20 February 2023 Development of a platform for broadband, spectra-fitted, tissue optical phantoms
Author Affiliations +
Abstract

Significance

Current methods of producing optical phantoms are incapable of accurately capturing the wavelength-dependent properties of tissue critical for many optical modalities.

Aim

We aim to introduce a method of producing solid, inorganic phantoms whose wavelength-dependent optical properties can be matched to those of tissue over the wavelength range of 370 to 950 nm.

Approach

The concentration-dependent optical properties of 20 pigments were characterized and used to determine combinations that result in optimal fits compared to the target properties over the full spectrum. Phantoms matching the optical properties of muscle and nerve, the diffuse reflectance of pale and melanistic skin, and the chromophore concentrations of a computational skin model with varying oxygen saturation (StO2) were made with this method.

Results

Both optical property phantoms were found to accurately mimic their respective tissues’ absorption and scattering properties across the entire spectrum. The diffuse reflectance phantoms were able to closely approximate skin reflectance regardless of skin type. All three computational skin phantoms were found to have emulated chromophore concentrations close to the model, with an average percent error for the StO2 of 4.31%.

Conclusions

This multipigment phantom platform represents a powerful tool for creating spectrally accurate tissue phantoms, which should increase the availability of standards for many optical techniques.

1.

Introduction

The development and validation of optical instrumentation for medical diagnosis, surgical guidance, and light-based therapeutics often require the use of well-characterized phantoms as standardized samples. Their uses include determining detection limits, assessing spectral and spatial resolution, validating postprocessing algorithms, and performing quality control over time and between different systems, among others.1,2 Although an ideal phantom would be one that could be used for all applications and modalities, this would at least require a solid phantom that has tunable optical properties with tissue-like wavelength dependence over a broad spectral range, while being stable over time and throughout multiple uses, something that current methods cannot achieve.

Phantoms with wavelength-dependent optical properties similar to tissue are usually achieved by directly using biological chromophores, such as whole blood or hemoglobin, to provide absorption and lipid emulsions, such as Intralipid, to provide scattering.37 Though simple aqueous solutions can be used, these additions can also be suspended in a biologically compatible gelatin- or agar-based hydrogel to produce solid, moldable phantoms.6,810 Although this can result in spectrally accurate phantoms for a wide range of tissue types, the organic nature of both the chromophores and matrix materials inherently limits the shelf-life of these phantoms to somewhere between a few hours and a couple weeks, depending on environmental conditions.1,5,6,9 As such, they are typically made for a single task and discarded after use, preventing them from serving as effective standards.

To make phantoms whose optical properties are stable over long periods of time, transparent, inorganic matrix materials, such as silicone or resin, are used as they are resistant to environmental effects and can result in effective shelf-lives of at least several years. However, this stability comes with the trade-off of these materials being incompatible with the inclusion of biological chromophores, preventing them from easily matching the spectral optical properties of tissue.1 Instead, the optical properties of these phantoms are typically controlled through the addition of a metal oxide powder, usually either titanium dioxide (TiO2) or aluminum oxide (Al2O3), or polystyrene microspheres to provide scattering and a carbon-based black pigment, usually in the form of either India ink or nigrosin, to provide flat absorption across the spectrum.8,9,1118 By varying the concentrations of these additions, nearly any combination of absorption and scattering coefficients can be achieved at a specified wavelength.

Although this approach works well when only a single wavelength is of interest, it is incapable of accurately capturing the wavelength-dependent scattering and absorption properties of tissue. Having such a dependence can be important for many optical modalities, such as fluorescence imaging, as both the properties at the excitation and emission wavelengths impact the collected emitted signal, chromophore extraction techniques, including pulse oximetry and near-infrared spectroscopy, which rely on the absorption at two or more distinct wavelength bands to approximate the concentrations of tissue chromophores and other spectroscopic techniques. Many groups have attempted to address this need through the addition of more specific absorbers with the inclusion of red ink being a relatively popular choice as it provides higher absorption in the green and blue regions of the spectrum while maintaining a low absorption in the red and NIR.19 Cosmetic powders matched to different skin-tones have also been used to provide similar spectral absorption differentials, though are less commonly used due to their comparatively high scattering.20 Different molecular dyes with relatively narrow absorption peaks have also been used in order to independently tune two separate wavelengths to the desired properties.15,21 However, to our knowledge, there has yet to be a phantom design, which is capable of mimicking the wavelength-dependence of an arbitrary tissue across the VIS-NIR spectrum.

To address this, we have developed a platform capable of producing solid, stable tissue phantoms whose absorption and scattering properties can be spectrally matched the desired tissue properties over a range 370 to 950 nm. This was accomplished by first characterizing 20 commercially available pigments which provide either scattering, baseline absorption, or absorption peaks throughout blue, green, and red wavelengths. By feeding the concentration-dependent properties of the pigments into a nonnegative, least squares optimization, the combination of pigments which results in a phantom with wavelength-dependent optical properties highly similar to those of the target tissue can be determined. To assess the accuracy and scope of this method, phantoms were made matching the optical properties of human tissues for either the full spectrum or selected wavelength bands within the full spectrum. The efficacy of the phantom platform was further assessed by making more complicated phantom designs including diffuse reflectance matched phantoms and chromophore mimicking phantoms.

2.

Materials and Methods

2.1.

Phantom Component Selection

Optical tissue phantoms typically consist of three basic components: a matrix material, an absorbing agent, and a scattering agent. For the matrix material, we chose to use a two-part epoxy resin (JANCHUN Crystal Clear Epoxy Resin Kit) due to its relatively quick cure time and inexpensive material cost compared to other common matrix materials, such as PDMS. The epoxy resin consists of a base material (part A) and a hardener (part B), which are mixed together using a 1A:1B volume ratio or 1A:0.9B weight ratio. After mixing, the epoxy has a pot life of 30 to 40 min where it can still easily be manipulated and will fully cure after 24 h at room temperature. To mitigate the risk of any additives settling, the cure time can be reduced to 30  min by elevating the temperature to 70°C.

To better mimic the complex and varied absorption and scattering spectra that can be found within tissue, we used 20 different pigments as shown in Table 1. For these pigments, we chose to utilize artist-grade acrylic paints whenever possible as opposed to the traditionally used powdered components. This was due to the fact that in our initial testing, we found that the included surfactants and stabilizers in acrylic paint made homogenizing the phantom easier to achieve without the need for sonication while also helping to prevent the scatterers from settling to the bottom of the epoxy, issues typically found when making solid, inorganic phantoms.

Table 1

Absorbing and scattering pigments used in the multipigment phantom platform.

