Open Access
26 April 2018 Study on verifying the angle measurement performance of the rotary-laser system
Jin Zhao, Yongjie Ren, Jiarui Lin, Shibin Yin, Jigui Zhu
Author Affiliations +
Abstract
An angle verification method to verify the angle measurement performance of the rotary-laser system was developed. Angle measurement performance has a great impact on measuring accuracy. Although there is some previous research on the verification of angle measuring uncertainty for the rotary-laser system, there are still some limitations. High-precision reference angles are used in the study of the method, and an integrated verification platform is set up to evaluate the performance of the system. This paper also probes the error that has biggest influence on the verification system. Some errors of the verification system are avoided via the experimental method, and some are compensated through the computational formula and curve fitting. Experimental results show that the angle measurement performance meets the requirement for coordinate measurement. The verification platform can evaluate the uncertainty of angle measurement for the rotary-laser system efficiently.

1.

Introduction

The rotary-laser system is a large-scale distributed system based on the intersection of laser beams from multiple stations. The rotary-laser system1 has been used accurately to measure large structures and has potential applications in robotics,2 aircraft, shipbuilding, the assembly of large components,3 and mobile platform calibrations.4 The system combines the principle of the photoelectric scanning with the concept of large-scale volume measurement.5 Coordinate calculation is conducted by the signal processor and terminal computer to obtain high-precision computation. The distributed measurement network allows multiple receivers to work simultaneously. The measurement range can be expanded by increasing the number of transmitters, and the expanded range can ensure that the uncertainty of local measurement meets that of the overall measurement at the same time. Accurate measurements could subsequently be used to automatically correct the processes and compensate for manufacturing errors.6

The rotary-laser system uses the principle of intersection to set up a plurality of plane equations through the angle acquired by the receiver. It is then possible for three-dimensional (3-D) coordinates to be calculated according to the space plane intersection constraint.5,7 In the absence of an accurate goniometric auxiliary device, angle measurement performance is difficult to assess.

There has been a large body of literature studying commercial rotary-laser systems, which include the indoor Global Positioning System (iGPS) of Nikon and the workshop Measurement Positioning System (wMPS) of Tianjin University. Studies have verified the performance of iGPS through comparison with points calibrated using a network of laser trackers8,9 and obtained coordinate measurement accuracy through the principle of isolating subsystems. Since there is reference significance for angle measurement performance of the rotary-laser system, there is an urgent need to establish a standard for the angle verification. Hirt et al.10 did some research on the accuracy of tilt measurements on a hexapod-based Digital Zenith Camera System (DZCS) and reached the conclusion that DZCS may be a suitable system for accurate tilt measurement. However, DZCS is unsuitable for the verification of the rotary-laser system. To analyze the angular uncertainties of rotary-laser automatic theodolite, the fundamental physics of the rotating head device may be used to propagate uncertainties using Monte Carlo simulation, and Muelaner et al.11 found that the uncertainty in azimuth is considerably higher than the uncertainty in elevation. Muelaner et al. carried out experiments to determine the actual uncertainty in the azimuth angle using the reference cylinder and dial gauge, but the dial gauge method is so complicated that the operator has to adjust the cylinder again and again when the dial gauge is moved to different positions. Furthermore, the verification accuracy can be influenced by a mechanical error of the cylinder. Geng et al.12 investigated the relationship among uncertainties of azimuth and the elevation angle and internal parameters of wMPS; the study provides a basis for the theoretical analysis of the angle verification. Chao et al.13 and Zhi et al.14 designed methods to verify the wMPS using the polyhedral prism and electronic autocollimator. The adjustment device added to the rotating head is so heavy that it may negatively influence the shafting, and the process is too complex to do in a single day to verify a transmitter. However, this previous research has been essential for informing this study.

The key to verifying the angle measurement performance is ensuring that the virtual axis of rotation of the rotor system is highly coaxial with the rotation axis of the angle standard device. Unfortunately, the work published11,13,14 for the verification of the rotary-laser system is not perfect enough. The objective of this paper is to devise an optimized verification method for the performance of the rotary-laser system. Multiple devices are integrated on a platform, and a verification system is devised. Additionally, this paper does in-depth research on the adjustment errors and system errors, finding that eccentricity contributes the most to the system through simulation. The error is compensated through the computational formula and curve fitting.

