25 April 2024 Margin assessment during breast conserving surgery using diffuse reflectance spectroscopy
Author Affiliations +
Abstract

Significance

During breast-conserving surgeries, it is essential to evaluate the resection margins (edges of breast specimen) to determine whether the tumor has been removed completely. In current surgical practice, there are no methods available to aid in accurate real-time margin evaluation.

Aim

In this study, we investigated the diagnostic accuracy of diffuse reflectance spectroscopy (DRS) combined with tissue classification models in discriminating tumorous tissue from healthy tissue up to 2 mm in depth on the actual resection margin of in vivo breast tissue.

Approach

We collected an extensive dataset of DRS measurements on ex vivo breast tissue and in vivo breast tissue, which we used to develop different classification models for tissue classification. Next, these models were used in vivo to evaluate the performance of DRS for tissue discrimination during breast conserving surgery. We investigated which training strategy yielded optimum results for the classification model with the highest performance.

Results

We achieved a Matthews correlation coefficient of 0.76, a sensitivity of 96.7% (95% CI 95.6% to 98.2%), a specificity of 90.6% (95% CI 86.3% to 97.9%) and an area under the curve of 0.98 by training the optimum model on a combination of ex vivo and in vivo DRS data.

Conclusions

DRS allows real-time margin assessment with a high sensitivity and specificity during breast-conserving surgeries.

1.

Introduction

Breast cancer is the most frequently diagnosed type of cancer among women globally, with an estimated incidence of 2.26 million cases in the year 2020.1 It is the leading cause of cancer death among females, with an estimated number of 685,000 deaths in the year 2020.1 The standard of care for early-stage breast cancer is breast-conserving therapy (BCT), which involves breast-conserving surgery (BCS) followed by adjuvant radiotherapy.2,3 In BCS, the surgeon attempts to excise the entire tumor along with a small margin of healthy tissue while providing a satisfactory cosmetic outcome. Multiple randomized clinical trials have demonstrated equivalent survival outcomes for mastectomy and BCT if clear resection margins (edges of a breast cancer specimen) are achieved during BCS.47 To assess whether the margins of a specimen are clear, and therefore no cancerous cells remain in the adjacent tissue, a histopathological examination is performed after surgery. If positive (tumor-involved) margins are found, the patient is usually recommended to undergo a re-excision and/or additional doses of radiotherapy,8 as tumor positive margins are associated with a twofold increase in ipsilateral breast tumor recurrence and an overall lower long-term survival.913 The required additional therapy in patients with positive margins may result in anxiety, increased clinical complications,14,15 unsatisfactory cosmetic outcome, and1619 increased health care costs.15,2022 Globally, there is a lack of consensus on what constitutes a “positive margin,” and the reported rates of positive margins after BCS in the literature vary from 9% to as high as 36% for invasive breast cancer and from 4% to as high as 23% for ductal carcinoma in situ (DCIS).23 According to Dutch guidelines, a positive margin is defined as invasive carcinoma (IC) cells reaching the inked margin over a trajectory >4  mm or DCIS cells reaching the inked margin over any trajectory.24 Focally positive margins are defined as IC cells reaching the inked margin over a trajectory 4  mm.24 Otherwise, the margins are considered negative. For the United States, according to the guidelines of the Society of Surgical Oncology (SSO) and the American Society for Radiation Oncology (ASTRO), a margin is considered tumor-positive when ink touches the IC or when there is DCIS present within 2 mm from the resection margin.9

During BCS, the surgeon has to rely on visual and haptic inspection of the tissue to obtain information about the surgical margins. The definitive margin status is only reported following the histopathological examination, which usually occurs a few days after surgery. Therefore, there is a clinical need for methods that are able to support the surgeon’s evaluation of resection margins in real time and accurately during BCS. Current techniques for intraoperative margin assessment involve specimen radiography, frozen section analysis, and touch preparation cytology. However, these techniques are not incorporated into the standard surgical workflow due to low diagnostic accuracy, low speed, and cumbersome workload for pathologists and/or radiologists.25,26 In light of this, a wide variety of approaches with the overarching goal of intraoperative margin assessment is being investigated.2729 These techniques include intraoperative ultrasonography,3032 radiofrequency spectroscopy,3336 bioimpedance spectroscopy,37 Raman spectroscopy,3840 digital breast tomosynthesis,41 fluorescence imaging,42 microcomputed tomography,43 optical coherence tomography,4446 quantitative micro-elastography,47 ultraviolet photoacoustic microscopy,48,49 intraoperative flow cytometry,50 and microscopy with ultraviolet surface excitation.51 Although some of the mentioned techniques seem promising, none of them have been translated into routine surgical practice yet. The reasons for this lack of implementation include low diagnostic accuracy,41,43,52 time-consuming process, difficult interpretation, high operator dependence, inability to perform over the entire margin,50 early stage of development, and/or undetermined cost-effectiveness.27,28,53,54

Fiber-optic diffuse reflectance spectroscopy (DRS) is an optical technique that is also being investigated for the purpose of margin assessment in cancer surgeries. In DRS, a beam of light is directed through an emitting fiber into the tissue, where it undergoes several light-tissue interactions, such as scattering and absorption, before the light is partially reflected back. The intensity of the diffusely reflected light at various wavelengths is recorded, producing a diffuse reflectance spectrum. By analyzing this spectrum, information can be obtained about the composition and optical properties of the illuminated tissue. This information could be used to distinguish different tissue types and thus potentially detect cancerous tissue during cancer surgeries.

In recent years, our research group has conducted multiple studies to investigate the combination of DRS with classification models to distinguish tumorous tissue from healthy tissue on sliced, ex vivo breast cancer specimens with excellent results, including a sensitivity and specificity of 100% in distinguishing tumorous breast tissue from healthy breast tissue.55,56 In these studies, DRS measurements were performed on pure healthy tissue locations and pure tumorous tissue locations on sliced breast cancer specimens. In our latest publication, we reported the results of our study on the diagnostic accuracy of DRS combined with classification models, as a margin assessment tool during BCS. We applied DRS on the actual resection margin of ex vivo breast cancer specimens rather than sliced specimens and found a sensitivity of 93% and a specificity of 75% for distinguishing tumorous tissue from healthy tissue up to 2 mm from the margin, when using five optical fibers.57

The next step is to move forward to real-time clinical application during breast-conserving surgeries and investigate if these classification models that are trained and tested on DRS data acquired from ex vivo breast tissue have a similar performance when applied to DRS data from in vivo breast tissue. This is relevant because acquiring data in vivo is more complex as it is time-consuming and usually limited to a small number of measurements per patient. Therefore, it is worthwhile to investigate if one could mainly use ex vivo data for training a classification model without compromising the performance when used in vivo.

