24 July 2012 Phase-sensitive optical coherence tomography characterization of pulse-induced trabecular meshwork displacement in ex vivo nonhuman primate eyes
Peng Li, Roberto Reif, Zhongwei Zhi, Tueng Shen, Ruikang K. Wang, Elizabeth Martin, Murray Johnstone
Author Affiliations +
Abstract
Glaucoma is a blinding disease for which intraocular pressure (IOP) is the only treatable risk factor. The mean IOP is regulated through the aqueous outflow system, which contains the trabecular meshwork (TM). Considerable evidence indicates that trabecular tissue movement regulates the aqueous outflow and becomes abnormal during glaucoma; however, such motion has thus far escaped detection. The purpose of this study is to describe anovel use of a phase-sensitive optical coherence tomography (PhS-OCT) method to assess pulse-dependent TM movement. For this study, we used enucleated monkey eyes, each mounted in an anterior segment holder. A perfusion system was used to control the mean IOP as well as to provide IOP sinusoidal transients (amplitude 3 mmHg, frequency 1 pulse/second) in all experiments. Measurements were carried out at seven graded mean IOPs (5, 8, 10, 20, 30, 40, and 50 mm Hg). We demonstrate that PhS-OCT is sensitive enough to image/visualize TM movement synchronous with the pulse-induced IOP transients, providing quantitative measurements of dynamic parameters such as velocity, displacement, and strain rate that are important for assessing the biomechanical compliance of the TM. We find that the largest TM displacement is in the area closest to Schlemm's canal (SC) endothelium. While maintaining constant ocular pulse amplitude, an increase of mean IOP results in a decrease of TM displacement and mean size of the SC. These results demonstrate that the PhS-OCT is a useful imaging technique capable of assessing functional properties necessary to maintain IOP in a healthy range, offering a new diagnostic alternative for glaucoma.

1.

Introduction

Glaucoma, the second leading cause of irreversible blindness,1 is defined as an optic neuropathy of unknown etiology characterized by optic nerve damage, which results in visual field loss. Currently, direct assessment of the damage to the visual system is the primary means for diagnosing glaucoma and determining the need for escalation of therapy.2 However, current treatment techniques are frequently associated with continued progression of visual system damage. For example, in eyes of patients managed with standard glaucoma therapy, one study with a 20-year follow-up found the risk of legal blindness to be 27%.3 In another study with a 15-year follow-up, 68% of eyes had progressive damage and 27% became legally blind.4

Deterioration in visual function is detected by visual fields. Progressive deterioration in the microstructures of the visual system can be detected by several modalities, with the least complex being the ophthalmoscope and photography. Optical coherence tomography (OCT) maps the contour of the optic nerve head, the peripapillary nerve fiber layer, and the ganglion cell complex.514 Scanning laser polarimetry is often used to assess polarization of retinal nerve fibers,11,1517 whereas, scanning laser tomography is used to measure the surface contour of the optic nerve head.1820 However, changes in the parameters measured represent evidence of permanent damage to the visual system.

The one treatable problem that minimizes or stops progression of the glaucoma process is the regulation of intraocular pressure (IOP) within a narrow range.21 Clinical measurements of IOP are typically done three to four times a year, with each measurement taking about 3 s that provide about 12 s of IOP sampling per year. Blinking and eye movements induce IOP transients of 10mmHg that occur numerous times each minute.22 Measurements of IOP do not capture these transients because they are acquired while the patient is looking straight ahead without blinking or eye movement. Marked diurnal fluctuations of IOP are also well recognized but not practical to measure.23,24 In addition, IOP spontaneously varies from day to day.25,26

The difficulty in gaining a true assessment of IOP fluctuation over time limits its ability to predict progressive glaucoma damage,27 leaving us with progressive visual field28 and structural damage29,30 to the visual system as a primary diagnostic tool for treatment guidance. Currently there is no method available to assess the functional properties of the trabecular tissue—properties that are necessary to maintain IOP in a narrow range.31,32 A more robust technique to assess the functionality of the tissues controlling IOP could enable early diagnosis and treatment for those at risk, reducing reliance on progressive visual system damage as a guide for management.

IOP is regarded as the major risk factor and the only treatable problem in glaucoma.2 The aqueous outflow system (AOS), the pathway for aqueous humor to leave the eye, involves passage of aqueous from the trabecular meshwork (TM) into the Schlemm’s canal (SC), and then into the collector channels (CCs). From the CCs, the aqueous enters the aqueous veins and then the episcleral veins. A recent model of the AOS indicates that the TM acts as a biomechanical pump, which induces pulsatile flow.3336 Failure of the pulsatile flow mechanism results in an increase of outflow resistance. Clinical and laboratory evidence3336 points to alterations in elasticity and compliance as changes responsible for increased outflow resistance in glaucoma patients. Current acceptance of the TM as the source of resistance is illustrated by the recent use of stents to bypass abnormal trabecular tissues in an attempt to re-establish normal rates of aqueous flow into the SC.37,38

A body of evidence indicates that the aqueous outflow from the SC is pulsatile and synchronous with IOP transients.33,34,3941 Transients such as the cardiac pulse cannot transfer pressure gradients directly from the anterior chamber to SC because of the complex labyrinthine structure of the TM. An explanation for the pulsatile flow in the SC may be due to the trabecular tissue movement in response to the cardiac pulse. However, no technologies have thus far identified such movement, probably because the TM is located deep within the corneo-scleral limbus, and its movement is too small (in the micron range) to be detected by any currently available technologies, either in vivo or in vitro. The current report meets this challenge by describing a noninvasive, yet highly sensitive method that is capable of characterizing TM movement in response to an experimentally induced ocular pulse in an ex vivo primate model.

OCT is a powerful noninvasive, noncontact, three-dimensional (3-D), real-time imaging modality with high spatial resolution (10 μm axially).10,4245 Studies have revealed the potential of using an anterior segment OCT to image the AOS38,44,4651 and to visualize the TM and the SC,44 with their appearances in the OCT images confirmed by histological correlations.46 A recent report proposed the use of a polarization-sensitive OCT to image the AOS with an enhanced ability to visualize the TM.47 By integration with a prototype endoscopic probe, OCT has also been reported to guide implantation of stents during SC surgery37 and to identify the CCs in human cadaver eyes.38

Due to its high resolution (10 μm) and high imaging speed (100-kHz line rate), it is now possible for OCT to provide quantifiable in vivo information about tissue structures with a size comparable with or greater than the OCT resolution. For example, a number of recent studies have reported the use of real-time OCT to quantify the geometrical sizes of the TM30 and SC46,48,49 as well as to map the surrounding microcirculation of the AOS.50,51 Regional differences of SC size48,49 as well as differences between normal and glaucomatous subjects48 have been explored. Furthermore, several methods have been developed to measure fundus pulsations using OCT52 or low-coherence interferometry.53,54 Notwithstanding, all these studies are based on OCT microstructural images to identify and track structural boundary positions (e.g. cornea and retina), which are, however, not feasible for measuring TM movement. This is because 1) the visibility of the TM boundary is often poor, and 2) the amount of TM movement is typically 2 μm, which is beyond the limits of resolution attainable with OCT structural measurements.

