12 September 2019 Multichannel fiber photometry for mapping axonal terminal activity in a restricted brain region in freely moving mice
Han Qin, Jian Lu, Wenjun Jin, Xiaowei Chen, Ling Fu
Author Affiliations +
Abstract

Fiber photometry has been increasingly popular in neuroscience research in freely behaving animals. In combination with genetically encoded calcium indicators, it allows for real-time monitoring of neural activity in neuronal somata, dendrites, and axonal terminals. We developed a multichannel fiber photometry device to map the activity of axonal terminals in a restricted, 100-μm-wide brain region in freely moving mice. This device consists of four bundled multimode fibers, each with a 50-μm core diameter and a scientific complementary metal-oxide semiconductor camera to simultaneously acquire fluorescence. We achieved a sampling rate of 100  frames  /  s and sufficient sensitivity to acquire data from axonal terminals. To avoid interference with neighboring channels, the recording depth of each channel was restricted to <250  μm. Furthermore, the small-core-diameter fibers did not restrict mouse locomotion. Using the Ca2  +   indicator GCaMP5G, we validated the system by recording Ca2  +   signals in axonal terminals from the medial entorhinal cortex layer II to the hippocampal dentate gyrus (DG) in freely moving mice. We detected spatially separated Ca2  +   signals at four different sites in the DG. Therefore, our multichannel fiber photometry device provides a simple but powerful method to functionally map axonal terminals in spatially confined brain areas of freely moving animals.

1.

Introduction

Axons transmit information from one neuron to another mainly through synapses at their terminals. Recording from axonal terminals in freely moving animals is an essential step to understand information processing during animal behaviors. A certain type of neuron in one area often projects to many different downstream regions with varying brain functions.1 The function and neural dynamics of the subdomains projected from the same cell population, with a distance of hundreds of micrometers, may also be different. For example, the retinal ganglion cells project to distant laminae zones of lateral geniculate nucleus and superior colliculus, which participate in different brain functions.2,3 Additionally, dopaminergic axons from the ventral tegmental area (VTA) project to both the nucleus accumbens core and shell, which play different roles in reward.4,5 Furthermore, the axonal terminals from the medial entorhinal cortex layer II (MECII) project to the entire region of the hippocampal dentate gyrus (DG), whose different subdomains may have different functions. Therefore, a method to record simultaneously from multiple sites in subdomains of an axonal projection target is greatly needed.

However, monitoring the activity of axonal terminals at multiple sites in freely behaving animals is challenging. Owing to their tiny sizes and high densities, recording the activity of axonal terminals using traditional electrophysiology is not feasible, especially in freely behaving animals.1,6,7 This situation changed when calcium (Ca2+) imaging techniques, such as two-photon Ca2+ imaging, were developed, which enabled in vivo Ca2+ activity to be monitored at the micrometer scale.8,9 With the help of transgenetic or viral infection-based expression of genetically encoded Ca2+ indicators,10,11 two-photon imaging is particularly suited to record from superficial brain regions, like cerebral cortex, in anesthetized or head-fixed animals.1214 Under these head-fixed conditions, two-photon imaging has been previously used to image neuronal activities at both axon buttons15,16 and dendritic spines.17,18 Two-photon imaging has also been adapted to record from multiple sites or large volumes;19,20 however, these applications are restricted to superficial brain regions of head-fixed animals as well. Recently developed miniaturized microscopes equipped with gradient refractive index (GRIN) lenses have been increasingly used in functional studies of deep brain regions in freely behaving animals. Dendrites and spines at significant tissue depths can even be recorded using a two-photon miniaturized microscope.21 However, this kind of technology has a restricted recording range in an individual brain region. Moreover, the use of the GRIN lens, with a typical diameter of 500  μm or 1 mm, causes larger invasion.22

Although it has poorer spatial resolution compared to the imaging approaches described above, fiber photometry can be used to record the activity of a group of neuronal cell bodies or axonal terminals in many brain regions in freely moving animals.1,7,2331 Combined with genetic tools, one can perform fiber photometry recordings in axonal terminals in a cell-type and long-term manner. For example, Gunaydin et al.1 showed that the activity patterns of the axonal terminals from VTA to nucleus accumbens encoded and predicted key features of social behavior. We recently used this method to measure axonal terminal activity in the MECII-DG projection during place learning tasks.7 However, we were unable to determine whether the spatially distributed signals in the subdomains of the MECII-DG projection were homogenous or heterogeneous. To answer this question, we need to simultaneously record from the subdomains of the axonal terminals in freely behaving animals.

We previously developed a multichannel fiber photometry device that allowed us to record from three freely moving mice simultaneously.32 A recent study extended the recording channel up to seven channels and used it to record simultaneously from different brain regions receiving VTA axonal projections.33 Even with these advances, there is no multichannel fiber photometry device that can be used to simultaneously record from the subdomains of an axonal projection within a distance of only a few hundreds of micrometers. These studies used separated fiber ferrules and implanted each channel independently, which made the minimal distance of each neighboring channels more than 1.25 mm. In addition, the fiber core diameter of these previous studies was 200  μm32 or 400  μm,33 which restricts the minimum distance without cross talk between neighboring channels. Furthermore, using optical fibers with larger diameters would cause greater mechanical stress and affect the locomotion of animals.

To enable simultaneous recordings from multiple axonal terminals in a restricted area such as the DG, we modified the fiber photometry device by adding a confined and integrated fiber probe with the neighboring channel close to 245  μm to map axonal terminals. Equipped with a 50-μm core diameter fiber, the recording depth was restricted to <250  μm and the mechanical constraint was reduced to a level that minimally affected mouse locomotion. In combination with a genetically encoded Ca2+ indicator, GCaMP5G, we then used this multichannel fiber photometry system to record neuronal activity from GCaMP5G-expressing axons projecting from MECII to the DG in freely moving mice. We were able to record from four sites in the MECII-DG projection. We demonstrate a long-term and stable recording of the neuronal activity from the axonal terminals of this projection when the mice are freely exploring. Therefore, this work provides an effective approach to functionally map axonal terminals within a spatially confined brain region. In the future, the multichannel fiber photometry device can easily be extended to more than four channels.

2.

Methods

2.1.

