Open Access
16 December 2020 Visualization of two-dimensional transverse blood flow direction using optical coherence tomography angiography
Author Affiliations +
Abstract

Significance: Evaluation of vessel patency and blood flow direction is important in various medical situations, including diagnosis and monitoring of ischemic diseases, and image-guided vascular surgeries. While optical coherence tomography angiography (OCTA) is the most widely used functional extension of optical coherence tomography that visualizes three-dimensional vasculature, inability to provide information of blood flow direction is one of its limitations.

Aim: We demonstrate two-dimensional (2D) transverse blood flow direction imaging in en face OCTA.

Approach: A series of triangular beam scans for the fast axis was implemented in the horizontal direction for the first volume scan and in the vertical direction for the following volume scan, and the inter A-line OCTA was performed for the blood flow direction imaging while the stepwise pattern was used for each slow axis scan. The decorrelation differences between the forward and the backward inter A-line OCTA were calculated for the horizontal and the vertical fast axis scans, and the ratio of the horizontal and the vertical decorrelation differences was utilized to show the 2D transverse flow direction information.

Results: OCTA flow direction imaging was verified using flow phantoms with various flow orientations and speeds, and we identified the flow speed range relative to the scan speed for reliable flow direction measurement. We demonstrated the visualization of 2D transverse blood flow orientations in mouse brain vascular networks in vivo.

Conclusions: The proposed OCTA imaging technique that provides information of 2D transverse flow direction can be utilized in various clinical applications and preclinical studies.

1.

Introduction

Assessment of vessel patency and blood flow direction is critical in various medical situations, including diagnosis and treatment monitoring of ischemic diseases18 and image-guided vascular surgeries.921 Ischemic diseases are typically characterized by the disruption of the blood flow, such as changes in blood flow speed and blood flow direction.11 In the treatment of ischemic diseases, the natural and procedurally promoted vascular remodeling of existing arterial interconnections to form adequate collateral development for restoration of perfusion to the organs that have suffered ischemia is closely related to the patient outcomes.27 Non-invasive monitoring of blood flow direction and occasional flow direction reversal in the treatment of ischemic diseases can provide important information to assess therapeutic effectiveness.15 In various vascular surgeries such as vascular bypass,1317 vascular shunt,5,1821 and aneurysm surgery,9,10,14,17 blood flow characteristics such as blood flow path and blood flow direction are intentionally modified. Imaging the blood flow direction during the surgeries that confirms the intended changes of the paths and directions of the blood flow while avoiding any unintended alteration in vessel patency and blood flow direction is highly desirable for monitoring of the immediate effects of the surgery and for quality control of the vascular intervention.5,9,10

While optical coherence tomography angiography (OCTA) is the most widely used functional form of optical coherence tomography (OCT) that visualizes three-dimensional vasculature without a need of exogenous contrast agent injection,2226 inability to provide information of blood flow direction is one of its limitations. Doppler ultrasound35,813,20 and Doppler OCT2729 have been main imaging tools used for measuring the flow direction. However, the Doppler techniques can only provide the axial flow direction, whether the flow moves toward or away from the imaging wave source, and therefore usually visualize the axial flow directions in cross-sectional images. Moreover, Doppler OCT typically requires additional phase stabilization or phase calibration because it is based on measurement of the Doppler phase difference. Laser speckle imaging and multiphoton laser scanning microscopy also showed flow direction imaging, but they required vessel by vessel scanning for tracking injected fluorescent-labeled red blood cells or fluorescent microbeads.1,2

In this work, we demonstrate an OCTA imaging technique for the visualization of two-dimensional (2D) transverse blood flow direction information. We implemented the inter A-line OCTA using an isosceles triangular fast axis beam scan and calculated the difference between OCTA decorrelation signals acquired from the forward and backward beam scan sections to determine the flow direction in the fast axis. Performing a pair of consecutive OCTA volume scans with the horizontal fast axis scan and the vertical fast axis scan, respectively, provided 2D transverse flow direction information. OCTA flow direction imaging was verified using flow phantoms with various orientations. We demonstrated the visualization of the blood flow orientations in mouse brain vascular networks in vivo.

2.