To measure such small tissue movement, one has to use the phase information embedded in OCT signals, rather than the amplitude information that is often used to reconstruct the OCT images. This is because the change of optical phase is much more sensitive to tissue movement than that of amplitude. This concept was first reported in 200655 where the phase changes between adjacent A-scans were used to measure the tissue movement, also called tissue Doppler OCT (tDOCT), a method similar to that of phase resolved OCT to measure blood flow.56 It was then quickly realized that the phase noise from such treatment is high due to the heterogeneous properties of the biological tissue, which causes deterioration of the spatial correlation between the adjacent A-scans.57 In addition, the time interval, τ, between the adjacent A-scans in real time OCT systems is short (in the microsecond range), making the measurement of slow tissue movement difficult. These factors combined prevent the tDOCT from measuring small tissue movements less than 1 µm.58 To mitigate this problem, a phase sensitive OCT (PhS-OCT) was developed in 2007 that uses the phase changes between adjacent B-scans to measure the tissue movement.58 The report demonstrated that PhS-OCT is extremely sensitive to tissue motion as small as 0.26 nm, a sensitivity made possible by the much increased time interval τ which is typically several hundred times that of tDOCT.

In this pilot study, we report for the first time the use of PhS-OCT to image/visualize TM movement with sensitivity at the nanometer scale. We used nonhuman primate eyes for this proof of concept. To achieve our objectives, we designed an anterior segment perfusion system that is capable of providing sinusoidal IOP transients, thus permitting the primate eyes to mimic the in vivo ocular pulse generated by the cardiac cycle. A real-time PhS-OCT system working at 92-kHz line rate was used to image the AOS of the primate eyes in response to experimentally induced IOP transients, enabling quantitative assessment of dynamic TM movement in situ.

2.

Materials and Methods

2.1.

System Setup and Methodology

The schematic of the experimental setup used to image/measure dynamic TM movements in response to IOP transients is shown in Fig. 1. The setup consists primarily of three units: a PhS-OCT system to image the AOS, a perfusion system to control the IOP, and a controlling unit to provide synchronization during data acquisition.

Fig. 1

Schematic diagram of the experimental setup for imaging the anterior segment (PhS-OCT) and controlling IOP within the anterior chamber. Reservoir 1 establishes a mean IOP in the anterior chamber of the eye; Reservoir 2 establishes a mean IOP in the posterior chamber of the eye and induces a sinusoidal IOP gradient change with a peak-to-peak amplitude of 3 mmHg at 1 pulse/second. A pressure transducer records IOP in real time; Trigger 1 triggers PhS-OCT for imaging; Trigger 2 triggers pressure transducer for recording. ROI indicates the OCT imaging area. OC: optical coupler; CIR: circulator; CL: collimating lens; FL: focusing lens; M: mirror; OL: objective lens; DG: diffraction grating.

JBO_17_7_076026_f001.png

The PhS-OCT system (Fig. 1) used in this study is based on a spectral domain OCT (SD-OCT) configuration, in which a broadband superluminescent diode with a central wavelength of 1340 nm and a spectral bandwidth of 110 nm is used to achieve a high axial resolution (theoretically 7.2 μm in air). In the sample arm, an objective lens (focal length 50 mm) is used to focus the probing light on the region of interest (ROI), i.e. the corneo-scleral limbus, yielding a measured lateral resolution of 16 μm. The spectrometer is equipped with an InGaAs line scan camera capable of 92-kHz A-line scan rate. The spectrometer has a spectral resolution of 0.14 nm providing a measured imaging depth of 3.0mm in air. The system dynamic range was measured at 105dB at the depth-position of 0.5 mm with an incident optical power of 2.5 mW in the sample arm.

Four ex vivo monkey eyes (nemestrina), having an outflow system similar to humans, were used in this study. The enucleated eyes were obtained from the University of Washington Primate Center and harvested within 0.5h of initiation of the experiments. A system for imaging the anterior segment by PhS-OCT was designed by mounting the eye in a device that encompassed a region about 6-mm posterior to the limbus (see Fig. 1). Since the mean IOP in vivo is 16mmHg and the mean episcleral venous pressure is 8mmHg, the pressure differential between the anterior chamber and episcleral vein under normal conditions is 8mmHg. Because ex vivo eyes do not have episcleral venous pressure, an IOP of 8 mmHg was used to approximate the pressure gradient present in vivo. To control the IOP, the height of two reservoirs leading to the anterior chamber (reservoir 1) and posterior chamber (reservoir 2) were adjusted. An anterior chamber needle connected to a pressure transducer (PowerLab, ML866) was used to continuously monitor the anterior chamber mean IOP and pulse amplitudes. Reservoir 2 was used to induce sinusoidal IOP transients that mimic the ocular pulse, with a peak-to-peak amplitude of 3mmHg and a frequency of 1pulse/second. Two trigger signals (trigger 1 to the OCT imaging system; trigger 2 to the pressure transducer) were generated by a computer to synchronize the data acquisition of the PhS-OCT system and the pressure transducer. Repeated OCT B-frames (MB-mode scan) were acquired at the same location at the limbal region of the anterior chamber (as indicated by ROI in Fig. 1). Each B-frame containing 512 A-lines and 500 repeated B-frames was acquired with a frame rate of 140fps, during 3.5s of IOP pulse transients. The experiments were replicated at seven graded mean IOPs (IOP=5, 8, 10, 20, 30, 40, 50 mmHg).

In spectral domain OCT, the interference signal between the reference and the scattering light from within a sample is spectrally resolved by a linear array detector (InGaAs camera in Fig. 1). By ignoring the terms that do not contribute to the OCT image formation, the OCT signal that localizes the scatter within a sample over a B scan at time instant t an be expressed as

Eq. (1)

f(x,z,t)=A(x,z,t)exp[iϕ(x,z,t)],
where (x,z) denotes the lateral and depth coordinates of the scatter position within the cross-section (B-scan) OCT image. Equation (1) is a complex function where the amplitude, A(x,z,t) is used to generate the traditional OCT structural images, whereas the phase term (x,z,t) generally random, but fixed for the static scatter at the position (x,z) However, if the scatter moves by an instantaneous distance Δd(x,z) ring a time interval τ between two successive B scans, a localized change in the measured phase of the reflected light will be induced, which is given by

Eq. (2)

Δ(x,z,t)=4πΔd(x,z,t)λ0,
where λ0 is the central wavelength of the light source, and n is the refractive index of the sample which is typically 1.38 for most soft tissues.59 In this study, the phase-frame Δ(x,z,t) as initially measured by calculating the phase shift between the successive B-frames, and then every fivephase-frames were averaged to minimize the effect of phase noise on the final results.Thusthe localized tissue velocity in the beam direction can be deduced,

Eq. (3)

v(x,z,t)=Δφ(x,z,t)λ04πnτ.

