30 August 2017 Sensitivity improvement of midinfrared gas sensing system using single-wavelength quantum cascade laser and hollow-waveguide gas cell
Takashi Katagiri, Keisuke Yaegashi, Yuji Matsuura
Author Affiliations +
Abstract
The configuration of a gas analysis system using the combination of a single-wavelength quantum cascade laser and a hollow-optical-fiber gas cell was optimized to improve measurement sensitivity. On the basis of theoretical calculation, a 6-m looped fiber gas cell was introduced into the system to more sensitively measure the concentration of nitric monoxide (NO) gas. High-precision measurement is achieved by precisely scanning the wavelength of the laser around the sharp absorption peak of NO gas. Concentrations of NO gas as low as 100 ppb were successfully measured.

For applications including environment monitoring, industrial quality control, and breath analysis for medical diagnosis, detecting gases with a concentration of sub-ppm to ppb order is frequently necessary. To satisfy this necessity, a variety of gas sensors based on semiconductor devices, microstructured materials, microgravimetric or calorimetric sensors, and so on have been proposed and developed.1,2

Among these gas sensors, optical sensors offer the advantages of fast response and high sensitivity with minimal drift.3 In place of conventional multipass gas cells, hollow-core waveguides such as photonic crystal fibers (PCF),46 photonic bandgap fibers (PBGF),79 and glass-capillary hollow-optical fibers (HOF) with an inner metal coating10,11 have been used as a gas cell because a long optical path is obtained with an extremely small sample volume. Likewise, midinfrared absorption spectroscopy for detecting fundamental molecular vibration has received attention because of the development of quantum cascade lasers (QCLs), which emit much higher power than the lamp sources used so far. Midinfrared QCLs enable the development of a robust, compact, and highly sensitive gas-sensing system.1214 Various groups have developed gas-sensing systems by combining a midinfrared QCL and hollow-core waveguides.15,16 Table 1 summarizes the data shown in the literatures. In this letter, we propose a gas sensing system using an HOF as a gas cell and a single-wavelength QCL as a light source for robust, inexpensive, and highly sensitive gas sensing.

Table 1

Data shown in literatures that use a hollow waveguide as gas cell.

Fiber typeLiteratureLight sourceWavelengthFiber size (dia., length)Sample gasConcentration
PCFRef. 5ASE1530 nm62  μm, 25 mC2H20.5%
PBGFRef. 7LED1550, 1300 nm10  μm, 10 mC2H210 mbar
Ref. 8Tunable LD1400 to 1700 nm12 mmCH410 ppm
Ref. 9Super-cont. source3 to 3.8  μm50  μm, 5 to 7 mC2H20.9 ppm
HOFRef. 10FT-IR5.2  μm2 mm, 2 mNO<1  ppm
Ref. 11FT-IR13 to 15  μm2 mm, 0.5 mC6H6, C6H5CH3, C6H4(CH3)2<100  ppb
Ref. 15External-cavity QCL7.7 to 8.2  μm2 mm, 0.5 mCH3CH2Cl, CHCl3ppb order
Ref. 16External-cavity QCL7 to 7.8  μm0.5 mm, 3 mVOC’sppm order
This letterDFB-QCL5.2  μm2 mm, 6 mNO100 ppb

A previous study showed that there is an optimum length for an HOF used as a gas cell. It is determined by the transmission loss of the fiber and the absorption coefficient of the sample gas.10 Sub-ppm concentrations of nitric monoxide (NO) gas were successfully measured with a system comprising a Fourier-transfer infrared spectrometer (FT-IR) and a hollow-optical-fiber gas cell with a length of 2 m. Although most of the QCL-based gas-sensing systems proposed so far have used wavelength tunable lasers to obtain the absorption spectra of the target gas, such lasers are usually very expensive, and ones having an external cavity sometimes have environmental disturbance problems such as vibration. In this letter, we report the development of a robust, inexpensive, and highly sensitive gas-sensing system that uses a monolithic QCL that emits a single wavelength in the midinfrared region. High-precision measurement is achieved by precisely scanning the QCL wavelength around the sharp absorption peak of the target gas.

