and Technology as an MS PhD candidate in Pohang, Republic of Korea, under the supervision of Prof. Chulhong Kim. His research focuses on the clinical translational applications of photoacoustic and ultrasound imaging systems.

Chulhong Kim is a fellow member of SPIE and currently serves as a professor in the fields of electrical engineering, convergence IT engineering, mechanical engineering, and medical science and engineering at POSTECH, Pohang, Republic of Korea. He is also the recipient of the prestigious Namgo Chair Professorship at POSTECH. He received his PhD in biomedical engineering from Washington University, St. Louis, Missouri, United States. His research is primarily focused on advancing preclinical and clinical translations of multiscale photoacoustic and ultrasound imaging techniques.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Donghyeon Oh, Hyunhee Kim, Minsik Sung, and Chulhong Kim "Video-rate endocavity photoacoustic/harmonic ultrasound imaging with miniaturized light delivery," Journal of Biomedical Optics 29(S1), S11528 (19 March 2024). https://doi.org/10.1117/1.JBO.29.S1.S11528
Received: 8 November 2023; Accepted: 28 February 2024; Published: 19 March 2024