For limited tissue motion, the correlation between the phase measurements of successive B scans is well preserved because the M-B mode of scanning is used in this study. Note that because the imaging speed of the system was 140 fps, the time interval was τ=7.1ms in this study. After the depth-resolved instantaneous displacement and velocity are derived, the amount of tissue movement over a time period of T can be obtained by integrating Eq. (3) over the time variable,

Eq. (4)

d(x,z)=0TΔϕ(x,z,t)λ04πnτdt.

With these dynamic parameters, the strain rate ε(x,z,t) can be calculated to quantify the local response of the tissue, such as the TM, to the IOP transients as follows,

Eq. (5)

ε(x,z,t)=dv(x,z,t)dz.

Alternatively, the strain rate can be evaluated by taking the slope of the velocity curve along the vertical axis (probe beam direction), which was the method used in this study. To improve the accuracy, we used a linear fitting algorithm reported in Ref. 60 to estimate the slope of the TM velocity curve.

It is known that the OCT phase measurement described above is only sensitive to tissue movement along the probe beam direction. In other words, the true tissue movement can be obtained only if the Doppler angle is known a priori. It should be noted that the TM is oriented circumferentially around the anterior segment, while the OCT imaging is performed radially in this study. Thus the TM surface is orthogonal to the scanning plane i.e. the B-frames and the pressure and velocity direction is perpendicular to the TM surface. As illustrated in Fig. 2, it is trivial to find that the Doppler angle, α, can be calculated as the angle between the TM surface and the horizontal plane (x-y plane). With this Doppler angle, the tissue velocity, movement and strain rate as described in Eq. (3) through Eq. (5), respectively, are modified by dividing the measured values by cos(α) to provide their true values.

Fig. 2

Schematic diagram of Doppler angle correction.

JBO_17_7_076026_f002.png

2.2.

Theoretical Considerations

According to the pulsatile flow model,3336 the increase in the IOP transients, due to the ocular pulse, blinking, and eye movement, displaces the SC endothelium (SCE) outward into the SC, placing the TM under tension. Outward TM movement narrows the SC, which as a result discharges the aqueous from the SC into the CCs and the aqueous veins. Concurrently, the TM stretches, allowing the aqueous in the anterior chamber to pass through the TM interstices and enter one-way aqueous valves. During the decrease of the IOP transient, the TM recoils, leading to a pressure reduction in the SC and the aqueous to flow from the aqueous valves into the SC.

It is useful to theoretically estimate the TM movement with each pulse wave to fully account for aqueous outflow in the primate eyes. To simplify the calculation, the elongated lumen SC oriented circumferentially around the anterior segment may be modeled as a rectangular box, with length (L)×width(W)×height(H), as illustrated in Fig. 3. Corneal diameter near the corneo-scleral junction is 9.5mm61 from which we estimate the SC diameter to be about 1 mm larger (10.5mm). Using a 10.5 mm diameter,the circumference of the SC is determined to be 33mm [refer to Fig. 3(a)], which is used as the length (L) of the rectangular box in our model. The anterior-posterior length of SC has been measured to be 230to30 μm,61 which is used as the width (W) of the rectangular box [refer to Fig. 3(b)]. The total height of the box is estimated to be 11 µm.61

Fig. 3

SC Deformation Model. (a) Schematic diagram of the front view of the eye. (b) Schematic diagram of the SC cross-section and (c) outward displacement of the SCE into the SC induced by an increase of IOP transient. (d) Schematic diagram of the rectangular box model that approximates the change in the SC volume. (e) Table summarizing the parameters and equations used in the theoretical estimation of the SCE movement.

JBO_17_7_076026_f003.png

Aqueous flow has been estimated to be 1.5 μl/min.62,63 For a 60-beats-per-minute heart rate, the aqueous stroke volume with each pulse wave, which corresponds to the change in SC volume (ΔV) per heartbeat, is therefore 0.025 µl. To cause a 0.025 µl change in the SC volume, a change in the height of the rectangular box (ΔH) should then be 2.3 to 3.3 µm [refer to Fig. 3(c) and 3(d)]. According to the pulsatile flow model, the change in the box height (ΔH) can be considered as a % change of SCE moving into SC, i.e., ΔH/H*100 or 21%–30%. During one aqueous pulse wave, the SC size change (ΔSCS) should also be 21% to 30%. The calculations have been summarized in Fig. 3(e).

3.

Results

Figure 4(a) is a schematic representation of structural relationships and pathways of the AOS. Aqueous humor, which is generated by the ciliary processes, travels from the posterior to the anterior chamber of the eye through the pupil, and exits the eye via the aqueous outflow pathway which is located at the corneo-scleral limbus and includes the TM, SC, CCs, and aqueous and episcleral veins.3336 The AOS structural elements are summarized in Fig. 4(b). The dashed square marked in Fig. 4(a) indicates the area for PhS-OCT imaging. Figure 4(c) is a representative OCT image captured from an enucleated nonhuman primate eye, where the important structural features such as the TM, SC and CC can be clearly identified. However, due to the limited imaging resolution of the system (7 μm axial and 16 μm lateral), it is difficult to visualize/measure the small movement of the TM induced by the IOP transients.

Fig. 4

PhS-OCT is capable of imaging important microstructural features of the AOS. (a) and (b) Schematic diagram and flow chart of the aqueous outflow pathway, respectively; (c) are presentative OCT microstructural cross-section of the corneo-scleral limbus [marked by the dashed blue square in (a)]; (d) enlarged view of the area marked by the dashed yellow square in (c). The schematic of (a) is modified from Refs. 33 and 34.

JBO_17_7_076026_f004.png

The potential of PhS-OCT to visualize/image the dynamic movement of the AOS, particularly the TM, is demonstrated in Fig. 5. With the mean IOP maintained at 8mmHg, the pulse amplitude of 3 mmHg and the pulse rate kept at 1 pulse/second, the typical tissue velocity maps of the scanned cross-section are given in Fig. 5(a) and 5(b) captured at the systolic (maximum) and diastolic (minimum) phases, respectively. The images demonstrate that the TM moves outward into the SC during systole [Fig. 5(a)], while it moves inward away from SC during diastole [Fig. 5(b)]. Figure 5(c) and 5(d) present the depth-dependent velocity profiles along the vertical dashed lines in Fig. 5(a) and 5(b). The largest movement is located in the area closest to SCE. The slopes of the TM velocity versus depth are marked by the thick lines in Fig. 5(c) and 5(d) and are considered to represent the strain rate of the corresponding tissue; i.e., the TM deformation speed, which was almost linear.

Fig. 5

Ex vivo pulse-induced TM movement and SC deformation in nonhuman primate eye at 8 mmHg mean IOP. Representative cross-sectional images of tissue velocity in the corneo-scleral limbus: (a) red corresponds to TM movement into SC during systole; (b) blue corresponds to TM movement away from SC during diastole; (c) and (d) depth-dependent velocity profiles along the vertical dashed lines in (a) and (b), respectively; (e) and (f) Corresponding OCT microstructural images from (a) and (b). (g) Enlarged view of the area marked by the dashed yellow square in (e). The closed white curve in (g) depicts the boundary of the SC. (h) Schematic of the SC endothelial attachment to the underlying trabecular lamellae (adopted from Refs. 33 and 34). The bold arrows in (H) indicate tissue responses to deforming forces induced by IOP transients. The horizontal lines are used to mark approximately the position of TM, facilitating comparison between figures given.