The first aim was to investigate the performance of DRS combined with classification algorithms, trained on DRS data acquired from ex vivo and/or in vivo breast tissue, and tested on data acquired from in vivo breast tissue. For this, we acquired an extensive dataset of DRS data from ex vivo and in vivo breast tissue. We built four different types of classification models, and evaluated their classification performance when trained on the different types of data (ex vivo/in vivo). The second aim was to select the optimum model based on the highest classification performance and investigate the model training strategy that yields the best performance for the selected classification model.

2.

Materials and Methods

2.1.

Diffuse Reflectance Spectroscopy Setup

The instrumentation of the optical measurement system consisted of a console with five light sources for tissue illumination, two different spectrometers, and an in-house developed, handheld fiber-optic DRS probe. The light sources were identical, halogen broadband lamps (Avantes, AvaLight-HAL, 360 to 2500 nm) with integrated shutters. Of both spectrometers, one covered a visible wavelength range from 200 to 1160 nm (Avantes, AVASPEC-HS2048XL-EVO), and the other covered a near-infrared wavelength range from 900 to 1750 nm (Avantes, AVASPECNIR256-1.7-RS). The DRS probe consisted of five peripheral illumination fibers placed in a circle around one receiving fiber in the center. Each light source was coupled to one illumination fiber. The source–detector fiber distance for all fibers was 2.0 mm. All diffuse reflectance spectra were calibrated on a white reference object of spectralon in a similar manner as described in Refs. 56 and 58. Therefore, any disparity between the light sources would be inconsequential. A schematic representation of the optical measurement system is given in Sec. 1 in the Supplementary Material.

2.2.

Study Design and Participants

This prospective non-randomized cohort study was conducted from 2019 to 2023 at the Netherlands Cancer Institute - Antoni van Leeuwenhoek Hospital (NKI-AvL). Ethical approval for the study protocol was granted by the Institutional Review Board. Patients undergoing breast-conserving surgery due to histologically proven IC and/or DCIS, with or without neoadjuvant therapy, were included. Patients with a complete radiologic response after neoadjuvant therapy were excluded. The study period consisted of two phases. During the first phase, patients were included for only ex vivo tissue measurements on BCS specimens. According to the medical research involving human subjects act, no written consent was required for these measurements. During the second phase of the study, patients were included for in vivo as well as ex vivo tissue measurements during BCS. All included participants for these measurements had provided written informed consent and were treated according to local standard protocols.

2.3.

Data Acquisition

A schematic representation of the data acquisition process is shown in Fig. 1. The following patient characteristics were collected from medical records: age at the time of surgery, administration of neoadjuvant therapy, pathological diagnosis of the resected specimen, and margin status.

Fig. 1

Overview of the data acquisition method for phase I (top left) and phase II (top right). In panel (a), the image displays a point-based DRS measurement on an ex vivo BCS specimen. This is followed by (d) (middle row) marking the measurement location with black pathology ink, (e) standard processing by the pathology department, including coloring, and (f) slicing the specimen. In phase II, the workflow starts with (b) point-based DRS measurements on a volume in vivo breast tissue that is meant to be resected, followed by (c) DRS measurement on an ex vivo BCS specimen. This is followed by the same steps (d)–(f) as explained for phase I. A schematic overview of the method for determining the tumor-margin distance and tissue area percentages in the corresponding H&E sections (bottom). In panel (g), the original H&E section with the annotated borders of the lesion in red and the black ink along the margin at the measurement location (blue arrows) and (h) the magnified image of the measured tissue location (blue arrows). The yellow arrows indicate the maximum, central, and minimum tumor-margin distance, determined by measuring the perpendicular distance from the surgical surface to the tumor. Finally, the percentage of tumorous tissue and healthy breast tissue was determined over a depth of 2 mm at the marked region (i), indicated by the blue box (h).

JBO_29_4_045006_f001.png

2.3.1.

Measurements phase I

Directly after surgery, the BCS specimen was collected from the surgical team, and DRS measurements were performed after an estimated time interval of a few minutes. We selected approximately three to five locations on the margins of the specimen for DRS measurements. A total of three consecutive DRS measurements were performed on each selected location while ensuring the surface of the probe tip was in contact with the tissue. After the measurements, the measured tissue location was marked with black pathology ink. Considering the overall limited number of measurements that were possible due to the histopathology protocol in our hospital, we performed ultrasound imaging on some of the specimens to localize the areas with a short distance from the tumor edge to the resection margin.

2.3.2.

Measurements phase II

During this phase, in vivo measurements as well as ex vivo measurements were performed. Prior to the surgery, the optical measurement system was set up. During the standard surgical procedure, optical measurements were obtained by the surgeon at one location on each of the six resection sides of the tissue volume meant to be resected (comparable to six sides of a cube). Each location was selected based on a suspected short distance between the margin and the tumor edge after visual and haptic inspection by the surgeon. Similar to the ex vivo measurements, three consecutive DRS spectra were acquired on each location. This was followed by the placement of a surgical clip to mark it. The acquisition time of each measurement was a few seconds. After surgery, we removed each surgical clip on the in vivo measurement locations and directly marked it again with black pathology ink. Furthermore, we performed DRS measurements on one or two additionally selected locations as per the measurement protocol of phase I.

2.3.3.

Specimen handling and ground truth labels

After DRS measurements, the specimen was delivered to the pathology department for standard pathology processing. This processing entailed inking the resection margins for orientation purposes, freezing the specimen, slicing the specimen in a serial manner, and processing the slices into cellular thin hematoxylin and eosin-stained (H&E) sections. The digitized pathological H&E sections of all measurement locations were examined by an experienced pathologist. During this process, the pathologist annotated all areas of IC and DCIS in the H&E images. Subsequently, a region up to 2 mm underneath the black ink mark was determined. The area percentage of tumor tissue within this region was determined based on the annotations, using image processing tools in MATLAB. The remaining healthy tissue in the region consisted of fat and connective tissue. The area percentages of both healthy tissue types were determined using a threshold value for the intensity in the green channel of the H&E image, as described in Kho et al.59 These tissue percentages together form the ground truth labels of the dataset. If any tumor tissue was present within the determined region up to a depth of 2 mm, the location was labeled as malignant. The rationale behind choosing this depth is the definition of a positive margin according to the SSO ASTRO guidelines mentioned earlier.9 If no tumor tissue was present in the region, the location was labeled as healthy. Additionally, if any tumor percentage was present within the region, the maximum, minimum, and central perpendicular distance from the black ink to the tumor edge was determined. The tumor-margin distance was calculated as the average of the aforementioned distances.

2.4.

Data Processing

For all data analyses, we used MATLAB version R2022a (MathWorks Inc., Natick, Massachusetts, United States). The entire data analysis process is displayed in Fig. 2.

Fig. 2

Overview of the workflow for data analysis. The process started with data preprocessing, which entailed spectrum normalization. The plotted spectra on the left are examples of characteristic spectra from locations with IC, DCIS, and fat, respectively, connective tissue. The plots on the right are the same spectra after normalization was applied. The next step was feature reduction, which involved feature extraction and feature selection. Subsequently, the selected features coupled with the ground truth labels derived from the histopathological analysis were used to train different classification models using fivefold cross-validation. The performance of each model was evaluated.