Figure 1 shows a schematic of the measurement setup. The QCL (Hamamatsu, L12015) is a pulsed distributed feedback (DFB) laser with an oscillating wavelength of 5.26  μm, an emission linewidth of 0.2  cm1, a peak power of 50 mW, and a repetition frequency of around 120 kHz. The emitting wavelength was calibrated using an FT-IR with a resolution of 0.125  cm1. The laser beam is focused on the input end of the HOF used as a gas cell by an aspheric ZnSe lens with a focal length of 4.8 mm, and the output beam from the fiber is detected by an InSb detector. The input end of the fiber is sealed by a metal sleeve with a ZnSe window, and the sample gas flows into the fiber through the sleeve with a flow rate of 100  ml/min and the bore of the fiber is held in atmospheric pressure.

Fig. 1

Schematic of measurement setup.

OE_56_8_080503_f001.png

We used an HOF with an inner diameter of 2 mm to reduce the coupling loss at the input end because the beam quality of a QCL is usually low, which causes a high coupling loss as a result of the generation of higher order modes in the fiber. The fiber uses flexible polycarbonate tubing as a base, and silver film is formed on the inside of the tubing.17 The fiber also has a dielectric thin film on the top of the silver film to reduce the loss at the target wavelengths caused by the optical interference effect of the dielectric thin layer. To form fibers longer than 2 m, we connected fibers with lengths of 1 and 2 m by using plastic sleeves, and the fiber was looped with a radius of 20 cm.

We first investigated the optimum length of the HOF. Because of the transmission loss of the fiber, there is an optimum length giving the highest signal-to-noise ratio (SNR) in a hollow-fiber-based gas-sensing system.10 Under the assumptions that the power of the input light is P0 and that the noise power spectrum is N, the SNR of the spectrum measured with an HOF with length l and an attenuation constant of αL is expressed as

Eq. (1)

SNR(λ)=P0(λ)[1e2α(λ)cl]e2αL(λ)lN(λ),
where α and c are the absorption constant and the concentration of sample gas, respectively. This equation shows that SNR depends on the absorption of the gas and the attenuation loss of the fiber because P0(λ)/N(λ) is inherent to the measurement system. From the results of a preliminary experiment, we found that the P0/N of our system was nearly constant in the 5.26-μm wavelength region and was around 1370. From this value and from the attenuation loss of the fiber, which is around 2.5  dB/m, we calculated the SNR as a function of the fiber length, and the result is shown in Fig. 2. In the calculation and the following experiments, we used NO as the sample gas because it is frequently used in atmospheric monitoring and breath analysis. We set the attenuation constant of NO at a wavelength of 5.26  μm to 5.68  cm1 on the basis of the results of the preliminary experiment. From this calculated result, we found that the maximum SNR is obtained with a fiber length of around 6.5 m. We also found that this optimum length does not change with the gas concentration when it is lower than 10 ppm. This optimum length should change if the attenuation constant αL is different from that of the fiber used in our system. For the fiber with lower attenuation constant, the optimum length should be larger.

Fig. 2

Theoretical SNR of analysis system for NO gas with 2-mm-diameter hollow-fiber gas cell as function of fiber length. Open circles are SNRs of measured spectra obtained with different fiber lengths.

