The field of diffuse optics has provided a rich set of neurophotonic tools to measure the human brain noninvasively. Interferometric detection is a recent, exciting methodological development in this field. The approach is especially promising for the measurement of diffuse fluctuation signals related to blood flow. Benefitting from inexpensive sensor arrays, the interferometric approach has already dramatically improved throughput, enabling the measurement of brain blood flow faster and deeper. The interferometric approach can also achieve time-of-flight resolution, improving the accuracy of acquired signals. We provide a historical perspective and summary of recent work in the nascent area of interferometric diffuse optics. We predict that the convergence of interferometric technology with existing economies of scale will propel many advances in the years to come. |
1.IntroductionDiffuse optics seeks to characterize and image biological tissue with multiply scattered near-infrared light. From the inception of this scientific field, the potential of near-infrared light to probe brain physiology was recognized.1 Diffuse optics is divided according to the measured signals, with light absorption and scattering for near-infrared spectroscopy (NIRS)2 or coherent light fluctuations (CLFs) for diffuse correlation spectroscopy (DCS),3 and according to detection methodology, with continuous wave (CW),4 time domain (TD),5 frequency domain (FD),6 or spatial frequency domain.7 CW NIRS was first demonstrated around 45 years ago (Fig. 1)1 and remains widely used for functional brain measurements.16,17 In the 1980s, researchers recognized the inherent ambiguities of CW measurements and explored novel dimensions, by resolving tissue responses in the TD18 and FD.19 These measurements are now considered to be quantitative benchmarks to disambiguate absorption and scattering and to achieve depth sensitivity. In the late 1980s, soft matter physicists characterized the temporal fluctuations of diffusely scattered light under the rubric of diffusing wave spectroscopy (DWS).20 This approach was later applied to measure the biological blood flow index (BFI) under the name DCS in the mid-1990s (Fig. 1).21 Though diffuse correlation functions provide coarse information about long and short paths,20,22 today DCS remains essentially a CW technique,3,23 with rare exceptions.10,24 Yet, as in NIRS, the potential benefits of time-resolved detection for DWS and DCS were appreciated early. For instance, nonlinear gating approaches could resolve intensity fluctuations according to time-of-flight (TOF),25 yet these early efforts were inefficient, and thus infeasible, in biological tissue. The early 1990s saw the development of an interferometric approach for measuring backscattered light from tissue, using light coherence to temporally gate or filter detected light.26 This new approach was called optical coherence tomography (OCT). The analogy between direct TOF detection and coherence gating of OCT was appreciated, and the literature is peppered with examples of a coherence gate aiding measurements of multiply scattered light.27–29 Yet these early implementations essentially replicated the measurement geometry of OCT, with a very narrow coherence gate and a single detector, resulting in excessive loss of potentially useful scattered light signal. Though the efforts yielded useful insight into the transition between single and diffusive dynamic scattering, they were not applied in vivo. Swept source approaches, inspired by Fourier domain OCT,30 for investigating colloidal suspensions were implemented but were either never applied in vivo31 or had limited speed and longevity.32,33 Meanwhile, in vivo CW DCS began to show potential in monitoring the human brain8 and became the method of choice for clinical studies.3,34,35 Later, CLFs were found to theoretically provide times more brain-specificity than absorption measured by NIRS at a given source–collector (S–C) separation or TOF (Fig. 2).22,38 Yet, for many years, this potential advantage was tempered by the inherently low-light throughput of conventional DCS/DWS, making it impractical to achieve such a high brain specificity in adults.3,23 Over the past 5 years, motivated in part by the successes of DCS, several groups have independently investigated interferometric detection for diffusely scattered light in living biological tissue.12–15,36,39–43 This area of investigation remains active. Already, several clear and tangible advantages of the interferometric approach have emerged. First, interferometric detection provides comparable or better performance than photon counting with a dramatically reduced cost per pixel.12,15,36,42 Second, modifying light temporal coherence is a powerful and flexible way to achieve TOF resolution or discrimination in an interferometric setup,39–41,43 removing ambiguities to accurate interpretation of signals. Third, as a consequence, interferometric detection has great potential for blood flow fluctuation signals,36 whereas conventional DCS is constrained by high cost and low throughput.44 In this work, we discuss core principles of interferometric detection, including the concept of heterodyne gain, signal-to-noise ratios (SNRs) of heterodyne and homodyne detection, choice of reference power, coherence gating, TOF-resolved detection, and 2D sensor approaches. Then we describe remaining issues to be understood and addressed, such as mitigating motion artifact, further improving the SNR and brain specificity, and reducing the light source cost. 2.Core Principles2.1.Heterodyne GainIn iDWS, heterodyne gain enables inexpensive sensors to measure weak sample light levels. Here we adopt the definition of “heterodyne” from dynamic light scattering,45 which differs from the definition in optical communications. In heterodyne DCS/DWS methods, a strong reference light field is interfered with a weak diffuse (sample) light field(s) : where is the measured power; is the homodyne sample power, which fluctuates over time due to sample dynamics; is the reference power, which is assumed to be constant; and is the heterodyne term, which includes the in-phase component of . The time-averaged reference and sample powers are and , respectively.For a simple signal comparison between homodyne and heterodyne detection, we assume that (1) all powers are given in units of photoelectrons per second; (2) single detector/pixel for single speckle, indicating homodyne coherence factor3 and heterodyne mutual coherence degree12 are both equal to 1; and (3) the detector exposure time is much shorter than the decorrelation time of , (i.e., ), ensuring no decorrelation during the exposure time.15 Thus the heterodyne gain is the heterodyne to homodyne signal ratio of (see Table 1). Table 1Signal and SNR for heterodyne and homodyne detection.