Multichannel Fiber Photometry Setup

The framework of our custom-built multichannel fiber photometry system is shown in Fig. 1(a). Four individual multimode fibers were integrated into a bundle and then glued into a subminiature version A connector. The excitation light from a 488-nm semiconductor laser (Coherent, OBIS 488 LX, 50 mW) was coupled into each fiber through an objective (Olympus, Plan N, 10×, NA 0.25) after being collimated with a beam expander [Fig. 1(a)]. The emitted fluorescence was separated from the excitation light through a dichroic mirror (Semrock, Di02-R488) and a bandpass emission filter (Semrock, FF01-535/22). The images of the fiber end faces were captured by a scientific complementary metal-oxide semiconductor (sCMOS) camera (Hamamatsu, C11440-22C) with a sampling frequency of 100  frames/s [Figs. 1(b) and 1(c)]. The image of each fiber end-face was circled as a region of interest (ROI) and the averaged fluorescent intensity of each ROI was acquired as Ca2+ signal for each channel. The control of light excitation and the data acquisition were implemented by a custom-written software on the LabVIEW platform (National Instruments).

Fig. 1

The construction of multichannel fiber photometry system. (a) Overview of the multichannel fiber photometry. Light beam from a 488-nm laser is coupled into all multimode fibers simultaneously by an objective. Then the emitted fluorescence is collected by the same optical fibers and detected by a sCMOS camera. DM, dichroic mirror. (b), (c) Images of the fiber end faces on sCMOS camera for the dual-channel mode (b) and four-channel mode (c).

NPH_6_3_035011_f001.png

2.2.

Fabrication of Fiber Probe

We utilized multimode 50-μm diameter fibers (Thorlabs, FG050UGA, NA: 0.22, core diameter: 50  μm, cladding diameter: 125  μm, and coating diameter: 245  μm), instead of the 200-μm diameter or 400-μm diameter ones that have been used previously.32,33 Optical fibers were cut to create flat end faces and arranged into the shape of a square. Then a metal canula (0.62-mm inner diameter and 0.9-mm outer diameter) was used to bundle the fibers together [Fig. 2(a)]. To test the set-up, the laser beams from the fiber probe were illuminated onto a cardboard coated with fluorescent dye [Fig. 2(b)]. The power from each channel of the fiber probe was measured to confirm a uniform power distribution among the four channels [Fig. 2(c)]. The mean±standard deviation of power was 27.5±1.5  μW with 3-mW illumination from laser.

Fig. 2

The fabrication of multichannel fiber probe: (a) photograph showing the four-channel fiber probe which was arrayed into a form of square, (b) fluorescence image showing the excitation of fluorescent dye by multichannel fiber photometry, and (c) output power from the fiber probe as a function of laser power. Output power at the end of each fiber was measured showing a liner function of laser power and only slight difference among different channels.

NPH_6_3_035011_f002.png

2.3.

Animals

C57/BL6J adult (2 to 5 months old) male mice were used in our experiments. Mice were housed in groups under a 12-h light/dark cycle with free access to food and water. Mice were deprived of food 24 h before open field exploration. All procedures were performed in accordance with institutional animal welfare guidelines and were approved by the Third Military Medical University Animal Care and Use Committee.

2.4.

Virus Injection

As described previously,7,34 mice were anesthetized with 1% to 2% isoflurane in pure oxygen and placed on a stereotactic frame. A warm pad was used to keep the animals at a proper body temperature (36.5°C to 37.5°C). A small craniotomy with a diameter of 0.5  mm was made above CA1 (AP=2.0  mm and ML=2.0  mm) or MEC (AP=5.1  mm and ML=3.25  mm). A glass micropipette with an inner tip diameter of 10 to 20  μm was lowered down to the CA1 pyramidal layer or dorsal MEC to deliver the AAV2/9-syn-GCaMP5G virus. Approximately 100 nL of virus solution were injected into the brain tissue in 5  min. The micropipette was held in place for 5 min before retraction. Tissue glue (3M Animal Care Products, Vetbond) was used for bonding the scalp incision.

2.5.

Histology and Fluorescence Imaging

After neuronal activity recording, mice were transcardially perfused with 4% paraformaldehyde. Brains were then dehydrated with 15% sucrose in PBS solution before being cut into sagittal sections (50  μm). Brain sections were then mounted on glass slides and covered by coverslips with mounting medium (Dako, C056330-2). Images were acquired with a wide-field fluorescence microscope (Olympus, BX43) to confirm virus expression and the position of the fibers.

2.6.

Data Analysis

Ca2+ recording data were analyzed similar to previous work.7,23,28 The data were filtered with a Savitzky–Golay FIR smoothing filter (five side points and three polynomial order). Then Ca2+ signals were calculated as

Eq. (1)

Δf/f0=(ff0)/f0,
where {f0} is the average fluorescent intensity of baseline segments without transients. And then the signal-to-noise ratio (SNR) is:35

Eq. (2)

SNR=(Δf/f0)peak/σf,
where σf is the standard deviation of the baseline period. Events that exceeded an amplitude of 3 times the standard deviation of baseline were identified as Ca2+ events.

The correlation coefficient was calculated to evaluate the cross correlation of Ca2+ signals among different channels:

Eq. (3)

rxy=i=1n(xix¯)(yiy¯)i=1n(xix¯)2i=1n(yiy¯)2,
where x and y represent the Ca2+ signals from different channels, rxy is the correlation coefficient, x¯ and y¯ are the mean values.

To evaluate asynchronized activity across channels, we calculated the proportions of local Ca2+ signals and global Ca2+ signals. Ca2+ events in two channels were paired in turn for comparison. Ca2+ events that were detected in a single channel were considered as local Ca2+ events, while ones detected in both channels were considered global Ca2+ events. The ratios of local Ca2+ events and global Ca2+ events to the total number of Ca2+ events were calculated.

Data were plotted as mean+SEM in figures. Nonparametric tests were used for data comparison, including Wilcoxon rank-sum test for unpaired two-group test, Wilcoxon signed-rank test for paired two-group test, Kruskal–Wallis test for unpaired multigroup test and Friedman test for paired multigroup test. Results were considered significant if p<0.05.

2.7.