JBO_17_7_076026_f005.png

In the corresponding microstructural OCT images, SC appeared as a low-intensity area adjacent to the TM [Fig. 5(e) and 5(f)]. Given the resolution of the current system, the measurement of the SC area within the scanned cross-section is feasible. We therefore segmented the boundary of the SC lumen, as shown in Fig. 5(g). A schematic diagram of the SCE attachment to the underlying trabecular lamellae is depicted in Fig. 5(h). SCE connects with the juxtacanalicular cells, which in turn attach to the trabecular lamellae via the cytoplasmic cell processes.33,34

After localizing the TM and SC in the structural OCT images (e.g., Fig. 5), the time-lapse parameters of the TM velocity, displacement, and strain rate can be obtained by the use of Eq. (3) through Eq. (5) described in Sec. 2. These parameters are important for a detailed investigation of the dynamic pulse-dependent TM behavior. Figure 6 demonstrates such a result, displaying the temporal relationship of the induced IOP pulsations [Fig. 6(a)] to the TM velocity [Fig. 6(b)], displacement [Fig. 6(c)], and strain rate [Fig. 6(d)] obtained from one eye. In Fig. 6(e) the changes of SC size normalized to the maximal SC dimension at the corresponding mean IOP level are shown. These results were obtained under the following conditions: 8-mmHg mean IOP, 3-mmHg IOP pulse amplitude, and 1 pulse/second. The oscillations in TM displacement and SC size were synchronous with the IOP pulse wave, but involved a phase delay. The trough of the SC size curve came immediately after the IOP pulse peak, while the peak of the TM displacement came later than the trough of the SC size curve.

Fig. 6

(a) Temporal plots over 3.5 s of IOP measurements provided by the in-line pressure transducer, (b) trabecular meshwork velocity (TMV); (c) trabecular meshwork displacement (TMD); (d) trabecular meshwork strain rate (TMSR) and (e) normalized SC size (SCS). The experimental conditions were mean IOP 8 mmHg, pulse amplitude 3 mmHg and 1 pulse/second. The dashed vertical lines indicate the time of the IOP pulse peaks.

JBO_17_7_076026_f006.png

At the different mean IOP levels of 8, 10, and 40 mmHg, structural SC cross-sections (obtained at the same spatial position and at the same phase of the IOP pulse wave) are shown along with the corresponding TM movement in Fig. 7. In the experiments, a pulse amplitude of 3 mm Hg and frequency of 1 pulse/second were induced at each mean IOP level. The images demonstrate that the increase of mean IOP was associated with progressive TM distention into SC, leading to TM apposition to SC external wall [comparing Fig. 7(a) and 7(b) with 7(c) acquired at the same phase of IOP transient], and was associated with the decrease of TM movement [Fig. 7(d) to 7(f)]. Calculations of IOP-dependent SC size changes (ΔSCS) were consistent with findings from the OCT images (see Fig. 8 for quantitative results).

Fig. 7

IOP-dependence of SC deformation and TM movement. (a) to (c) Images of the same SC cross-sections at an IOP of 8, 10, and 40 mm Hg, respectively, (d) to (f) corresponding plots of TM displacement versus time. While the IOP transients (dashed lines) are relatively unchanging at three mean IOP levels, the ability of TM movement (solid lines) is markedly reduced with the increase of the mean IOP level.

JBO_17_7_076026_f007.png

Normalized mean SC size and peak-to-peak TM displacement (ppTMD) in response to a 3-mm pulse amplitude were quantified while maintaining mean IOPs of 5, 8, 10, 20, 30, 40, and 50 mmHg, respectively. Results are summarized in Figs. 8 and 9, respectively. Both the mSCA and the ppTMD decreased exponentially as the mean IOP increased. The standard deviation for TM displacement decreased as the mean IOP increased. At 8 mmHg mean IOP, the peak-to-peak SC size change (ΔSCS) was 30% of the maximal SC dimension and the ppTMD was 2.28 μm during one IOP transient.

Fig. 8

Mean and standard deviation of the normalized mean SC dimension as a function of the mean IOP. The data was fit to an exponential curve. The inserted table lists the peak-to-peak SC size change (ΔSCS) at 5 and 8 mmHg mean IOPs (each mean IOP has two values).

JBO_17_7_076026_f008.png

Fig. 9

Mean and standard deviation of the ppTMD as a function of the mean IOP. The data was fit to an exponential curve.

JBO_17_7_076026_f009.png

4.

Discussion

The PhS-OCT technique described in this report provides quantitative ex vivo evidence of synchronous TM displacement in response to IOP transients that simulate the normal in vivo ocular pulse. The ex vivo model enables the study of the pulse-induced TM displacement while simultaneously providing the ability to correlate TM displacement changes with SC dimension changes.

Ideally, laboratory studies exploring IOP regulatory mechanisms should provide insights directly translatable to improved glaucoma management. One study limitation is the use of a nonhuman primate model that may not fully reflect behavior in the human eye. However, anatomic features are remarkably similar to the human eye.33 A second limitation is the lack of normal episcleral venous pressure or afterload.51 An attempt has been made to minimize the problem by choosing pressure gradients that are reflective of the normal pressure difference experienced in the outflow system. A related limitation is the absence of normal ciliary body tone that provides tensioning of the scleral spur and TM in vivo. In addition, the resolution of the OCT system is not sufficient to study structural elements in SC that appear to subserve a valve-like function.33 However, the measurement of TM displacement and visualization of SC dimension changes in this study provide a useful mean to correlate pulse-dependent outflow system behavior and global SC size changes.

Displacement of the TM structural elements detected by PhS-OCT varied regionally within the TM tissue. Displacement of the TM tissue elements seemed greater in the regions closer to the endothelium lining of the SC, and progressively diminishing in the tissues closer to the anterior chamber, as shown in Fig. 5(c) and 5(d). Histological evidence is consistent with the finding from PhS-OCT. The entire monolayer of the endothelium lining distends outward into the SC in response to the IOP-induced forces, with the distention being limited by extensive tethering of endothelium to the trabecular lamellae.33,34 As a result, the greatest TM distention occurs in the region closest to the epithelial layer of the SC as shown in Fig. 5(h).33,34

The IOP pulse peak could be expected to correspond to the peak of the TM displacement at which time the dimension of the SC becomes the smallest (i.e., trough in the curve). Figure 6 demonstrates that, after the IOP pulse peak, the trough of SC curve came earlier than the peak of TM velocity and TM displacement. The delay most likely suggests that the force of IOP transients affects the endothelium layer earlier than the trabecular lamellae, which are the elements primarily measured by PhS-OCT, due to their larger size. The delay in responses between the SC and the TM is consistent with the pressure dependent endothelium displacement occurring first followed by force transmission to the trabecular beams through the tethering mechanism observed histologically.64,65

As the mean IOP increases (refer to Figs. 8 and 9), the ability of TM and SC to respond to the IOP transients decreases. One explanation for the reduced ability of TM to move may be the apposition between the TM and SC external wall [refer to Figs. 7(a) to 7(c) and 8]. Apposition of SC walls was correlated with a reduction in TM displacement in this study [refer to Figs. 7(d) to 7(f) and 9]. Reduced TM excursions in response to the ocular pulse as IOP increases will necessarily reduce the amount of aqueous that can be discharged with each transient pulse wave, thus leading to a further increase in mean IOP and potentially introducing a deleterious feedback loop.