σshot2: shot noise variance; σtot2: total noise variance; σcam2: camera/detector noise variance; SNRshot: signal-to-shot-noise ratio; and SNRtot: signal-to-total-noise ratio, where P¯RT≫σcam2≫P¯ST for low sample power, high reference power, and noisy camera/detector. Heterodyne gain or amplification is particularly beneficial for lower sample powers (Fig. 3). With an appropriate detector, we can achieve . Boosting the sample signal in this manner has three important advantages. First, it enables digitizing the signal of interest with more bits and thus higher fidelity [Fig. 3(b)]. Second, it boosts the associated signal and shot noise above the detector noise [Fig. 3(c)], so it is possible to reach the shot noise limit, as described in the next section. Third, it entails measuring the sample field, or at least its in-phase part, as opposed to the sample power. Thus the heterodyne method is sensitive to sample intensity and phase fluctuations, whereas the homodyne method is just sensitive to intensity fluctuations.46 2.2.Shot-Noise Limited SNRThough heterodyne gain is a useful concept, it does not account for noise. For instance, a large reference power increases both the signal and the shot noise. For this reason, SNR is a more appropriate metric to compare detection approaches. First, we consider a noiseless detector in which the only noise source is shot noise (). If for weak sample light, we find that for heterodyne detection (Table 1), double that of homodyne detection. The factor of 2 in arises from the heterodyne signal term in Eq. (1). Next, we consider detector or camera noise (i.e., total noise given by ). If for weak sample light detected by a noisy sensor, we find that is still obtained for heterodyne detection (Table 1), whereas for homodyne detection, much lower than that of case 1. This is because heterodyne detection can achieve the shot noise limit even for a noisy camera/detector (i.e., ), enabling us to “see” a small sample signal , which is otherwise buried in the camera noise [compare “heterodyne” and “sample only” in Fig. 3(c)]. 2.3.Choice of Reference PowerThe above considerations suggest that a reference count that is much larger than the sample count is beneficial. This ensures that the homodyne autocorrelation term is much smaller than the heterodyne autocorrelation term, which simplifies the interpretation of data. An ideal reference power should enable all pixels to reach the shot noise limit12,15,36 but, obviously, not cause the detector to saturate. A high reference power also places stringent requirements on stability so as not to induce additional spurious decorrelation dynamics from reference fluctuations. We have not observed excess noise that would preclude high reference powers in our experiments36 with isolated and temperature-controlled DFB and DBR lasers. Others have proposed signal processing strategies to deal with noise from mode hopping.47 Despite the benefits of a large reference count, several works have operated in a regime in which the reference count is comparable to the sample count.48 Whether this is a necessary consequence of using detectors with a low saturation level at short S–C separation, or whether there are actually advantages to working in this regime, remains unclear. Early work suggested the choice of an optimal reference power approximately equal to the sample power for measuring low-correlation levels,49 but a high reference power was later shown to be optimal for large S–C separations over much of the autocorrelation decay;42 it was demonstrated that relative variability of fitted BFIs monotonically decreased by up to 80% for a reference to total power ratio, , increasing from 0 to . Data analysis in the case of comparable reference and sample powers can be handled by a modified Siegert relationship that includes a static term,12,42,46,48 which is expressed as where and are the normalized intensity and field autocorrelations, respectively, is the reference to total power ratio, is the heterodyne mutual coherence degree,12 is the homodyne coherence factor, and the standard Siegert relation of is obtained by setting . Note that the middle term of Eq. (2) is the heterodyne term of iDWS.12 In the case that is real and , Eq. (2) reduces to more common forms.422.4.Coherence GatingIn homodyne DCS, all optical paths must be coherent with each other to maximize the available fluctuation signal (i.e., of intensity autocorrelation). Reduced coherence superposes independent speckles and effectively reduces [Figs. 4(a) and 4(c)]. Low coherence is undesirable in CW DCS and is unavoidable in time-domain DCS.10,50 In iDWS with a large heterodyne gain, the predominant signal is formed by the interference of sample paths with the reference path. A low-coherence light source can, therefore, create a TOF filter, attenuating paths that