Ca2+ Event Detection

We first assessed the Ca2+ event detection efficiency of our multichannel fiber photometry system using 50-μm diameter fibers. Owing to weaker signals in axonal terminals than cell bodies, we compared the detection of signals recorded from axonal terminals with that from cell bodies by our system.1 To test this, we carried out simultaneous recording from cell bodies and axonal terminals in mouse CA1 pyramidal layer (Fig. 3). Adeno-associated virus (AAV) carrying GCaMP5G gene was injected into the CA1 pyramidal layer. One month after AAV-GCaMP5G injection, mice were anesthetized with 1% to 2% isoflurane in pure oxygen and placed on a stereotactic frame. Fiber #1 was implanted above the virus injection site (AP: 2.0  mm, ML: 2.0 mm, and DV: 1.2 mm) to record Ca2+ signals from cell bodies and fiber #2 was implanted 1  mm posterior to the injection site (AP: 3.0  mm, ML: 2.0 mm, DV: 1.2 mm) to record Ca2+ signals from axonal terminals [Fig. 3(a)]. The expression of virus and positions of fibers were confirmed by post hoc histology [Fig. 3(b)]. The Ca2+ signals from cell bodies and axonal terminals in freely behaving animals are shown in Fig. 3(c). The SNR of Ca2+ signals recorded from cell bodies and axonal terminals was not significantly different (13.3±1.5 for cell body and 15.3±2.0 for axon, p=0.79), suggesting that the sensitivity of our multichannel fiber photometry was sufficient for Ca2+ recording in axonal terminals [Fig. 3(d)].

Fig. 3

The SNR of multichannel fiber photometry system in axonal terminals recording. (a) Diagram showing virus injection in CA1 pyramidal layer and Ca2+ recording at the virus injection site as well as the axonal terminals. R, rostral; C, caudal; D, dorsal; V, ventral; M, medial; and L, lateral. (b) Post hoc histological image showing GCaMP5G expression and the fiber tracks. Scale bars, 100  μm. Inset: the magnification of the dashed white box area. (c) Examples of Ca2+ signals recorded from cell bodies and axons in freely behaving mouse. Bottom panel showing the magnification of the light blue area in the upper panel. (d) Summary of the SNR of Ca2+ signals recorded from cell bodies and axonal terminals (Wilcoxon rank-sum test, z=0.27, p=0.79, and n=41 trials).

NPH_6_3_035011_f003.png

2.8.

Recording under Different Laser Power

Photobleaching occurs when using too much excitation power. Thus a suitable excitation power intensity should be chosen to balance the SNR and photobleaching. We recorded Ca2+ signals from axonal terminals under different excitation powers. Continuous recordings of Ca2+ activities from axonal terminals under different power densities at fiber tip (15.1, 10.0, 4.8, 2.2, and 15.1  mW/mm2 in sequence) are shown in Fig. 4(a). SNR of Ca2+ activities increased with increasing excitation power densities [Fig. 4(b)]. Signal started to be distinguished from noise with a power density of 4.8  mW/mm2 at fiber tip (with an amplitude larger than 3*SD of baseline and mean±SEM:7.3±2.5). Then we tested the photobleaching effect under a power density of 15.1  mW/mm2. The recording was conducted continuously for 40 to 60 min [Fig. 4(c)]. The baseline decreased only 2.8% after a 40-min continuous recording. The amplitudes and frequency of Ca2+ transients remained stable [χ2=0.3, p=0.84; Fig. 4(d) and χ2=1.6, p=0.46; Fig. 4(e)]. Therefore, a power density of 15.1  mW/mm2 was chosen since it gave high SNR with minimal photobleaching.

Fig. 4

The SNR and photobleaching effect of the multichannel fiber photometry system under different excitation powers. (a) Examples of Ca2+ signals under different excitation power intensities. (b) Summary of the SNR of Ca2+ signals under different power densities (Wilcoxon signed-rank test, z=3.8, ***p<0.001, and n=20). Gray dashed line, 3 times standard deviation of baseline. (c) The photobleaching effect of the multichannel fiber photometry. A continuous recording of more than 40 min was made under a power density of 15.1  mW/mm2 at fiber tip. Three traces of Ca2+ signals with 20 min interval were plotted showing that the baseline decreased only 2.8% after a 40-min recording. Data were normalized to the baseline of the first trace. (d) Summary of the maximum ΔF in the three recording periods. Data were normalized to the first trace. (Friedman test, χ2=0.3, p=0.84, and n=20). (e) Summary of the frequency of Ca2+ transients in the three recording periods. (Friedman test, χ2=1.6, p=0.46, and n=20).

NPH_6_3_035011_f004.png

2.9.

Recording Range of Single Channel

We used 50-μm core diameter fibers for recording to improve the mapping accuracy and reduce the recording range. We evaluated the recording range of our multichannel fiber photometry by theoretical calculation and experimental measurement.

We simplified the fiber probe from four channels into two channels for theoretical calculation. First light beams from two fibers propagated as shown in the schematic diagram in Fig. 5(a). The minimum distance of neighboring channels was 245  μm, the coating diameter of fiber. Therefore, we calculated the propagating distance from fiber tips before two beams overlapped using the function below:

Eq. (4)

S=(D/2r0/2)/{tan[sin1(NAfiber/ntis)]},
where D was the distance of two fibers, r0 was the core diameter of fiber, NAfiber was the numerical aperture of fiber, and ntis=1.36 was the refractive index of gray matter.36 The calculation results indicated that two light beams overlapped after a propagation of 600  μm.

Fig. 5

Light propagating path from fiber tip and the cross talk from neighboring channel. (a) Diagram showing the propagating of laser beam from two neighboring fibers. (b) Geometric structure of laser beam after transmitted from fiber tip. (c) Attenuation curve of light intensity in different depths from tip (50-μm core diameter and NA=0.22). (d) Diagram showing the fiber probe configuration for cross talk measurement. (e) Color map showing the fiber core images from three channels in the two illumination modes. Control group, illumination provided by multichannel fiber photometry system. Single-channel illumination group, illumination provided by an individual fiber. (f) Summary of the fluorescent intensity in the two illumination modes at three different illumination powers. Data were normalized to the control group.