At 8 mmHg corresponding to a physiologic pressure gradient, the measured values of the peak-to-peak change of the SC dimension and the peak-to-peak movement of TM are 30% and 2.27 μm, respectively, which are values close to the theoretical estimates of 21% to 30% and 2.3 to 3.3 µm (see Sec. 2.2). The standard deviation of the measured movement of the TM was larger at lower mean IOPs, which may be due to the higher elasticity of the tissue at lower pressures.

Changes in the SC dimensions were quantified based on the segmentation of the structural images with a resolution of 1 pixel (4 to 5 µm). This limits its quantification for higher mean IOPs where the changes in the size of the SC are small. This resolution limitation permitted obtaining two values for the measurement of the changes in SC dimension only at the mean IOPs of 5 and 8 mmHg. In contrast, quantification of TM displacement by the PhS-OCT method described in this paper is capable of submicron resolution.

The highly sensitive measurement of the TM movement afforded by a relatively low spatial resolution PhS-OCT system resulted in a large available dynamic range with the resultant ability to detect submicron TM movement. Such high sensitivity offers the potential to make PhS-OCT an effective screening tool for identifying patients at risk for IOP elevation in glaucoma. More importantly, the dynamic range offers the possibility of developing a sensitive, objective technique for identifying progressive TM functional abnormalities associated with the loss of the ability to maintain IOP within a normal range.

This PhS-OCT study documents a reduction in the magnitude of TM movement in response to an increase of the mean IOP (refer to Figs. 7 and 9). Ex vivo experimental counterparts34 include evidence that the TM progressively distends into the SC, and eventually resting against SC external wall as the mean IOP increases.33 Apposition of the walls of the SC necessarily leads to progressive limitation of the TM displacements. A corresponding increase in resistance due to aqueous outflow is well documented.34 In vivo counterparts include evidence that the incidence of visible pulsatile aqueous outflow decreases as the mean IOP increases; visible pulsatile flow is generally absent at an IOP above 28mmHg.36

In vivo evidence is also consistent with a loss of TM compliance and elasticity in glaucoma.34 Pressure reversal studies during gonioscopy have documented that blood rapidly enters and fills the SC and leaves when normal gradients are restored in subjects without glaucoma. As the glaucoma process develops, entry and exit of blood into the SC slows down, eventually leading to only patchy filling of the canal with blood. In advanced diseases, blood does not enter the SC. Investigators have concluded that as glaucoma progresses, the trabecular tissues gradually stiffen, preventing rapid movement of blood into and out of SC with TM tissues eventually in permanent contact with SC external wall completely preventing blood reflux.34

The TM displacement documented in this study results from movement that is determined by compliance and elasticity of the TM tissue. Such tissue properties in turn determine the functional behavior of the TM, properties that also may be reasonably expected to play an important role in maintaining normal IOP homeostasis. The sensitivity of PhS-OCT to tissue movement provides a large dynamic range, offering the potential for a technique capable of monitoring the functional properties of the outflow system in vivo.

A sensitive objective noninvasive test that characterizes IOP regulatory mechanisms may improve the diagnosis and treatment of glaucoma. Although office-based tonometry is a useful technique, it is dependent on corneal properties that often lead to measurement errors.6669 Furthermore, testing only involves brief measurements four times a year and the measurements do not adequately reflect diurnal and longer-term IOP variability.

In the treatment of glaucoma, some widely used glaucoma medications, such as prostaglandins and adrenergics, are found to increase aqueous outflow by increasing pulsatile flow from the SC into the aqueous veins, a response that occurs within minutes of instillation. The increases in pulsatile flow from the SC should increase the TM displacement. Using PhS-OCT, medication nonresponders might be rapidly identified through monitoring of medication-dependent alterations of the TM displacement. By identifying nonresponders, patients would be spared the cost and inconvenience of what for them is an ineffective medication.

Current study used the proposed PhS-OCT system to track small TM motion in the enucleated primate monkey eyes beyond the imaging system resolution. Challenges ahead will be to translate this technique to image in vivo human eyes, where bulk tissue motion, other than that of the TM, is inevitable. Active research is currently underway to develop motion-compensation algorithms; for example, using speckle tracking and/or phase-resolved Doppler approaches70,71 in order to minimize/eliminate the bulk tissue movement due to involuntary subject movement.

5.

Conclusion

This study describes a new sensitive PhS-OCT method to assess pulse-dependent TM movement in an ex vivo primate model system, with an ability to measure the small tissue movement beyond its structural imaging resolution. The technique provides quantitative evidence of synchronous TM displacement in response to IOP transients that simulate the normal in vivo ocular pulse. In this model system, the pulse-induced TM displacement was dependent on mean IOP; as IOP increased, the TM displacement decreased markedly. At the same time, the quantitative measurement of SC demonstrated a corresponding marked reduction in the SC dimension.

Evidence of TM movement obtained with the PhS-OCT technique is consistent with findings from histological and clinical studies. Such studies indicate that the trabecular tissue elasticity and compliance are the properties important for IOP regulation, the only treatable problem in glaucoma. Since PhS-OCT is a highly sensitive technique to assess functional properties associated with TM movement, the technique may provide new information to aid decision making in glaucoma management.

Acknowledgments

This research was supported in part by an Innovative Ophthalmic Research Award from Research to Prevent Blindness.

References

1. 

H. A. QuigleyA. T. Broman, “The number of people with glaucoma worldwide in 2010 and 2020,” Br. J. Ophthalmol., 90 (3), 262 –267 (2006). http://dx.doi.org/10.1136/bjo.2005.081224 BJOPAL 0007-1161 Google Scholar

2. 

Preferred Practice Pattern® Guidelines. Primary Open-Angle Glaucoma. San Francisco, CA: American Academy of Ophthalmology; 2010. Available at: www.aao.org/ppp Google Scholar

3. 

M. G. Hattenhaueret al., “The probability of blindness from open-angle glaucoma,” Ophthalmology, 105 (11), 2099 –2104 (1998). http://dx.doi.org/10.1016/S0161-6420(98)91133-2 OPANEW 0743-751X Google Scholar

4. 

Y. H. Kwonet al., “Rate of visual field loss and long-term visual outcome in primary open-angle glaucoma,” Am. J. Ophthalmol., 132 (1), 47 –56 (2001). http://dx.doi.org/10.1016/S0002-9394(01)00912-6 AJOPAA 0002-9394 Google Scholar

5. 