Methods

In OCTA, the magnitude of the decorrelation signal is proportional to the displacement of the scatterers in the flow during the time interval between two measurements when the imaging beam is at the same location for the two measurements.3032 When the imaging beam moves between the two measurements, the inter A-line OCTA decorrelation signal is determined by the relative displacement of the scatterers with respect to the location of the imaging beam in each of the two measurements. For the same flow, the effective displacement of the scatterers becomes smaller when the imaging beam scans in the same direction of the flow than when it scans in the opposite direction of the flow as depicted in Fig. 1. Therefore, the decorrelation signal is also measured as a smaller value when the imaging beam scans in the same direction of the flow than when it scans in the opposite direction of the flow, and the one-dimensional flow direction information can be obtained by comparing the OCTA signals acquired during the forward and the backward imaging beam scans. The 2D flow direction information is obtained by performing a pair of the inter A-line OCTA volume scans with the fast axis scans in the horizontal direction and the vertical direction, respectively.

Fig. 1

The relative blood flow speed with respect to the imaging beam scan speed with (a) forward scan (imaging beam scans in the same direction of the flow) and (b) backward scan (imaging beam scans in the opposite direction of the flow).

JBO_25_12_126003_f001.png

An OCTA system with a lab-built short cavity wavelength-swept laser operating at 1300 nm as a light source was used in this work.33,34 The axial and transverse resolutions were measured to be 11.3 and 8.2  μm in air, respectively, and the system sensitivity was measured to be 102 dB at an A-line rate of 220.4 kHz. Figure 2 illustrates the beam scanning protocol for flow direction OCTA imaging. It consists of a pair of consecutive inter A-line OCTA volume scans. A volume scan with horizontal direction as the fast axis is performed, followed by another volume scan with vertical direction as the fast axis. At each B-scan position in the vertical direction, an isosceles triangular scan was performed to provide both forward and backward fast axis scans with the same scan speed. Upon completion of each isosceles triangular scan, the imaging beam was subsequently stepped to the next B-scan position. The same volume scan is repeated but with the fast axis isosceles triangular scan in the vertical direction and the slow axis stepwise scan in the horizontal direction.

Fig. 2

Beam scanning protocol of the inter A-line OCTA for 2D transverse flow direction imaging.

JBO_25_12_126003_f002.png

The decorrelation signal from the inter A-line OCTA is proportional to the displacement of the scatterers in the flow during the inter A-line time interval and can be expressed as a linear function of the displacement, such as D=α×Δx=α×v×Δt, when the displacement is smaller than approximately 22% of the resolution,31 where D is the decorrelation,32 Δx is the displacement of moving scatterers in the flow, α is a proportional coefficient, v is the flow speed relative to the beam scan speed, and Δt is the inter A-line time interval. With the horizontal (x) fast axis scan, the decorrelation signal can be expressed as Dx,f=αΔt|vflowcosθvscan|+Do for the forward scan section and Dx,b=αΔt|vflowcosθ+vscan|+Do for the backward scan section, where θ is the angle between the flow direction and the horizontal forward direction, vflow is the flow speed, vscan is the fast axis scan speed, and Do is the baseline decorrelation signal with no directional flow. If the scan speed is higher than the horizontal flow speed, the difference between decorrelation signals obtained from the forward scan and the backward scan is given by ΔDx=Dx,bDx,f=2αΔtvflowcosθ. With the vertical (y) fast axis scan, the decorrelation signal difference between the forward and backward OCTA can be expressed as ΔDy=Dy,bDy,f=2αΔtvflowsinθ. Therefore, the flow direction information can be obtained as θ=arctan (ΔDy/ΔDx).

3.

Results and Discussions

We first demonstrated the visualization of the 2D transverse flow direction using flow phantoms oriented in various angles. The flow phantom consisted of a transparent tube with an inner diameter of 150  μm (CAP360-150P, LabSmith) embedded in a polydimethylsiloxane block with TiO2 added as scattering particles. Milk was pumped using a syringe pump (NE-300, New Era Pump Systems, Inc.) to form flow in the tube. The average flow speed was set at 50 mm/s, and an area over 2  mm×2  mm was imaged with 2048 A-lines ×1024 B-scans. Figure 3 shows a clear visualization of the 2D transverse flow directions inside the tube phantoms oriented eight different directions with a 45-deg interval.

Fig. 3

Flow phantom images showing transverse flow direction measurements. Arrows inside the tube flow phantom indicate the flow directions. Flow speed was set at 50 mm/s. Scale bar: 500  μm.