NPH_6_3_035011_f005.png

Light intensity attenuates when propagated in tissue owing to light absorption and light scattering. Light scattering was mainly considered here because only minimal light absorption occurs in brain tissue at the wavelength range of 400 to 900 nm.37 The light intensity after propagating in brain tissue could be simulated with the Kubelka–Munk model.36,38 The transmission coefficient T is a function of the tissue depth z, as shown:

Eq. (5)

T=1μsz+1,
where μs is the scatter coefficient, given as 11.2  mm1 here.36 The power density at the depth z is based on the energy conservation and geometry [Fig. 5(b)]:

Eq. (6)

IzI0=T(r0r0+ztanα)2,
where Iz and I0 are the power density at the depth of z and 0 from fiber tip, respectively, r0 is the core diameter of fiber, α=sin1(NAfiber/ntis) is the divergence angle of optical fiber. The attenuation curve was calculated and shown in Fig. 5(c). The calculation predicted that light intensity would decrease to 0.5% of that at the tip after a propagation of overlapping distance (600  μm).

Photons scatter in brain tissue which means that the fluorescence photon excited by one fiber may be scattered into the second fiber. Therefore, we conducted an in vivo experiment to directly measure the cross talk from neighboring channels. The experiment diagram is shown in Fig. 5(d). The fiber probe was slightly changed from previously described. Three channels were still connected to the multichannel fiber photometry system, whereas the last channel was individually attached to another 488 nm laser. The power at fiber tip was measured to guarantee a uniform light intensity. The fluorescent dye Oregon green 488 BAPTA-1 (OGB-1, 500  μM, O6807, Molecular Probes) was injected into C57 mouse cortex at the depth of 300  μm (300  nL). The fiber probe was implanted above the injection site. A normal recording where illumination was provided by the multichannel fiber photometry system was considered as control. The single-channel illumination group consisted of recording illumination only provided by the individual fiber. The images of three-fiber core end faces captured in the two illumination modes are shown in Fig. 5(e). The fluorescent intensities at three illumination power intensities are summarized in Fig. 5(f). The fluorescent intensity in the single-channel illumination group was 3.5%±0.6% (mean±standard deviation) compared to the control, indicating that the cross talk among channels was negligible.

We also tried to find the recording depth by in vivo Ca2+ recording in the axonal terminals of GCaMP5G-expressing mice. We carried out the recording as the fiber probe was advancing to the target in anaesthetized mice. The fiber probe was inserted into the brain tissue from the cranial window. The dura was carefully removed before implantation by fine forceps to confirm a smooth entry. Fiber probe was advanced with a speed of 50  μm per min. At the depth of 1.2 mm from dura, we started to record the Ca2+ signal. After a 10-min continuous recording, fiber probe was advanced by a depth of 50  μm. The accurate distance to the target was measured with post hoc histology [Fig. 6(a)]. A power density of 15  mW/mm2 at fiber tip was used. An example of recording at different depths was shown in Fig. 6(b), corresponding to the morphology in Fig. 6(a). The SNR of Ca2+ signals recorded at different depths is summarized [Fig. 6(c), n=7 mice]. The SNR increased as the fiber tip approached the target. Ca2+ signals could be distinguished from noise (with the amplitude larger than 3×SD) at 200 to 250  μm. Thus the Ca2+ recording depth was 200 to 250  μm in our recording condition. For contrast, we also measured the recording depth of the previously developed single-channel fiber photometry system [Fig. 6(d), n=7 mice]. We used a 200-μm diameter and 0.48 NA multimode fiber for detection. The power density at fiber tip was about 0.22  mW/mm2. The recording depth was measured to be 400 to 500  μm, which was similar to the results in previous studies.23,25 Then the recording volume of each single-channel can be calculated according to the volume of frustum of a cone [Fig. 5(b)]:

Eq. (7)

V=13π[{r0/2+ztan[sin1(NAfiber/ntis)]}2(z+z0)r02z0/4],
where z0=r0/2/tan[sin1(NAfiber/ntis)]. Calculations showed that the single-channel recording volume of our multichannel fiber photometry system was 1.13 to 1.73×106  μm3. Furthermore, the recording volumes of previously developed fiber photometry systems were 4.11 to 6.40×107  μm3 for 200-μm diameter, 0.48 NA fiber and 9.78 to 14.1×107  μm3 for 400-μm diameter, 0.48 NA fiber. Therefore, the recording volume of our multichannel fiber photometry system decreased to 1% to 3% of the previously developed ones.

Fig. 6

The recording depth of multichannel fiber photometry system. (a) Post hoc histology after Ca2+ recording; (b) examples of Ca2+ signals in four different depths [shown in (a)] relative to the axonal terminals in anesthetized state. (c) Summary of SNR of Ca2+ signals recorded by multichannel fiber photometry system in different depths. Gray dashed line, 3 times standard deviation of baseline (Kruskal–Wallis test, χ2=84.5, ***p<0.001, and n=28 recording sites from 7 mice). (d) Summary of SNR of Ca2+ signals recorded by single-channel fiber photometry system in different depths (200-μm diameter, 0.48 NA fiber, Kruskal–Wallis test, χ2=34.5, ***p<0.001, and n=7 mice).

NPH_6_3_035011_f006.png

2.10.

Mechanical Constraint

The optical fibers could impose a mechanical constraint to freely moving mice, affecting their ability to freely access the entire recording environment. To test for this, we recorded from freely behaving mice exploring a 1.2-m diameter round basin [Fig. 7(a)] with the Ca2+ signals and behavior video recorded simultaneously. The optical fibers we used were 4- to 5-m long, and the recording system was put at the north side of the basins [see the azimuth diagram in Fig. 7(e) inset]. Videos were captured at a frame rate of 50 Hz under infrared light illumination (Cannon, XA-25, 1920×1080  pixels). The location of the mouse was tracked using a custom MATLAB program [Fig. 7(b)] and 20-min recording of each mouse was used for analyses. Mice implanted with our multichannel fiber probe were formed multichannel group, whereas mice implanted with single-fiber patch cable (200  μm and 0.48 NA) were single-channel group. First, the mean locomotion speed was compared between multichannel group and single-channel group, showing a lower speed in the multichannel group [multichannel group: 2.7±0.16  cm/s, n=9 mice; single-channel group: 3.6±0.24  cm/s, n=10 mice, z=2.8, p<0.01, Fig. 7(c)]. To quantify the restriction by the optical fibers, we divided the basin into three regions with different diameters but the same area and named these: inner part, middle part, and outer part [Fig. 7(b)]. The time that mice stayed in these three regions is summarized in Fig. 7(d). We found no significant difference in the time spent in each of the three regions from multichannel group (black, χ2=2.0 and p=0.37). Additionally, mice in the single-channel group also showed similar time distribution in the three divided parts (gray, χ2=0.8 and p=0.67). We then divided the basin space into four quadrants located in four directions [Fig. 7(e), inset]. There was no significant difference in the time spent in each of the four quadrants in the multichannel group [χ2=4.1, p=0.25, Fig. 7(e) black bar] nor the control single-channel group [χ2=4.0, p=0.27, Fig. 7(e) gray bar]. This indicated that the fibers had no visible effect on the moving range of mouse.