J. S. Schumanet al., “Quantification of nerve fiber layer thickness in normal and glaucomatous eyes using optical coherence tomography,” Arch. Ophthalmol., 113 (5), 586 –596 (1995). http://dx.doi.org/10.1001/archopht.1995.01100050054031 AROPAW 0003-9950 Google Scholar

6. 

B. Censeet al., “Thickness and birefringence of healthy retinal nerve fiber layer tissue measured with polarization-sensitive optical coherence tomography,” Invest. Ophthalmol. Vis. Sci., 45 (8), 2606 –2612 (2004). http://dx.doi.org/10.1167/iovs.03-1160 IOVSDA 0146-0404 Google Scholar

7. 

F. A. Medeiroset al., “Evaluation of retinal nerve fiber layer, optic nerve head, and macular thickness measurements for glaucoma detection using optical coherence tomography,” Am. J. Ophthalmol., 139 (1), 44 –55 (2005). http://dx.doi.org/10.1016/j.ajo.2004.08.069 AJOPAA 0002-9394 Google Scholar

8. 

J. E. DeLeon Ortegaet al., “Effect of glaucomatous damage on repeatability of confocal scanning laser ophthalmoscope, scanning laser polarimetry, and optical coherence tomography,” Invest. Ophthalmol. Vis. Sci., 48 (3), 1156 –1163 (2007). http://dx.doi.org/10.1167/iovs.06-0921 IOVSDA 0146-0404 Google Scholar

9. 

B. Povazayet al., “Minimum distance mapping using three-dimensional optical coherence tomography for glaucoma diagnosis,” J. Biomed. Opt., 12 (4), 041204 (2007). http://dx.doi.org/10.1117/1.2773736 JBOPFO 1083-3668 Google Scholar

10. 

T. Mumcuogluet al., “Improved visualization of glaucomatous retinal damage using high-speed ultrahigh-resolution optical coherence tomography,” Ophthalmology, 115 (5), 782 –789 (2008). http://dx.doi.org/10.1016/j.ophtha.2007.06.038 OPANEW 0743-751X Google Scholar

11. 

M. Yamanariet al., “Phase retardation measurement of retinal nerve fiber layer by polarization-sensitive spectral-domain optical coherence tomography and scanning laser polarimetry,” J. Biomed. Opt., 13 (1), 014013 (2008). http://dx.doi.org/10.1117/1.2841024 JBOPFO 1083-3668 Google Scholar

12. 

M. PircherC. K. HitzenbergerU. Schmidt-Erfurth, “Polarization sensitive optical coherence tomography in the human eye,” Prog. Retin. Eye. Res., 30 (6), 431 –451 (2011). http://dx.doi.org/10.1016/j.preteyeres.2011.06.003 PRTRES 1350-9462 Google Scholar

13. 

T. Shojiet al., “Assessment of glaucomatous changes in subjects with high myopia using spectral domain optical coherence tomography,” Invest. Ophthalmol. Vis. Sci., 52 (2), 1098 –1102 (2011). http://dx.doi.org/10.1167/iovs.10-5922 IOVSDA 0146-0404 Google Scholar

14. 

N. G. Strouthidiset al., “Longitudinal change detected by spectral domain optical coherence tomography in the optic nerve head and peripapillary retina in experimental glaucoma,” Invest. Ophthalmol. Vis. Sci., 52 (3), 1206 –1219 (2011). http://dx.doi.org/10.1167/iovs.10-5599 IOVSDA 0146-0404 Google Scholar

15. 

R. N. WeinrebS. ShakibaL. Zangwill, “Scanning laser polarimetry to measure the nerve fiber layer of normal and glaucomatous eyes,” Am. J. Ophthalmol., 119 (5), 627 –636 (1995). AJOPAA 0002-9394 Google Scholar

16. 

N. T. Choplinet al., “Retinal nerve fiber layer measurements do not change after LASIK for high myopia as measured by scanning laser polarimetry with custom compensation,” Ophthalmology, 112 (1), 92 –97 (2005). http://dx.doi.org/10.1016/j.ophtha.2004.07.027 OPANEW 0743-751X Google Scholar

17. 

N. T. ChoplinQ. ZhouR. W. Knighton, “Effect of individualized compensation for anterior segment birefringence on retinal nerve fiber layer assessments as determined by scanning laser polarimetry,” Ophthalmology, 110 (4), 719 –725 (2003). http://dx.doi.org/10.1016/S0161-6420(02)01899-7 OPANEW 0743-751X Google Scholar

18. 

W. V. Hatchet al., “Laser scanning tomography of the optic nerve head in ocular hypertension and glaucoma,” Br. J. Ophthalmol., 81 (10), 871 –876 (1997). http://dx.doi.org/10.1136/bjo.81.10.871 BJOPAL 0007-1161 Google Scholar

19. 

S. MelamedH. Levkovitch-Verbin, “Laser scanning tomography and angiography of the optic nerve head for the diagnosis and follow-up of glaucoma,” Curr. Opin. Ophthalmol., 8 (2), 7 –12 (1997). COOTEF Google Scholar

20. 

J. C. TanR. A. Hitchings, “Approach for identifying glaucomatous optic nerve progression by scanning laser tomography,” Invest. Ophthalmol. Vis. Sci., 44 (6), 2621 –2626 (2003). http://dx.doi.org/10.1167/iovs.02-0850 IOVSDA 0146-0404 Google Scholar

21. 

M. B. Waxet al., “Emerging perspectives in glaucoma: optimizing 24-h control of intraocular pressure,” Am. J. Ophthalmol., 133 (6), S1 –S10 (2002). AJOPAA 0002-9394 Google Scholar

22. 

D. J. ColemanS. Trokel, “Direct-recorded intraocular pressure variations in a human subject,” Arch. Ophthalmol., 82 (5), 637 –640 (1969). http://dx.doi.org/10.1001/archopht.1969.00990020633011 AROPAW 0003-9950 Google Scholar

23. 

S. MosaedJ. H. LiuR. N. Weinreb, “Correlation between office and peak nocturnal intraocular pressures in healthy subjects and glaucoma patients,” Am. J. Ophthalmol., 139 (2), 320 –324 (2005). http://dx.doi.org/10.1016/j.ajo.2004.09.062 AJOPAA 0002-9394 Google Scholar

24. 

A. J. Sitet al., “Sustained effect of travoprost on diurnal and nocturnal intraocular pressure,” Am. J. Ophthalmol., 141 (6), 1131 –1133 (2006). http://dx.doi.org/10.1016/j.ajo.2006.01.049 AJOPAA 0002-9394 Google Scholar

25. 

T. RealiniR. N. WeinrebS. R. Wisniewski, “Diurnal intraocular pressure patterns are not repeatable in the short term in healthy individuals,” Ophthalmology, 117 (9), 1700 –1704 (2010). http://dx.doi.org/10.1016/j.ophtha.2010.01.044 OPANEW 0743-751X Google Scholar

26. 

T. RealiniR. N. WeinrebS. Wisniewski, “Short-term repeatability of diurnal intraocular pressure patterns in glaucomatous individuals,” Ophthalmology, 118 (1), 47 –51 (2011). http://dx.doi.org/10.1016/j.ophtha.2010.04.027 OPANEW 0743-751X Google Scholar

27. 