JBO_25_12_126003_f003.png

To identify the flow speed range that provides a reliable flow direction measurement, we imaged the flow phantom of various speeds ranging from 5 to 200 mm/s using two different fast axis scan speeds. Figures 4(a) and 4(b) show color-encoded transverse flow direction OCTA images of a 2  mm×2  mm area of the flow phantom acquired with 2048 A-lines/B-scan and 6144 A-lines/B-scan, which correspond to the fast axis scan speed of 215 and 71.7 mm/s, respectively, for four different linear flows and a circular flow with various speeds. Figures 4(c) and 4(d) show the measured mean flow directions and their standard deviations within the ROI of 120  μm width and 1.2 mm length along the center of the tube for the four linear flow phantoms. The flow directions were well-determined with the standard deviations smaller than 5 deg when the flows were faster than 22 mm/s with 2048 A-lines/B-scan (vscan=215  mm/s) and 7 mm/s with 6144 A-lines/B-scan (vscan=71.7  mm/s), respectively, which approximately correspond to 10% of the beam scan speed. As the flow direction is determined by the difference in relative flow speeds with respect to the beam scan speeds acquired during the forward and backward beam scan periods, the accurate flow direction measurement is hampered when the flow speed is too small compared to the scan speed. To maintain accurate flow direction measurements for slow flows, the scan speed needs to be slowed down by oversampling at the expense of the imaging speed.

Fig. 4

Flow phantom images over a 2  mm×2  mm area with various flow speeds ranging from 5 to 200 mm/s. (a) Images acquired using 2048 A-lines/B-scan. (b) Images acquired using 6144 A-lines/B-scan. (c) Mean flow directions measured with the linear flow phantoms at various flow speeds. (d) Standard deviations of the flow direction measurements for the linear flow phantoms at various flow speeds. Scale bar: 500  μm.

JBO_25_12_126003_f004.png

It is also noteworthy that the standard deviation of the flow direction measurement was the minimum at a flow speed of approximately 50 mm/s and increases as the flow speed increases with the fast axis scan speed of 71.7 mm/s (6144 A-lines/B-scan). For the fast axis scan speed of 215 mm/s (2048 A-lines/B-scan), the standard deviation of the flow direction measurement monotonically decreased to a flow speed of 150 mm/s, but a slight increase was observed as the flow speed reached approximately 200 mm/s. For the determination of the flow direction information, we assumed that the fast axis scan speed is higher than the flow speed component parallel to the fast axis. While the fast axis scan speed is always higher than the fastest flow speed used in the experiment with the fast axis scan speed of 215 mm/s, the flow speed component parallel to the fast axis can be higher than the scan speed with the fast axis scan speed of 71.7 mm/s (6144 A-lines/B-scan), which contributed the increased standard deviation of the flow direction measurement at the higher flow speeds when imaging with 6144 A-lines/B-scan.

In vivo OCTA imaging of blood flow direction was demonstrated in the cranial windows implanted on the mouse brains. The animal experiments were performed under the approval of the Institutional Animal Care and Use Committee of Korea Advanced Institute of Science and Technology (KAIST).35 Male C57BL/6 mice (>8 weeks) were anesthetized with 1.5% to 3% (v/v) of isoflurane during surgical and imaging procedures. The head of a mouse was fixed in a stereotaxic frame and a circular cranial window was implanted on the cortex for imaging. The mouse brain over a field of 2  mm×2  mm (6144 A-lines ×1024 B-scans) was imaged with the proposed beam scanning protocol for the flow direction imaging. Figures 5(a), 5(c), and 5(e) show the en face mean projections of the inter B-scan OCTA volume images. Figure 5(c) shows the en face OCTA image acquired by rotating the same mouse cranial window used for the imaging of Fig. 5(a) by 180 deg. Figure 5(e) shows a cranial window OCTA image of another mouse. In Figs. 5(b), 5(d), and 5(f), the color-encoded information of blood flow directions was overlaid on the en face OCTA images. Cross-sectional OCTA images and color-encoded blood flow direction images at locations indicated by dashed lines in the en face OCTA images are shown in Figs. 5(e1), 5(e2), 5(f1), and 5(f2). The blood flow directions in relatively large vessels with the blood flow speed higher than the slowest direction-detectable flow speed, which is approximately 7 mm/s in the image with 6144 A-lines×1024 B-scans for a 2  mm×2  mm area, were well visualized. Flow directions in the capillaries were not detected because the blood flow speed in most capillaries is lower than the slowest direction-detectable flow speed. While the direction of the slow blood flows in capillaries may be detected by decreasing the fast axis scan speed through dense oversampling, reliable flow direction measurements can be limited for flows faster than the fast axis scan speed.

