1 March 2019 Optimizing voxel-wise dose calculations in cone-beam computed tomography
Author Affiliations +
Abstract
An analytical algorithm for the estimation of the patient-specific dose distributions in cone-beam computed tomography is introduced. The developed dose estimation method requires the reconstructed voxel data in values of linear attenuation coefficients and the scanning protocol. The algorithm first calculates the dose distribution due to the primary beam attenuation along the beam path between the source and each reconstructed voxel in conjunction with the solid angle subtended by given voxel. Then, this primary dose voxel map becomes the source for the dose distribution due to the scattered photons. For the pre-calculated primary dose value in a given voxel, the scatter dose values to all the other voxels are similarly calculated as the primary. The developed algorithm shows a good agreement with the Monte Carlo (MC) simulation for an anthropomorphic head-and-neck phantom. The accuracy of the analytical method is investigated by comparing estimates with the MC estimates and the strategy for computational acceleration is discussed in terms of the number of projections used for reconstruction, the number of spectral bins of the incident x-ray spectrum, the number of voxels, and the extent of scattering ranges.
© (2019) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jinwoo Kim, Jonghee Yun, Seungwoo Ha, Hanbean Youn, Hosang Jeon, and Ho Kyung Kim "Optimizing voxel-wise dose calculations in cone-beam computed tomography", Proc. SPIE 10948, Medical Imaging 2019: Physics of Medical Imaging, 1094860 (1 March 2019); https://doi.org/10.1117/12.2512325
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Algorithm development

X-rays

Scattering

Computed tomography

Monte Carlo methods

X-ray computed tomography

Back to Top