2.

Experimental Methods

This test was carried out on the platform for verification. An optimized method was used to reach the goal of the experiment.

The multitooth indexing rotary table was used to provide the angular standard, and the polyhedral prism, electronic autocollimator,15 and tilt-adjustment device were used for assistance. There was an even number of working faces on the polyhedral prism. The polyhedral prism was put on the transmitter carefully via a light-fixing device, and a locking device was used to make sure the rotating head of the transmitter did not rotate. The tilt-adjustment device had two rotational degrees of freedom. All the devices were integrated on a unit of measurement platform (see Fig. 1).

Fig. 1

Design of azimuth accuracy verification platform.

OE_57_4_044106_f001.png

The uncertainty in azimuth was obtained by comparing the measured azimuth angles with the rotation angles of the rotary table, and the uncertainty in elevation was obtained through the relationship between the azimuth angle and elevation angle. The maximum angle error of the rotary table was 0.5  arcsec, the manufacturing tolerance of the polyhedral prism was ±1  arcsec, and the angle measurement accuracy of the electronic autocollimator (TRIOPTICS TriAngle) was 0.4  arcsec.

The first procedure conducted was presetting the measurement device for the view angle of the electronic autocollimator to be <10  arcsec. The transmitter replica, which was the same height as the real transmitter, was placed onto the tilt-adjustment device. The pitch of the electronic autocollimator and the dip of the tilt-adjustment device were adjusted slightly so that the electronic autocollimator could get reflection images clearly at each working surface of the polyhedral prism.

The second procedure was to adjust the pitch of the autocollimator. The polyhedral prism was simply placed on the head of the transmitter via a connector. Since this may have caused the polyhedral prism and the rotating head of the transmitter not to be parallel to each other, the pitch angle was adjusted slightly until the horizontal-direction angle readings in the opposite working surfaces of the polyhedral prism were equal and the vertical-direction angle readings were opposite when the rotary table turned 180 deg (see Fig. 2) without changing the dip of the tilt-adjustment device. After rotating 90 deg and fine-tuning the pitch, the angle readings in each working surface of the polyhedral prism were in tolerance (shown in the tips in the end of this section). The vector of the electronic autocollimator projection and the axis of rotation of the rotary table were perpendicular to each other at the end of this procedure [see Fig. 3(a)].

Fig. 2

Diagram of exit rays of autocollimator.

OE_57_4_044106_f002.png

Fig. 3

Schematic diagram of axis adjustment process: (a) procedure two: pitch adjust and (b) procedure three: tilt adjust.

OE_57_4_044106_f003.png

The third procedure was replacing the transmitter replica with the real transmitter. The pitch of the electronic autocollimator was unchanged. The dip of the tilt-adjustment device was adjusted until the angle reading in the electronic autocollimator was the same as the second procedure. At this point, the axis of rotation of the rotary table and the axis of rotation of the transmitter are parallel to each other [see Fig. 3(b)].

The final procedure was placing the receiver about 10 m away from the transmitter and approximately in the x-O-y plane of the transmitter coordinate (z-value equals to zero). The data stream is provided by the processor.

Repeating the last two procedures was necessary only when another transmitter’s accuracy needed to be verified.

This experimental process is shown in Fig. 4.

Fig. 4

Image of experimental process: (a) electronic autocollimator pitch adjust and (b) transmitter tilt adjust.

OE_57_4_044106_f004.png

Tips: In the second and third procedures, when the horizontal-direction angle readings of the electronic autocollimator are <5  arcsec in the opposite working surfaces of the polyhedral prism, regard them as “equal.” When the sums of the vertical-direction angle readings are between 10  arcsec and 10  arcsec, regard them as “opposite.”

3.

Error Analysis

In Sec. 2, we described the procedure for adjusting the axis of rotation of the transmitter. However, the prework may not be perfect and the axis of rotation of the transmitter was not strictly parallel to the axis of the rotary table. Adjustment errors, such as axis tilt errors and eccentricity errors, may influence the final measurement uncertainty. When the verification experiment is carried out, the height of the receiver also influences the result.

3.1.