do not fall within the coherence function [Fig. 4(b)].51 TOF filtering reduces the height of , but is beneficial, on balance, as superficial paths can be preferentially attenuated.52 Thus coherence reduction is a simple approach to achieving TOF filtering or discrimination, which can attenuate scalp photons to enhance brain sensitivity at small S–C separations [Fig. 4(d)].39 2.5.TOF-Resolved Interferometric Diffuse OpticsGoing beyond mere TOF discrimination, TOF resolution aims for a full temporal point spread function (TPSF)53 and concordant TOF-resolved dynamics.10,24 iNIRS is an interferometric diffuse optical approach that measures superposed interference spectra, which are Fourier transformed to yield TOF-resolved complex optical field time courses, from which TOF-resolved intensity and field autocorrelations are determined (Fig. 5). iNIRS can separate static and dynamic components in the TPSF,41 providing rich information about multiple scattered light paths. TPSFs provide optical properties, which facilitate recovering absolute BFI.3,23,35 Admittedly, any technique providing TOF-resolved CLFs10,40,51 can also measure TPSFs and, in principle, with sufficient TOF resolution, quantify optical properties. However, unlike TD-DCS, iNIRS does not suffer from the tradeoff between coherence and TOF resolution54 and can achieve TOF resolutions of picoseconds41 with a high dynamic range through spectral shaping.54 The major drawback of iNIRS is its intrinsically low-light throughput, resulting from the high cost of channels digitized at hundreds of MHz. As argued elsewhere, TOF-discrimination is compatible with highly parallel detection by less expensive sensor arrays. 2.6.2D Camera Interferometric ApproachesRecently, interferometric diffuse optical approaches have been adapted for low frame rate 2D cameras. Conventional iDWS uses a fast camera with a short exposure time relative to the decorrelation time .12 MiDWS relaxes this requirement, instead probing the autocorrelation indirectly by varying exposure time of a 2D camera with low frame rate.15 Other 2D approaches probe multiple heterodyne frequencies13 or even use a single exposure and measure the sample power separately.14 2D approaches can theoretically achieve orders of magnitude lower cost per speckle than iDWS due to the massive number of available pixels ( to ) on 2D cameras. In benefitting from 2D cameras, these approaches are akin to SCOS/DSCA (diffuse speckle contrast analysis), which measures BFI from homodyne speckle contrast(s).11,55–58 MiDWS and SCOS/DSCA are compared for human brain BFI measurements in Table 2. Table 2Comparisons of key features between SCOS/DSCA and MiDWS/iSVS for the human brain.
κ, speckle contrast55 and ⟨UAC2⟩, mean square of the heterodyne signal.15 Is it possible to obtain TOF information while benefitting from the scalability of 2D interferometric approaches? A brute force approach that borrows concepts from iNIRS and iDWS is to employ a high-speed and expensive camera for parallelization of iNIRS.59 However, to achieve sufficient sampling of the interference spectrum for reconstruction of a full TPSF41 with this approach, even at frame rates of MHz, maximal sweep rates are, at best, in the tens of kHz range. An alternative strategy is to forego TOF resolution in favor TOF discrimination,39 which relaxes the frame rate requirements and is (in theory, at least) compatible with highly parallel multiexposure approaches. 3.Limitations, Questions, and Issues to be SolvedNotwithstanding their potential, interferometric diffuse optical techniques come with unique limitations that must be understood and addressed. 3.1.Motion ArtifactsInterferometric detection, which depends on fluctuations of both light intensity and phase, is more susceptible to motion artifacts than conventional intensity-based DCS/DWS. Moreover, many iDWS systems collect light with MMFs, the motion or vibration of which could induce dynamics beyond those intrinsic to the sample. We generally find that iDWS measurements are not that susceptible to moderate motion of a few meter collection MMF, given that the motional decorrelation dynamics are much slower than intrinsic sample decorrelation dynamics.36 To further mitigate MMF motion artifacts, an MMF bundle with a smaller core size and NA could be employed,60 as suggested by related work.14 Another issue for noncontact interferometric detection41,43 is axial motion between the subject and probe, which causes Doppler phase drifts resulting in overestimated sample dynamics. The iNIRS method solves this issue by measuring TOF-resolved complex optical fields, from which the Doppler shift can be quantified from static light at short TOFs, and such artifacts can be corrected.41 3.2.SNR or Brain SpecificityEven with recent advances, state-of-the-art iDWS has not yet reached the