JBO_29_4_045006_f002.png

2.4.1.

Spectra normalization

The three consecutively collected diffuse reflectance spectra of each measurement location were averaged, and the spectra from both wavelength ranges were stitched together. Moreover, the extremities of the spectral range were removed due to a large level of noise relative to the signal being measured. These extremities consisted of the spectral range between 350 and 400 nm and the spectral range between 1600 and 1700 nm. As a result, the remaining data were of higher quality and more reliable for further processing. A multiplicative scatter correction (MSC) normalization approach was applied60 to correct the spectra in a manner that they are as close as possible to the mean spectrum of the dataset. By applying this technique to the dataset, intensity differences due to variability in light scattering effects and path length were corrected. Furthermore, random inter-patient variations and random variations in background light and signal quality were also reasonably reduced.

2.4.2.

Feature extraction and selection

Each normalized spectrum consisted of the reflection intensities at the wavelength range 400 to 1600 nm, meaning 1200 different wavelengths. Each reflection intensity represents a feature that could be used for training classification models. To reduce overfitting and improve classification performance, we performed feature extraction and selection as described in de Boer et al.56 The extracted features were a set of visible spectral features in the near-infrared wavelength range, including the slopes of spectra between the maximum intensities in the first 25% and the last 25% of particular wavelength ranges, the maximum difference between the slope and the spectrum, the corresponding wavelength at the point of maximum difference, and the inflection points left and right of the point of maximum difference, as described by de Boer et al.56 This set included 80 different features over the spectral range that we measured. Because we measured using 5 different fibers, in total 400 features could be extracted (80 features per fiber). This was followed by feature selection, which entailed applying a minimum redundancy maximum relevance (MRMR) feature selection algorithm.61 MRMR is used to rank the importance of all quantified features for the binary classification of each spectrum into belonging to either tumorous tissue or healthy tissue. Features with an importance score 0.015 were selected.

2.5.

Classification

Four different types of supervised machine learning classification models were built: (1) linear support vector machine (SVM); (2) quadratic SVM, (3) weighted K-nearest neighbors (KNNs) algorithm, and (4) ensemble random under sampling boosted tree (RUSBoost). We selected these particular model types because our research group has built similar models for DRS data classification during earlier studies, which showed high classification performance.56,57,6264 Three classification models per model type were built, each trained on one of the following types of data: (1) spectral features from ex vivo tissue locations, (2) spectral features from in vivo tissue locations, and (3) spectral features from ex vivo and in vivo tissue locations. Then, each classification model was tested on spectral features from in vivo tissue locations. For each experiment, the data were divided into training and test sets using a repeated 5-fold cross-validation method, with 20 iterations. The data from each patient were assigned to either the training set or the test set during each iteration to avoid bias.

2.6.

Performance Evaluation

First, we evaluated the performance of each classification model using the following metrics, averaged over 20 iterations: Matthews correlation coefficient (MCC) with corresponding standard deviation (SD). The MCC ranges between −1 and 1 and indicates the agreement between the actual and predicted labels. This metric considers the true negative (N+) rate, the true positive (P+) rate, the false negative (N) rate, and the false positive (P) rate and is calculated as follows:

Eq. (1)

MCC=(P+×N+P×N)(P++P)(P++N)(N++P)(N++N).

It is a robust measure that only produces a high score if the prediction has good rates in all four categories. This metric takes imbalance between classes into account. We compared the MCC of the different classification models, when trained on the three different types of data. Subsequently, we selected the model type with the overall highest MCC for further analysis. We plotted receiver operating characteristic (ROC) curves of the selected models trained on the three different types of data. Subsequently, we selected the optimal operating point for each ROC curve, which was defined as the point with the minimum average misclassification cost. The sensitivity and specificity with the corresponding 95% confidence intervals (CI) at each of these points were determined. The threshold for calculating sensitivity and specificity was selected based on the highest MCC value. Finally, we plotted the tumor-margin distance against the tumor percentage of all correctly and incorrectly classified tumor locations. We investigated the correlation between these parameters and the misclassification of tumor locations using the aforementioned plot.

3.

Results

3.1.

Characteristics of Patient Cohort

A total of 228 patients were recruited into this study. An overview of the patient and tumor characteristics as well as measurement locations of all included patients in phase I (N=128) and phase II (N=100) are given in Table 1. The median age of the patients in phase I was 60.5 years (SD = 11.9) compared with 62.0 years (SD = 13.1) in phase II (Table 1). The median lesion diameter was 13.5 mm (SD = 12.6) in phase I compared with 14.0 (SD = 13.5) in phase II (Table 1). Both the median age and the median lesion diameter were comparable in both groups. Regarding neoadjuvant treatment (chemotherapy or endocrine therapy), both patient groups had a comparable rate of patients without neoadjuvant treatment (63.3% of patients in phase I compared with 62% of patients in phase II) (Table 1). In phase I, a slightly higher percentage of patients had received neoadjuvant chemotherapy (27.3%) compared with phase II (22%) (Table 1). In phase II, a slightly higher percentage of patients had received neoadjuvant endocrine therapy (16%) compared with phase I (9.4%) (Table 1). It should be mentioned that many different (combinations of) malignant pathological diagnoses have been included in both phases and the distribution was quite comparable, as can be seen in Table 1. The majority of all patients in phase I (42,2%) and phase II (39%) had an IC NST (IC of no special type). This was followed by IC NST combined with DCIS (28.1% in phase I and 36% in phase II) (Table 1). When looking at margin status, according to Dutch guidelines, 14,0% of patients in phase I had positive margins compared with 12% of patients in phase II (Table 1). In total, patients in phase I had a slightly higher rate of patients with focally positive margins (18%) compared with patients in phase II (10%) (Table 1).

Table 1

Characteristics of patient cohort and measurement locations.

Phase I: Ex vivo (N=128)Phase II: In vivo and ex vivo (N=100)
Age (years) (median, SD)60,5 (11,9)62,0 (13,1)
Lesion diameter (mm) (median, SD)13,5 (12,6)14,0 (13,5)
Neoadjuvant treatment
 Chemotherapy35 (27,3%)22 (22%)
 Endocrine therapy12 (9,4%)16 (16%)
 None81 (63,3%)62 (62%)
Histological tumor type
 IC NST54 (42,2%)39 (39%)
 IC NST + DCIS36 (28,1%)36 (36%)
 IC NST + LCIS4 (3,1%)1 (1%)
 ILC13 (10,2%)11 (11%)
 ILC + LCIS6 (4,7%)2 (2%)
 ILC + DCIS1 (0.8%)0
 DCIS13 (10,1%)8 (8%)
 LCIS1 (0.8%)2 (2%)
 DCIS + LCIS01 (1%)
Margin status
 Negative87 (68,0%)78 (78%)
 Focally positive23 (18,0%)10 (10%)
 Positive18 (14,0%)12 (12%)
Measurement locations
 Tumor locations in vivo31
 Tumor locations ex vivo10646
 Healthy locations in vivo483
 Healthy locations ex vivo44780
IC NST, invasive carcinoma of no special type; DCIS, ductal carcinoma in situ; ILC, invasive lobular carcinoma; and LCIS, lobular carcinoma in situ.