Axis Tilt Error

Axis tilt error represents the axis of rotation of the rotary table when it is not strictly parallel to the axis of rotation of the transmitter. The tilt angle of the transmitter rotation axis has an effect on the measurement azimuth angle. A base coordinate in the initial position of the rotary table is established as shown in Fig. 5, where α0 and β0 represent the horizontal angle and vertical angle of the transmitter rotation axis, respectively. The rotation matrix between the base coordinate and the transmitter coordinate is calculated as

Fig. 5

Schematic diagram of axis inclination of transmitter.

OE_57_4_044106_f005.png

Eq. (1)

RT=[RxRyRz]=RY·Rz=[cos(π2β0)0sin(π2β0)010sin(π2β0)0cos(π2β0)]·[cosα0sinα00sinα0cosα00001]=[cos(π2β0)·cosα0cos(π2β0)·sinα0sin(π2β0)sinα0cosα00sin(π2β0)·cosα0sin(π2β0)·sinα0cos(π2β0)]

Assume that the receiver’s position under the base coordinate is P=[XYZ]T and α0=0. The 3-D coordinates of the receiver in the transmitter coordinate are calculated as

Eq. (2)

P0=RT·P=[cos(π2β0)0sin(π2β0)010sin(π2β0)0cos(π2β0)]·[XYZ].

If the rotary table rotates at an angle θ, the transmitter coordinate also rotates at an angle θ around the z-axis of the base coordinate. The 3-D coordinates of the receiver under the transmitter coordinate are calculated as

Eq. (3)

P1=RT·[cosθsinθ0sinθcosθ0001]·P=[X1Y1Z1].

The measured azimuth angle of the transmitter can be calculated easily by geometric projection and is calculated as

Eq. (4)

α=arctan(Y1X1).

Assuming that the receiver is in the x-O-y plane and 10 m away from the transmitter, the angle error can be calculated when the tilt angle is given. As shown in Fig. 6, when the tilt angle of the transmitter rotation axis is 30  arcsec, the peak-to-valley error caused by the tilt is <0.003  arcsec. When the angle is reduced to 10  arcsec, the error becomes 0.0003  arcsec.

Fig. 6

Effects of the tilt angle.

OE_57_4_044106_f006.png

As a result, if the tilt angle is small enough, the azimuth angle error can be ignored. It is acceptable to adjust the tilt angle of the transmitter rotation axis to a level <10  arcsec.

3.2.

Eccentricity Error

The eccentricity error is caused by the axis of rotation of the transmitter, and the axis of rotation of the rotary table is not adjusted to a coaxial position.

This error is shown in Fig. 7(a). Eccentricity error may be compensated for when the mathematical model of the error is given according to the geometric relationship [see Fig. 7(b)].

Fig. 7

Effect of the eccentricity error: (a) stereo model of axis eccentricity and (b) schematic diagram of axis eccentricity.

OE_57_4_044106_f007.png

Assuming that the receiver is R meters away from the axis of rotation of the rotary table, the eccentricity between the axis of rotation of the rotary table and the axis of rotation of transmitter is d. The rotary table has rotated to an angle θ relative to its initial setting.

The vector from the center of the rotary table to the receiver is given by

Eq. (5)

O1P¯=[RcosθRsinθ].

The vector from the center of the transmitter to the receiver is given by

Eq. (6)

O2P¯=[RcosθdRsinθ].

It is straightforward to calculate the azimuth angle error caused by eccentricity; the error |Δθ| (see Fig. 8) is calculated as

Eq. (7)

|Δθ|=arccos(O1P¯ΔO2P¯|O1P¯||O2P¯|)=arccos(RdcosθR22Rdcosθ+d2).

Fig. 8

Change regulation caused by the eccentricity error.

OE_57_4_044106_f008.png

As shown in Fig. 7, when the eccentricity d is 0.1 mm, the maximum azimuth angle error caused by eccentricity is up to 2.06  arcsec. This is too large, and it must be compensated for.

3.3.

Error Caused by the Height of the Receiver

The error caused by the height of the receiver (z-value) does not influence the result independently. In other words, when the receiver is not in the x-O-y plane and the axis tilt angle and eccentricity are zero, the result is not influenced. However, when the z-value becomes larger, the error analyzed in Secs. 3.1 and 3.2 increases. This error is related to the other two errors.

3.3.1.

Influence on the axis tilt error

The error level analyzed in Sec. 3.1 is premised on the receiver being in the x-O-y plane and the z-value being zero. However, the larger the height is, the larger the azimuth angle error is in the same rotation angle of the rotary table. In this test, a simulation was carried out regarding this problem.