PigmentBrandColor nameLightfastness
Yellow
PY3Liquitex: Basics acrylicCadmium yellow light hueII
PY65Winsor & Newton: Galeria acrylicCadmium yellow deep hueI
PY74Liquitex: Basics acrylicPrimary yellowI
PY100Wilton: Gel icing colorLemon yellow
PY154Golden: Fluid acrylicsBenzimidazolone yellow mediumI
PY175Golden: Heavy body acrylicsBenzimidazolone yellow lightI
Magenta
PR122Golden: Heavy body acrylicsQuinacridone magentaI
PR170Winsor & Newton: Galeria acrylicCrimsonI
PR202Plaza Art: Artists’ acrylicQuinacridone fuchsiaI
PV15Plaza Art: Artists’ acrylicUltramarine violetI
PV19Golden: Heavy body acrylicsQuinacridone violetI
PV23Plaza Art: Artists’ acrylicDioxazine purpleII
Miscellaneous
PG7Golden: Heavy body acrylicsPhthalo green (blue shade)I
PB15:3Golden: Heavy body acrylicsPhthalo blue (green shade)I
PO73Golden: Heavy body acrylicsPyrrole orangeI
White
PW4Golden: Heavy body acrylicsZinc whiteI
PW6Plaza Art: Artists’ acrylicTitanium whiteI
Al2O3The Rock ShedAl2O3 polish (2  μm)
Black
Black 2.0Culture HustleBlack 2.0
India inkSpeedballSuper black

Additionally, the pigments used in artist-grade acrylic paint typically have had their lightfastness, or relative degree in change due to light exposure, characterized using the ASTM standard test methods.22 This process exposes the material to either direct solar illumination or a Xenon arc lamp with a similar spectral curve for a total irradiance of 1260  MJ/m2 and assess the degree of change which occurs within the CIE 1976 color space. Depending on the magnitude of the change, a pigment can receive a rating of either I, excellent; II, good; or III, fair. Where applicable, the pigments chosen for this work were all rated as having a lightfastness of I, with only two being rated as having a rating of II (see Table 1). As such, any phantoms made using the chosen pigments are being expected to have high stability over time and usage.

To match the variable scattering slope of different tissue types, we chose to use three of the most common scattering agents used in solid phantoms, titanium dioxide (pigment white 6; PW6), zinc oxide (pigment white 4; PW4), and Al2O3.8,1216,23,24 Due to differences in refractive index and average particle size, these three white pigments have been shown to provide different scattering powers and slopes which can be combined to better match that of tissue.16,23

The typical absorption spectrum of tissue consists of three main regions that a broadband phantom would need to match using three different categories of absorbing pigments. First, there is a general baseline absorption arising from all present chromophores, which is observed in the red and near-infrared regions of the spectra. This can be matched using black pigments that have broadband absorption across the entire spectra. We chose to use two different black pigments for this work, India ink, which has become a standard absorber for optical phantoms, and Black 2.0, to have some variance in the shape of the baseline absorption.

On top of the baseline, tissue has two major absorption peaks due to presence of hemoglobin, one in the blue region and another in the green region of the spectrum. To match these two absorption peaks, we used two classes of colored pigments: yellow pigments which primarily absorb blue wavelengths and magenta pigments which primarily absorb green wavelengths. Although the absorption coefficients of most pigments have not before been reported, many studies have analyzed their diffuse reflectance spectra.2528 Using reported reflectance spectra, we selected six yellow and six magenta pigments that have absorption peaks around 420 and 550 nm, respectively. Finally, to account to any absorption features not covered by these three categories, we included three miscellaneous pigments. Pigment orange 73 (PO73) was included as its absorption should fall between the yellow and magenta pigments, whereas pigment green 7 (PG7) and pigment blue 15:3 (PB15:3) were chosen as their reported reflectance spectra include significant red and NIR absorption which none of the other included pigments have.

2.2.

Phantom Fabrication

Each epoxy phantom was made using the same general procedure with each component being measured by weight. To begin, the desired amounts of any of the yellow, magenta, and miscellaneous pigments are weighed out in a plastic weigh boat. Following this, water weighing 10% of the desired amount of base epoxy is added to thin and rehydrate the acrylic paints. The water is then mixed with the pigments using a glass stir rod until it becomes homogenous in color. After this, the weigh boat is placed back onto the analytical balance and the desired amount of part A of the epoxy is weighed out. The added part A is then mixed by hand with the thinned pigments until a smooth emulsion is achieved. To remove the added water, the mixture is placed in a vacuum chamber and brought to 750  mmHg until bubbles stop forming. This process is then repeated in a separate weigh boat to combine part B of the epoxy with the black and white pigments, making sure that the amount of water used is at least 30% more than the weight of the white pigments.

After removing the water, the part A mixture is poured into the part B one and mixed by hand until homogeneous. After mixing, an initial degassing is performed by placing the weigh boat into a vacuum chamber and bringing the pressure to 750  mmHg until the bubbles stop forming or 10 min pass. The resin tends to expand by a large degree during the initial portion of degassing, so it is important to use a large enough vessel. After the initial degassing, the epoxy mixture is poured into silicone molds of the desired shape and size. Silicone molds were chosen to provide smooth surfaces to the phantoms so that the effects of diffuse reflectance from the sample surface are prevented from confounding the optical property measurements. These molds are then placed back into the vacuum chamber for a final degassing using the same procedure previously described. After all the remnant air bubbles have been removed, the epoxy is left to cure in an oven set to 70°C for at least 30 min, after which it is solid enough to safely demold.

2.3.

Optical Property Measurement

The optical properties of the epoxy materials were determined between 370 and 950 nm using the single integrating sphere, inverse adding–doubling (IAD) approach.29,30 IAD solves for the optical properties of a material using reflectance and transmittance measurements as well as the thickness and refractive index of the sample. It accomplishes this by iteratively solving the radiative transport equation using the adding–doubling method for different pairs of optical properties until a set of absorption and reduced scattering coefficients are found produce a reflectance and transmittance which match the measurements. The thickness of each sample was measured using digital calipers, whereas the refractive index of epoxy was estimated from the literature to be n=1.56.16 Additionally, as we did not collect any collimated transmission spectra, we assumed a fixed scattering anisotropy of g=0.8 for every sample based on the average refractive index and size distributions of the different scattering particles.16,23

The required diffuse reflectance and total transmittance measurements were taken using a dual-beam spectrophotometer (Cary 5000, Agilent) with an integrating sphere attachment (DRA-2500, Agilent). Sample reflection and transmission measurements were converted into reflectance and transmittance using a zero correction and baseline normalization. For reflectance, the baseline measurement is taken using a 99% diffuse reflectance standard at the sample port (USRS-99-020, LabSphere), whereas the zero measurement is taken by allowing the light into the sphere but leaving the sample reflectance port open, accounting for any light that may be incident on the sphere wall before the sample. For the transmittance, the zero measurement is taken by blocking the sample beam from entering the sphere, whereas the baseline is determined in the same manner as for reflectance. The diameter of the spot size at the sample was maintained at 2 mm for both measurements, which was always no more than 5% of the total area of the sample port to minimize the loss of laterally scattered light.

To obtain samples that were thin enough for the required transmission measurements, a 50×50  mm square silicone mold was used, adding a fraction of the total amount of epoxy such that the bottom of the mold was lightly coated. After curing, the thin samples were demolded and a 25×25  mm square cut from the center to mitigate any effects that a meniscus would have on the thickness. Although not accurate enough to produce a sample of a specified thickness, this process consistently resulted in thicknesses between 0.5 and 1.0 mm.

Although IAD has been shown to produce accurate determination of optical properties from integrating sphere measurements, a substantial crosstalk effect has been known to occur in the IAD calculated reduced scattering coefficient for wavelength ranges corresponding to significant absorption peak of the material. To address this, we utilized the two-stage IAD process as proposed by Xie et al. for all of our measurements.31,32 The first stage of this process is to run IAD as previously described. The second stage takes the resulting reduced scattering coefficient and fits it to a Mie–Rayleigh power law equation. IAD is then run a second time using the same reflectance and transmittance data with the scattering constrained to the values obtained from the fitted model. This process has been shown to not only correct the shape of the scattering slope but also correct the absorption peaks to their expected values.

2.4.

