18 November 2022 Geometry design for a fully insertable glucose biosensor with multimodal optical readout
Author Affiliations +
Abstract

Significance

Insertable optical continuous glucose monitors (CGMs) with wearable readers are a strong option for monitoring individuals with diabetes. However, a fully insertable CGM requires a small form factor while still delivering sufficient signal to be read through tissue by an external device. Previous work has suggested that a multimodal repeating unit (barcode) approach may meet these requirements, but the biosensor geometry must be optimized to meet performance criteria.

Aim

This work details in silico trials conducted to evaluate the geometry of a fully insertable multimodal optical biosensor with respect to both optical output and species diffusion in vivo.

Approach

Monte Carlo modeling is used to evaluate the luminescent output of three presupposed biosensor designs based on size constraints for an injectable and logical placement of the bar code compartments. Specifically, the sensitivity of the luminescent output to displacement of the biosensor in the X and Y directions, overall size of the selected design, and size of an individual repeating unit are analyzed. Further, an experimentally validated multiphysics model is used to evaluate the diffusion and reaction of glucose and oxygen within the biosensor to estimate the occurrence of chemical crosstalk between the assay components.

Results

A stacked cylinder multimodal biosensor 4.4 mm in length with repeating units 0.36 mm in length was found to yield a greater luminescent output than the current “barcode” biosensor design. In addition, it was found that a biosensor with enzymatic elements does not significantly deplete glucose locally and thus does not impact the diffusion profile of glucose in adjacent compartments containing nonenzymatic assays.

Conclusions

Computational modeling was used to design the geometry of a multimodal, insertable, and optical CGM to ensure that the optical output and chemical diffusion profile are sufficient for this device to function in vivo.

1.

Introduction

Since the development of the first biosensor in modern history, a biological sample glucometer by Clark and Lyons in 1962,1 biosensors have expanded to be a ubiquitous component of acute and chronic disease management. In 2020, biosensors had grown to a global market size of $22,400,000,000 USD and were the largest number of commercial medical devices.1,2 With biosensors providing useful diagnostic and prognostic information related to physiological parameters, many diseases, such as cardiovascular disease and diabetes, can be treated and managed.3,4 A growing research space is that of continuous and skin-integrated biosensors. Often referred to as implantable, insertable, or injectable, these devices rest on or within a patient and enable precision medicine by providing real-time measurement of physiological biomarkers with minimal inconvenience to the patient.5,6 Due to its prevalence, morbidity, and mortality, diabetes has been the focus of much research for continuous biosensors and personalized/precision medicine.7,8 Continuous glucose monitors (CGMs) have been proven to improve diabetes management and reduce negative outcomes affiliated with disease progression, and they can provide care to the over 122 million American adults that are estimated to be diabetic or prediabetic.911 Moreover, although CGMs were first approved by the Food and Drug Administration (FDA) in 1999, recent work is being done to integrate CGMs holistically into diabetes management to supplement essential interventional components, such as diet monitoring.12,13 Thus, as CGMs become more integral to diabetes management best practices, it is key to continue research and development in new CGM devices.

Although becoming more integral to diabetes management, current commercial CGMs—with the three market leaders being Dexcom (Dexcom G6), Medtronic (Medtronic Guardian/MiniMed 670G), and Abbott (Abbott Freestyle Libre Pro)—are relatively expensive, create risk for infection, and have a short lifetime.14,15 The only fully implantable glucose biosensor, Senseonic’s Eversense, requires a physician to perform implantation and explantation and is cost prohibitive. As described previously, several researchers are developing an inexpensive, multimodal, and fully insertable glucose biosensor, which is unique in its small form factor (6.5  mm×0.7  mm×0.9  mm, enabling injection through a hypodermic needle).16,17 In addition, this biosensor has repeating unit functionality, wherein two separate assays independently provide glucose predictions, and optical signal transduction, which enables a combination of the biosensor’s recognition and transduction elements. Specifically, this biosensor uses a phosphorescence lifetime decay within an enzymatic schema (assay 1) and Förster resonance energy transfer within a competitive binding schema (assay 2). Previous work demonstrated the use of the Monte Carlo (MC) method to establish that the anticipated optical signal strength of this biosensor is great enough to be detected by a photodiode as a significant concern with optical and indwelling biosensors is the inability of the signal to propagate through tissue.18