3.2.

Measurement Locations

Of all 553 measurement locations acquired in phase I, there were 106 (19.2%) ex vivo tumor locations (tumor tissue within 2 mm from the margin) and 447 (80.8%) ex vivo healthy locations (healthy tissue within 2 mm from the margin) (Table 1). Among the 640 measurement locations acquired in phase II, there were 46 (7.2%) ex vivo tumor locations, 31 (4.8%) in vivo tumor locations, 80 (12.5%) ex vivo healthy locations, and 483 (75.5%) in vivo healthy locations (Table 1).

Figure 3 displays the distribution of the area percentages of all tissue types in the correlated H&E region of each ex vivo measurement location (top) and each in vivo measurement location (bottom). The tissue type distributions are quite similar for both groups, except for four notable differences. The first difference is that the percentage of measurement locations with tumor tissue (either IC or DCIS) is higher in the ex vivo group (22%) compared with the in vivo group (6%) (Fig. 3). The second difference is that the average percentage of tumor tissue in the tumor locations is higher among the in vivo locations compared with the ex vivo locations (Fig. 3). The third difference is that a higher percentage of tumor locations in the ex vivo group contain DCIS (22%) compared with the tumor locations in the in vivo group (10%) (Fig. 3). The fourth difference is that the locations in the in vivo group on average have a slightly higher area percentage of fat tissue compared with the ex vivo group (Fig. 3).

Fig. 3

(a) Distribution of area percentages per tissue type for all ex vivo measurement locations. (b) Distribution of area percentages per tissue type for all in vivo measurement locations.

JBO_29_4_045006_f003.png

3.3.

Selected Features

The MRMR analysis yielded 21 optimum features per spectrum. Table 2 gives an overview of all selected optimum features, ranked by feature importance score. It is important to mention that all of the described features are in the near-infrared wavelength range. In this particular range, fat, water, and collagen are the primary absorbers in tissue that affect the interaction with light.

Table 2

Optimum features according to MRMR analysis, ranked by feature importance score.

RankSelected feature
1Slope of the measured spectrum between 1213 and 1248 nm
2Slope of the measured spectrum between 931 and 1195 nm
3Wavelength of the left inflection point of the peak between 1224 and 1331 nm
4Maximum difference between the slope and the measured spectrum between 1142 and 1225 nm
5Maximum difference between the slope and the measured spectrum between 1021 and 1102 nm
6Slope of the measured spectrum between 999 and 1034 nm
7Slope of the measured spectrum between 1395 and 1430 nm
8Wavelength of the right inflection point of the peak between 1404 and 1437 nm
9Slope of the measured spectrum between 1467 and 1502 nm
10Wavelength of the right inflection point of the peak between 1142 and 1225 nm
11Slope of the measured spectrum between 1121 and 1401 nm
12Slope of the measured spectrum between 932 and 967 nm
13Slope of the measured spectrum between 1112 and 1147 nm
14Wavelength of the right inflection point of the peak between 1224 and 1331 nm
15Slope of the measured spectrum between 1201 and 1236 nm
16Wavelength of the right inflection point of the peak between 1382 and 1574 nm
17Wavelength of the right inflection point of the peak between 1404 and 1437 nm
18Slope of the measured spectrum between 883 and 1149 nm
19Maximum difference between the slope and the measured spectrum between 1224 and 1329 nm
20Slope of the measured spectrum between 932 and 967 nm
21Slope of the measured spectrum between 926 and 1149 nm

3.4.

Classification Performance of Four Types of Classification Models Trained on Different Types of Data

Table 3 gives an overview of the MCCs for the discrimination of tumor locations from healthy locations for four types of classification models, trained on three different types of data and tested on the in vivo dataset. From this table, it is apparent that training on DRS data acquired from ex vivo tissue yields similar performance results compared with training on DRS data acquired from in vivo tissue, regardless of model type. Furthermore, it is noticeable that the RUSBoost has the overall best classification performance on any type of training data compared with the other model types (Table 3). Overall, the highest classification performance is achieved by the RUSBoost trained on a combination of in vivo and ex vivo data, with an MCC of 0.76 (Table 3). The quadratic SVM and weighted KNN have the lowest performance with an MCC ranging between 0.57 and 0.59 and 0.58 and 0.61, respectively, depending on the type of training data (Table 3). For the remainder of the study, we selected the RUSBoost as the optimum model type for further analysis.

Table 3

Classification performance for four different types of classification models, when trained on different types of DRS data and tested on in vivo data.

Type of training dataLinear SVMQuadratic SVMWeighted KNNRUSBoost
MCC (SD)Ex vivo0.67 (0.007)0.57 (0.016)0.59 (0.007)0.73 (0.010)
In vivo0.62 (0.055)0.58 (0.048)0.58 (0.036)0.74 (0.029)
In vivo + Ex vivo0.71 (0.037)0.59 (0.055)0.61 (0.021)0.76 (0.028)

3.5.

Effect of Different Types of Training Data on Classification Performance of Optimum Model Type

Figure 4 shows the ROC curves and the area under the curve (AUC) for the RUSBoost, when trained on different types of training data. It is noteworthy to mention that the AUC of all differently trained models is similarly high, indicating a high classification performance (Fig. 4).

Fig. 4

ROC curves for RUSBoost model when trained on different types of data.

JBO_29_4_045006_f004.png

Table 4 displays the sensitivity and specificity with the 95% CI at the optimal operating point of each of the respective ROC curves. Training on a combination of ex vivo and in vivo data yields a slightly higher sensitivity (96.7% (95% CI 95.6% to 98.2%) and specificity (90.6% (95% CI 86.3% to 97.9%) compared with models trained on either ex vivo or in vivo data. This finding is in accordance with the marginally higher MCC (Table 3) and AUC (Fig. 4) of this model.

Table 4

Sensitivity and specificity of the RUSBoost model for distinguishing tumorous tissue from healthy breast tissue in vivo, when trained on different types of data.

Sensitivity (%) (95% CI)Specificity (%) (95% CI)
Trained onin vivodata96.1 (94.6 to 97.9)87.6 (70.2 to 93.4)
Trained onex vivodata96.0 (94.5 to 97.3)85.7 (77.1 to 89.2)
Trained onex vivoandin vivodata96.7 (95.6 to 98.2)90.6 (86.3 to 97.9)

3.6.