light throughput of widely used CW-NIRS, and measurements at large S–C separations (), at late TOFs, or in regions with overlying hair remain challenging.36 How do we further improve SNR by additional 1 to 2 orders-of-magnitude and better achieve theoretical advantages of BFI for brain specificity (Fig. 2)? Here we describe three potential solutions, not mutually exclusive, that are being investigated: first, enhancing SNR by further increasing the speckle number, based on parallelizing multiple cameras for iDWS/MiDWS, is a straightforward way to achieve larger S–C separation (i.e., higher brain specificity); second, combining iDWS with coherence gating, to filter out extracerebral photons with short TOFs at a small S–C separation, can improve both SNR (more cerebral photons) and brain specificity;39,41,43 third, the intrinsic advantages of SNR and high source power for DCS at 1064 nm can be carried over to interferometric detection,61,62 in which extremely costly 1064 nm photon counting detectors63 can be replaced by InGaAs cameras,64 which are less costly. 3.3.Light Source CostWith detector technologies supported by economies of scale, we project that source cost will soon become a limiting factor in scaling interferometric diffuse optics. Benefiting from the high pixel counts of CMOS sensors, interferometric detection has reduced the cost of a single detected speckle by orders of magnitude36 compared with conventional multispeckle DCS/DWS.44 This trend is likely to continue. For instance, the 2D CMOS camera used in Fourier domain DCS13,47 only costs hundreds of dollars, which is dozens of times less expensive than the source. State-of-the-art iDWS36 has used similar light sources as state-of-the-art DCS/DWS44 to date. We believe that more systematic iDWS investigations with lower cost laser sources65 are warranted. Indeed, we have found that a lower temporal coherence, characteristic of lower cost sources, can actually be helpful to reducing the impact of unwanted system reflections in iDWS. Finally, although iDWS at 1064 nm can also reduce the laser source cost,64 InGaAs sensors unfortunately are expensive. Though not strictly a limitation, we project that lowering the light source cost, perhaps in conjunction with noise mitigation strategies,47 will be important to enabling new applications in the coming years. 4.OutlookFunctional NIRS developed from a single channel4 to high-density diffuse optical tomography (DOT)16 in two decades. Because photon counting does not easily scale, DCS/DWS has not seen similar progress, almost two decades since the first functional measurements.8 Technologies on the horizon such as SPAD arrays may eventually improve performance but have not yet done so.66,67 Interferometric detection presents one way to enhance both the SNR and performance-to-cost ratio to fully realize the advantages of BFI.22,38 Moreover, iDWS may scale to multiple detection channels with a lower cost than other technologies. Looking ahead, an interferometric diffuse optical tomography system could achieve more brain-specific functional mapping with same S–C pairs as conventional DOT, or higher spatial resolution (shorter S–C separation) with the same brain specificity as conventional DOT. With the rapid progress over the past five years, we predict that these exciting developments, which could benefit neurointensive care, functional neuroimaging, and brain–computer interfaces, are just on the horizon. AcknowledgmentsThis work was supported by the National Institutes of Health (Nos. R01EB029747, R01EY031469, P41EB032840, and R01NS094681) and the National Natural Science Foundation of China (No. 62105315). ReferencesF. F. Jöbsis,
“Noninvasive, infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters,”
Science, 198 1264
–1267 https://doi.org/10.1126/science.929199 SCIEAS 0036-8075
(1977).
Google Scholar
M. Ferrari and V. Quaresima,
“A brief review on the history of human functional near-infrared spectroscopy (fNIRS) development and fields of application,”
NeuroImage, 63 921
–935 https://doi.org/10.1016/j.neuroimage.2012.03.049 NEIMEF 1053-8119
(2012).
Google Scholar
T. Durduran and A. G. Yodh,
“Diffuse correlation spectroscopy for non-invasive, micro-vascular cerebral blood flow measurement,”
NeuroImage, 85 51
–63 https://doi.org/10.1016/j.neuroimage.2013.06.017 NEIMEF 1053-8119
(2014).
Google Scholar
A. Villringer et al.,
“Near infrared spectroscopy (NIRS): a new tool to study hemodynamic changes during activation of brain function in human adults,”
Neurosci. Lett., 154 101
–104 https://doi.org/10.1016/0304-3940(93)90181-J NELED5 0304-3940
(1993).
Google Scholar
D. T. Delpy et al.,
“Estimation of optical pathlength through tissue from direct time of flight measurement,”
Phys. Med. Biol., 33 1433
–1442 https://doi.org/10.1088/0031-9155/33/12/008 PHMBA7 0031-9155
(1988).