Fig. 7

Mechanical constraint imposed by fibers. (a) Recording scene in a large circular basin with a diameter of 1.2 m. (b) Example of mouse trajectory after a freely exploring of 30  min. The circular basin was divided into three parts with an equal area but different diameters named as inner part, middle part, and outer part. (c) Summary of the locomotion speed in multichannel and single-channel group. Black, multichannel group, n=9 mice; gray, single-channel group, n=10 mice (Wilcoxon rank-sum test, z=2.8, **p<0.01). (d) Summary of the time distribution in the three different parts of multichannel and single-channel group (Friedman test, multichannel group: χ2=2.0, p=0.37, and n=9 mice; single-channel group: χ2=0.8, p=0.67, and n=10 mice). (e) Summary of the time distribution in the four quadrants of multichannel and single-channel fiber implantation group. Inset: diagram showing the quadrant division (Friedman test, multichannel group: χ2=2.0, p=0.37, and n=9 mice; single-channel group: χ2=0.8, p=0.67, and n=10 mice).

NPH_6_3_035011_f007.png

3.

Results

The MEC provides information to hippocampus from cerebral cortex and transmits processed messages back to cerebral cortex.39,40 The MECII-DG projection, the initial phase of the trisynaptic circuit (MECII-DG-CA3-CA1), transmits spatial information to DG.7,41,42 The neural representation at different subdomains of the MECII-DG projection during free exploration remains unknown. To address this question, we simultaneously recorded Ca2+ signals at multiple MECII-DG projection sites with our multichannel fiber photometry in freely exploring mice.

To monitor the Ca2+ activities in the MECII-DG projection, we first injected AAV-syn-GCaMP5G into MEC to label this pathway. A fiber probe with four channels was then implanted above the MECII-DG projection [Fig. 8(a)]. The expression of virus in the MECII-DG projection was confirmed by post hoc histology [Figs. 8(b) and 8(c)]. Most of the cells expressing GCaMP5G in the injection site were in MEC layer II [Fig. 8(b)]. The axonal terminals in DG were also labeled with GCaMP5G [Fig. 8(c)]. The configuration of the axonal terminals in DG from MECII was a U-shape in rostral-caudal direction [Fig. 8(c)]. We measured the width of the dorsal side of the MECII-DG projection (sagittal section), which was 64±2.4  μm [Fig. 8(d), n=10 mice]. The distance between the dorsal and ventral side of the projection was 260±22.2  μm [Fig. 8(e), n=7 mice]. Therefore, Ca2+ signals on the ventral side of the DG could not be detected if the fiber tip was above the dorsal side due to the recording depth of 200 to 250  μm.

Fig. 8

MECII-DG pathway labeled by GCaMP5G through virus injection. (a) Diagram showing virus injection in MECII and multiple sites Ca2+ recording at the axonal terminals site of stratum molecular of DG (ML, stratum molecular of DG). (b), (c) Post hoc histology showing GCaMP5G expression in the injection site MEC (b) and axonal terminals in DG (c). (d) Distribution of the width of ML (Gaussian fit, mean=64  μm, R2=0.96, and n=95 from 10 mice). (e) Distribution of distance from the dorsal edge to ventral edge of ML (n=25 from 7 mice).

NPH_6_3_035011_f008.png

One month after AAV-GCaMP5G injection, the prepared fiber probe was implanted through a cranial window to the DG (AP: 2.0  mm and ML: 2.0 mm) at a depth of 1.8  mm below the pia surface. The locations of the four individual channels were confirmed by post hoc histology [Fig. 9(a)]. After the multifiber probe was implanted above the MECII-DG projection, we recorded Ca2+ signals when the mouse was freely exploring in the open field. The open field test was performed in a 1.2-m diameter round basin [Fig. 7(a)]. Examples of Ca2+ signals recorded from four channels during free exploration in the open field are shown in Fig. 9(b). Ca2+ transients were observed in three of the four channels; the channel without a detectable signal was inserted through the projection [fiber #1 in Fig. 9(a)]. Asynchronized Ca2+ signals were detected in channel #3 and shown in Fig. 9(b).

Fig. 9

Multichannel fiber recording in MECII-DG projection in freely exploring mice. (a) Post hoc histology after multichannel fiber recording showing the positions of four fibers, scale bar 200  μm. Inset: the relative positions of recording sites. (b) Example showing Ca2+ signals of four individual channels in MECII-DG projection during freely exploring. (c) Comparison of the correlation coefficient recorded under anesthetized and freely exploring (Wilcoxon rank-sum test, z=6.0, ***p<0.001, and n=24 recording sites from 6 mice). (d) Pie plot showing the fractions of local events and global events from 6 mice. R, rostral; C, caudal; D, dorsal; V, ventral; M, medial; and L, lateral.

NPH_6_3_035011_f009.png

We calculated the correlation coefficient of the Ca2+ signals recorded from different channels in the anesthetized state and freely exploring state. We found that the correlation coefficient significantly decreased in the freely exploring state (0.65±0.04) compared to the anesthetized state (0.90±0.01, z=6.0, and p<0.001). These data coincided the previous results that neurons in neocortex, thalamus, and hippocampus show synchronous activity under anesthesia.4345 We then sought out the local Ca2+ events and global Ca2+ events from four channels (see details in Sec. 2.6). Data from six mice, 24 recording sites in total, were used for analysis. The mouse with only three active channels was not used for analyses. Local Ca2+ events accounted for 61% of all the events, while the remaining events were global [Fig. 9(c)]. These results demonstrated a heterogeneous activity pattern in the subdomains of the MECII-DG projection during free exploration. These results validate the use of our multichannel fiber photometry device for functional recording in the subdomains of the MECII-DG projection.

4.