Pigment Characterization

Before being used for tissue phantoms, the optical properties of each of the 20 pigments (Table 1) while suspended in epoxy resin were characterized. To accomplish this, the phantom fabrication protocol as previously described was followed to add individual pigments into 9.5 g of epoxy. This process was repeated three times for every pigment, using a different mass fraction, which we defined as the amount of pigment in milligrams divided by the total mass in grams, each time. As the relative absorption and scattering strengths of the pigments differed, different sets of mass fractions were used for each pigment. The absorption and reduced scattering coefficients of the individual pigment samples were determined using the approach previously described. The results were then normalized to their respective mass fractions and averaged together for each given pigment.

2.5.

Tissue Optical Property Phantoms

Using the concentration-dependent optical properties of the different pigments, phantoms were made to match the optical properties of muscle and nerve from human cadavers, which our group previously reported.33 These two tissue types were chosen as they represent two possible extremes, with muscle having relatively high absorption and low scattering and nerve having the opposite.

To determine the amount of each pigment required to best match one of the sets of optical properties between 370 and 950 nm, a two-step, nonnegative least squares optimization was used. The first step of this process determined the concentration of yellow, magenta, black, and miscellaneous pigments, which would result in a phantom absorption spectrum that most closely matched the desired values using the following equation:

Eq. (1)

argminxλ=370950[w(λ)μatissue(λ)w(λ)Cabsx(λ)x]2,
where x is the concentrations of the absorbing pigments, w(λ) is a weighting function, Cabsx(λ) is the concentration-dependent absorption coefficients of those pigments, and μatissue(λ) is the target absorption coefficient spectrum. However, as the absorption of tissue can vary over at least two orders of magnitude across the visible and NIR regions of the spectrum, any solution to this process would prioritize fitting to the absorption peaks in the blue and green portions of the spectrum over the lower values in the red and NIR. To account for this, we applied a wavelength-dependent weight to the optimization based on the target absorption spectrum. The weight w(λ) at each wavelength was set as the maximum absorption coefficient divided by the absorption coefficient at that wavelength. This corrects for any differences in magnitude across the spectrum so that each wavelength is treated equally in the least-squares optimization.

The second step of the optimization process involved a second nonnegative least squares optimization for the white pigments and the reduced scattering coefficient:

Eq. (2)

argminzλ=370950[(μstissue(λ)Cscatx(λ)x)Cscatz(λ)z]2,
where z is the concentrations of the white pigments, Cscatz(λ) is their concentration-dependent reduced scattering coefficients, and μstissue(λ) is the target reduced scattering coefficient spectrum. Although the weighting approach as used for the absorption coefficient was not required, the amount of scattering that the non-white pigments would contribute was an important consideration. As such, the predicted amount of scattering caused by the absorbing pigments, calculated from their concentrations x and concentration-dependent reduced scattering coefficients Cscatx(λ) was subtracted from the target scattering beforehand.

Additionally, many phantoms may only require matching properties at a number of discreet wavelength bands instead of across the entire spectrum. To determine the pigment concentrations required for such a phantom, a single optimization step was used, in which sum of the squared residuals for each property is minimized using the following equation:

Eq. (3)

argminqλ=370950[wa(λ)(μatissue(λ)Cabs(λ)q)2+ws(λ)(μstissue(λ)Cscat(λ)q)2],  
where q is the concentrations of all of the pigments, Cabs(λ) and Cscat(λ) are their concentration-dependent absorption and reduced scattering coefficients, and μatissue(λ) and μstissue(λ) are the target optical properties. To select the bands of interest, strong weighting functions (wa(λ) and ws(λ)) are applied to both sets of properties at the specified wavelengths while the weights outside those bands are set to 1. These weights can be manually adjusted to account for differences in magnitude between the bands and to preferentially fit to certain bands of interest over others. A phantom matching the optical properties of muscle at four bands was made. The chosen bands were 385 to 405 nm, 530 to 570 nm, 615 to 635 nm, and 825 to 865 nm as these regions spanned the full range of that the absorption coefficient of that muscle exhibits.

After determining the desired pigment concentrations using either of these methods (see Table S1 in the Supplementary Material), phantoms were made using the previously described fabrication process. Each of the phantoms were made using 38 g of epoxy to allow for lower concentrations of some of the pigments to fall above the 5 mg minimum of the analytical balance. To account for the orders of magnitude difference in absorption coefficient that these phantoms would have across the spectrum, each phantom was made into two thin samples of different thicknesses, one around 0.5 mm and the other between 1.5 and 2.5 mm. The optical properties of both were then determined. The first sample is thin enough to allow for measurable transmittance and reflectance at the blue absorption peak, whereas the second is thick enough for the low absorption in the NIR to have a significant effect on the measurements. From the results of these two samples, a composite absorption coefficient spectrum was made. The absorption from the thinner sample was used from 370 to 560 nm, and the thicker sample’s absorption was used from 620 to 950 nm. Between these two regions, the absorption coefficients were blended together using a sigmoid-weighted average.

2.6.

Diffuse Reflectance Phantoms

In some cases, the optical properties of a tissue of interest may be unknown making it difficult to make a phantom to accurately calibrate or test the performance of an optical system. Additionally, many modalities rely on diffuse reflectance, which, while dependent on the absorption and scattering properties, is also dependent on the refractive index. With the refractive index of epoxy resin, and other common phantom matrix materials, being significantly higher than that of tissue, tissue phantoms that mimic optical properties are likely to not match the corresponding diffuse reflectance. This is especially relevant for many systems that are used for in vivo applications. To determine the feasibility of making optical phantoms directly matched to tissue diffuse reflectance, phantoms matching different skin types were made. From a NIST study, which assessed the diffuse reflectance of the inner forearm in vivo, two spectra were selected, one representing pale (NIST#74) and the other melanistic skin (NIST#44).34

To determine the pigment concentrations that would result in phantoms matched to the desired diffuse reflectance spectra R(λ), a nonlinear, nonnegative, least-squares model was used. This process utilizes a white Monte Carlo (wMC) to predict the phantom diffuse reflectance given the refractive index of the material (n=1.56) and the absorption and reduced scattering coefficients, calculated using the concentration-dependent properties of the pigments, Cabs(λ) and Cscat(λ).35,36 The optimization iterates the pigment concentrations q, which results in new wavelength-dependent optical properties, which are fed into the wMC to produce diffuse reflectance spectra which are then compared to the target spectrum in a least-squares manner:

Eq. (4)

argminqλ=370950[(R(λ)wMC(Cabs(λ)q,Cscat(λ)q,n)]2.

The diffuse reflectance phantoms were made using the determined pigment concentrations (see Table S2 in the Supplementary Material) and 38 g of epoxy. Using a 35×35  mm mold, this resulted in phantoms with an approximate thickness of 17 mm, which was found to be sufficiently optically thick for all wavelengths. The diffuse reflectance of the phantoms was measured using an integrating sphere and spectrophotometer in a similar manner as for the optical property measurements. However, to maximize the capture of deep, laterally scattered photons a larger, 25-mm-diameter aperture was used.

2.7.

Oxygenation Saturation Phantoms

A common modality that uses the optical properties of tissue at multiple spectral bands is the determination of tissue hemoglobin oxygen saturation (StO2). By measuring the absorption coefficient at two or more wavelengths, tissue chromophore concentrations, namely oxy- and deoxyhemoglobin, can be estimated by solving a series of equations.3740 This reliance on multiple wavelengths has made using inorganic phantoms to accurately emulate tissue chromophores difficult. As such, we wanted to determine if the multipigment approach could create phantoms that accurately emulate physiological chromophore concentrations given different levels of StO2.