Classifiction of Tumor Locations In Vivo

Figure 5 displays the correctly and incorrectly classified tumor locations (IC and DCIS) with their corresponding tumor-margin distance and area percentage of tumor for the in vivo measurements. Overall, there appears to be a negative, linear correlation between tumor percentage and the distance to the tumor margin. This was found in our earlier research as well.65 This occurs because the tumor percentage for each measurement location is evaluated within a two-dimensional region extending up to 2 mm from the margin. When the average distance to the tumor is greater, it is likely that there will be a lower percentage of tumor tissue in that area. In total, there were 31 in vivo tumor locations in the entire dataset (Table 1). As can be seen in Fig. 5, most of the tumor locations (28) were identified correctly, and only 3 tumor locations were misclassified. All misclassified locations have a distance >1.0  mm from the resection margin, which according to the Dutch guidelines is not considered a positive margin.24 In earlier research, we found that similar DRS classification models are more prone to misclassifying ex vivo tumor locations with a higher tumor-margin distance and lower tumor area percentage.65 A similar trend can be seen in Fig. 5.

Fig. 5

Tumor-margin distance compared with tumor area percentage of all correctly and incorrectly classified IC and DCIS locations in vivo.

JBO_29_4_045006_f005.png

4.

Discussion

In breast-conserving surgeries, it is essential to ensure negative resection margins to minimize the risk of local tumor recurrence and the potential need for subsequent surgeries or additional radiotherapy. However, no standard method is employed yet to support margin assessment by surgeons. With the ability to discriminate tumorous tissue from healthy tissue with high accuracy on the margin of ex vivo BCS specimens,65 DRS has the potential to be applied as a real-time margin assessment tool during breast-conserving surgeries. However, optimal strategies for selecting training data and the performance of different DRS classification models on breast tissue, which can subsequently be applied in vivo during BCS, has not been fully evaluated yet.

In this study, we therefore built an extensive dataset of DRS measurements on ex vivo breast tissue and in vivo breast tissue, which we used to develop four different types of classification models to evaluate the classification performance on DRS data acquired from in vivo breast tissue. We investigated which type of training data yielded optimum results for the classification model with the highest performance.

For all built model types, the classification performance when trained on DRS data acquired from ex vivo tissue was comparable to the classification performance when trained on DRS data acquired from in vivo tissue. This could be explained by the fact that all selected features were situated in the near-infrared wavelength range, where fat and water are the primary absorbers. In an earlier study of our group, it was found that the fat fraction in combination with the total volume of fat and water provided the best discrimination between tumorous breast tissue and healthy breast tissue.55 This optical parameter did not change significantly when comparing ex vivo tissue measurements with in vivo tissue measurements.56,66 On the other hand, the rapid change in ratio of oxygenated to deoxygenated hemoglobin post surgery substantially impacts the optical absorption in the visual wavelength range. Therefore, based on our previous results, no features were extracted from the visual wavelength range.

The RUSBoost has the highest classification performance on any type of training data compared with the other investigated models. The highest performance is achieved when the RUSBoost is trained on a combination of DRS data acquired from ex vivo tissue and in vivo tissue, with a MCC of 0.76. The explanation for the high performance of this particular model is that the RUSBoost model alleviates the imbalance of the in vivo dataset using random under sampling and boosting.67

When analyzing the three ROC curves for the RUSBoost models trained on DRS data from (1) ex vivo tissue, (2) in vivo tissue, and (3) ex vivo and in vivo tissue, it was evident that the AUCs of all models were comparably high. Similarly, all models had a high sensitivity and specificity at the optimal operating point. The sensitivity of all models is higher compared with the specificity, which is desirable because missing any tumor tissue on the margin has a more negative impact on the clinical outcome of a patient. Thus our data show that it is possible to mainly use ex vivo data for training the models. The results showed that training on a combination of ex vivo and in vivo data yields the highest performance metrics, with an AUC of 0.98, sensitivity of 96.7% (95% CI 95.6% to 98.2%), and specificity of 90.6% (95% CI 86.3% to 97.9%). A probable explanation for the slightly higher performance is the higher amount of available data for training compared with the models trained on only ex vivo or only in vivo data.

It should be noted that during this research, we carefully correlated the measured tissue locations to the corresponding region in the H&E sections. However, the area percentages of tissue types and the tumor-margin distances could only be calculated from the 2D H&E region derived from a few cell layers, whereas the actual probed volume is a bigger 3D volume. Therefore, the calculated parameters in the H&E section will not completely correspond with the probed tissue volume. Furthermore, it should be mentioned that the tissue percentage distribution among the ex vivo tissue locations is slightly different from the in vivo tissue locations. A possible explanation for this difference is that the data acquisition on ex vivo tissue could be performed in a more controlled method, and US imaging was available to localize tumor areas in some specimens. Because the data distribution is slightly different for the in vivo locations, including tumor locations with higher tumor percentages and healthy locations with higher fat percentages, this dataset could possibly lead to less misclassifications compared with the ex vivo dataset. This could partially explain the finding in a previous study in which we built an RUSBoost model on an ex vivo dataset to distinguish tumorous tissue from healthy tissue up to 2 mm from the margin ex vivo and found a sensitivity of 93% and a specificity of 75%, which were lower performance metrics compared with the findings in this study (sensitivity of 97% and specificity of 91%).65 An underexposed point in our manuscript would be the influence of different patient and tumor characteristics on the classification accuracy of the models. Unfortunately, the individual groups in our study were too small to perform any subgroup analyses. However, this would be an interesting point to study in the future.

Our ultimate goal is the clinical application of DRS for real-time margin assessment during BCS to improve surgical outcomes. For this purpose, it is important to perform the DRS measurements at the exact tissue locations with a high probability of a (focally) positive margin. There are two possible options to solve the issue of finding these locations: (1) using a second modality to find the measurement locations and (2) applying continuous DRS measurements. For the first option, US imaging combined with deep learning based tumor segmentation could be a valuable addition. In an earlier study, we found a median dice score of 0.88 when using an ensemble approach for tumor segmentation in US images of BCS specimens,65 which shows great potential to be applied as a modality to find the correct locations for margin assessment. In the second option, the DRS probe could be moved over the entire margin, permitting quick measurements and classification in a continuous manner. The feasibility and diagnostic accuracy of both options should be investigated in a follow-up study.

5.

Conclusion

In this study, we have advanced toward the use of fiber-optic DRS for intra-operative margin assessment during BCS. We built an extensive set of DRS data acquired from ex vivo and in vivo breast tissue and built several classification models to discriminate tumorous tissue from healthy tissue up to 2 mm from the resection margin. The optimum classification model had a similarly high performance for classifying DRS data from in vivo tissue, when trained on ex vivo and/or in vivo data. By training on a combination of ex vivo and in vivo data, we achieved an MCC of 0.76, a sensitivity of 96.7% (95% CI 95.6% to 98.2%), a specificity of 90.6% (95% CI 86.3% to 97.9%), and an AUC of 0.98. These results show the potential of DRS to be applied as a tool to support margin assessment in real time during BCS. To achieve this ultimate goal, additional research needs to be performed regarding the optimum method to find the tissue locations with a high probability of a (focally) positive margin during BCS.