OE_56_8_080503_f002.png

In the gas analysis experiment, to obtain the spectral shape of the absorption peak of NO gas, the oscillation wavenumber of the QCL was first scanned from 1903.1 to 1903.2  cm1 by changing the operating temperature of the Peltier-cooled QCL from 23.3°C to 24.3°C in 0.02°C steps. By using this wavelength-scan-based method, one can perform more precise measurement compared to common QCL-based methods using a fixed single wavelength because one can collect information on such as baseline fluctuation. Figure 3 shows the absorption peak spectra of 1-ppm NO gas measured with 1- and 6-m fibers and with nitrogen gas used as a background spectrum. The dots are measured data and the solid line is a Gaussian curve that was fitted to the measured data and it took around 3 min to obtain a single spectrum in the measurement. With this scanning range, only one absorption peak was observed at around 1903.14  cm1 for both fibers. We confirmed that this peak was one of the absorption peaks of NO by measurement using FT-IR. By comparing the data shown in Figs. 3(a) and 3(b), it is found that the SNR was much higher with the 6-m fiber. The measured SNRs for different fiber lengths are shown in Fig. 2. The SNR was maximized with a fiber length of 6 m, as expected from theoretical calculation although the measured SNRs were somewhat lower than the theoretical ones. This discrepancy may be due to the bending losses of the fibers and the coupling losses between the fibers connected by plastic sleeves.

Fig. 3

Absorption peak spectra of 1-ppm NO gas measured with fiber gas cells with lengths of (a) 1 and (b) 6 m.

OE_56_8_080503_f003.png

Figure 4 shows the calibration plots for NO gas measurement with different fiber lengths. The open circles are average intensities of the absorption peak for three measurements and it is seen that the absorbance linearly increased with the fiber length. From this result, we found that the measurement error for a 1-ppm concentration with the 6-m fiber gas cell was around 10%.

Fig. 4

Calibration plots for NO gas measurement with different fiber lengths.

OE_56_8_080503_f004.png

Figure 5 shows the absorption peak of 0.1-ppm NO gas measured with the 6-m gas cell. Although the SNR of measured data was lower than those shown in Fig. 3, a peak is seen at around 1903.15  cm1. We confirmed that the concentration as low as 100 ppb was successfully detected and this was the lowest detection limit of our measurement system.

Fig. 5

Absorption peak of 0.1-ppm NO gas measured with 6-m gas cell.

OE_56_8_080503_f005.png

In conclusion, optimization of the configuration of a gas analysis system using the combination of a single-wavelength DFB-QCL and a hollow-optical-fiber gas cell improved measurement sensitivity. The introduction of a 6-m looped fiber gas cell enabled measurement of sub-ppm concentrations of NO gas. High-precision measurement is achieved by precisely scanning the QCL wavelength around the sharp absorption peak of the target gas. Concentrations of NO gas as low as 100 ppb were successfully measured. However, in our present system, it takes around 3 min to obtain a single spectrum. This is mainly because the system should wait for the temperature being stabilized after changing it stepwise. Therefore, it is expected that the measuring time is drastically shortened by changing the temperature continuously. This is possible because the time for obtaining data for detecting power is very short (<10  ms) and we are now working on improvement of our system.

References

1. 

Gas Sensing Fundamentals, Springer Series on Chemical Sensors and Biosensors, Springer-Verlag, Heidelberg (2014). Google Scholar

2. 

Solid State Gas Sensing, Springer, New York (2009). Google Scholar

3. 

J. Hodgkinson and R. P. Tatam, “Optical gas sensing: a review,” Meas. Sci. Technol., 24 (1), 012004 (2013). http://dx.doi.org/10.1088/0957-0233/24/1/012004 MSTCEP 0957-0233 Google Scholar

4. 

M. Morshed et al., “Microstructure core photonic crystal fiber for gas sensing applications,” Appl. Opt., 54 (29), 8637 –8643 (2015). http://dx.doi.org/10.1364/AO.54.008637 APOPAI 0003-6935 Google Scholar

5. 

S. H. Kassani et al., “Suspended ring-core photonic crystal fiber gas sensor with high sensitivity and fast response,” IEEE Photon. J., 7 (1), 1 –9 (2015). http://dx.doi.org/10.1109/JPHOT.2015.2396121 Google Scholar

6. 

Q. He et al., “DLAS-WMS based near-infrared methane sensor system using hollow-core photonic crystal fiber as gas-chamber,” Opt. Quant. Electron., 49 115 (2017). http://dx.doi.org/10.1007/s11082-017-0946-2 OQELDI 0306-8919 Google Scholar

7. 