To accomplish this, a computational model of skin absorption was made using blood, melanin, water, and fat as the primary chromophores. The absorption coefficient spectrum of the theoretical skin was determined through a summation of the absorption coefficients of the pure chromophores, obtained from the literature, such that

Eq. (5)

μa(λ)=StO2BHμaHbO2(λ)+(1StO2)BHμaHb(λ)+Wμawater(λ)+Fμafat(λ)+Mμamel(λ),
where H is the concentration of hemoglobin within blood, StO2 is the ratio of oxyhemoglobin to total hemoglobin, and B, W, F, and M are the volume fractions of blood, water, fat, and melanosomes, respectively, within the tissue.4145

Using fixed volume fractions, tissue absorption spectra were produced between 470 and 950 nm using three different values for StO2: 30%, 50%, and 80%. The volume fractions for water and fat were set to 0.5 and 0.02, respectively, based on the values used in similar models.38,39 A low melanosome volume fraction of 0.0255% was selected so that the theoretical tissue would have an absorption spectrum representative of pale, minimally pigmented skin.4447 The volume fraction of total blood was set to be 0.047 which, given a blood hemoglobin concentration of 2.3  mmol/L, corresponded to a total hemoglobin concentration of 110  μM, which falls within reported physiological ranges.37,39 Although not as important for StO2 measurements, the reduced scattering coefficient of the tissue model was set to be similar to be similar to that of skin. This was accomplished through a Mie-Rayleigh power law equation with values derived from the literature.44,47

Multipigment phantoms were made to match the theoretical absorption coefficients for each of the three StO2 values using 38 g of total epoxy. The required pigment concentrations (see Table S3 in the Supplementary Material) were determined using the previously described [Eq. (3)] weighted-band approach with bands at 540 to 560 nm, 625 to 645 nm, and 820 to 840 nm, which correspond to wavelength regions often used to measure StO2 in tissue.39,40 After measuring the optical properties of the oxygenation phantoms, the absorption coefficients at the three bands were used along with Eq. (5) to set up a system of equations to solve for the equivalent melanosome volume fraction, blood volume fraction, and StO2, assuming a water volume fraction of 0.5 and fat volume fraction of 0.02.

2.8.

Statistical Analysis

Spectral similarity is a nontrivial comparison with many different assessment methods existing, each with their own pros and cons.48,49 For our study, two different metrics were selected to compare the target spectra to the measured spectra of the optical property and diffuse reflectance phantoms.

The mean absolute error (MAE) was chosen due to being intuitively interpretable as well as not being overly sensitive to outliers, such as with squared-error metrics, and is not skewed when expected values are near zero, as with percent-error-based metrics. As its name implies, MAE is calculated as the mean of the absolute differences between the measured and expected spectra at each wavelength. For absorption coefficient comparisons, the measured and target spectra were log10-transformed prior to calculating the MAE to bring the values, which will vary over multiple orders of magnitude, onto the same relative scale.

The other chosen metric was the spectral angle mapper (SAM) between the measured and reference spectra, which is a common classification method used in remote sensing and hyperspectral imaging.48,49 SAM represents each spectrum as vectors with dimensionality equal to the number of spectral bands and finds the angle between those two vectors, with smaller angles representing spectra that are more similar in shape to one another and angles closer to 90 deg representing more dissimilar spectra:

Eq. (6)

SAM(R,S)=cos1R·SRS=cos1RiSiRi2Si2.
Mathematically, this is determined by taking the inverse cosine of the dot product of the two spectra R and S divided by their root sum of squares. As SAM is an angular measurement, it reflects only on how similar in shape the two spectra are disregarding their relative magnitudes.

The code used to determine the pigment concentrations necessary for the broadband and multiband and diffuse reflectance phantoms have been made available on GitHub along with the normalized pigment properties ( https://github.com/AlecWalter/MultipigmentPhantoms).

3.

Results and Discussion

3.1.

Pigment Characterization

Figure 1 shows the concentration-dependent absorption coefficient spectra of the yellow, magenta, black, and miscellaneous absorbing pigments. As expected, the yellow and magenta pigments were found to have absorption peaks falling between 370 to 480 nm and 500 to 600 nm, respectively. Additionally, it was found that for most of these pigments, the shape of their absorption peaks was similar to what would be expected for tissue in that wavelength region. The yellow pigments mainly have a Gaussian-like shape to their absorption peaks, which are generally similar to the shape of the blue absorption peak of hemoglobin. However, as the Gaussian peaks of the yellow pigments tend to be more broadband as compared to that of hemoglobin, fully matching to that part of the absorption spectrum will be difficult. The magenta pigments, on the other hand, are more closely matched to their tissue counterpart, with most having a bimodal absorption peak similar to that of oxyhemoglobin. Two pigments deviate from this trend, with PV23 having a third absorption peak around 625 nm and PV15 providing a more broadband, but relatively weak, baseline absorption in this region of the spectrum.

Fig. 1

Absorption characterization of colored pigments. Mean measured absorption coefficients of (a) yellow, (b) magenta, (c) miscellaneous, and (d) black pigments normalized to the mass fraction included in epoxy resin. Shaded regions indicate the 95% confidence interval (n=3).

JBO_28_2_025001_f001.png

Although not being matched to specific tissue spectral features, the three miscellaneous pigments were found to have absorption profiles close to what was desired when they were selected. PO73 was found to have a peak absorption between the yellow and magenta pigments near 500 nm and both PG7 and PB15:3 were found to primarily absorb red wavelengths with significant absorption in the NIR out to 750 nm. For the black pigments, the absorption of India ink was found to be similar to what has been reported previously, decreasing monotonically with wavelength. Interestingly, the absorption of black 2.0 does not follow this same trend, having a nearly flat absorption coefficient between 370 and 950 nm. Although not too major, this difference should allow for better fitting to baseline tissue absorption in the NIR without overly impacting the absorption coefficients in other parts of the spectrum.

Although primarily acting as absorbing agents, the colored pigments were found to contribute measurable amounts of scattering (Fig. S1 in the Supplementary Material). Although the concentration-dependent scattering cannot be neglected, within the range of mass fractions expected to be used the absorbing pigments should only induce small variations in the final reduced scattering coefficients. The majority of the strength and shape of the scattering curve will come from the white pigments, whose concentration-dependent reduced scattering coefficients can be seen in Fig. 2(a).

Fig. 2

Optical property characterization of white pigments. Mean measured (a) reduced scattering coefficients and (b) absorption coefficients of the white pigments normalized to the mass fraction included in epoxy resin. Shaded regions indicate a 95% confidence interval (n=3). Inset showcases the UV absorption of ZnO and TiO2.

JBO_28_2_025001_f002.png

The relative scattering strength of the three white pigments was, as expected, found to be based on their refractive index mismatch with the epoxy (n=1.56). As seen in Fig. 2(a), PW6 is the most strongly scattering of the white pigments with titanium dioxide having a refractive index of 2.6 at 600 nm. This is followed by PW4, with zinc oxide having a refractive index of 2.0 and Al2O3 with a refractive index of 1.76. However, as scattering strength can be overcome with increased concentration, the more important difference between the three white pigments is the differences in the shape of their scattering slopes caused by them having varying proportions of Mie and Rayleigh scattering, due to the differences in refractive index and size distribution of the particles. As with the absorbing pigments, these differences should allow for the three while pigments to be combined at different ratios to tweak the shape of the phantom scattering curve to best match that of the target tissue.