Disclosures

The authors declare no conflicts of interest.

Code and Data Availability

Data underlying the results presented in this paper are not publicly available at the time but may be obtained from the authors upon reasonable request.

Acknowledgments

The authors thank all employees from the NKI-AvL Core Facility Molecular Pathology and Biobanking (CFMPB), all surgeons and nurses from the Department of Surgery, and all pathologists and pathologist assistants from the Department of Pathology for their assistance. This research received no external funding.

References

1. 

H. Sung et al., “Global cancer statistics 2020: Globocan estimates of incidence and mortality worldwide for 36 cancers in 185 countries,” CA: A Cancer J. Clin., 71 (3), 209 –249 https://doi.org/10.3322/caac.21660 (2021). Google Scholar

2. 

T.-A. Moo et al., “Overview of breast cancer therapy,” PET Clin., 13 (3), 339 –354 https://doi.org/10.1016/j.cpet.2018.02.006 (2018). Google Scholar

3. 

M. Morrow et al., “Standard for breast conservation therapy in the management of invasive breast carcinoma,” CA: A Cancer J. Clin., 52 (5), 277 –300 https://doi.org/10.3322/canjclin.52.5.277 (2002). Google Scholar

4. 

U. Veronesi et al., “Twenty-year follow-up of a randomized study comparing breast-conserving surgery with radical mastectomy for early breast cancer,” N. Engl. J. Med., 347 (16), 1227 –1232 https://doi.org/10.1056/NEJMoa020989 NEJMBH (2002). Google Scholar

5. 

B. Fisher et al., “Twenty-year follow-up of a randomized trial comparing total mastectomy, lumpectomy, and lumpectomy plus irradiation for the treatment of invasive breast cancer,” N. Engl. J. Med., 347 (16), 1233 –1241 https://doi.org/10.1056/NEJMoa022152 NEJMBH (2002). Google Scholar

6. 

B. K. Goh and W.-S. Yong, “Eighteen-year results in the treatment of early breast carcinoma with mastectomy versus breast conservation therapy,” Cancer: Interdiscipl. Int. J. Am. Cancer Soc., 100 (8), 1766 https://doi.org/10.1002/cncr.20173 (2004). Google Scholar

7. 

A. C. Voogd et al., “Differences in risk factors for local and distant recurrence after breast-conserving therapy or mastectomy for stage I and II breast cancer: pooled results of two large european randomized trials,” J. Clin. Oncol., 19 (6), 1688 –1697 https://doi.org/10.1200/JCO.2001.19.6.1688 JCONDN 0732-183X (2001). Google Scholar

8. 

A. Nayyar, K. K. Gallagher and K. P. McGuire, “Definition and management of positive margins for invasive breast cancer,” Surg. Clin., 98 (4), 761 –771 https://doi.org/10.1016/j.suc.2018.03.008 (2018). Google Scholar

9. 

M. S. Moran et al., “Society of Surgical Oncology–American Society for Radiation Oncology consensus guideline on margins for breast-conserving surgery with whole-breast irradiation in stages I and II invasive breast cancer,” Int. J. Radiat. Oncol. Biol. Phys., 88 (3), 553 –564 https://doi.org/10.1016/j.ijrobp.2013.11.012 PHYSGM 1943-2879 (2014). Google Scholar

10. 

N. Houssami et al., “The association of surgical margins and local recurrence in women with early-stage invasive breast cancer treated with breast-conserving therapy: a meta-analysis,” Ann. Surg. Oncol., 21 717 –730 https://doi.org/10.1245/s10434-014-3480-5 (2014). Google Scholar

11. 

C. Leong et al., “Effect of margins on ipsilateral breast tumor recurrence after breast conservation therapy for lymph node-negative breast carcinoma,” Cancer: Interdiscipl. Int. J. Am. Cancer Soc., 100 (9), 1823 –1832 https://doi.org/10.1002/cncr.20153 (2004). Google Scholar

12. 

F. Meric et al., “Positive surgical margins and ipsilateral breast tumor recurrence predict disease-specific survival after breast-conserving therapy,” Cancer: Interdiscipl. Int. J. Am. Cancer Soc., 97 (4), 926 –933 https://doi.org/10.1002/cncr.11222 (2003). Google Scholar

13. 

S. J. DiBiase et al., “The number of positive margins influences the outcome of women treated with breast preservation for early stage breast carcinoma,” Cancer: Interdiscipl. Int. J. Am. Cancer Soc., 82 (11), 2212 –2220 https://doi.org/10.1002/(SICI)1097-0142(19980601)82:11<2212::AID-CNCR16>3.0.CO;2-X (1998). Google Scholar

14. 

M. A. Olsen et al., “Increased risk of surgical site infection among breast-conserving surgery re-excisions,” Ann. Surg. Oncol., 22 2003 –2009 https://doi.org/10.1245/s10434-014-4200-x (2015). Google Scholar

15. 

L. N. Metcalfe et al., “Beyond the margins-economic costs and complications associated with repeated breast-conserving surgeries,” JAMA Surg., 152 (11), 1084 –1086 https://doi.org/10.1001/jamasurg.2017.2661 (2017). Google Scholar

16. 

J. F. Waljee et al., “Predictors of breast asymmetry after breast-conserving operation for breast cancer,” J. Am. Coll. Surg., 206 (2), 274 –280 https://doi.org/10.1016/j.jamcollsurg.2007.07.029 JACSEX 1072-7515 (2008). Google Scholar

17. 

S. Collette et al., “Predictors of the risk of fibrosis at 10 years after breast conserving therapy for early breast cancer–a study based on the eortc trial 22881–10882 ‘boost versus no boost’,” Eur. J. Cancer, 44 (17), 2587 –2599 https://doi.org/10.1016/j.ejca.2008.07.032 (2008). Google Scholar

18. 

J. Heil et al., “Do reexcisions impair aesthetic outcome in breast conservation surgery?,” Ann. Surg. Oncol., 19 (2), 541 –547 https://doi.org/10.1245/s10434-011-1947-1 (2012). Google Scholar

19. 

D. E. Wazer et al., “Factors influencing cosmetic outcome and complication risk after conservative surgery and radiotherapy for early-stage breast carcinoma.,” J. Clin. Oncol., 10 (3), 356 –363 https://doi.org/10.1200/JCO.1992.10.3.356 JCONDN 0732-183X (1992). Google Scholar

20. 

R. A. Greenup et al., “Cost implications of the sso-astro consensus guideline on margins for breast-conserving surgery with whole breast irradiation in stage I and II invasive breast cancer,” Ann. Surg. Oncol., 21 1512 –1514 https://doi.org/10.1245/s10434-014-3605-x (2014). Google Scholar

21. 

