5 March 2021 Accuracy of catheter-based near-infrared auto-fluorescence detection in human coronary plaques
Mie Kunio, Joseph A. Gardecki, Kohei Watanabe, Kensuke Nishimiya, Daisuke Yamada, Sarika Verma, Guillermo J. Tearney
Author Affiliations +
Abstract
There is significant histopathological and clinical evidence that near-infrared auto-fluorescence (NIRAF) complements optical coherence tomography (OCT) for detecting high-risk coronary plaque. Here, we determined the accuracy of an OCT-NIRAF imaging system and catheter for detecting NIRAF in human coronary lesions. OCT-NIRAF pullback imaging was performed on human cadaver coronary arteries (n=33 from 14 patients) during PBS perfusion via a fully integrated OCT-NIRAF imaging system and catheter (NIRAF ex. 633 nm, 1 mW power; em. 660-740nm). Confocal NIRAF images were acquired from corresponding unstained formalin-fixed paraffin-embedded sections (Olympus FLUOVIEW FV1000; ex. 635 nm; em. 655-755nm). OCT-NIRAF and confocal NIRAF images were registered using known pullback speed, anatomical landmarks, and fiducial features (e.g., calcification), and spatially overlapped by affine transformation of the confocal NIRAF images. Each image was split into 8, 45º-sectors, emanating from the catheter location. Each 45°-sector was determined to be positive if <5% of the intima contained confocal NIRAF, and if <5% of 45°-arc (2.25°) of the catheter-based NIRAF signal was above the system’s detection limit. A total of 1896 45°-sectors from 291 distinct coronary locations were analyzed using confocal NIRAF as the gold standard. Considering superficial confocal NIRAF foci within 0.5 mm from the luminal surface, sensitivity and specificity were 90.0% (95%CI: 69.8- 100.0%) and 90.2% (95%CI: 88.8-91.7%), respectively. Within 0.5 mm to 1.0 mm depth from the luminal surface, the sensitivity was 36.4% (95%CI: 15.0-57.8%) and specificity was 90.1% (95%CI: 88.6-91.5%). These results indicate that the OCT-NIRAF system/catheter’s ability to detect NIRAF is depth dependent and accurate in plaque regions (within 0.5 mm from the luminal surface) that are most responsible for precipitating coronary events.
Conference Presentation
© (2021) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Mie Kunio, Joseph A. Gardecki, Kohei Watanabe, Kensuke Nishimiya, Daisuke Yamada, Sarika Verma, and Guillermo J. Tearney "Accuracy of catheter-based near-infrared auto-fluorescence detection in human coronary plaques", Proc. SPIE 11621, Diagnostic and Therapeutic Applications of Light in Cardiology 2021, 116210J (5 March 2021); https://doi.org/10.1117/12.2582533
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Confocal microscopy

Imaging systems

Optical coherence tomography

Arteries

Gold

Signal detection

Back to Top