Google Scholar
G. Gratton et al.,
“Shades of gray matter: noninvasive optical images of human brain reponses during visual stimulation,”
Psychophysiology, 32 505
–509 https://doi.org/10.1111/j.1469-8986.1995.tb02102.x PSPHAF 0048-5772
(1995).
Google Scholar
D. J. Cuccia et al.,
“Modulated imaging: quantitative analysis and tomography of turbid media in the spatial-frequency domain,”
Opt. Lett., 30 1354
–1356 https://doi.org/10.1364/OL.30.001354 OPLEDP 0146-9592
(2005).
Google Scholar
T. Durduran et al.,
“Diffuse optical measurement of blood flow, blood oxygenation, and metabolism in a human brain during sensorimotor cortex activation,”
Opt. Lett., 29 1766
–1768 https://doi.org/10.1364/OL.29.001766 OPLEDP 0146-9592
(2004).
Google Scholar
D. Borycki et al.,
“Interferometric near-infrared spectroscopy (iNIRS) for determination of optical and dynamical properties of turbid media,”
Opt. Express, 24 329
–354 https://doi.org/10.1364/OE.24.000329 OPEXFF 1094-4087
(2016).
Google Scholar
J. Sutin et al.,
“Time-domain diffuse correlation spectroscopy,”
Optica, 3 1006
–1013 https://doi.org/10.1364/OPTICA.3.001006
(2016).
Google Scholar
T. Dragojević et al.,
“Compact, multi-exposure speckle contrast optical spectroscopy (SCOS) device for measuring deep tissue blood flow,”
Biomed. Opt. Express, 10 322
–334 https://doi.org/10.1364/BOE.9.000322 BOEICL 2156-7085
(2018).
Google Scholar
W. Zhou et al.,
“Highly parallel, interferometric diffusing wave spectroscopy for monitoring cerebral blood flow dynamics,”
Optica, 5 518
–527 https://doi.org/10.1364/OPTICA.5.000518
(2018).
Google Scholar
E. James and S. Powell,
“Fourier domain diffuse correlation spectroscopy with heterodyne holographic detection,”
Biomed. Opt. Express, 11 6755
–6779 https://doi.org/10.1364/BOE.400525 BOEICL 2156-7085
(2020).
Google Scholar
J. Xu et al.,
“Interferometric speckle visibility spectroscopy (ISVS) for human cerebral blood flow monitoring,”
APL Photonics, 5 126102 https://doi.org/10.1063/5.0021988
(2020).
Google Scholar
W. Zhou et al.,
“Multi-exposure interferometric diffusing wave spectroscopy,”
Opt. Lett., 46 4498
–4501 https://doi.org/10.1364/OL.427746 OPLEDP 0146-9592
(2021).
Google Scholar
A. T. Eggebrecht et al.,
“Mapping distributed brain function and networks with diffuse optical tomography,”
Nat. Photonics, 8 448
–454 https://doi.org/10.1038/nphoton.2014.107 NPAHBY 1749-4885
(2014).
Google Scholar
M. D. Wheelock, J. P. Culver and A. T. Eggebrecht,
“High-density diffuse optical tomography for imaging human brain function,”
Rev. Sci. Instrum., 90 051101 https://doi.org/10.1063/1.5086809 RSINAK 0034-6748
(2019).
Google Scholar
M. S. Patterson, B. Chance and B. C. Wilson,
“Time resolved reflectance and transmittance for the noninvasive measurement of tissue optical properties,”
Appl. Opt., 28 2331
–2336 https://doi.org/10.1364/AO.28.002331 APOPAI 0003-6935
(1989).
Google Scholar
S. Fantini et al.,
“Frequency-domain multichannel optical detector for noninvasive tissue spectroscopy and oximetry,”
Opt. Eng., 34 32
–42 https://doi.org/10.1117/12.183988
(1995).
Google Scholar
D. J. Pine et al.,
“Diffusing wave spectroscopy,”
Phys. Rev. Lett., 60 1134
–1137 https://doi.org/10.1103/PhysRevLett.60.1134 PRLTAO 0031-9007
(1988).
Google Scholar
D. A. Boas and A. G. Yodh,
“Spatially varying dynamical properties of turbid media probed with diffusing temporal light correlation,”
J. Opt. Soc. Am. A, 14 192
–215 https://doi.org/10.1364/JOSAA.14.000192 JOAOD6 0740-3232
(1997).
Google Scholar
W. B. Baker et al.,
“Modified Beer–Lambert law for blood flow,”
Biomed. Opt. Express, 5 4053
–4075 https://doi.org/10.1364/BOE.5.004053 BOEICL 2156-7085
(2014).
Google Scholar
E. Buckley et al.,
“Diffuse correlation spectroscopy for measurement of cerebral blood flow: future prospects,”
Neurophotonics, 1 011009 https://doi.org/10.1117/1.NPh.1.1.011009
(2014).