However, there remains a need to further consider the geometry of this biosensor to maximize performance, namely modifications to the form factor to increase optical output and ensure satisfactory chemical functionality within the subcutaneous space is of great importance. Computational modeling is a valuable tool for evaluating biosensor geometry as in silico methods can provide accurate results more efficiently with respect to time and cost when properly verified and validated. Specifically, MC methods have frequently been used to estimate photon propagation through turbid media.1921

There is also a need to validate and create a model that can be used to estimate reaction kinetics and molecular interactions between the biosensor and the in vivo environment as a function of the biosensor geometry. Specifically, whether the two assays interact chemically need to be investigated to explore whether the luminescent output of one assay is affected by the diffusion gradients or local conditions created by the other assay. This model does not currently exist; thus, we present the use of COMSOL Multiphysics to build such a model and validate its use experimentally. Multiphysics platforms, such as COMSOL®, are effective tools for evaluating complex systems, such as biosensors, due to their ability to simulate complex conditions.22 For example, multiphysics platforms can use computational fluid dynamics methods to evaluate the deformation of a biosensor from fluid flow and understand how that deformation affects the diffusion profiles of species into the biosensor.23,24 Here, we present the use of MC modeling to evaluate luminescent output of a biosensor as it depends on geometry, as well as develop and validate a multiphysics model to evaluate chemical functionality.

2.

Materials and Methods

2.1.

MC Model

2.1.1.

Model geometry and properties

The MC modeling of photon propagation is a scientifically appropriate option for evaluating the biosensor size, design, and subsequent impact on luminescent output. Thus, an MC model created via MCmatlab, previously described elsewhere, was used within this work.25 In short, a three-layer skin model (epidermis, dermis, and hypodermis) of the dorsal wrist 17  mm×17  mm×4  mm (length × width × height) was created, as shown in Fig. 1. Optical properties for tissue were derived from Jacques et al.,26 and thickness values were derived from Chen et al.,27 Sandby-Moller et al.,28 and Van Mulder et al.29 Optical properties used for this model are found in Table 1. The refractive index for all components is set to 1 for refractive index matching, which is a needed step to simulate geometries that are heterogenous in the Z direction within MCmatlab.21 Within this tissue model is a biosensor of variable location and size, as described in Sec. 2.1.2. The simulated light source is located at the origin (0,0,0) and is light emitting diode (LED)-like with a top-hat distribution in the near field and a Lambertian distribution in the far field with a side length of 1.75 mm and a wavelength of 680 nm. These parameters were chosen to match commercial offerings from ThorLabs single-color visible LEDs because these are assumed to be high-probability options for designing the associated wearable in future work. Luminescent emission was simulated at 800 nm, and the power yield of the fluorophores was set to 1 to enable a comparison of results between simulations while minimizing computational time. Each voxel was a 0.05-mm cube. Each simulation was repeated three times with 1e7  photons, which was shown to be sufficient through parametric analysis from previous work.17 To better analyze the biosensor response, the saved output of each simulation is the emission light that reached the top surface of the geometry as opposed to a photodiode or detector. All simulations were completed in MATLAB 2019b on a Lenovo Legion Y720b (Windows 10) and were parallelized to an NVIDIA GeForce GTX 1060.

Fig. 1

MC model geometry.

JBO_27_11_117001_f001.png

Table 1

The MC model optical properties.