R. N. WeinrebP. L. Kaufman, “Glaucoma research community and FDA look to the future, II: NEI/FDA Glaucoma Clinical Trial Design and Endpoints Symposium: measures of structural change and visual function,” Invest. Ophthalmol. Vis. Sci., 52 (11), 7842 –7851 (2011). http://dx.doi.org/10.1167/iovs.11-7895 IOVSDA 0146-0404 Google Scholar

28. 

F. A. Medeiroset al., “Combining structural and functional measurements to improve estimates of rates of glaucomatous progression,” Am. J. Ophthalmol., 153 (6), 1197 –1205 (2012). http://dx.doi.org/10.1016/j.ajo.2011.11.015 AJOPAA 0002-9394 Google Scholar

29. 

C. K. Leunget al., “Retinal nerve fiber layer imaging with spectral-domain optical coherence tomography: patterns of retinal nerve fiber layer progression,” Ophthalmology, (2012) http://dx.doi.org/10.1016/j.ophtha.2012.03.044 Google Scholar

30. 

K. Mansouriet al., “Assessment of rates of structural change in glaucoma using imaging technologies,” Eye, 25 (3), 269 –277 (2011). http://dx.doi.org/10.1038/eye.2010.202 12ZYAS 0950-222X Google Scholar

31. 

W. M. Grant, “Further studies on facility of flow through the trabecular meshwork,” Arch. Ophthal., 60 (4), 523 –533 (1958). http://dx.doi.org/10.1001/archopht.1958.00940080541001 AROPAW 0003-9950 Google Scholar

32. 

W. M. Grant, “Facility of flow through the trabecular meshwork,” Arch. Ophthal., 54 (2), 245 –248 (1955). http://dx.doi.org/10.1001/archopht.1955.00930020251012 AROPAW 0003-9950 Google Scholar

33. 

M. A. Johnstone, “The aqueous outflow system as a mechanical pump: evidence from examination of tissue and aqueous movement in human and non-human primates,” J. Glaucoma., 13 (5), 421 –438 (2004). JOGLES Google Scholar

34. 

M. A. Johnstone, “A new model describes an aqueous outflow pump and explores causes of pump failure in glaucoma,” Glaucoma. Essentials in Ophthalmology, 3 –34 Springer, Berlin, Heidelberg (2006). Google Scholar

35. 

M. A. Johnstone, “Aqueous humor outflow system overview,” Becker-Shaffer’s Diagnosis and Therapy of the Glaucomas, 25 –46 Springer, Berlin, Heidelberg St. Louis, Mo (2009). Google Scholar

36. 

M. JohnstoneE. MartinA. Jamil, “Pulsatile flow into the aqueous veins: Manifestations in normal and glaucomatous eyes,” Exper. Eye Res., 92 (5), 318 –327 (2011). http://dx.doi.org/10.1016/j.exer.2011.03.011 EXERA6 0014-4835 Google Scholar

37. 

C. K. Bahleret al., “Trabecular bypass stents decrease intraocular pressure in cultured human anterior segments,” Am. J. Ophthalmol., 138 (6), 988 –994 (2004). http://dx.doi.org/10.1016/j.ajo.2004.07.035 AJOPAA 0002-9394 Google Scholar

38. 

J. Renet al., “Ex vivo optical coherence tomography imaging of collector channels with a scanning endoscopic probe,” Invest. Ophthalmol. Vis. Sci., 52 (7), 3921 –3925 (2011). http://dx.doi.org/10.1167/iovs.10-6744 IOVSDA 0146-0404 Google Scholar

39. 

K. W. Ascher, “Glaucoma and the aqueous veins,” Am. J. Ophthalmol., 25 (11), 1309 –1315 (1942). AJOPAA 0002-9394 Google Scholar

40. 

H. Goldmann, “Abfluss des Kammerwassers beim Menschen,” Ophthalmologica, 111 (2–3), 146 –152 (1946). http://dx.doi.org/10.1159/000300317 OPHTAD 0030-3755 Google Scholar

41. 

R. StegmannA. PienaarD. Miller, “Viscocanalostomy for open-angle glaucoma in black African patients,” J. Cataract. Refract. Surg., 25 (3), 316 –322 (1999). http://dx.doi.org/10.1016/S0886-3350(99)80078-9 JCSUEV 0886-3350 Google Scholar

42. 

M. R. Heeet al., “Optical coherence tomography of the human retina,” Arch. Ophthalmol, 113 (3), 325 –332 (1995). http://dx.doi.org/10.1001/archopht.1995.01100030081025 AROPAW 0003-9950 Google Scholar

43. 

M. E. van Velthovenet al., “Recent developments in optical coherence tomography for imaging the retina,” Prog. Retin. Eye Res., 26 (1), 57 –77 (2007). http://dx.doi.org/10.1016/j.preteyeres.2006.10.002 PRTRES 1350-9462 Google Scholar

44. 

S. Asraniet al., “Detailed visualization of the anterior segment using fourier-domain optical coherence tomography,” Arch. Ophthalmol., 126 (6), 765 –771 (2008). http://dx.doi.org/10.1001/archopht.126.6.765 AROPAW 0003-9950 Google Scholar

45. 

P. H. TomlinsR. K. Wang, “Theory, developments and applications of optical coherence tomography,” J. Phys. D Appl. Phys., 38 (15), 2519 –2535 (2005). http://dx.doi.org/10.1088/0022-3727/38/15/002 JPAPBE 0022-3727 Google Scholar

46. 

T. Usuiet al., “Identification of Schlemm′s canal and its surrounding tissues by anterior segment fourier domain optical coherence tomography,” Invest. Ophthalmol. Vis. Sci., 52 (9), 6934 –6939 (2011). http://dx.doi.org/10.1167/iovs.10-7009 IOVSDA 0146-0404 Google Scholar

47. 

Y. Yasunoet al., “Visibility of trabecular meshwork by standard and polarization-sensitive optical coherence tomography,” J. Biomed. Opt., 15 (6), 061705 (2010). http://dx.doi.org/10.1117/1.3499421 JBOPFO 1083-3668 Google Scholar

48. 

L. Kagemannet al., “Identification and assessment of Schlemm's canal by spectral-domain optical coherence tomography,” Invest. Ophthalmol. Vis. Sci., 51 (8), 4054 –4059 (2010). http://dx.doi.org/10.1167/iovs.09-4559 IOVSDA 0146-0404 Google Scholar

49. 

G. Shiet al., “Morphometric measurement of Schlemm′s canal in normal human eye using anterior segment swept source optical coherence tomography,” J. Bio. Optics, 17 (1), 016016 (2012). http://dx.doi.org/10.1117/1.JBO.17.1.016016 JBOPFO 1083-3668 Google Scholar

50. 

L. Kagemannet al., “3D visualization of aqueous humor outflow structures in-situ in humans,” Exp. Eye Res., 93 (3), 308 –315 (2011). EXERA6 0014-4835 Google Scholar

51. 