Outside of scattering, the absorption coefficients of the white pigments are another major factor of consideration. Although all three were found to have a minimal absorption coefficient throughout the visible and NIR, as seen in Fig. 2(b), both PW4 and PW6 exhibited relatively strong absorption of blue and UV wavelengths, with PW6 absorbing wavelengths shorter than 410 nm and PW4 absorbing wavelengths shorter than 380 nm. Although this may not be relevant for most wavelengths and applications, these absorption peaks limit the ability to use titanium dioxide or zinc oxide as scattering agents when accurate absorption in the blue is required, and especially when high scattering is also required.

Although the pigment characterizations were found to be internally consistent, variations between manufacturers and between batches may result in different concentration-dependent optical properties, as has been shown for India ink.18 However, any potential variations that do occur should only be due to amount of pigment originally included in the acrylic paint, as the specified pigments themselves are chemically unique and should thus provide the same absorption peak location and overall shape across batches. Due to this unknown variance, characterizing any materials used prior to use will increase the accuracy of any resulting phantoms and is a generally good practice if possible.

3.2.

Tissue Optical Property Phantoms

Figures 3(a) and 3(b) show the measured absorption and reduced scattering coefficients of the multipigment phantom made to match the optical properties of muscle. Figures 3(c) and 3(d) show the same comparison for the nerve matched phantom. From a visual comparison, it can be seen that the multipigment phantoms are capable approximating the shape of both properties over the entire spectral range of 370 to 950 nm, matching the general scattering curves of both low and high scattering tissue as well as having a nonmonotonic absorption coefficient spectrum that ranges over two to three orders of magnitude.

Fig. 3

Broadband optical property phantoms. Multipigment phantoms matching the broadband absorption and reduced scattering coefficients of [(a), (b)] muscle and [(c), (d)] nerve tissue from human cadavers.33 The spectral optical properties for each were compared to the theoretical best fit of a traditional phantom made using only TiO2 and India ink.

JBO_28_2_025001_f003.png

The quantitative assessment of the similarity of the phantom optical properties to that of their respective tissues can be found in Table 2. From at the analysis of the SAMs, it can be seen that the phantoms are better able to match the shape of the reduced scattering coefficient as compared to absorption coefficient. This is despite the inability of the white pigments to fully match the proportion of Rayleigh scattering present within the tissue scattering spectra. As is apparent from both the SAM and MAE trends, this mismatch results in a more poorly matched scattering spectrum as the average target scattering increases.

Table 2

Spectral similarity metrics of the broadband optical property phantoms.

MuscleNerve
MultipigmentTraditional fitMultipigmentTraditional fit
Absorption
MAE0.0550.6760.0500.535
SAM (°)22.4954.2733.8646.87
Scattering
MAE0.0920.3210.6751.447
SAM (°)1.6119.4711.5127.04

The major reason for the worse overall fit of the absorption coefficient is the discrepancies that occur around the blue peak due to the yellow pigments having a more broadband absorption as compared to the hemoglobin absorption peak. This results in the overall best fit undershooting the maximum absorption value to better match the rest of the values of the slopes of the peak. Additionally, the UV absorption from the white pigments results in further discrepancies in this part of the spectrum, especially for the higher scattering nerve tissue phantom, an issue which was identified during the characterization process.

To contextualize the improvements that the multipigment approach brings to broadband and multispectral phantoms, the muscle and nerve tissue properties were also compared to the theoretical properties that a traditional phantom, consisting only of India ink and titanium dioxide, would have if matched to the full spectrum in the same manner that the multipigment phantoms were. As seen in Fig. 3, the traditional phantom is very limited in the range of values both its absorption and reduced scattering coefficients can take across the spectrum. Although this could be used to roughly approximate the properties of tissue in the NIR, the variable and much greater properties in the visible portion of the spectrum result in an overall poor fit. This limitation is directly reflected in the quantitative similarity metrics of the traditional phantoms. Compared to those of the multipigment phantoms, the theoretical traditional phantoms had MAEs and SAMs that were two to six times greater, representing the high levels of dissimilarity in both spectral shape and in value.

In contrast with the broadband optical property phantoms, Fig. 4 shows the measured properties of the phantom made to match the absorption and reduced scattering coefficients of muscle at four distinct wavelength bands across the blue, green, red, and NIR regions of the spectra. Although the absorption coefficient of this phantom is still broadly shaped like that of tissue, many points outside of the bands deviate away from the target. This mainly occurs where a specific pigment was used to better match the absorption with the band. The two notable examples within this phantom are for the red band, where PB15:3 was used to better match the shape of the shoulder present within the band while also inadvertently introducing an absorption peak around 700 nm, and the blue band, where PY74 was used to try and match the rising slope of the absorption peak despite an overall redshift in the peak location.

Fig. 4

Multiband optical property phantom. Multipigment phantom matching the (a) absorption and (b) reduced scattering coefficients of muscle tissue at four spectral bands in the blue, green, red, and NIR regions of the spectrum, highlighted in gray.

JBO_28_2_025001_f004.png

When analyzing the quantitative metrics for the band fitted phantom, found in Table 3, the benefits of spectral selection can be seen. For all four bands, the MAE of both the absorption and reduced scattering are very low showcasing the ability of multipigment phantoms to accurately match the optical properties of tissue over multiple orders of magnitude. Additionally, each of these bands were found to have a very small SAM indicating that any wavelength-dependent trends present within the bands can be accurately emulated as well.

Table 3

Spectral similarity metrics of the multiband optical property phantom for human muscle tissue.

AbsorptionReduced scattering
MAESAM (°)MAESAM (°)
Blue0.0585.930.2090.02
Green0.0213.660.0150.29
Red0.0132.140.0620.71
NIR0.0081.030.0900.26
All bands0.0226.140.0803.92

Overall, the results from both the broadband and discreet band phantoms showcase that the multipigment approach to phantom design provides more tissue-like optical properties over a broad spectral range. With the ability to independently adjust the optical properties at different regions of the spectrum, multipigment phantoms showcase an increased ability to be tailored toward specific needs and applications.

3.3.

Diffuse Reflectance Phantoms

Figures 5(a) and 5(b) show the measured diffuse reflectance of the reflectance-matched phantoms compared to the NIST reported diffuse reflectance of pale (NIST #74) and melanistic (NIST #44) skin.34 From a visual inspection, it can be seen that the multipigment approach, coupled with the wMC model, was able to produce phantoms accurately matching the target diffuse reflectance over most of the spectrum. This is reflected in the quantitative similarity metrics found in Table 4. For both the pale and melanistic skin phantoms, the SAM and the MAE values were close to zero indicating that both the shapes and values of the measured reflectances are close to their targets, despite some apparent deviations.

Fig. 5

Broadband diffuse reflectance phantoms. Multipigment phantoms matching the broadband diffuse reflectance of (a) pale and (b) melanistic inner forearm skin reported by NIST.34 The spectral diffuse reflectances were also compared to the theoretical best fit of a traditional phantom made using only TiO2 and India ink.

JBO_28_2_025001_f005.png

Table 4

Similarity metrics for the multipigment diffuse reflectance phantoms.