Featureμa (680/800) cm−1μs (680/800) cm−1ngQuantum yield
Epidermis0.78/0.46294.1/250.010.90
Dermis0.48/0.30195.3/140.810.90
Poly(ethylene glycol) diacrylate hydrogel1e6/1e65e4/5e410.750
Assay 15/1e61e6/1e610.751
Assay 25/1e61e6/1e610.751

2.1.2.

MC simulations

A simulated biosensor was inserted within the tissue model described in Sec 2.1.1 to evaluate luminescent performance of both assays under different conditions. As shown in Fig. 2, three different designs were considered. Each sensor was required to have the following design constraints: (1) be 6.5 mm in length, (2) have repeating units for the two sensing assays, (3) enable 17-gauge hypodermic needle insertion, and (4) have visible areas with both sensing chemistries across its four sides. These designs were tested to quantify overall luminescent output by simulating photon interactions in the case in which the center of the biosensor rests at 0  mm×0  mm×2  mm and all emission light reaching the top surface of the skin is assumed to be captured. However, in practice, patient motion will potentially result in offsets being between the center illumination source axis and the insertable biosensor axis. Thus, sensitivity analyses of these designs to offsets are required.

Fig. 2

Presupposed biosensor designs (mm). Green and purple domains represent alternating sensing assays. Created with Ref. 30.

JBO_27_11_117001_f002.png

To conduct an analysis of the biosensor luminescent output sensitivity to displacement between the illumination source and the biosensor, simulations were conducted wherein each iteration of the biosensor was centered at 5, 3, 0, 3, and 5  mm in the X and Y dimensions as well as 1.25, 2, and 2.75 in the Z dimension. This work determined which biosensor has the least variation in luminescent output as a function of X, Y, and Z dimension offsets and overall which biosensor design yields the greatest luminescent output.

The next step is to assess the length of the longest dimension and the thickness of the individual repeating unit for the biosensor design. The length of the optimized biosensor can be found by analyzing the full width half maximum (FWHM) of excitation light that originates from the LED at the target implant depth of 2 mm. Then, the ideal thickness of a single sensing compartment was found by completing iterations of simulations wherein the 6.5-mm biosensor is divided into 1, 3, 5, 7, 9, 11, 13, and 15 compartments of each assay and located 5 mm in the positive and negative X and Y axes. It is hypothesized that the luminescent output of these simulations will eventually converge once the number of each sensing assays increases to a given amount. The assay compartment thickness wherein luminescent output converges is the optimal suggested size of the sensing compartment because then it can be said that a typical excitation beam will be exciting both assays relatively equally. This prevents one assay from failing due to insufficient excitation. From these MC simulations, the final output is a biosensor that will be determined based on its strong luminescent output with respect to the limitation of being able to be inserted via a hypodermic needle.

2.2.

COMSOL Multiphysics Model

2.2.1.

Validation of species diffusion, reaction, and phosphorescence decay

A two-dimensional multiphysics model (validation model) was constructed in COMSOL Multiphysics 5.2 (COMSOL, Inc., Sweden) to (1) represent the diffusion of glucose and oxygen into a phosphorescence lifetime decay-based glucose biosensor and (2) represent the reaction of glucose and oxygen. Subsequently, this model also demonstrated the quenching of a phosphorescent dye, such as Pd-meso-tetra (sulfophenyl) tetrabenzoporphyrin (HULK, Frontier Scientific), by calculating the phosphorescence lifetime as a function of the Stern–Volmer relationship (ksv=0.023 and τ0=201  μs).31 The validation model was constructed to represent the testing of phosphorescence lifetime decay enzymatic biosensors within a “flow-through” system as described elsewhere and was performed with a single sample biosensor.32,33 Briefly, the biosensor itself consists of alginate microparticles contained within an alginate hydrogel. Within the microparticles are HULK, glucose oxidase (GOx, 146.2 mmol, Aspergillus niger, BBI Solutions, Cardiff, Wales, United Kingdom), and Catalase (0.61 mol, bovine liver, Calzyme Laboratories, Inc., San Luis Obispo, California, United States). Using layer-by-layer assembly, the diffusive properties of this gel can be controlled such that, when in vivo, the diffusive rate of glucose and oxygen into the sensor is within a known range.34 As oxygen enters the biosensor, the oxygen dynamically quenches the biosensor; however, the presence of glucose causes the consumption of oxygen via enzymatically driven glucose oxidation, as described in the following equations:

Eq. (1)

E+GX1E+gluconate,

Eq. (2)

E+O2X2E+H2O2,
where E is oxidized GOx, G is glucose, X1 is enzyme-substrate complex 1, E is reduced GOx, and X2 is enzyme-substrate complex 2.35 The reaction rate for the reversible formation of enzyme-substrate complex 1 is k1 and k1, the reaction rate for the formation of gluconate and reduced GOx is k2, the reaction rate for the reversible formation of enzyme-substrate complex 2 is k3 and k3, and the reaction rate for the formation of hydrogen peroxide is k4.

Thus, the phosphorescence lifetime of the biosensor may be indirectly correlated to glucose. The flow-through system is a bench top evaluation system wherein sensors are placed in Delrin channels with glass windows, and a glucose/phosphate-buffered saline solution is pumped through at rates and concentrations controlled by LabVIEW (National Instruments, Austin, Texas, United States). A version of this system used for lactate sensing is described and illustrated in Ref. 32. Directly above the sensor is a “reader head,” which excites with a pulsed LED and collects phosphorescence emission from the biosensor at a high sampling frequency, enabling the determination of phosphorescence lifetime by fitting the exponential decay.

The COMSOL representation of this testing system is limited to the flow cell chamber wherein the biosensor rests, as shown in Fig. 3. The geometry of the vertical cross section is a 7.6  mm×2.8  mm (length × height) rectangle with a 4  mm×0.5  mm (length × height). The biosensor is centered on the bottom surface with a geometrical fillet to replicate the cylindrical representation of the biosensor. The glucose solution enters from the left boundary with a velocity of 0.009  m/s and exits through the right boundary via the laminar flow COMSOL physics module. The chemical transport and reaction properties of this biosensor are viewable in Table 2, and the reaction rates are described as35

Eq. (3)

Gt=DG2Gz2k1GE+k1X1,

Eq. (4)

O2t=DO2O2z2k3O2E+k3X2,

Eq. (5)

Et=k1GE+k1X1+k4X2,

Eq. (6)

Et=k2X1k3O2E+k3X2,

Eq. (7)

X1t=k1GE(k1+k2)X1,

Eq. (8)

X2t=k3O2E(k3+k4)X2.

Fig. 3

Meshed COMSOL cross-section model of flow-through system (cm).

JBO_27_11_117001_f003.png

Table 2

Biosensor diffusion and reaction properties.

ParameterValueUnitDescriptionReferences
k1100m3/(s*mol)Reaction rate35
k10.003s1Reaction rate35
k2300s1Reaction rate35
k31000m3/(s*mol)Reaction rate35
k3150s1Reaction rate35
k450s1Reaction rate35
DG_S2.64e10m2/sDiffusion of glucose in subcutaneous space36
DO2_S1.5e9m2/sDiffusion of oxygen in subcutaneous space37
ICG2.78mol/m3Initial concentration of glucose38
ICO20.009mol/m3Initial concentration of oxygen39
RG0.027mol/(m3*s)Reaction of glucose consumption by cells40 and 41
RO21.04e5mol/(m3*s)Reaction of oxygen consumption by cells42
Et0.1463mol/m3Enzyme concentrationn/a
Pw(t)Figure 5mol/m3Concentration of glucose from capillaries at time t38 and 43
CO20.009mol/m3Constant concentration of oxygen from capillaries39
DG_Assay14.0e10m2/sDiffusion of glucose in hydrogel35
DO2_Assay12.45e9m2/sDiffusion of oxygen in hydrogel35

To maximize computational efficiency and obtain a large average element quality, the biosensor was modeled with a triangular mesh with element sizes between 0.004 and 0.037 mm. The remainder of the domain—the “flow-cell”—is a triangular mesh with element sizes between 0.001 and 0.282 mm. The final average element quality or skewness is defined as a scalar measure of equiangular skew, where a value of 0 represents poor element size regularity that will prevent model convergence and a value of 1 is ideal. The skewness value was found to be sufficiently high at 0.9531.