Discussion

Two-photon Ca2+ imaging, which allows for subcellular structure resolution, is often used for projection-specific functional mapping.15 However, the mapping region is restricted to less than 200  μm×200  μm.15,46 Although the recording site can be extended using a dual-axis mode, the minimum distance of the sites was restricted to 1  mm by the mechanical components.19 In addition, two-photon microscopy is mainly restricted to brain regions on the surface of a head-fixed animal.

Our method provides a simple way to do real-time functional monitoring in axonal terminals at multiple sites in freely behaving mice. The detector in our multichannel fiber photometry system is highly sensitive. The sCMOS camera with low noise is as sensitive as a photodetector equipped with a lock-in amplifier for recording Ca2+ signals in vivo.33 Multichannel fiber photometry using a camera as the detector, rather than a photodetector, is much simpler and easier to extend compared with other kinds of multichannel fiber photometry.26,32 Multichannel fiber photometry in the early phase is a simple mixture of single-channel fiber photometry.26 This method needs one photodetector for each channel and a more complex light splitting and coupling path. This makes it difficult to extend. Scan-based multichannel fiber photometry requires a galvanometer to scan among different channels to couple light into different channels.32 Different channel number designs require different scanning modes and data acquisition programs. However, our image-based multichannel fiber photometry is much easier to extend because the coupling of different channels is simultaneous. When we increase the number of recording channels, there is no need to make any other changes except for replacing the fiber probe.

The ability of multimode fiber to collect fluorescence increases with the increase of core diameter due to a larger NA. In general, fibers with larger core diameters are more sensitive for Ca2+ recording. But the recording range, mechanical constraint and invasiveness increases with larger core diameter fiber. Here our system used optical fibers with a core diameter of 50  μm. This change led to a decrease in recording depth to <250  μm. However, mechanical constraint was reduced to a level that had little influence on the locomotion of the mice. Furthermore, the reduction of fiber core diameter still provided enough sensitivity and SNR to record from axonal terminals.

We also designed a fiber probe with four fibers glued in a single-metal canula. This design decreased the distance of neighboring fibers to the coating diameter of 245  μm. This design allowed for the recording of multiple sites within a close distance. By designing the probes according to the geometry of the recording site and creating denser fiber bundles, recording from dozens of channels is possible. The small size of the probe also saves space and lightens the weight of carrying.

Recently, researchers reported methods for multimode fiber-based imaging techniques.4749 However, there are several limitations that prevent the application of these methods to brain tissue. Owing to randomization, thick scattering samples such as brain tissue are not suitable for this method. Furthermore, the fiber must be kept static through the whole experiment, which is impossible when measuring from freely moving animals. A fiber bundle-based imaging technique has also been used in brain imaging.50,51 However, the invasion caused by implantation was extensive and the fiber bundle imposed a significant mechanical constraint on the animal. Despite the poor spatial resolution of fiber photometry, the ability to measure structures deep in the brain in freely behaving animals has made it a popular choice in neuroscience research.

5.

Conclusion

We designed a new version of multichannel fiber photometry that can be used for Ca2+ recording in subdomains of axonal terminals. Our method of using small-core diameter fibers reduced the recording range. It allowed the mice to freely explore the space without mechanical constraints and was sufficiently sensitive to detect Ca2+ from axonal terminals. The simultaneous light coupling strategy and compact design of the fiber probe make the system easy to extend. Finally, we validated the system by performing four-channel Ca2+ recording in the MECII-DG projection in freely moving mice. We found a heterogeneous neuronal activity pattern in the different subdomains of the MECII-DG projection during exploration. This system could have multiple applications in neuroscience research.

Disclosures

All authors declare no conflicts of interest.

Acknowledgments

This work was supported by the National Natural Science Foundation of China (Nos. 61890952, 81671106, and 31700933); “973Program” (No. 2015CB759500); the National Key Research and Development Program of China (No. 2016YFA0201403); and Science Fund for Creative Research Group of China (No. 61721092).

References

1. 

L. A. Gunaydin et al., “Natural neural projection dynamics underlying social behavior,” Cell, 157 1535 –1551 (2014). https://doi.org/10.1016/j.cell.2014.05.017 CELLB5 0092-8674 Google Scholar

2. 

J. J. Nassi and E. M. Callaway, “Parallel processing strategies of the primate visual system,” Nat. Rev. Neurosci., 10 360 –372 (2009). https://doi.org/10.1038/nrn2619 NRNAAN 1471-003X Google Scholar

3. 

I. J. Kim et al., “Laminar restriction of retinal ganglion cell dendrites and axons: subtype-specific developmental patterns revealed with transgenic markers,” J. Neurosci., 30 1452 –1462 (2010). https://doi.org/10.1523/JNEUROSCI.4779-09.2010 JNRSDS 0270-6474 Google Scholar

4. 

D. S. Zahm and J. S. Brog, “On the significance of subterritories in the ‘accumbens’ part of the rat ventral striatum,” Neuroscience, 50 751 –767 (1992). https://doi.org/10.1016/0306-4522(92)90202-D Google Scholar

5. 

X. Han et al., “Role of dopamine projections from ventral tegmental area to nucleus accumbens and medial prefrontal cortex in reinforcement behaviors assessed using optogenetic manipulation,” Metab. Brain Dis., 32 1491 –1502 (2017). https://doi.org/10.1007/s11011-017-0023-3 MBDIEE 0885-7490 Google Scholar

6. 

E. A. Rancz et al., “High-fidelity transmission of sensory information by single cerebellar mossy fibre boutons,” Nature, 450 1245 –1248 (2007). https://doi.org/10.1038/nature05995 Google Scholar

7. 

H. Qin et al., “A visual-cue-dependent memory circuit for place navigation,” Neuron, 99 47 –55.e4 (2018). https://doi.org/10.1016/j.neuron.2018.05.021 NERNET 0896-6273 Google Scholar

8. 

W. Denk, J. H. Strickler and W. W. Webb, “Two-photon laser scanning fluorescence microscopy,” Science, 248 73 –76 (1990). https://doi.org/10.1126/science.2321027 SCIEAS 0036-8075 Google Scholar

9. 

R. Yuste and W. Denk, “Dendritic spines as basic functional units of neuronal integration,” Nature, 375 682 –684 (1995). https://doi.org/10.1038/375682a0 Google Scholar

10. 