P. Liet al., “In vivo microstructural and microvascular imaging of the human corneo-scleral limbus using optical coherence tomography,” Biomed. Opt. Express, 2 (11), 3109 –3118 (2011). http://dx.doi.org/10.1364/BOE.2.003109 BOEICL 2156-7085 Google Scholar

52. 

K. Singh, “Measurement of ocular fundus pulsation in healthy subjects using a novel Fourier-domain optical coherence tomography,” Invest. Ophthalmol. Vis. Sci., 52 (12), 8927 –8932 (2011). http://dx.doi.org/10.1167/iovs.11-7854 IOVSDA 0146-0404 Google Scholar

53. 

N. Dragostinoffet al., “Depth-resolved measurement of ocular fundus pulsations by low-coherence tissue interferometry,” J. Bio. Optics, 14 (5), 054047 (2009). http://dx.doi.org/10.1117/1.3251049 JBOPFO 1083-3668 Google Scholar

54. 

K. Singhet al., “Development of a novel instrument to measure the pulsatile movement of ocular tissues,” Exp. Eye Res., 91 (1), 63 –68 (2010). http://dx.doi.org/10.1016/j.exer.2010.03.022 EXERA6 0014-4835 Google Scholar

55. 

R. K. WangZ. MaS. J. Kirkpatrick, “Tissue Doppler optical coherence elastography for real time strain rate and strain mapping of soft tissue,” Appl. Phys. Lett., 89 (14), 144103 (2006). http://dx.doi.org/10.1063/1.2357854 APPLAB 0003-6951 Google Scholar

56. 

Y. Zhaoet al., “Phase-resolved optical coherence tomography and optical Doppler tomography for imaging blood flow in human skin with fast scanning speed and high velocity sensitivity,” Opt. Lett., 25 (2), 114 –116 (2000). http://dx.doi.org/10.1364/OL.25.000114 OPLEDP 0146-9592 Google Scholar

57. 

R. K. WangZ. Ma, “Real-time flow imaging by removing texture pattern artifacts in spectral-domain optical Doppler tomography,” Opt. Lett., 31 (20), 3001 –3003 (2006). http://dx.doi.org/10.1364/OL.31.003001 OPLEDP 0146-9592 Google Scholar

58. 

R. K. WangS. KirkpatrickM. Hinds, “Phase-sensitive optical coherence elastography for mapping tissue microstrains in real time,” Appl Phys. Lett., 90 (16), 164105 (2007). http://dx.doi.org/10.1063/1.2724920 APPLAB 0003-6951 Google Scholar

59. 

R. K. Wang, “Modelling optical properties of soft tissue by fractal distribution of scatterers,” J. Modern Optics, 47 (1), 103 –120 (2000). JMOPEW 0950-0340 Google Scholar

60. 

P. Liet al., “Assessment of strain and strain rate in embryonic chick heart in vivo using tissue Doppler optical coherence tomography,” Phys. Med. Biol., 56 (22), 7081 –7092 (2011). http://dx.doi.org/10.1088/0031-9155/56/22/006 PHMBA7 0031-9155 Google Scholar

61. 

B. A. SmitM. A. Johnstone, “Effects of viscoelastic injection into Schlemm's canal in primate and human eyes: potential relevance to viscocanalostomy,” Ophthalmology, 109 (4), 786 –792 (2002). http://dx.doi.org/10.1016/S0161-6420(01)01006-5 OPANEW 0743-751X Google Scholar

62. 

D. F. WoodwardA. H.-P. KraussS. F. E. Nilsson, “Bimatoprost effects on aqueous humor dynamics in monkeys,” J. Ophthalmol., 2010 926192 (2010). http://dx.doi.org/10.1155/2010/926192 JOOPA8 2090-004X Google Scholar

63. 

C. B. Toriset al., “Aqueous humor dynamics in monkeys with laser-induced glaucoma,” J Ocular Pharm Therapeutics: the Official Journal of the Association for Ocular Pharmacology and Therapeutics, 16 (1), 19 –27 (2000). Google Scholar

64. 

M. A. JohnstoneW. G. Grant, “Pressure-dependent changes in structures of the aqueous outflow system of human and monkey eyes,” Am. J. Ophthalmol., 75 (3), 365 –383 (1973). AJOPAA 0002-9394 Google Scholar

65. 

M. A. Johnstone, “Pressure-dependent changes in nuclei and the process origins of the endothelial cells lining Schlemm's canal,” Invest. Ophthalmol. Vis. Sci., 18 (1), 44 –51 (1979). IOVSDA 0146-0404 Google Scholar

66. 

N. EhlersT. BramsenS. Sperling, “Applanation tonometry and central corneal thickness,” Acta. Ophthalmol., 53 (1), 34 –43 (1975). http://dx.doi.org/10.1111/j.1755-3768.1975.tb01135.x AOOUAQ 1755-375X Google Scholar

67. 

M. C. Grieshaberet al., “Effect of central corneal thickness on dynamic contour tonometry and Goldmann applanation tonometry in primary open-angle glaucoma,” Arch. Ophthalmol., 125 (6), 740 –744 (2007). http://dx.doi.org/10.1001/archopht.125.6.740 AROPAW 0003-9950 Google Scholar

68. 

Y. Gelawet al., “The influence of central corneal thickness on intraocular pressure measured by goldmann applanation tonometry among selected Ethiopian communities,” J. Glaucoma, 19 514 –518 (2010). http://dx.doi.org/10.1097/IJG.0b013e3181ca7708 1057-0829 Google Scholar

69. 

P. Fogagnoloet al., “Test-retest variability of intraocular pressure and ocular pulse amplitude for dynamic contour tonometry: a multicentre study,” Br. J. Ophthalmol., 94 (4), 419 –423 (2010). http://dx.doi.org/10.1136/bjo.2009.165142 BJOPAL 0007-1161 Google Scholar

70. 

R. K. Wanget al., “Three dimensional optical angiography,” Opt. Express, 15 (7), 4083 –4097 (2007). http://dx.doi.org/10.1364/OE.15.004083 OPEXFF 1094-4087 Google Scholar

71. 

L. AnT. T. ShenR. K. Wang, “Using ultrahigh sensitive optical microangiography to achieve comprehensive depth resolved microvasculature mapping for human retina,” J. Biomed. Optics, 16 (10), 106013 (2011). http://dx.doi.org/10.1117/1.3642638 JBOPFO 1083-3668 Google Scholar
© 2012 Society of Photo-Optical Instrumentation Engineers (SPIE) 0091-3286/2012/$25.00 © 2012 SPIE
Peng Li, Roberto Reif, Zhongwei Zhi, Tueng Shen, Ruikang K. Wang, Elizabeth Martin, and Murray Johnstone "Phase-sensitive optical coherence tomography characterization of pulse-induced trabecular meshwork displacement in ex vivo nonhuman primate eyes," Journal of Biomedical Optics 17(7), 076026 (24 July 2012). https://doi.org/10.1117/1.JBO.17.7.076026
Published: 24 July 2012
Lens.org Logo
CITATIONS
Cited by 59 scholarly publications and 2 patents.