Google Scholar
S. Samaei et al.,
“Time-domain diffuse correlation spectroscopy (TD-DCS) for noninvasive, depth-dependent blood flow quantification in human tissue in vivo,”
Sci. Rep., 11 1817 https://doi.org/10.1038/s41598-021-81448-5 SRCEC3 2045-2322
(2021).
Google Scholar
A. G. Yodh, P. D. Kaplan and D. J. Pine,
“Pulsed diffusing-wave spectroscopy: high resolution through nonlinear optical gating,”
Phys. Rev. B, 42 4744
–4747 https://doi.org/10.1103/PhysRevB.42.4744
(1990).
Google Scholar
D. Huang et al.,
“Optical coherence tomography,”
Science, 254 1178
–1181 https://doi.org/10.1126/science.1957169 SCIEAS 0036-8075
(1991).
Google Scholar
K. K. Bizheva, A. M. Siegel and D. A. Boas,
“Path-length-resolved dynamic light scattering in highly scattering random media: the transition to diffusing wave spectroscopy,”
Phys. Rev. E, 58 7664
–7667 https://doi.org/10.1103/PhysRevE.58.7664
(1998).
Google Scholar
G. Popescu and A. Dogariu,
“Optical path-length spectroscopy of wave propagation in random media,”
Opt. Lett., 24 442
–444 https://doi.org/10.1364/OL.24.000442 OPLEDP 0146-9592
(1999).
Google Scholar
A. Wax et al.,
“Path-length-resolved dynamic light scattering: modeling the transition from single to diffusive scattering,”
Appl. Opt., 40 4222
–4227 https://doi.org/10.1364/AO.40.004222 APOPAI 0003-6935
(2001).
Google Scholar
M. A. Choma et al.,
“Sensitivity advantage of swept source and Fourier domain optical coherence tomography,”
Opt. Express, 11 2183
–2189 https://doi.org/10.1364/OE.11.002183 OPEXFF 1094-4087
(2003).
Google Scholar
L. Mei, G. Somesfalean and S. Svanberg,
“Frequency-modulated light scattering interferometry employed for optical properties and dynamics studies of turbid media,”
Biomed. Opt. Express, 5 2810
–2822 https://doi.org/10.1364/BOE.5.002810 BOEICL 2156-7085
(2014).
Google Scholar
J. M. Tualle et al.,
“Time-resolved measurements from speckle interferometry,”
Opt. Lett., 30 50
–52 https://doi.org/10.1364/OL.30.000050 OPLEDP 0146-9592
(2005).
Google Scholar
K. Zarychta et al.,
“Time-resolved diffusing wave spectroscopy with a CCD camera,”
Opt. Express, 18 16289
–16301 https://doi.org/10.1364/OE.18.016289 OPEXFF 1094-4087
(2010).
Google Scholar
C. Gregori-Pla et al.,
“Microvascular cerebral blood flow fluctuations in association with apneas and hypopneas in acute ischemic stroke,”
Neurophotonics, 6 025004 https://doi.org/10.1117/1.NPh.6.2.025004
(2019).
Google Scholar
D. R. Busch et al.,
“Detection of brain hypoxia based on noninvasive optical monitoring of cerebral blood flow with diffuse correlation spectroscopy,”
Neurocrit. Care, 30 72
–80 https://doi.org/10.1007/s12028-018-0573-1
(2019).
Google Scholar
W. Zhou et al.,
“Functional interferometric diffusing wave spectroscopy of the human brain,”
Sci. Adv., 7 eabe0150 https://doi.org/10.1126/sciadv.abe0150 STAMCV 1468-6996
(2021).
Google Scholar
M. Wu et al,
“Improved accuracy of cerebral blood flow quantification in the presence of systemic physiology cross-talk using multi-layer Monte Carlo modeling,”
Neurophotonics, 8 015001 https://doi.org/10.1117/1.NPh.8.1.015001
(2021).
Google Scholar
J. Selb et al.,
“Sensitivity of near-infrared spectroscopy and diffuse correlation spectroscopy to brain hemodynamics: simulations and experimental findings during hypercapnia,”
Neurophotonics, 1 015005 https://doi.org/10.1117/1.NPh.1.1.015005
(2014).
Google Scholar
M. Zhao et al.,
“Parallel interferometric diffusing wave spectroscopy (iDWS) with time-of-flight discrimination,”
in Biophotonics Congr.: Biomed. Opt. 2022 (Translational, Microscopy, OCT, OTS, BRAIN), Technical Digest Series,
BW1C.3
(2022). https://doi.org/10.1364/BRAIN.2022.BW1C.3 Google Scholar
D. Borycki, O. Kholiqov and V. J. Srinivasan,
“Reflectance-mode interferometric near-infrared spectroscopy quantifies brain absorption, scattering, and blood flow index in vivo,”
Opt. Lett., 42 591
–594 https://doi.org/10.1364/OL.42.000591 OPLEDP 0146-9592
(2017).