T. Ritari et al., “Gas sensing using air-guiding photonic bandgap fibers,” Opt. Express, 12 (17), 4080 –4087 (2004). http://dx.doi.org/10.1364/OPEX.12.004080 OPEXFF 1094-4087 Google Scholar

8. 

A. M. Cubillas et al., “Methane detection at 1670-nm band using a hollow-core photonic bandgap fiber and a multiline algorithm,” Opt. Express, 15 (26), 17570 –17576 (2007). http://dx.doi.org/10.1364/OE.15.017570 OPEXFF 1094-4087 Google Scholar

9. 

M. N. Petrovich et al., “High sensitivity gas detection using hollow core photonic bandgap fibres designed for mid-IR operation,” in Proc. IEEE Sensors, (2014). http://dx.doi.org/10.1109/ICSENS.2014.6985144 Google Scholar

10. 

Y. Takahashi et al., “Design and optimization of hollow-optical-fiber gas cell for infrared spectroscopy,” Opt. Eng., 52 (1), 013601 (2013). http://dx.doi.org/10.1117/1.OE.52.1.013601 Google Scholar

11. 

C. R. Young et al., “Infrared hollow waveguide sensors for simultaneous gas phase detection of benzene, toluene, and xylenes in field environments,” Anal. Chem., 83 (16), 6141 –6147 (2011). http://dx.doi.org/10.1021/ac1031034 ANCHAM 0003-2700 Google Scholar

12. 

L. Zhang et al., “Applications of absorption spectroscopy using quantum cascade lasers,” Appl. Spectrosc., 68 (10), 1095 –1107 (2014). http://dx.doi.org/10.1366/14-00001 APSPA4 0003-7028 Google Scholar

13. 

J. Li, U. Parchatka and H. Fischer, “Development of field-deployable real time QCL spectrometer for simultaneous detection of ambient N2O and CO,” Sens. Actuators B, 182 659 –667 (2013). http://dx.doi.org/10.1016/j.snb.2013.03.073 SABCEB 0925-4005 Google Scholar

14. 

Y. Cao et al., “Simultaneous atmospheric nitrous oxide, methane and water vapor detection with a single continuous wave quantum cascade laser,” Opt. Express, 23 (3), 2121 –2132 (2015). http://dx.doi.org/10.1364/OE.23.002121 OPEXFF 1094-4087 Google Scholar

15. 

C. Young et al., “External cavity widely tunable quantum cascade laser based hollow waveguide gas sensors for multianalyte detection,” Sens. Actuators B, 140 (1), 24 –28 (2009). http://dx.doi.org/10.1016/j.snb.2009.03.023 SABCEB 0925-4005 Google Scholar

16. 

J. M. Kriesel et al., “Versatile, ultra-low sample volume gas analyzer using a rapid, broad-tuning ECQCL and a hollow fiber gas cell,” Proc. SPIE, 10210 1021003 (2017). http://dx.doi.org/10.1117/12.2262612 PSISDG 0277-786X Google Scholar

17. 

M. Nakazawa et al., “Hollow polycarbonate fiber for Er:YAG laser light delivery,” Opt. Lett., 31 (10), 1373 –1375 (2006). http://dx.doi.org/10.1364/OL.31.001373 OPLEDP 0146-9592 Google Scholar
CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Takashi Katagiri, Keisuke Yaegashi, and Yuji Matsuura "Sensitivity improvement of midinfrared gas sensing system using single-wavelength quantum cascade laser and hollow-waveguide gas cell," Optical Engineering 56(8), 080503 (30 August 2017). https://doi.org/10.1117/1.OE.56.8.080503
Received: 7 June 2017; Accepted: 14 August 2017; Published: 30 August 2017
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
KEYWORDS
Quantum cascade lasers

Absorption

Signal to noise ratio

Gas lasers

Sensing systems

Signal attenuation

FT-IR spectroscopy

Back to Top