NIST #74NIST #44
MultipigmentTraditional fitMultipigmentTraditional fit
MAE0.0210.1260.0120.152
SAM3.37°14.09°2.54°30.84°

For the pale skin phantom, the major deviations occur at the ends of the spectrum, with the absorption of the scattering agents decreasing the reflectance at the UV end while the lack of variety in absorbers for the NIR prevented accurate matching for that portion of the spectrum. Although the issue in the UV cannot easily be overcome, finding and including new pigments that absorb wavelengths longer than 750 nm would improve the matching in the NIR. In contrast to the pale skin phantom, the main deviations of the melanistic skin phantom occur in the middle of the spectrum. Here the three miscellaneous pigments were unable to bring enough variation outside the yellow and magenta pigments to be combined into an approximation of the smooth absorption profile of melanin, resulting in undesired fluctuations in the measured diffuse reflectance. As such, including more pigments with absorption peaks outside the blue and green regions of the spectrum should result in better emulation of broadband absorbers like melanin.

As with the optical property phantoms, the performance of the multipigment approach when creating broadband diffuse reflectance-matched phantoms was contextualized using the predicted diffuse reflectance of traditional phantoms matched to the full spectrum of the NIST skin reflectances. As seen in Fig. 5, traditional phantoms are expected to have relatively flat diffuse reflectance over the entire spectra, with a sharp decrease in the UV due to the absorption of titanium dioxide. As such, while traditional phantoms could be used to mimic the reflectance of pale skin in the NIR, they are overall limited to matching the value, but not the shape, of the target spectra at a single wavelength.

This limitation is reflected when comparing the quantitative similarity metrics found in Table 4. As compared to the multipigment phantoms, the predicted traditional phantoms were found to have MAEs and SAMs at least 4 times greater, indicating a much greater degree of dissimilarity with the target spectra. This difference is especially apparent for the melanistic skin with the predicted traditional phantom having both an MAE and SAM over 12 times greater than those of the multipigment phantom.

The overall similarity of the phantom and tissue reflectance spectra demonstrates the overall feasibility of using a nonlinear, least squares optimization coupled with a wMC model to produce tissue phantoms based on diffuse reflectance alone. Additionally, these results signify just how poorly traditional phantoms are at mimicking the diffuse reflectance of skin, especially that of darker skin tones. With the multipigment phantom approach’s ability to produce highly accurate, broadband, reflectance-matched phantoms for all skin types, they will be able to better serve as standards when designing and validation optical modalities that minimize racial or ethnic biases due to the presence of melanin.

3.4.

Oxygenation Saturation Phantoms

Figures 6(a)6(c) show the absorption coefficients of the theoretical skin model with blood StO2 values of 30%, 50%, and 80%, respectively, compared to the measured absorption coefficients of their corresponding phantoms. The comparison of reduced scattering coefficients can be found in Fig. S2 in the Supplementary Material. From a direct comparison, it can be seen that the phantom absorption coefficients are generally similar to their respective tissue models, with this being especially true within the three selected bands of 540 to 560 nm, 625 to 645 nm, and 820 to 840 nm. Using the absorption coefficients of the phantoms within these three bands, the respective StO2, blood volume fraction, and melanosome volume fraction were determined using fixed volume fractions of 0.5 and 0.02 for water and fat. The calculated values were compared to the expected values used to generate the theoretical skin optical properties in Table 5.

Fig. 6

Hemoglobin oxygen saturation phantoms. Multipigment phantoms matching the absorption coefficients of a computational skin model with StO2 values of (a) 30%, (b) 50%, and (c) 80% at three wavelength bands centered at 550, 635, and 830 nm, highlighted in gray.

JBO_28_2_025001_f006.png

Table 5

Measured chromophore concentrations of the multipigment oxygen saturation phantoms.

Expected (%)Measured (%)Error (%)
StO230.0032.949.81
Blood volume fraction4.784.62−3.44
Melanosome volume fraction0.02550.0209−18.04
StO250.0049.20−1.60
Blood volume fraction4.784.58−4.25
Melanosome volume fraction0.02550.0720182.35
StO280.0081.211.51
Blood volume fraction4.784.72−1.29
Melanosome volume fraction0.02550.02778.63

Corresponding to the similar optical properties within the three bands, the calculated chromophore values for the three phantoms were found to be physiologically relevant and relatively similar to the expected values from the tissue model. The parameter most prone to error in all three phantoms was found to be the melanosome volume fraction. With the skin model having a minimal amount of melanin, small inaccuracies in the red and NIR absorption bands resulted in small absolute, but large relative, changes in the calculated melanin levels. The worst case of this was for the 50% StO2 phantom where the absorption coefficients at 635 and 830 nm being 0.01 and 0.005  mm1 too high, respectively, resulted in a melanosome volume fraction 180% greater than the target value. However, the original volume fraction was set low enough that even this large increase kept the phantom properties within the bounds of pale skin.

Despite this variance in the emulated melanin content of the oxygenation phantoms, the overall blood volume fraction and the StO2 were found to be minimally divergent from the target values. The total blood volume fraction was found to be consistent across all three phantoms, tending to be slightly lower than the target value within a 5% margin of error. More importantly, a similar level of accuracy was also observed in the emulated StO2 of each phantom, despite the value varying between phantoms, with the 50% and 80% phantoms falling within a 2% error margin and the 30% phantom falling within a 10% error margin. As with the melanosome volume fraction, even with the largest discrepancies the phantoms maintained realistic values that fell within the expected variance of most pulse oximetry systems.50 This ability to produce phantoms that can mimic physiologically relevant concentrations of biological chromophores highlights the versatility of the multipigment phantom approach and opens up the possibility to further increase the number of optical modalities that can effectively use tissue phantoms to characterize and validate the performance of their systems.

4.

Conclusion

In this work, we have outlined a new platform for producing optical phantoms with wavelength-dependent properties that can be adjusted to match those of a desired tissue type over a broad spectral range of 370 to 950 nm through the combination of multiple absorbing and scattering pigments. This opens up the capabilities for solid, shelf-stable phantoms to serve as standards for a wide range of optical techniques which rely on multiple spectral bands or broadband signals. To demonstrate the capability of this approach, we produced multipigment phantoms matched to the optical properties of both human nerve and muscle tissue over the entire wavelength range with greater accuracy as what can be achieved using traditional phantom methods. We also showcased that through the addition of a reflectance model, broadband diffuse reflectance phantoms can be made for both pale and melanistic skin types without knowledge of the underlying tissue optical properties. Finally, we demonstrated that the spectral matching of this approach is accurate enough that inorganic phantoms can be made which emulate physiologically relevant concentrations of tissue chromophores, including representing the same tissue under varying states of hemoglobin StO2, a feat that, to our knowledge, had yet to been accomplished. Overall, we believe that this multipigment phantom platform represents a powerful new tool which should increase the availability of solid, stable, and accurate standards for many optical techniques including chromophore imaging, fluorescence imaging, and diffuse reflectance spectroscopy.

Disclosure

The authors declare that they have no relevant financial interests regarding the materials discussed in this paper and no potential conflicts of interest to disclose.

Acknowledgments

The authors would like to thank Ezekiel Haugen and Dr. Graham Throckmorton for their help in the early development of this project as well as Dr. Sylvain Gioux for invaluable suggestions regarding the StO2 phantoms. Financial support for this project was provided by a Vanderbilt University Seeding Success (Grant No. FF_300152).

Code, Data, and Materials Availability

The normalized pigment properties and the code used to create the phantom fits have been made available at GitHub: https://github.com/AlecWalter/MultipigmentPhantoms.

References

1. 

B. W. Pogue and M. S. Patterson, “Review of tissue simulating phantoms for optical spectroscopy, imaging and dosimetry,” J. Biomed. Opt., 11 (4), 041102 https://doi.org/10.1117/1.2335429 JBOPFO 1083-3668 (2006). Google Scholar

2. 