If the tilt angle of the axis of rotation of the transmitter is 10  arcsec and the height of the receiver changes from 20 to 520 mm, the result is as shown in Fig. 9. When the height of the receiver is <20  mm, the maximum angle error caused by axis tilt error and the height of the receiver is <0.02  arcsec. It is better to place the receiver near the x-O-y plane within 20 mm.

Fig. 9

Influence on the axis tilt error of height of the receiver.

OE_57_4_044106_f009.png

3.3.2.

Influence on the eccentricity error

The error analyzed in Sec. 3.2 is also calculated when the z-value equals zero. When the z-value increases, the error model caused by the height of the receiver is as shown in Fig. 10.

Fig. 10

Influence on the eccentricity error of the height of the receiver.

OE_57_4_044106_f010.png

Point P1 is in the x-O-y plane and point P2 is h=20  mm higher than the x-O-y plane.

Assuming that the eccentricity d=0.1  mm and R=10  m, the maximum eccentricity error is Δθ1=2.0626  arcsec when the receiver is placed in point P1. When the receiver is placed in point P2, the eccentricity error is Δθ1=2.0622  arcsec, according to geometrical relations.

As a result, when the height of the receiver is <20  mm, the error is small enough to be ignored.

3.4.

Machining Error

Machining error is caused by the axis of rotation of the transmitter and does not strictly coincide with the axis of the rotating head of the transmitter. In the process of this experiment, the rotating head of the transmitter was locked onto the stationary body. The axis of rotation of the transmitter was used as the target of the experimental adjusting axis, which introduces machining error, because the transmitter is symmetrical. In fact, the axis of the rotating head should be adjusted, but it is more complicated to make the axis of the rotating head rotate the same angle as that of the rotary table. However, the transmitter was assembled with high precision, and the deviation of the axis of the rotation of the transmitter is small. Furthermore, the machining error analyzed in this section conforms to the axis tilt error model and the eccentricity error model supported by the theoretical analysis in Secs. 3.1 and 3.2. This error can be compensated along with the eccentricity error.

4.

Results

4.1.

Data Processing

The data stream includes t1, t2, and T, where t1 and t2 represent the moment that the two scanning fans reach the receiver and T represents the transmitter’s rotation period.

The measurement data were obtained from rotating the rotary table 10 deg counterclockwise each time and 540 deg in total. In the test, there were 54 packages of measurement data and 1530 measurement instances in each package. The average of each package was used to find the angle in each position.

The angle calculation formula is calculated as

Eq. (8)

{Δφ1(i)=θ1(i+1)θ1  i=(t1(i+1)t1(i))*2  πΔφ2(i)=θ2(i+1)θ2  i=(t2(i+1)t2(i))*2  π,
where i=1,2,,53 and t1(i) and t2(i) represent the moment that two scanning fans reached the receiver at the angle 10·(i1)  deg. Δφ1(i) and Δφ2(i) represent the difference between the two measured rotation angles at 10·(i1)  deg and 10·i  deg, which were obtained by the two fans, respectively.

The receiver is placed approximately in the horizontal plane of the transmitter coordinate x-O-y; the z-value (the elevation angle) may not be equal to zero. Though the measured rotation angle is not strictly equal to the azimuth angle, the change in the angle is the same. Finally, the azimuth angle error is obtained by the subtraction of the measured rotation angle.

The following discussion is based on the data regarding the first fan of the transmitter Δφ1(i).

It is clear that the data (see in Fig. 11) comply with the obvious law of sinusoidal variation and the law of the eccentricity error model discussed in Sec. 3 in this paper.

Fig.11

Raw data curve and residual curve.

OE_57_4_044106_f011.png

The sinusoidal equation can be determined according to the curve fitting method when processing the data. The fitting equation is given by

Eq. (9)

f(x)=a1sin(b1x+c1)+a2sin(b2x+c2).

The result of this fitting in MATLAB™ is calculated as Eq. (10). The RMSE is 0.82, and the red curve in Fig. 11 is the residual, which is the difference between the raw data and the fitted data. The model’s results were in good agreement with the experimental data.

Eq. (10)

f(x)=4.515*sin(0.1926*x+0.1506)+37.46*sin(0.172*x+0.9758).

4.2.