To validate the ability of this model to represent (1) diffusion/reaction of species and (2) biosensor quenching and phosphorescence lifetime, simulation and real experiments modeling exposure to different glucose concentrations were performed. Specifically, five different concentrations ranging from 0 to 11.10  mol/m3 (0, 2.78, 5.55, 8.34, and 11.1  mol/m3) were used as inputs as they cover the physiological range of glucose concentration for individuals with diabetes.44 These concentrations with four biosensors were completed and replicated in silico. To represent biosensor quenching and phosphorescence lifetime, the lifetime output of the model was compared with the experimental lifetime in three biosensors at each glucose concentration. To compare the diffusion/reaction of the species, the time to steady-state (TSS) lifetime of the experiment and computational counterpart, defined as the time between 5% greater than the initial lifetime and 5% of the next steady-state lifetime, were also compared.

2.2.2.

Simulated biosensor performance when inserted

A second COMSOL model, representing the interstitial space 2  mm below the surface of the volar forearm, was developed with information sourced from the literature. This model geometry is shown in Fig. 4. The model is 1 mm in height, and the length is determined by the size of the biosensor being simulated, which was variable in this case and ranges from 4.53 to 15.42 mm. This was done to simulate biosensors with variable “spacers” from 0.01 to 0.1 cm while maintaining the domain thickness of 0.036 cm to evaluate the impact of the glucose diffusion profile caused by the enzymatic sensing domains on the nonenzymatic domains. The sensing domains have triangular mesh elements ranging from 0.0001 to 0.05 mm in size, and the remainder of the geometry has triangular mesh elements ranging in size from 0.006 to 0.184 mm. Overall, this yields an average element quality from 0.94 to 0.98. In this simulation, glucose and oxygen are sourced from the top and bottom boundaries—representing capillaries—that are 0.05 mm from the biosensor in each direction, as determined by the capillary density in the subcutis. Glucose has a variable concentration within the physiological range, and oxygen is supplied at 0.009  mol/m3. In addition to the glucose and oxygen consumption that occurs by the biosensor, the background tissue is consuming these species in the form of a 0th order reaction with rates 0.027 and 1.04e5  mol/(m3*s), as shown in Refs. 4041.42. To evaluate the impact of the glucose diffusion profile caused by the enzymatic biosensor on the nonenzymatic domains, the concentration of glucose was evaluated within the nonenzymatic domains for all biosensors with different spacer sizes.

Fig. 4

(a) COMSOL multiphysics model of interstitial space and biosensor and (b) capillary glucose concentration as a function of time “pw(t).”

JBO_27_11_117001_f004.png

3.

Results and Discussion

3.1.