J. Hwang, J. C. Ramella-Roman and R. Nordstrom, “Introduction: feature issue on phantoms for the performance evaluation and validation of optical medical imaging devices,” Biomed. Opt. Express, 3 (6), 1399 https://doi.org/10.1364/BOE.3.001399 BOEICL 2156-7085 (2012). Google Scholar

3. 

K. E. Michaelsen et al., “Anthropomorphic breast phantoms with physiological water, lipid, and hemoglobin content for near-infrared spectral tomography,” J. Biomed. Opt., 19 (2), 026012 https://doi.org/10.1117/1.JBO.19.2.026012 JBOPFO 1083-3668 (2014). Google Scholar

4. 

J. C. Finlay and T. H. Foster, “Hemoglobin oxygen saturations in phantoms and in vivo from measurements of steady-state diffuse reflectance at a single, short source-detector separation,” Med. Phys., 31 (7), 1949 –1959 https://doi.org/10.1118/1.1760188 MPHYA6 0094-2405 (2004). Google Scholar

5. 

M. Majedy et al., “Spectral characterization of liquid hemoglobin phantoms with varying oxygenation states,” J. Biomed. Opt., 27 (7), 074708 https://doi.org/10.1117/1.JBO.27.7.074708 JBOPFO 1083-3668 (2021). Google Scholar

6. 

H. Jonasson, C. D. Anderson and R. B. Saager, “Water and hemoglobin modulated gelatin-based phantoms to spectrally mimic inflamed tissue in the validation of biomedical techniques and the modeling of microdialysis data,” J. Biomed. Opt., 27 (7), 074712 https://doi.org/10.1117/1.JBO.27.7.074712 JBOPFO 1083-3668 (2022). Google Scholar

7. 

M. T. Ghijsen et al., “Quantitative real-time optical imaging of the tissue metabolic rate of oxygen consumption,” J. Biomed. Opt., 23 (3), 036013 https://doi.org/10.1117/1.JBO.23.3.036013 JBOPFO 1083-3668 (2018). Google Scholar

8. 

L. Ntombela, B. Adeleye and N. Chetty, “Low-cost fabrication of optical tissue phantoms for use in biomedical imaging,” Heliyon, 6 (3), e03602 https://doi.org/10.1016/j.heliyon.2020.e03602 (2020). Google Scholar

9. 

G. Wagnières et al., “An optical phantom with tissue-like properties in the visible for use in PDT and fluorescence spectroscopy,” Phys. Med. Biol., 42 (7), 1415 –1426 https://doi.org/10.1088/0031-9155/42/7/014 PHMBA7 0031-9155 (1997). Google Scholar

10. 

P. Lai, X. Xu and L. v. Wang, “Dependence of optical scattering from Intralipid in gelatin-gel based tissue-mimicking phantoms on mixing temperature and time,” J. Biomed. Opt., 19 (3), 035002 https://doi.org/10.1117/1.JBO.19.3.035002 JBOPFO 1083-3668 (2014). Google Scholar

11. 

L. Spinelli et al., “Determination of reference values for optical properties of liquid phantoms based on Intralipid and India ink,” Biomed. Opt. Express, 5 (7), 2037 https://doi.org/10.1364/BOE.5.002037 BOEICL 2156-7085 (2014). Google Scholar

12. 

F. Ayers et al., “Fabrication and characterization of silicone-based tissue phantoms with tunable optical properties in the visible and near infrared domain,” Proc. SPIE, 6870 687007 https://doi.org/10.1117/12.764969 (2008). Google Scholar

13. 

G. J. Greening et al., “Characterization of thin poly(dimethylsiloxane)-based tissue-simulating phantoms with tunable reduced scattering and absorption coefficients at visible and near-infrared wavelengths,” J. Biomed. Opt., 19 (11), 115002 https://doi.org/10.1117/1.JBO.19.11.115002 JBOPFO 1083-3668 (2014). Google Scholar

14. 

A. M. Goldfain et al., “Polydimethylsiloxane tissue-mimicking phantoms with tunable optical properties,” J. Biomed. Opt., 27 (7), 074706 https://doi.org/10.1117/1.JBO.27.7.074706 JBOPFO 1083-3668 (2021). Google Scholar

15. 

J. Pekar and M. S. Patterson, “Fabrication and characterization of phantoms with tissue-like optical properties from 500 to 700 nm,” Med. Laser Appl., 25 (3), 147 –153 https://doi.org/10.1016/j.mla.2010.04.002 (2010). Google Scholar

16. 

P. Krauter et al., “Optical phantoms with adjustable subdiffusive scattering parameters,” J. Biomed. Opt., 20 (10), 105008 https://doi.org/10.1117/1.JBO.20.10.105008 JBOPFO 1083-3668 (2015). Google Scholar

17. 

D. Passos et al., “Tissue phantom for optical diagnostics based on a suspension of microspheres with a fractal size distribution,” J. Biomed. Opt., 10 (6), 064036 https://doi.org/10.1117/1.2139971 JBOPFO 1083-3668 (2005). Google Scholar

18. 

P. di Ninni, F. Martelli and G. Zaccanti, “The use of India ink in tissue-simulating phantoms,” Opt. Express, 18 (26), 26854 https://doi.org/10.1364/OE.18.026854 OPEXFF 1094-4087 (2010). Google Scholar

19. 

D. A. Loginova et al., “Liquid optical phantoms mimicking spectral characteristics of laboratory mouse biotissues,” Quantum Electron., 46 (6), 528 –533 https://doi.org/10.1070/QEL16133 (2016). Google Scholar

20. 

M. Lualdi et al., “A phantom with tissue-like optical properties in the visible and near infrared for use in photomedicine,” Lasers Surg. Med., 28 (3), 237 –243 https://doi.org/10.1002/lsm.1044 LSMEDI 0196-8092 (2001). Google Scholar

21. 

T. Moffitt, Y.-C. Chen and S. A. Prahl, “Preparation and characterization of polyurethane optical phantoms,” J. Biomed. Opt., 11 (4), 041103 https://doi.org/10.1117/1.2240972 JBOPFO 1083-3668 (2006). Google Scholar

22. 

ASTM, “Standard test methods for lightfastness of colorants used in artists’ materials,” (2022). Google Scholar

23. 

M. S. Wróbel et al., “Measurements of fundamental properties of homogeneous tissue phantoms,” J. Biomed. Opt., 20 (4), 045004 https://doi.org/10.1117/1.JBO.20.4.045004 JBOPFO 1083-3668 (2015). Google Scholar

24. 

M. S. Wróbel et al., “Multi-layered tissue head phantoms for noninvasive optical diagnostics,” J. Innov. Opt. Health Sci., 8 (3), 1541005 https://doi.org/10.1142/S1793545815410059 (2015). Google Scholar

25. 

A. Cosentino, “FORS spectral database of historical pigments in different binders,” e-Conserv. J., 2 54 –65 https://doi.org/10.18236/econs2.201410 (2014). Google Scholar

26. 

Y. Zhao and R. S. Berns, “Predicting the spectral reflectance factor of translucent paints using Kubelka–Munk turbid media theory: review and evaluation,” Color Res. Appl., 34 (6), 417 –431 https://doi.org/10.1002/col.20525 CREADU 0361-2317 (2009). Google Scholar

27. 