Error Compensation

In the test, the distance R from the receiver to the axis of rotation of the rotary table was 9.9222  m. The maximum azimuth angle error can be calculated using the fitting curve and |ΔφiMAX|=41.75  arcsec. According to Eq. (7), the eccentricity is d=2.0085  mm, and the eccentricity error needs to be compensated for.

The raw data curve is moved horizontally so that the zero of the raw data and the zero in Eq. (7) coincide. The adjusted azimuth angle error is calculated by subtracting the raw data from the calculated system error in Eq. (7). The result is shown in Fig. 12.

Fig.12

Measurement uncertainty of the azimuth angle.

OE_57_4_044106_f012.png

5.

Discussions

5.1.

Uncertainty in Azimuth

The factors that contribute to the uncertainty are as follows: (a) the uncertainty of curve fitting when compensating the eccentricity error, (b) the uncertainty caused by adjustment error, which means the axis tilt error, and (c) the measurement uncertainties of the instruments caused by the rotary table, the polyhedral prism, and the autocollimator.

In all tests, it was found that the z-value of the receiver was <20 mm and the axis tilt error was adjusted to <10  arcsec. The maximum azimuth angle error is 0.04  arcsec, and, therefore, the uncertainty caused by the axis tilt error is u1=0.04  arcsec.

The system errors were ignored after compensation, but the uncertainty of curve fitting was introduced. According to the curve fitting evaluation, the RMSE is 0.82. RMSE represents the level of the deviation between the measurement value and the true value. In other words, it also represents the level of azimuth angle error. The level of RMSE (0.82) is conformable to reality in engineering applications of the rotary-laser systems. As a result, the uncertainty of the curve fitting can be ignored.

The uncertainties of the polyhedral prism and electronic autocollimator have little influence on the adjustment error and may be <0.01  arcsec. This error can be ignored. The uncertainty of the rotary table should use the assessment method of the type B uncertainty component. The maximum angle error of the rotary table is 0.5  arcsec. As a result, the uncertainty caused by the instruments is u2=0.5  arcsec/3=0.35  arcsec, assuming that the error obeys uniform distribution

Eq. (11)

S=i=1n(xix¯)2n1=(x1x¯)2+(x2x¯)2++(xnx¯)2n1.

According to the Bessel formula [see Eq. (11)], the standard deviation of the adjusted azimuth angle error is 0.98  arcsec, and the standard uncertainty of the adjusted azimuth angle error is uφ=S=0.98  arcsec. The relationship between the adjusted azimuth angle error and the measured angle is calculated as Eq. (8). According to the guide to the expression of uncertainty in measurement,16 the relationship of uncertainty between them is given by

Eq. (12)

uφ2=uθ1(i+1)2+uθ1(i)2.

Combined with uθ1(i+1)=uθ1  i, the standard uncertainty in azimuth may be arranged as

Eq. (13)

u3=uφ/2=0.69  arcsec.

The combined uncertainty of the azimuth angle is calculated as

Eq. (14)

uc1=u12+u22+u32=0.77  arcsec.

5.2.

Uncertainty in Elevation

The uncertainty of the elevation angle is obtained through the relationship between the azimuth angle and elevation angle and shows that relationship between uncertainties of the two angles can be calculated.1719 The elevation angle is calculated as

Eq. (15)

β=acrtan[sin(θ1α)tanφ1].

A differential operation is applied to Eq. (15) and calculated as

Eq. (16)

dβ=cos(θ1α)tanφ1dθ1cos(θ1α)tanφ1dαsin(θ1α)secφ12dφ1tanφ12+sin(θ1α)2.

The internal parameter φ1 remained unchanged when the verification method was carried out, and the covariance of β is calculated as

Eq. (17)

σβ=cos(θ1α)tanφ1σαtanφ12+sin(θ1α)2.

The rotation angle θ1 approximately equals the azimuth angle α for the receiver and is placed near the x-O-y plane, meaning that cos(θ1α)1 and sin(θ1α)0. Combined with φ1=0.8319, the covariance of the elevation angle is σβ=0.55  arcsec. The uncertainty in elevation is uc2=σβ=0.70  arcsec.

As a fact, cos(θ1α)<1 and sin(θ1α)0; uncertainty of the elevation angle is <0.55  arcsec.