R. Pataky and C. Baliski, “Reoperation costs in attempted breast-conserving surgery: a decision analysis,” Curr. Oncol., 23 (5), 314 –321 https://doi.org/10.3747/co.23.2989 (2016). Google Scholar

22. 

S. E. Abe et al., “Margin re-excision and local recurrence in invasive breast cancer: a cost analysis using a decision tree model,” J. Surg. Oncol., 112 (4), 443 –448 https://doi.org/10.1002/jso.23990 JSONAU 0022-4790 (2015). Google Scholar

23. 

S. G. B. de Koning et al., “Tumor resection margin definitions in breast-conserving surgery: systematic review and meta-analysis of the current literature,” Clin. Breast Cancer, 18 (4), e595 –e600 https://doi.org/10.1016/j.clbc.2018.04.004 (2018). Google Scholar

24. 

N. B. O. Nederland, “Nabon (2012). mammacarcinoom. landelijke richtlijn, versie: 2.0,” (2018). Google Scholar

25. 

C. Lin et al., “Specimen mammography for intraoperative margin assessment in breast conserving surgery: a meta-analysis,” Sci. Rep., 12 (1), 18440 https://doi.org/10.1038/s41598-022-23234-5 (2022). Google Scholar

26. 

K. Esbona, Z. Li and L. G. Wilke, “Intraoperative imprint cytology and frozen section pathology for margin assessment in breast conservation surgery: a systematic review,” Ann. Surg. Oncol., 19 3236 –3245 https://doi.org/10.1245/s10434-012-2492-2 (2012). Google Scholar

27. 

A. R. Pradipta et al., “Emerging technologies for real-time intraoperative margin assessment in future breast-conserving surgery,” Adv. Sci., 7 (9), 1901519 https://doi.org/10.1002/advs.201901519 1936-6612 (2020). Google Scholar

28. 

J. J. Keating et al., “Advances in intraoperative margin assessment for breast cancer,” Curr. Surg. Rep., 4 (4), 1 –8 https://doi.org/10.1007/s40137-016-0136-3 (2016). Google Scholar

29. 

J. Schwarz and H. Schmidt, “Technology for intraoperative margin assessment in breast cancer,” Ann. Surg. Oncol., 27 2278 –2287 https://doi.org/10.1245/s10434-020-08483-w (2020). Google Scholar

30. 

N. M. Krekel et al., “Intraoperative ultrasound guidance for palpable breast cancer excision (cobalt trial): a multicentre, randomised controlled trial,” Lancet Oncol., 14 (1), 48 –54 https://doi.org/10.1016/S1470-2045(12)70527-2 LOANBN 1470-2045 (2013). Google Scholar

31. 

M. Moschetta et al., “Role of specimen us for predicting resection margin status in breast conserving therapy,” G. Chir., 36 (5), 201 https://doi.org/10.11138/gchir/2015.36.5.201 (2015). Google Scholar

32. 

T. E. Doyle et al., “High-frequency ultrasound for intraoperative margin assessments in breast conservation surgery: a feasibility study,” BMC Cancer, 11 1 –15 https://doi.org/10.1186/1471-2407-11-444 BCMACL 1471-2407 (2011). Google Scholar

33. 

F. Schnabel et al., “A randomized prospective study of lumpectomy margin assessment with use of marginprobe in patients with nonpalpable breast malignancies,” Ann. Surg. Oncol., 21 1589 –1595 https://doi.org/10.1245/s10434-014-3602-0 (2014). Google Scholar

34. 

M. Thill et al., “Intraoperative assessment of surgical margins during breast conserving surgery of ductal carcinoma in situ by use of radiofrequency spectroscopy,” The Breast, 20 (6), 579 –580 https://doi.org/10.1016/j.breast.2011.08.134 BRJOFK 1075-122X (2011). Google Scholar

35. 

I. Pappo et al., “Diagnostic performance of a novel device for real-time margin assessment in lumpectomy specimens,” J. Surg. Res., 160 (2), 277 –281 https://doi.org/10.1016/j.jss.2009.02.025 (2010). Google Scholar

36. 

A. Kupstas et al., “A novel modality for intraoperative margin assessment and its impact on re-excision rates in breast conserving surgery,” Am. J. Surg., 215 (3), 400 –403 https://doi.org/10.1016/j.amjsurg.2017.11.023 AJSUAB 0002-9610 (2018). Google Scholar

37. 

J. M. Dixon et al., “Intra-operative assessment of excised breast tumour margins using clearedge imaging device,” Eur. J. Surg. Oncol., 42 (12), 1834 –1840 https://doi.org/10.1016/j.ejso.2016.07.141 (2016). Google Scholar

38. 

D. W. Shipp et al., “Intra-operative spectroscopic assessment of surgical margins during breast conserving surgery,” Breast Cancer Res., 20 (1), 1 –14 https://doi.org/10.1186/s13058-018-1002-2 BCTRD6 (2018). Google Scholar

39. 

A. S. Haka et al., “Diagnosing breast cancer using Raman spectroscopy: prospective analysis,” J. Biomed. Opt., 14 (5), 054023 https://doi.org/10.1117/1.3247154 JBOPFO 1083-3668 (2009). Google Scholar

40. 

G. Thomas et al., “Evaluating feasibility of an automated 3-dimensional scanner using Raman spectroscopy for intraoperative breast margin assessment,” Sci. Rep., 7 (1), 13548 https://doi.org/10.1038/s41598-017-13237-y (2017). Google Scholar

41. 

K. U. Park et al., “Digital breast tomosynthesis for intraoperative margin assessment during breast-conserving surgery,” Ann. Surg. Oncol., 26 1720 –1728 https://doi.org/10.1245/s10434-019-07226-w (2019). Google Scholar

42. 

R. Pleijhuis et al., “Near-infrared fluorescence (NIRF) imaging in breast-conserving surgery: assessing intraoperative techniques in tissue-simulating breast phantoms,” Eur. J. Surg. Oncol., 37 (1), 32 –39 https://doi.org/10.1016/j.ejso.2010.10.006 (2011). Google Scholar

43. 

S.-Q. Qiu et al., “Micro-computed tomography (micro-CT) for intraoperative surgical margin assessment of breast cancer: a feasibility study in breast conserving surgery,” Eur. J. Surg. Oncol., 44 (11), 1708 –1713 https://doi.org/10.1016/j.ejso.2018.06.022 (2018). Google Scholar

44. 

A. M. Zysk et al., “Intraoperative assessment of final margins with a handheld optical imaging probe during breast-conserving surgery may reduce the reoperation rate: results of a multicenter study,” Ann. Surg. Oncol., 22 3356 –3362 https://doi.org/10.1245/s10434-015-4665-2 (2015). Google Scholar

45. 

R. Ha et al., “Optical coherence tomography: a novel imaging method for post-lumpectomy breast margin assessment-a multi-reader study,” Acad. Radiol., 25 (3), 279 –287 https://doi.org/10.1016/j.acra.2017.09.018 (2018). Google Scholar

46. 