M. Longoni et al., “Non-invasive identification of synthetic organic pigments in contemporary art paints by visible-excited spectrofluorimetry and visible reflectance spectroscopy,” Spectrochim. Acta A Mol. Biomol. Spectrosc., 229 117907 https://doi.org/10.1016/j.saa.2019.117907s (2020). Google Scholar

28. 

M. Mohammadi et al., “Pigment selection using Kubelka–Munk turbid media theory and non-negative least square technique,” (2004). Google Scholar

29. 

S. A. Prahl, M. J. C. van Gemert and A. J. Welch, “Determining the optical properties of turbid media by using the adding–doubling method,” Appl. Opt., 32 (4), 559 –568 https://doi.org/10.1364/AO.32.000559 APOPAI 0003-6935 (1993). Google Scholar

30. 

S. A. Prahl, “Everything I think you should know about inverse adding-doubling,” https://omlc.org/software/iad/manual.pdf (2011). Google Scholar

31. 

Y. Xie et al., “Methodology development and evaluation of optical properties characterisation of small size tissue samples,” (2021). https://doi.org/10.48550/arXiv.2107.04909 Google Scholar

32. 

J. Shapey et al., “Optical properties of human brain and tumour tissue: an ex vivo study spanning the visible range to beyond the second near‐infrared window,” J. Biophotonics, 15 (4), e202100072 https://doi.org/10.1002/jbio.202100072 (2022). Google Scholar

33. 

E. Haugen et al., “Ex vivo determination of nerve and surrounding tissue optical properties in rats and human cadavers for nerve-sparing surgical applications,” Proc. SPIE, PC11949 PC119490D https://doi.org/10.1117/12.2608860 (2022). Google Scholar

34. 

C. C. Cooksey, D. W. Allen and B. K. Tsai, “Reference data set of human skin reflectance,” J. Res. Natl. Inst. Stand. Technol., 122 (26), 26 https://doi.org/10.6028/jres.122.026 JRITEF 1044-677X (2017). Google Scholar

35. 

E. Alerstam, S. Andersson-Engels and T. Svensson, “White Monte Carlo for time-resolved photon migration,” J. Biomed. Opt., 13 (4), 041304 https://doi.org/10.1117/1.2950319 JBOPFO 1083-3668 (2008). Google Scholar

36. 

D. J. Cuccia et al., “Quantitation and mapping of tissue optical properties using modulated imaging,” J. Biomed. Opt., 14 (2), 024012 https://doi.org/10.1117/1.3088140 JBOPFO 1083-3668 (2009). Google Scholar

37. 

H.-W. Wang et al., “Broadband reflectance measurements of light penetration, blood oxygenation, hemoglobin concentration, and drug concentration in human intraperitoneal tissues before and after photodynamic therapy,” J. Biomed. Opt., 10 (1), 014004 https://doi.org/10.1117/1.1854679 JBOPFO 1083-3668 (2005). Google Scholar

38. 

E. Aguénounon et al., “Real-time optical properties and oxygenation imaging using custom parallel processing in the spatial frequency domain,” Biomed. Opt. Express, 10 (8), 3916 https://doi.org/10.1364/BOE.10.003916 BOEICL 2156-7085 (2019). Google Scholar

39. 

A. Mazhar et al., “Wavelength optimization for rapid chromophore mapping using spatial frequency domain imaging,” J. Biomed. Opt., 15 (6), 061716 https://doi.org/10.1117/1.3523373 JBOPFO 1083-3668 (2010). Google Scholar

40. 

X. Chen et al., “In vivo real-time imaging of cutaneous hemoglobin concentration, oxygen saturation, scattering properties, melanin content, and epidermal thickness with visible spatially modulated light,” Biomed. Opt. Express, 8 (12), 5468 https://doi.org/10.1364/BOE.8.005468 BOEICL 2156-7085 (2017). Google Scholar

41. 

R. L. P. van Veen et al., “Determination of VIS-NIR absorption coefficients of mammalian fat, with time and spatially resolved diffuse reflectance and transmission spectroscopy,” in Biomed. Top. Meet., SF4 (2004). Google Scholar

42. 

R. M. Pope and E. S. Fry, “Absorption spectrum (380–700 nm) of pure water. II. Integrating cavity measurements,” Appl. Opt., 36 (33), 8710 –8723 https://doi.org/10.1364/AO.36.008710 APOPAI 0003-6935 (1997). Google Scholar

43. 

G. M. Hale and M. R. Querry, “Optical constants of water in the 200-nm to 200-μm wavelength region,” Appl. Opt., 12 (3), 555 –563 https://doi.org/10.1364/AO.12.000555 APOPAI 0003-6935 (1973). Google Scholar

44. 

S. L. Jacques, “Optical properties of biological tissues: a review,” Phys. Med. Biol., 58 (11), R37 –R61 https://doi.org/10.1088/0031-9155/58/11/R37 PHMBA7 0031-9155 (2013). Google Scholar

45. 

I. v. Meglinski and S. J. Matcher, “Quantitative assessment of skin layers absorption and skin reflectance spectra simulation in the visible and near-infrared spectral regions,” Physiol. Meas., 23 (4), 741 –753 https://doi.org/10.1088/0967-3334/23/4/312 PMEAE3 0967-3334 (2002). Google Scholar

46. 

T. E. Travis et al., “A multimodal assessment of melanin and melanocyte activity in abnormally pigmented hypertrophic scar,” J. Burn Care Res., 36 (1), 77 –86 https://doi.org/10.1097/BCR.0000000000000154 (2015). Google Scholar

47. 

T. Lister, P. A. Wright and P. H. Chappell, “Optical properties of human skin,” J. Biomed. Opt., 17 (9), 090901 https://doi.org/10.1117/1.JBO.17.9.090901 JBOPFO 1083-3668 (2012). Google Scholar

48. 

M. Agarla et al., “An analysis of spectral similarity measures,” Color Imaging Conf., 29 (1), 300 –305 https://doi.org/10.2352/issn.2169-2629.2021.29.300 (2021). Google Scholar

49. 

H. Deborah, N. Richard and J. Y. Hardeberg, “A comprehensive evaluation of spectral distance functions and metrics for hyperspectral image processing,” IEEE J. Sel. Top. Appl. Earth Obs. Remote Sens., 8 (6), 3224 –3234 https://doi.org/10.1109/JSTARS.2015.2403257 (2015). Google Scholar

50. 

M. Nitzan, A. Romem and R. Koppel, “Pulse oximetry: fundamentals and technology update,” Med. Devices: Evid. Res., 7 (1), 231 https://doi.org/10.2147/MDER.S47319 (2014). Google Scholar

Biography

Alec B. Walter received his BS degree in electrical engineering and biology from the University of Arkansas in 2016 and his MS degree in biomedical engineering from Vanderbilt University in 2019. He is currently a PhD candidate in the Department of Biomedical Engineering at Vanderbilt University, focusing on antimicrobial photodynamic therapy, real-time dosimetry, and spatial frequency domain imaging.

E. Duco Jansen received his MS degree in medical sciences from the University of Utrecht and his MS and PhD degrees in biomedical engineering from the University of Texas at Austin. He is currently a professor of biomedical engineering and neurosurgery at Vanderbilt University and is the senior associate dean for graduate education and faculty affairs in the School of Engineering. His research focus is on application of light, lasers, and optical technology in medicine and biology.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Alec B. Walter and E. Duco Jansen "Development of a platform for broadband, spectra-fitted, tissue optical phantoms," Journal of Biomedical Optics 28(2), 025001 (20 February 2023). https://doi.org/10.1117/1.JBO.28.2.025001