Google Scholar
O. Kholiqov et al.,
“Time-of-flight resolved light field fluctuations reveal deep human tissue physiology,”
Nat. Commun., 11 391 https://doi.org/10.1038/s41467-019-14228-5 NCAOBW 2041-1723
(2020).
Google Scholar
M. Robinson et al.,
“Interferometric diffuse correlation spectroscopy improves measurements at long source–detector separation and low photon count rate,”
J. Biomed. Opt., 25 097004 https://doi.org/10.1117/1.JBO.25.9.097004 JBOPFO 1083-3668
(2020).
Google Scholar
O. Kholiqov et al.,
“Scanning interferometric near-infrared spectroscopy,”
Opt. Lett., 47 110
–113 https://doi.org/10.1364/OL.443533 OPLEDP 0146-9592
(2022).
Google Scholar
G. Dietsche et al.,
“Fiber-based multispeckle detection for time-resolved diffusing-wave spectroscopy: characterization and application to blood flow detection in deep tissue,”
Appl. Opt., 46 8506
–8514 https://doi.org/10.1364/AO.46.008506 APOPAI 0003-6935
(2007).
Google Scholar
P. N. Pusey and W. Van Megen,
“Dynamic light scattering by non-ergodic media,”
Phys. A: Stat. Mech. Appl., 157 705
–741 https://doi.org/10.1016/0378-4371(89)90063-0 PHYADX 0378-4371
(1989).
Google Scholar
D. Borycki, O. Kholiqov and V. J. Srinivasan,
“Interferometric near-infrared spectroscopy directly quantifies optical field dynamics in turbid media,”
Optica, 3 1471
–1476 https://doi.org/10.1364/OPTICA.3.001471
(2016).
Google Scholar
E. James, S. Powell and P. Munro,
“Performance optimisation of a holographic Fourier domain diffuse correlation spectroscopy instrument,”
Biomed. Opt. Express, 13 3836
–3853 https://doi.org/10.1364/BOE.454346 BOEICL 2156-7085
(2022).
Google Scholar
A. M. Safi and A. B. Parthasarathy,
“Quantitative blood flow imaging at photon shot noise limit with heterodyne synthetic multi-exposure laser speckle imaging,”
in Biophotonics Congr.: Biomed. Opt. 2022 (Translational, Microscopy, OCT, OTS, BRAIN), Technical Digest Series,
OM4D.5
(2022). https://doi.org/10.1364/OTS.2022.OM4D.5 Google Scholar
R. Höhler, V. Labiausse and S. Cohen-Addad,
“High-resolution diffusing-wave spectroscopy using optimized heterodyne detection,”
J. Opt. Soc. Am. A, 20 2179
–2184 https://doi.org/10.1364/JOSAA.20.002179 JOAOD6 0740-3232
(2003).
Google Scholar
X. Cheng et al.,
“Time domain diffuse correlation spectroscopy: modeling the effects of laser coherence length and instrument response function,”
Opt. Lett., 43 2756
–2759 https://doi.org/10.1364/OL.43.002756 OPLEDP 0146-9592
(2018).
Google Scholar
A. M. Safi et al.,
“Quantitative measurement of static and dynamic tissue optical properties with continuous wave pathlength resolved diffuse correlation spectroscopy,”
in Biophotonics Congress 2021, OSA Technical Digest,
BTh1B.6
(2021). https://doi.org/10.1364/BRAIN.2021.BTh1B.6 Google Scholar
A. Torricelli et al.,
“Time-resolved reflectance at null source-detector separation: improving contrast and resolution in diffuse optical imaging,”
Phys. Rev. Lett., 95 078101 https://doi.org/10.1103/PhysRevLett.95.078101 PRLTAO 0031-9007
(2005).
Google Scholar
A. Torricelli et al.,
“Time domain functional NIRS imaging for human brain mapping,”
NeuroImage, 85 28
–50 https://doi.org/10.1016/j.neuroimage.2013.05.106 NEIMEF 1053-8119
(2014).
Google Scholar
O. Kholiqov, D. Borycki and V. J. Srinivasan,
“Interferometric near-infrared spectroscopy (iNIRS): performance tradeoffs and optimization,”
Opt. Express, 25 28567
–28589 https://doi.org/10.1364/OE.25.028567 OPEXFF 1094-4087
(2017).
Google Scholar
C. P. Valdes et al.,
“Speckle contrast optical spectroscopy, a non-invasive, diffuse optical method for measuring microvascular blood flow in tissue,”
Biomed. Opt. Express, 5 2769
–2784 https://doi.org/10.1364/BOE.5.002769 BOEICL 2156-7085
(2014).
