Paper
13 March 2009 The effect of lag on image quality for a digital breast tomosynthesis system
Author Affiliations +
Proceedings Volume 7258, Medical Imaging 2009: Physics of Medical Imaging; 72580R (2009) https://doi.org/10.1117/12.811889
Event: SPIE Medical Imaging, 2009, Lake Buena Vista (Orlando Area), Florida, United States
Abstract
Digital breast tomosynthesis (DBT) is a limited-view, limited-angle computed tomography (CT) technique that has the potential to yield improved lesion conspicuity over that of standard digital mammography. To maintain short acquisition time, the detector must have a rapid temporal response. Transient effects like lag and ghosting have been noted previously in digital mammography systems, but for the times between successive views (approx. 1 minute), their impact on image quality is generally negligible. However, tomosynthesis imaging requires much shorter times between projection images (< 1 s). Under these conditions, detectors that may have been acceptable for digital mammography may not be suitable for tomosynthesis. Transient effects will generally cause both a loss of signal and an increase in image noise. A cascaded systems analysis is used to determine the effect of lag on image quality in a DBT system. It is shown that in the projection images, lag results in artifacts appearing as a "trail" of prior exposures. The effect of lag on image quality is also evaluated with a simple Monte Carlo simulation of a cone-beam tomosynthesis image formation incorporating a filtered back-projection algorithm.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
James G. Mainprize, Xinying Wang, and Martin J. Yaffe "The effect of lag on image quality for a digital breast tomosynthesis system", Proc. SPIE 7258, Medical Imaging 2009: Physics of Medical Imaging, 72580R (13 March 2009); https://doi.org/10.1117/12.811889
Lens.org Logo
CITATIONS
Cited by 6 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Monte Carlo methods

X-rays

Image quality

Systems modeling

X-ray detectors

Digital breast tomosynthesis