As a result, the expanded uncertainty of the azimuth angle is 1.5  arcsec and the elevation angle is <1.4  arcsec. This is calculated using a coverage factor of 2 (which is related to a level of confidence of 95.45%).

6.

Conclusions

In this paper, a unit of verification platform was designed to verify the angle performance of the rotary-laser system. This platform can be easily used for the verification of the next transmitter without rebuilding the experimental environment. The adjustment error and system error models were analyzed in detail. The axis tilt error and the error caused by the height of the receiver have little influence on the results when the receiver is placed in the proper position. The concentric adjustment, which introduces system errors, was omitted through compensation and the result was as expected. The experiments indicate that the level of uncertainty of the azimuth angle was 1.5  arcsec and the uncertainty of the elevation angle was <1.4  arcsec at a 95.45% level of confidence. The results indicate that the transmitters meet the measurement requirements. This verification unit may gradually form a standard to verify the performance of the angle measurement of rotary-laser systems.

Future research will focus on adjusting the axis of the rotating head of the transmitter coaxial with the axis of rotation of the rotary table to verify the performance of the rotary-laser system more precisely. A study to verify the angle performance of the rotary-laser system will be carried on and the measurement accuracy of the rotary-laser system will be improved in the future.

Acknowledgments

This work was supported by the National Natural Science Foundation of China (Grant Nos. 51475329 and 51775380), the National Key Research and Development Project of China (Grant No. 2017YFF0204802), the Foundation for Innovative Research Groups of the National Natural Science Foundation of China (Grant No. 51721003), and the Young Elite Scientists Sponsorship Program by China Association for Science and Technology (Grant No. 2016QNRC001).

References

1. 

C. Depenthal and J. Schwendemann, “iGPS—a new system for static and kinematic measurements,” Opt. 3-D Meas. Tech. IX, 1 (1–3), 131 –140 (2009). Google Scholar

2. 

R. Schmitt et al., “Performance evaluation of iGPS for industrial applications,” in Int. Conf. on Indoor Positioning and Indoor Navigation (IPIN), (2010). https://doi.org/10.1109/IPIN.2010.5647630 Google Scholar

3. 

R. Schmitt et al., “Investigation of the applicability of an iGPS metrology system to automate the assembly in motion of a truck cabin,” Appl. Mech. Mater., 840 58 –65 (2016). https://doi.org/10.4028/www.scientific.net/AMM.840 Google Scholar

4. 

S. Kang and D. Tesar, “Indoor GPS metrology system with 3D probe for precision applications,” in Proc. of ASME Int. Mechanical Engineering Congress, 1 –4 (2004). Google Scholar

5. 

Z. Zhao et al., “Transmitter parameter calibration of the workspace measurement and positioning system by using precise three-dimensional coordinate control network,” Opt. Eng., 53 (8), 084108 (2014). https://doi.org/10.1117/1.OE.53.8.084108 Google Scholar

6. 

J. E. Muelaner et al., “Verification of the indoor GPS system by comparison with points calibrated using a network of laser tracker measurements,” in Proc. of 6th CIRP-Sponsored Int. Conf. on Digital Enterprise Technology, 607 –619 (2010). https://doi.org/10.1007/978-3-642-10430-5_47 Google Scholar

7. 

J. E. Muelaner et al., “Estimation of uncertainty in three-dimensional coordinate measurement by comparison with calibrated points,” Meas. Sci. Technol., 21 (2), 025106 (2010). https://doi.org/10.1088/0957-0233/21/2/025106 MSTCEP 0957-0233 Google Scholar

8. 

Y. Chen et al., “A structured light sensor measuring method based on wMPS global controlling,” Proc. SPIE, 9671 967105 (2015). https://doi.org/10.1117/12.2197076 PSISDG 0277-786X Google Scholar

9. 

D. Djurdjanovic and J. Ni, “Online stochastic control of dimensional quality in multistation manufacturing systems,” Proc. Inst. Mech. Eng. Part B J. Eng. Manuf., 221 (5), 865 –880 (2007). https://doi.org/10.1243/09544054JEM458 Google Scholar

10. 

C. Hirt et al., “Expected accuracy of tilt measurements on a novel hexapod-based digital zenith camera system: a Monte-Carlo simulation study,” Meas. Sci. Technol., 25 (8), 1408 –1424 (2015). https://doi.org/10.1088/0957-0233/25/8/085004 MSTCEP 0957-0233 Google Scholar

11. 