J. Akerboom et al., “Optimization of a GCaMP calcium indicator for neural activity imaging,” J. Neurosci., 32 13819 –13840 (2012). https://doi.org/10.1523/JNEUROSCI.2601-12.2012 JNRSDS 0270-6474 Google Scholar

11. 

T. W. Chen et al., “Ultrasensitive fluorescent proteins for imaging neuronal activity,” Nature, 499 295 –300 (2013). https://doi.org/10.1038/nature12354 Google Scholar

12. 

H. Jia et al., “Dendritic organization of sensory input to cortical neurons in vivo,” Nature, 464 1307 –1312 (2010). https://doi.org/10.1038/nature08947 Google Scholar

13. 

T. Komiyama et al., “Learning-related fine-scale specificity imaged in motor cortex circuits of behaving mice,” Nature, 464 1182 –1186 (2010). https://doi.org/10.1038/nature08897 Google Scholar

14. 

S. Sakadzic et al., “Two-photon microscopy measurement of cerebral metabolic rate of oxygen using periarteriolar oxygen concentration gradients,” Neurophotonics, 3 045005 (2016). https://doi.org/10.1117/1.NPh.3.4.045005 Google Scholar

15. 

L. Petreanu et al., “Activity in motor-sensory projections reveals distributed coding in somatosensation,” Nature, 489 299 –303 (2012). https://doi.org/10.1038/nature11321 Google Scholar

16. 

C. L. Cox et al., “Action potentials reliably invade axonal arbors of rat neocortical neurons,” Proc. Natl. Acad. Sci. U. S. A., 97 9724 –9728 (2000). https://doi.org/10.1073/pnas.170278697 Google Scholar

17. 

X. Chen et al., “Functional mapping of single spines in cortical neurons in vivo,” Nature, 475 501 –505 (2011). https://doi.org/10.1038/nature10193 Google Scholar

18. 

L. Guo et al., “Dynamic rewiring of neural circuits in the motor cortex in mouse models of Parkinson’s disease,” Nat. Neurosci., 18 1299 –1309 (2015). https://doi.org/10.1038/nn.4082 NANEFN 1097-6256 Google Scholar

19. 

J. Lecoq et al., “Visualizing mammalian brain area interactions by dual-axis two-photon calcium imaging,” Nat. Neurosci., 17 1825 –1829 (2014). https://doi.org/10.1038/nn.3867 NANEFN 1097-6256 Google Scholar

20. 

J. N. Stirman et al., “Wide field-of-view, multi-region, two-photon imaging of neuronal activity in the mammalian brain,” Nat. Biotechnol., 34 857 –862 (2016). https://doi.org/10.1038/nbt.3594 NABIF9 1087-0156 Google Scholar

21. 

W. Zong et al., “Fast high-resolution miniature two-photon microscopy for brain imaging in freely behaving mice,” Nat. Methods, 14 713 –719 (2017). https://doi.org/10.1038/nmeth.4305 1548-7091 Google Scholar

22. 

S. L. Resendez et al., “Visualization of cortical, subcortical and deep brain neural circuit dynamics during naturalistic mammalian behavior with head-mounted microscopes and chronically implanted lenses,” Nat. Protoc., 11 566 –597 (2016). https://doi.org/10.1038/nprot.2016.021 1754-2189 Google Scholar

23. 

H. Adelsberger, O. Garaschuk and A. Konnerth, “Cortical calcium waves in resting newborn mice,” Nat. Neurosci., 8 988 –990 (2005). https://doi.org/10.1038/nn1502 NANEFN 1097-6256 Google Scholar

24. 

C. Grienberger et al., “Sound-evoked network calcium transients in mouse auditory cortex in vivo,” J. Physiol., 590 899 –918 (2012). https://doi.org/10.1113/jphysiol.2011.222513 JPHYA7 0022-3751 Google Scholar

25. 

G. Cui et al., “Concurrent activation of striatal direct and indirect pathways during action initiation,” Nature, 494 238 –242 (2013). https://doi.org/10.1038/nature11846 Google Scholar

26. 

H. Adelsberger et al., “Local domains of motor cortical activity revealed by fiber-optic calcium recordings in behaving nonhuman primates,” Proc. Natl. Acad. Sci. U. S. A., 111 463 –468 (2014). https://doi.org/10.1073/pnas.1321612111 Google Scholar

27. 

Y. Chen et al., “Sensory detection of food rapidly modulates arcuate feeding circuits,” Cell, 160 829 –841 (2015). https://doi.org/10.1016/j.cell.2015.01.033 CELLB5 0092-8674 Google Scholar

28. 

Q. Zhang et al., “Locomotion-related population cortical Ca2+ transients in freely behaving mice,” Front. Neural Circuits, 11 24 (2017). https://doi.org/10.3389/fncir.2017.00024 Google Scholar

29. 

J. Yao et al., “A corticopontine circuit for initiation of urination,” Nat. Neurosci., 21 1541 –1550 (2018). https://doi.org/10.1038/s41593-018-0256-4 NANEFN 1097-6256 Google Scholar

30. 

L. Li et al., “In vivo fiber photometry of neural activity in response to optogenetically manipulated inputs in freely moving mice,” J. Innov. Opt. Health Sci., 10 1743001 (2017). https://doi.org/10.1142/S1793545817430015 Google Scholar

31. 

K. Simone et al., “Open-source, cost-effective system for low-light in vivo fiber photometry,” Neurophotonics, 5 025006 (2018). https://doi.org/10.1117/1.NPh.5.2.025006 Google Scholar

32. 

Q. Guo et al., “Multi-channel fiber photometry for population neuronal activity recording,” Biomed. Opt. Express, 6 3919 –3931 (2015). https://doi.org/10.1364/BOE.6.003919 BOEICL 2156-7085 Google Scholar

33. 

C. K. Kim et al., “Simultaneous fast measurement of circuit dynamics at multiple sites across the mammalian brain,” Nat. Methods, 13 325 –328 (2016). https://doi.org/10.1038/nmeth.3770 1548-7091 Google Scholar

34. 

J. Li et al., “Primary auditory cortex is required for anticipatory motor response,” Cereb. Cortex, 27 3254 –3271 (2017). https://doi.org/10.1093/cercor/bhx079 53OPAV 1047-3211 Google Scholar

35. 

G. J. Broussard et al., “In vivo measurement of afferent activity with axon-specific calcium imaging,” Nat. Neurosci., 21 1272 –1280 (2018). https://doi.org/10.1038/s41593-018-0211-4 NANEFN 1097-6256 Google Scholar

36. 