Fig. 5

In vivo 2D transverse flow direction images of mouse brains. (a), (c), and (e) En face mean projections of the inter B-scan OCTA volume images, and (b), (d), and (f) color-encoded blood flow direction information overlaid on the corresponding en face OCTA images. (e1), (e2), (f1), and (f2) Cross-sectional OCTA images and color-encoded blood flow direction images at locations indicated by dashed lines in the en face OCTA images. Scale bar: 500  μm

JBO_25_12_126003_f005.png

4.

Conclusion

We demonstrate an OCTA imaging technique that provides information of 2D transverse flow direction. Implementation of a pair of consecutive inter A-line OCTA volume scans using triangular fast axis beam scans in the horizontal direction for the first volume and in the vertical direction for the following volume, respectively, enabled determination of the 2D transverse flow direction through calculation of difference between OCTA decorrelation signals acquired during the forward and backward fast axis beam scans. Through flow phantom experiments, we verified the OCTA flow direction imaging and identified the flow speed range relative to the scan speed for reliable flow direction measurement. In vivo OCTA for 2D transverse blood flow direction was presented in the mouse cranial window imaging. The proposed OCTA imaging technique may be very useful in various applications where the assessment of the transverse blood flow direction information is important such as diagnosis and treatment monitoring of ischemic diseases, image-guided vascular surgeries, and preclinical studies using animal vascular disease models.

Disclosures

The authors declare that no competing interests exist.

Acknowledgments

This work was supported by the Ministry of Health and Welfare of Korea (No. HI15C0001) and the National Research Foundation of Korea (NRF) (Nos. 2016M3C7A1913844 and 2020R1A2C3009667).

References

1. 

H. Levy, D. Ringuette and O. Levi, “Rapid monitoring of cerebral ischemia dynamics using laser-based optical imaging of blood oxygenation and flow,” Biomed. Opt. Express, 3 (4), 777 –791 (2012). https://doi.org/10.1364/BOE.3.000777 BOEICL 2156-7085 Google Scholar

2. 

J. K. Meisner et al., “Trans-illuminated laser speckle imaging of collateral artery blood flow in ischemic mouse hindlimb,” J. Biomed. Opt., 18 (9), 096011 (2013). https://doi.org/10.1117/1.JBO.18.9.096011 JBOPFO 1083-3668 Google Scholar

3. 

A. V. Alexandrov, C. F. Bladin and J. W. Norris, “Intracranial blood flow velocities in acute ischemic stroke,” Stroke, 25 (7), 1378 –1383 (1994). https://doi.org/10.1161/01.STR.25.7.1378 SJCCA7 0039-2499 Google Scholar

4. 

L. Li et al., “Evaluation of cerebral blood flow changes in focal cerebral ischemia rats by using transcranial Doppler ultrasonography,” Ultrasound Med. Biol., 36 (4), 595 –603 (2010). https://doi.org/10.1016/j.ultrasmedbio.2010.01.005 USMBA3 0301-5629 Google Scholar

5. 

A. M. Lacy et al., “Long-term effects of distal splenorneal shunt on hepatic hemodynamics and liver function in patients with cirrhosis: importance of reversal of portal blood flow,” Hepatology, 15 (4), 616 –622 (1992). https://doi.org/10.1002/hep.1840150411 HEPADF 0161-0538 Google Scholar

6. 

H. K. Shin et al., “Vasoconstrictive neurovascular coupling during focal ischemic depolarizations,” J. Cereb. Blood Flow Metab., 26 (8), 1018 –1030 (2006). https://doi.org/10.1038/sj.jcbfm.9600252 Google Scholar

7. 

M. Choy et al., “The chronic vascular and haemodynamic response after permanent bilateral common carotid occlusion in newborn and adult rats,” J. Cereb. Blood Flow Metab., 26 (8), 1066 –1075 (2006). https://doi.org/10.1038/sj.jcbfm.9600259 Google Scholar

8. 

R. Aaslid, T. Markwalder and H. Nornes, “Noninvasive transcranial Doppler ultrasound recording of glow velocities in basal cerebral arteries,” J. Neurosurg., 57 (6), 769 –774 (1982). https://doi.org/10.3171/jns.1982.57.6.0769 JONSAC 0022-3085 Google Scholar

9. 

S. A- Hanjani et al., “The utility of intraoperative blood flow measurement during aneurysm surgery using an ultrasonic perivascular flow probe,” Oper. Neurosurg., 58 (4), 305 –312 (2006). https://doi.org/10.1227/01.NEU.0000209339.47929.34 Google Scholar

10. 