J. E. Muelaner et al., “Study of the uncertainty of angle measurement for a rotary-laser automatic theodolite (R-LAT),” Proc. Inst. Mech. Eng. Part B J. Eng. Manuf., 223 (3), 217 –229 (2009). https://doi.org/10.1243/09544054JEM1272 Google Scholar

12. 

L. Geng et al., “Analysis of angle measurement uncertainty for wMPS,” Proc. SPIE, 7544 754403 (2010). https://doi.org/10.1117/12.885389 PSISDG 0277-786X Google Scholar

13. 

D. Chao et al., “Evaluation method of angle measurement accuracy for wMPS,” Nano. Precis. Eng., 15 (6), 466 –472 (2018). https://doi.org/10.3969/j.issn.1004-1699.2012.02.018 Google Scholar

14. 

X. Zhi et al., “Verification of angle measuring uncertainty for workspace measuring and positioning system,” Chin. J. Sens. Actuators, 25 (2), 229 –235 (2012). https://doi.org/10.13494/j.npe.20160057 Google Scholar

15. 

J. Yuan and X. Long, “CCD-area-based autocollimator for precision small-angle measurement,” Rev. Sci. Instrum., 74 (3), 1362 –1365 (2003). https://doi.org/10.1063/1.1539896 RSINAK 0034-6748 Google Scholar

16. 

Guide to the expression of uncertainty in measurement (GUM) general metrology,” (1995). Google Scholar

17. 

Z. Liu et al., “A single-station multi-tasking 3D coordinate measurement method for large-scale metrology based on rotary-laser scanning,” Meas. Sci. Technol., 24 (10), 105004 (2013). https://doi.org/10.1088/0957-0233/24/10/105004 MSTCEP 0957-0233 Google Scholar

18. 

Y. Xie et al., “A new single-station wMPS measurement method with distance measurement,” Proc. SPIE, 9677 96771Z (2015). https://doi.org/10.1117/12.2200923 PSISDG 0277-786X Google Scholar

19. 

N. Zhou et al., “iGPS measurement network multi-station arrangement design,” Appl. Mech. Mater., 443 223 –227 (2013). https://doi.org/10.4028/www.scientific.net/AMM.443 Google Scholar

Biography

Jin Zhao is a student at Tianjin University. He received his BS degree in measurement and control technology and instruments from Tianjin University in 2016, and he will receive his MS degree from Tianjin University in 2019. His current research interests include large-scale measurement and precision vision measurement.

Yongjie Ren received his PhD from Tianjin University in 2007. Currently, he is an associate professor at the State Key Laboratory of Precision Measurement Technology and Instruments, Tianjin University. His research interests are focused on industrial robot as well as vision measurement and photoelectric measuring technology.

Jiarui Lin received his PhD from Tianjin University in 2012. Currently, he is an associate professor at the State Key Laboratory of Precision Measurement Technology and Instruments, Tianjin University. His research interests include vision measurement and photoelectric measuring technology.

Shibin Yin received his PhD from Tianjin University in 2015. Currently, he is a lecturer at the State Key Laboratory of Precision Measurement Technology and Instruments, Tianjin University. His research interests include vision measurement and photoelectric measuring technology.

Jigui Zhu received his BS and MS degrees from the National University of Defense Science and Technology of China in 1991 and 1994, respectively, and his PhD from Tianjin University, China, in 1997. Currently, he is a professor at the State Key Laboratory of Precision Measurement Technology and Instruments, Tianjin University. His research interests include laser and photoelectric measuring technology, such as industrial online measurement and large-scale precision metrology.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Jin Zhao, Yongjie Ren, Jiarui Lin, Shibin Yin, and Jigui Zhu "Study on verifying the angle measurement performance of the rotary-laser system," Optical Engineering 57(4), 044106 (26 April 2018). https://doi.org/10.1117/1.OE.57.4.044106
Received: 21 February 2018; Accepted: 23 March 2018; Published: 26 April 2018
Lens.org Logo
CITATIONS
Cited by 5 scholarly publications.
Advertisement
Advertisement
KEYWORDS
Transmitters

Receivers

Autocollimators

Prisms

Error analysis

Head

Optical engineering

RELATED CONTENT


Back to Top