MC-Based Biosensor Geometry Optimization

Key parameters in choosing the best performing biosensor geometry are overall luminescent intensity and sensitivity of the intensity as a function of relative changes in position between the excitation source and the biosensor in all three dimensions. Figure 5(a) shows the luminescent output of both assays for the three presupposed designs in arbitrary units. On average, design 2 yields the greatest luminescent output, being 1.97 times the luminescent output of design 1 and 3.22 times the luminescent output of design 3. This is possible due to the cylindrical shape of design 2, which enables it to have a greater volume and still meet the size requirement of insertion with a hypodermic needle. Figures 5(b)5(d) indicate the relative luminescent intensity of each biosensor design when translated 10 mm in the X direction (5 to 5 mm), 10 mm in the Y direction (5 to 5 mm), and 1.6 mm in the Z direction (1.2 to 2.8 mm). In the Y and Z direction, each biosensor design performs similarly. All designs experience an 70% reduction in luminescent intensity as they are shifted 3 mm in the Y direction and a 95% reduction in luminescent intensity as they are shifted 5 mm in the Y direction. The required rotational symmetry of the biosensors is one reason for the symmetric drop in luminescent intensity being observed in the positive and negative Y coordinates. It was anticipated that designs 1 and 3 would be more robust than design 2 in this sensitivity analysis as both have surfaces that are perpendicular to the excitation light; however, this was not found to be the case. In the Z direction, the intensity decreases 70% as implantation depth increases from 1.2 to 2 mm and 90% as implantation depth increases from 1.2 to 2.8 mm. As expected, all designs perform similarly. In the X direction, designs 1 and 2 have an assay that retains 90% of luminescent intensity between 3 and 3 mm, whereas design 3 has a relative luminescent intensity of 75% at 3 and 3 mm. In addition, at a maximum displacement of 5 and 5 mm, designs 1 and 2 have an assay that retains 40% of their maximum luminescent output. This is because designs 1 and 2 have repeating units along the long axis of the biosensor, whereas design 3 has longer and fewer repeating units that extend over the length of the biosensor. This allows designs 1 and 2 to maintain strong luminescent output of at least one assay over 6 mm of displacement. This is a favorable outcome compared with design 3 because it is not likely that the biosensor and wearable will be consistently colocated when movement of the wearable and biosensor occur. Overall, it was found that designs 1 and 2 have comparably similar performances in the sensitivity analysis; however, the overall luminescence in design 2 is the greatest, and subsequently design 2 is the strongest option. However, it can still be further optimized; specifically, the overall length of the biosensor and the length of individual repeating units can be optimized to maximize luminescent output.

Fig. 5

(a) Luminescent intensity of three presupposed designs, (b) sensitivity of relative luminescent intensity of three designs in the X axis, (c) sensitivity of relative luminescent intensity of three designs in the Y axis, and (d) sensitivity of relative luminescent intensity of three designs in the Z axis. Error bars are 1 standard deviation. Vertical line at X=0.16  cm in panel (d) indicates the position (depth) of the dermal/subcutaneous junction. Error bars are 1 standard deviation.

JBO_27_11_117001_f005.png

The ideal overall length of the biosensor is the FWHM of the excitation beam at the target insertion depth of 2 mm. This ensures that the biosensor is long enough to make full use of the LEDs excitation despite divergence, but not too long such that some of the biosensor remains unexcited. To determine this ideal length, MC simulations were carried out with the same geometry and properties as the previous component of this work, which analyzed the profile of excitation light within the geometry. At a depth of 2 mm, it was found that the FWHM in the X and Y directions was 4.3 mm, as shown in Fig. 6(a). Thus, for this simulated light source, the ideal biosensor length should be 4.3 mm. To determine the ideal length of one repeating unit, and consequently the total number of repeating units, 64 MC simulations with a biosensor placed with 5 and 5 mm offsets in the X and Y directions for each of the two assays were completed. The biosensor geometry was changed such that the number of repeating units in the biosensor was varied from 3.25 to 0.21 mm. The results of these simulations, as shown in Figs. 6(b) and 6(c), illustrate that, if the biosensor is only subdivided into one of each sensing compartment and the biosensor is not directly under the excitation light, one of the two assays will not strongly luminesce, depending on the positioning of the biosensor. As the unit thickness of a repeating unit is decreased and more total repeating units are included, the luminescence outputs of both assays converge. It was then found that using 0.36-mm-thick repeating units allows enough domains of each assay to exist such that luminescence is largely consistent across sensing chemistries. This was determined because there is a <5% change in luminescent intensity as the repeating unit thickness decreases. This leads to a final compartment thickness of 0.36 mm and 12 total sensing chemistry domains within a 4.3-mm biosensor.