F. Lindseth et al., “Blood flow imaging: an angle-independent ultrasound modality for intraoperative assessment of flow dynamics in neurovascular surgery,” Oper. Neurosurg., 65 (Suppl_6), 149 –157 (2009). https://doi.org/10.1227/01.NEU.0000345945.92559.C5 Google Scholar

11. 

J. E. Bailes et al., “Intraoperative microvascular Doppler sonography in aneurysm surgery,” Neurosurgery, 40 (5), 965 –970 (1997). https://doi.org/10.1097/00006123-199705000-00018 NEQUEB Google Scholar

12. 

R. Firsching, H. J. Synowitz and J. Hanebeck, “Practicability of intraoperative microvascular Doppler sonography in aneurysm surgery,” Minimum Invasive Neurosurg., 43 (3), 144 –148 (2000). https://doi.org/10.1055/s-2000-12261 Google Scholar

13. 

Y. H. Park et al., “Intraoperative trans-fontanellar cerebral ultrasonography in infants during cardiac surgery under cardiopulmonary bypass: an observational study,” J. Clin. Monit. Comput., 31 159 –165 (2017). https://doi.org/10.1007/s10877-015-9815-3 Google Scholar

14. 

Y. Murai et al., “The feasibility of detecting cerebral blood flow direction using the indocyanine green video angiography,” Neurosurg. Rev., 39 685 –690 (2016). https://doi.org/10.1007/s10143-016-0726-7 NSREDV 1437-2320 Google Scholar

15. 

A. Rüffer et al., “Comparable cerebral blood flow in both hemispheres during regional cerebral perfusion in infant aortic arch surgery,” Ann. Thorac. Surg., 103 (1), 178 –185 (2017). https://doi.org/10.1016/j.athoracsur.2016.05.088 Google Scholar

16. 

A. Nagm et al., “Intraoperative evaluation of reverse bypass using a naturally formed ‘Bonnet’ superficial temporal artery: technical note,” World Neurosurg., 88 603 –608 (2016). https://doi.org/10.1016/j.wneu.2015.10.087 Google Scholar

17. 

E. Orita et al., “Four-dimensional flow MRI analysis of cerebral blood flow before and after high-flow extracranial–intracranial bypass surgery with internal carotid artery ligation,” Neurosurgery, 85 (1), 58 –64 (2019). https://doi.org/10.1093/neuros/nyy192 NEQUEB Google Scholar

18. 

J. Belghiti et al., “Long-term loss of Warren’s shunt selectivity: angiographic demonstration,” Arch. Surg., 116 (9), 1121 –1124 (1981). https://doi.org/10.1001/archsurg.1981.01380210009002 Google Scholar

19. 

E. G. Grant et al., “Color Doppler imaging of portosystemic shunts,” Am. J. Roentgen, 154 (2), 393 –397 (1990). https://doi.org/10.2214/ajr.154.2.2105035 Google Scholar

20. 

L. Bolondi et al., “Morphological and hemodynamic changes in the portal venous system after distal splenorenal shunt: an ultrasound and pulsed Doppler study,” Hepatology, 8 (3), 652 –657 (1988). https://doi.org/10.1002/hep.1840080336 HEPADF 0161-0538 Google Scholar

21. 

K. Ohnishi et al., “Direction of splenic venous flow assessed by pulsed Doppler flowmetry in patients with a large splenorenal shunt: relation to spontaneous hepatic encephalopathy,” Gastroenterology, 89 (1), 180 –185 (1985). https://doi.org/10.1016/0016-5085(85)90760-7 GASTAB 0016-5085 Google Scholar

22. 

J. Vakoc et al, “Three-dimensional microscopy of the tumor microenvironment in vivo using optical frequency domain imaging,” Nat. Med., 15 (10), 1219 –1223 (2009). https://doi.org/10.1038/nm.1971 1078-8956 Google Scholar

23. 

A. Szkulmowska et al., “Three-dimensional quantitative imaging of retina and choroidal blood flow velocity using joint spectral and time domain optical coherence tomography,” Opt. Express, 17 (13), 10584 –10598 (2009). https://doi.org/10.1364/OE.17.010584 OPEXFF 1094-4087 Google Scholar

24. 

B. Braaf et al., “Angiography of the retina and the choroid with phase-resolved OCT using interval-optimized backstitched B-scans,” Opt. Express, 20 (18), 20516 –20534 (2012). https://doi.org/10.1364/OE.20.020516 OPEXFF 1094-4087 Google Scholar

25. 