A. M. Aravanis et al., “An optical neural interface: in vivo control of rodent motor cortex with integrated fiberoptic and optogenetic technology,” J. Neural Eng., 4 S143 –S156 (2007). https://doi.org/10.1088/1741-2560/4/3/S02 1741-2560 Google Scholar

37. 

A. N. Yaroslavsky et al., “Optical properties of selected native and coagulated human brain tissues in vitro in the visible and near infrared spectral range,” Phys. Med. Biol., 47 2059 –2073 (2002). https://doi.org/10.1088/0031-9155/47/12/305 PHMBA7 0031-9155 Google Scholar

38. 

E. Stark, T. Koos and G. Buzsaki, “Diode probes for spatiotemporal optical control of multiple neurons in freely moving animals,” J. Neurophysiol., 108 349 –363 (2012). https://doi.org/10.1152/jn.00153.2012 JONEA4 0022-3077 Google Scholar

39. 

G. Buzsaki and E. I. Moser, “Memory, navigation and theta rhythm in the hippocampal-entorhinal system,” Nat. Neurosci., 16 130 –138 (2013). https://doi.org/10.1038/nn.3304 NANEFN 1097-6256 Google Scholar

40. 

M. B. Moser, D. C. Rowland and E. I. Moser, “Place cells, grid cells, and memory,” Cold Spring Harbor Perspect. Biol., 7 a021808 (2015). https://doi.org/10.1101/cshperspect.a021808 1943-0264 Google Scholar

41. 

N. M. van Strien, N. L. M. Cappaert and M. P. Witter, “The anatomy of memory: an interactive overview of the parahippocampal–hippocampal network,” Nat. Rev. Neurosci., 10 272 –282 (2009). https://doi.org/10.1038/nrn2614 NRNAAN 1471-003X Google Scholar

42. 

S.-J. Zhang et al., “Optogenetic dissection of entorhinal–hippocampal functional connectivity,” Science, 340 1232627 (2013). https://doi.org/10.1126/science.1232627 SCIEAS 0036-8075 Google Scholar

43. 

A. Luczak et al., “Sequential structure of neocortical spontaneous activity in vivo,” Proc. Natl. Acad. Sci. U. S. A., 104 347 –352 (2007). https://doi.org/10.1073/pnas.0605643104 Google Scholar

44. 

A. Stroh et al., “Making waves: initiation and propagation of corticothalamic Ca2+ waves in vivo,” Neuron, 77 1136 –1150 (2013). https://doi.org/10.1016/j.neuron.2013.01.031 NERNET 0896-6273 Google Scholar

45. 

M. A. Busche et al., “Rescue of long-range circuit dysfunction in Alzheimer’s disease models,” Nat. Neurosci., 18 1623 –1630 (2015). https://doi.org/10.1038/nn.4137 NANEFN 1097-6256 Google Scholar

46. 

N. J. Sofroniew et al., “A large field of view two-photon mesoscope with subcellular resolution for in vivo imaging,” eLife, 5 e14472 (2016). https://doi.org/10.7554/eLife.14472 Google Scholar

47. 

Y. Choi et al., “Scanner-free and wide-field endoscopic imaging by using a single multimode optical fiber,” Phys. Rev. Lett., 109 203901 (2012). https://doi.org/10.1103/PhysRevLett.109.203901 PRLTAO 0031-9007 Google Scholar

48. 

T. Cizmar and K. Dholakia, “Exploiting multimode waveguides for pure fibre-based imaging,” Nat. Commun., 3 1027 (2012). https://doi.org/10.1038/ncomms2024 NCAOBW 2041-1723 Google Scholar

49. 

M. Plöschner et al., “Multimode fibre: light-sheet microscopy at the tip of a needle,” Sci. Rep., 5 18050 (2015). https://doi.org/10.1038/srep18050 SRCEC3 2045-2322 Google Scholar

50. 

B. A. Flusberg et al., “High-speed, miniaturized fluorescence microscopy in freely moving mice,” Nat. Methods, 5 935 –938 (2008). https://doi.org/10.1038/nmeth.1256 1548-7091 Google Scholar

51. 

P. Miao et al., “Laser speckle contrast imaging of cerebral blood flow in freely moving animals,” J. Biomed. Opt., 16 090502 (2011). https://doi.org/10.1117/1.3625231 JBOPFO 1083-3668 Google Scholar

Biography

Han Qin received his undergraduate degree in optical information science and technology from Huazhong University of Science and Technology in 2013. He is a PhD candidate in optical engineering at Huazhong University of Science and Technology. His current research interests include bioinstrumentation, neurophotonics, and cognitive neuroscience.

Jian Lu received his MD in physiology from Third Military Medical University. His research interests include spatial navigation, learning, and memory.

Wenjun Jin received his PhD in physiology and biophysics from Peking University. His research interests include spatial navigation, learning, and memory. Major brain regions concerned are entorhinal cortex, hippocampus, and prefrontal cortex. For research, he mainly uses techniques like fiber recording, electrophysiology recording, and optogenetics.

Xiaowei Chen is a professor of physiology and neurobiology at Third Military Medical University. His group uses methods such as Ca2+ imaging, fiber recording, electrophysiology recording, and optogenetics manipulating to probe the neural circuits in cortexes and hippocampus during learning and memory.

Ling Fu is a professor of biomedical photonics at Huazhong University of Science and Technology. Her group develops two-photon high-resolution microscopy, fiber-based microendoscopy for brain imaging, and clinical diagnosis. Her lab also builds customer-designed Ca2+ imaging or recording systems for the neuroscience research demands.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Han Qin, Jian Lu, Wenjun Jin, Xiaowei Chen, and Ling Fu "Multichannel fiber photometry for mapping axonal terminal activity in a restricted brain region in freely moving mice," Neurophotonics 6(3), 035011 (12 September 2019). https://doi.org/10.1117/1.NPh.6.3.035011
Received: 1 April 2019; Accepted: 5 August 2019; Published: 12 September 2019
Lens.org Logo
CITATIONS
Cited by 14 scholarly publications and 1 patent.
Advertisement
Advertisement
KEYWORDS
Calcium

Photometry

Brain

Signal to noise ratio

Brain mapping

Signal detection

Luminescence

Back to Top