Google Scholar
K. Murali and H. M. Varma,
“Multi-speckle diffuse correlation spectroscopy to measure cerebral blood flow,”
Biomed. Opt. Express, 11 6699
–6709 https://doi.org/10.1364/BOE.401702 BOEICL 2156-7085
(2020).
Google Scholar
R. Bi et al.,
“Fast pulsatile blood flow measurement in deep tissue through a multimode detection fiber,”
J. Biomed. Opt., 25 055003 https://doi.org/10.1117/1.JBO.25.5.055003 JBOPFO 1083-3668
(2020).
Google Scholar
E. G. Abu Jawdeh et al.,
“Noncontact optical imaging of brain hemodynamics in preterm infants: a preliminary study,”
Phys. Med. Biol., 65 245009 https://doi.org/10.1088/1361-6560/abc5a7 PHMBA7 0031-9155
(2020).
Google Scholar
S. Samaei, K. Nowacka and D. Borycki,
“Parallel interferometric near-infrared spectroscopy (piNIRS) for high temporal resolution blood flow monitoring,”
in Biophotonics Congr.: Biomed. Opt. 2022 (Transl., Microsc., OCT, OTS, BRAIN), Technical Digest Series,
OTu4D.4
(2022). https://doi.org/10.1364/OTS.2022.OTu4D.4 Google Scholar
F. T. S. Yu et al.,
“Analysis of a fiber specklegram sensor by using coupled-mode theory,”
Appl. Opt., 34 3018
–3023 https://doi.org/10.1364/AO.34.003018 APOPAI 0003-6935
(1995).
Google Scholar
S. Carp et al.,
“Diffuse correlation spectroscopy measurements of blood flow using 1064 nm light,”
J. Biomed. Opt., 25 097003 https://doi.org/10.1117/1.JBO.25.9.097003 JBOPFO 1083-3668
(2020).
Google Scholar
N. Ozana et al.,
“Superconducting nanowire single-photon sensing of cerebral blood flow,”
Neurophotonics, 8 035006 https://doi.org/10.1117/1.NPh.8.3.035006
(2021).
Google Scholar
C.-S. Poon et al.,
“First-in-clinical application of a time-gated diffuse correlation spectroscopy system at 1064 nm using superconducting nanowire single photon detectors in a neuro intensive care unit,”
Biomed. Opt. Express, 13 1344
–1356 https://doi.org/10.1364/BOE.448135 BOEICL 2156-7085
(2022).
Google Scholar
M. Robinson et al.,
“High framerate, InGaAs camera for interferometric diffuse correlation spectroscopy (iDCS) beyond the water peak,”
Proc. SPIE, 11641 116410O https://doi.org/10.1117/12.2578894 PSISDG 0277-786X
(2021).
Google Scholar
A. Biswas et al.,
“Fast diffuse correlation spectroscopy with a low-cost, fiber-less embedded diode laser,”
Biomed. Opt. Express, 12 6686
–6700 https://doi.org/10.1364/BOE.435136 BOEICL 2156-7085
(2021).
Google Scholar
W. Liu et al.,
“Fast and sensitive diffuse correlation spectroscopy with highly parallelized single photon detection,”
APL Photonics, 6 026106 https://doi.org/10.1063/5.0031225
(2021).
Google Scholar
E. Sie et al.,
“High-sensitivity multispeckle diffuse correlation spectroscopy,”
Neurophotonics, 7 035010 https://doi.org/10.1117/1.NPh.7.3.035010
(2020).
Google Scholar
BiographyWenjun Zhou is a professor of optical engineering at the College of Optical and Electronic Technology of China Jiliang University. From 2016 to 2021, he worked with Prof. Vivek Srinivasan in the Department of Biomedical Engineering at the University of California Davis as a postdoctoral fellow and an assistant project scientist. His current research interests include interferometric diffuse optics and fiber-optic biomedical sensors. Mingjun Zhao received her BE degree and PhD in biomedical engineering from Tianjin University in 2012 and the University of Kentucky in 2019, respectively. She is a postdoctoral fellow at New York University Langone Health. From 2020 to 2021, she worked at the Department of Biomedical Engineering at the University of California Davis as a postdoctoral fellow. Her current research interests include interferometric diffuse optics and its applications in human brain measurements. Vivek J. Srinivasan is an associate professor of radiology and ophthalmology at the NYU Grossman School of Medicine in the Tech4Health Institute. His group develops new light-based technologies for in vivo imaging and sensing of the brain and eye. Starting with a firm grounding in neurophysiology and biomedical engineering, his group employs innovative photonic technologies and approaches to accomplish this goal. |