C. Blatter et al., “In situ structural and microangiographic assessment of human skin lesions with high-speed OCT,” Biomed. Opt. Express, 3 (10), 2636 –2646 (2012). https://doi.org/10.1364/BOE.3.002636 BOEICL 2156-7085 Google Scholar

26. 

G. Liu et al., “High-resolution imaging of microvasculature in human skin in-vivo with optical coherence tomography,” Opt. Express, 20 (7), 7694 –7705 (2012). https://doi.org/10.1364/OE.20.007694 OPEXFF 1094-4087 Google Scholar

27. 

J. A. Izatt et al, “In vivo bidirectional color Doppler flow imaging of picoliter blood volumes using optical coherence tomography,” Opt. Lett., 22 (18), 1439 –1441 (1997). https://doi.org/10.1364/OL.22.001439 OPLEDP 0146-9592 Google Scholar

28. 

R. A. Leitgeb et al., “Real-time measurement of in vitro flow by Fourier-domain color Doppler optical coherence tomography,” Opt. Lett., 29 (2), 171 –173 (2004). https://doi.org/10.1364/OL.29.000171 OPLEDP 0146-9592 Google Scholar

29. 

I. Grulkowski et al, “Scanning protocols dedicated to smart velocity ranging in spectral OCT,” Opt. Express, 17 (26), 23736 –23754 (2009). https://doi.org/10.1364/OE.17.023736 OPEXFF 1094-4087 Google Scholar

30. 

S. B. Ploner et al., “Toward quantitative optical coherence tomography angiography: visualizing blood flow speeds in ocular pathology using variable interscan time analysis (VISTA),” Retina, 36 S118 –S126 (2016). https://doi.org/10.1097/IAE.0000000000001328 RETIDX 0275-004X Google Scholar

31. 

T. J. Park et al., “Wide dynamic range high-speed three-dimensional quantitative OCT angiography with a hybrid-beam scan,” Opt. Lett., 43 (10), 2237 –2240 (2018). https://doi.org/10.1364/OL.43.002237 OPLEDP 0146-9592 Google Scholar

32. 

A. S. Nam, I. Chico-Calero and B. J. Vakoc, “Complex differential variance algorithm for optical coherence tomography angiography,” Biomed. Opt. Express, 5 (11), 3822 –3832 (2014). https://doi.org/10.1364/BOE.5.003822 BOEICL 2156-7085 Google Scholar

33. 

H. S. Cho et al., “High frame-rate intravascular optical frequency domain imaging in vivo,” Biomed. Opt. Express, 5 (1), 223 –232 (2014). https://doi.org/10.1364/BOE.5.000223 BOEICL 2156-7085 Google Scholar

34. 

C. Jun et al, “All-fiber wavelength swept ring laser based on Fabry–Perot filter for optical frequency domain imaging,” Opt. Express, 22 (21), 25805 –25814 (2014). https://doi.org/10.1364/OE.22.025805 OPEXFF 1094-4087 Google Scholar

35. 

P. Shin et al., “Quantitative hemodynamic analysis of cerebral blood flow and neurovascular coupling using optical coherence tomography,” J. Cereb. Blood Flow Metab., 39 (10), 1983 –1994 (2019). https://doi.org/10.1177/0271678X18773432 Google Scholar

Biography

Inho Shin received his PhD in mechanical engineering from KAIST in 2020. His research interests include the development of OCT and its applications to mouse ischemic disease models and other biomedical studies.

Wang-Yuhl Oh received his PhD in physics from KAIST in 1997. He was an instructor/research fellow at Harvard Medical School and Massachusetts General Hospital, Wellman Center for Photomedicine. In 2009, he joined KAIST as an associate professor of the mechanical engineering. His research interests include development of novel optical imaging systems and ultrafast light sources for biomedical applications, metrology, and inspection.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Inho Shin and Wang-Yuhl Oh "Visualization of two-dimensional transverse blood flow direction using optical coherence tomography angiography," Journal of Biomedical Optics 25(12), 126003 (16 December 2020). https://doi.org/10.1117/1.JBO.25.12.126003
Received: 6 August 2020; Accepted: 24 November 2020; Published: 16 December 2020
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
KEYWORDS
Blood circulation

Visualization

Optical coherence tomography

Angiography

Surgery

In vivo imaging

Brain

Back to Top