K. Y. Foo et al., “Optical palpation for tumor margin assessment in breast-conserving surgery,” Biomed. Opt. Express, 12 (3), 1666 –1682 https://doi.org/10.1364/BOE.415888 BOEICL 2156-7085 (2021). Google Scholar

47. 

K. M. Kennedy et al., “Diagnostic accuracy of quantitative micro-elastography for margin assessment in breast-conserving surgery,” Cancer Res., 80 (8), 1773 –1783 https://doi.org/10.1158/0008-5472.CAN-19-1240 CNREA8 0008-5472 (2020). Google Scholar

48. 

T. T. Wong et al., “Fast label-free multilayered histology-like imaging of human breast cancer by photoacoustic microscopy,” Sci. Adv., 3 (5), e1602168 https://doi.org/10.1126/sciadv.1602168 (2017). Google Scholar

49. 

R. Li et al., “Assessing breast tumor margin by multispectral photoacoustic tomography,” Biomed. Opt. Express, 6 (4), 1273 –1281 https://doi.org/10.1364/BOE.6.001273 BOEICL 2156-7085 (2015). Google Scholar

50. 

G. Vartholomatos et al., “Rapid assessment of resection margins during breast conserving surgery using intraoperative flow cytometry,” Clin. Breast Cancer, 21 (5), e602 –e610 https://doi.org/10.1016/j.clbc.2021.03.002 (2021). Google Scholar

51. 

F. Fereidouni et al., “Microscopy with ultraviolet surface excitation for rapid slide-free histology,” Nat. Biomed. Eng., 1 (12), 957 –966 https://doi.org/10.1038/s41551-017-0165-y (2017). Google Scholar

52. 

T. Karni et al., “A device for real-time, intraoperative margin assessment in breast-conservation surgery,” Am. J. Surg., 194 (4), 467 –473 https://doi.org/10.1016/j.amjsurg.2007.06.013 AJSUAB 0002-9610 (2007). Google Scholar

53. 

E. R. St John et al., “Diagnostic accuracy of intraoperative techniques for margin assessment in breast cancer surgery,” Ann. Surg., 265 (2), 300 –310 https://doi.org/10.1097/SLA.0000000000001897 ANSUA5 0003-4932 (2017). Google Scholar

54. 

J. Heidkamp et al., “Novel imaging techniques for intraoperative margin assessment in surgical oncology: a systematic review,” Int. J. Cancer, 149 (3), 635 –645 https://doi.org/10.1002/ijc.33570 (2021). Google Scholar

55. 

L. De Boer et al., “Fat/water ratios measured with diffuse reflectance spectroscopy to detect breast tumor boundaries,” Breast Cancer Res. Treat., 152 509 –518 https://doi.org/10.1007/s10549-015-3487-z BCTRD6 (2015). Google Scholar

56. 

L. L. de Boer et al., “Optical tissue measurements of invasive carcinoma and ductal carcinoma in situ for surgical guidance,” Breast Cancer Res., 23 (1), 1 –15 https://doi.org/10.1186/s13058-021-01436-5 BCTRD6 (2021). Google Scholar

57. 

D. Veluponnar et al., “Diffuse reflectance spectroscopy for accurate margin assessment in breast-conserving surgeries: importance of an optimal number of fibers,” Biomed. Opt. Express, 14 (8), 4017 –4036 https://doi.org/10.1364/BOE.493179 BOEICL 2156-7085 (2023). Google Scholar

58. 

F. Geldof et al., “Diffuse reflection spectroscopy at the fingertip: design and performance of a compact side-firing probe for tissue discrimination during colorectal cancer surgery,” Biomed. Opt. Express, 14 (1), 128 –147 https://doi.org/10.1364/BOE.476242 BOEICL 2156-7085 (2023). Google Scholar

59. 

E. Kho et al., “Broadband hyperspectral imaging for breast tumor detection using spectral and spatial information,” Biomed. Opt. Express, 10 (9), 4496 –4515 https://doi.org/10.1364/BOE.10.004496 BOEICL 2156-7085 (2019). Google Scholar

60. 

P. Geladi, D. MacDougall and H. Martens, “Linearization and scatter-correction for near-infrared reflectance spectra of meat,” Appl. Spectrosc., 39 (3), 491 –500 https://doi.org/10.1366/0003702854248656 APSPA4 0003-7028 (1985). Google Scholar

61. 

H. Peng et al., “Feature selection based on mutual information criteria of max-dependency, max-relevance, and min-redundancy,” IEEE Transactions on Pattern Anal. Mach. Intell., 27 (8), 1226 –1238 (2005). Google Scholar

62. 

S. G. Brouwer de Koning et al., “Toward complete oral cavity cancer resection using a handheld diffuse reflectance spectroscopy probe,” J. Biomed. Opt., 23 (12), –121611 https://doi.org/10.1117/1.JBO.23.12.121611 JBOPFO 1083-3668 (2018). Google Scholar

63. 

E. J. Baltussen et al., “Using diffuse reflectance spectroscopy to distinguish tumor tissue from fibrosis in rectal cancer patients as a guide to surgery,” Lasers Surg. Med., 52 (7), 604 –611 https://doi.org/10.1002/lsm.23196 LSMEDI 0196-8092 (2020). Google Scholar

64. 

F. Geldof et al., “Layer thickness prediction and tissue classification in two-layered tissue structures using diffuse reflectance spectroscopy,” Sci. Rep., 12 (1), 1698 https://doi.org/10.1038/s41598-022-05751-5 (2022). Google Scholar

65. 

D. Veluponnar et al., “Toward intraoperative margin assessment using a deep learning-based approach for automatic tumor segmentation in breast lumpectomy ultrasound images,” Cancers, 15 (6), 1652 https://doi.org/10.3390/cancers15061652 (2023). Google Scholar

66. 

L. L. De Boer et al., “Towards the use of diffuse reflectance spectroscopy for real-time in vivo detection of breast cancer during surgery,” J. Transl. Med., 16 1 –14 https://doi.org/10.1186/s12967-018-1747-5 (2018). Google Scholar

67. 

C. Seiffert et al., “Rusboost: a hybrid approach to alleviating class imbalance,” IEEE Trans. Syst. Man Cybern. Part A Syst. Hum., 40 (1), 185 –197 https://doi.org/10.1109/TSMCA.2009.2029559 (2009). Google Scholar

Biographies of the authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Dinusha Veluponnar, Lisanne L. de Boer, Behdad Dashtbozorg, Lynn-Jade S. Jong, Freija Geldof, Marcos Da Silva Guimaraes, Henricus J. C. M. Sterenborg, Marie-Jeanne T. F. D. Vrancken-Peeters, Frederieke van Duijnhoven, and Theo Ruers "Margin assessment during breast conserving surgery using diffuse reflectance spectroscopy," Journal of Biomedical Optics 29(4), 045006 (25 April 2024). https://doi.org/10.1117/1.JBO.29.4.045006