Fig. 6

(a) FWHM of an LED in tissue, (b) luminescent intensity versus sensing assay length for displacement in the Y axis, (c) luminescent intensity versus sensing assay length for displacement in the X axis. The blue circles on panels (b) and (c) indicate the optimal size.

JBO_27_11_117001_f006.png

3.2.

COMSOL Multiphysics Biosensor Analysis

To first predict biosensor performance in vivo with a computational model, a valid model must be established. Here, the model was validated by matching the simulated phosphorescence lifetime of a biosensor within the phosphorescence lifetime in an experimental flow cell, confirming the capability to simulate phosphorescence lifetime as well as diffusion and reaction of oxygen and glucose. Figure 7(a) illustrates the comparison of lifetimes measured in computational and experimental work from 0  mol/m3 of glucose to 11.1  mol/m3 of glucose. Overall, the average percent error of the simulated lifetime and experimentally determined lifetime is 2.5%, ranging from 0.70% when the glucose concentration was 0  mg/dL to 6.28% when glucose concentration was 200  mg/dL. This is represented by the black bar in Fig. 7(a). With the strong agreement between the experimental phosphorescence lifetimes and their in silico counterparts, the representation of glucose oxidation in this computational model is validated. However, an important consideration is the speed at which glucose and oxygen diffuse within the biosensor. Thus, further work needs to be done to validate the ability of this model to represent the diffusive properties of an in vivo insertable biosensor.

Fig. 7

Validation of (a) species reaction and (b) species diffusion. Error bars are 1 standard deviation. Created with Ref. 30.

JBO_27_11_117001_f007.png

Figure 7(b) illustrates the comparison of the transition time that it took for the experimental setup to change lifetimes as glucose concentration changed and the time that it took for the simulations to perform the same change. In vitro, four biosensors had a TSS of 21.41 min as the flowing glucose concentration was changed in four even intervals from 0 to 11.1  mol/m3. In silico, the average TSS was found to be 21.08 min. This agreement is very strong; however, the in vitro data exhibit a negative trend in TSS as glucose concentration increases, which was not found in the in silico data. It is hypothesized that this negative trend is a result of a shell effect wherein the glucose and oxygen will have access to more phosphors once the outer layer of particles is quenched. However, given that this mechanism is a result of phosphorescence quenching properties as opposed to diffusion properties, the agreement in average TSS is taken to mean the model is able to recreate diffusion of glucose and oxygen within the biosensor. Thus, with a validated computational model, the diffusive gradient of the biosensor can be explored to determine if the enzymatic sensing domain negatively impacts the glucose concentration that reaches the nonenzymatic sensing domains.

Figure 8(a) depicts the average glucose concentration observed in each of the six nonenzymatic sensing domains for all biosensor versions with different spacer sizes. These data are enlarged in Fig. 8(b). On the right axis of Fig. 8(a) is data detailing the percent difference in glucose concentration across the 1- and 0.01-mm spacer. The percent difference ranges from 1% to 2% and thus is not likely to significantly affect the functionality of the nonenzymatic sensing domains in any case.

Fig. 8

(a) Effect of variable thickness spacer (spacer) size on glucose concentration. Vertical bars indicate the center of glucose concentration transition. (b) Magnified view of the inset from panel (a). Created with Ref. 30.

JBO_27_11_117001_f008.png

4.

Conclusion

This work is an example of using multiple advanced modeling methods to design medical devices. Specifically, an MC model of photon propagation and a COMSOL multiphysics model of diffusion–reaction were developed and used in tandem to design a geometry for a fully insertable and multimodal optical biosensor for glucose monitoring. This is the first case, to the authors’ knowledge, of specifically utilizing multiple computation models to design an insertable chemooptical device. The MC simulations were used effectively to evaluate and optimize different geometric designs, yielding a barcode with an ideal total length of 4.3 mm with 0.36-mm-thick repeating units. Multiphysics diffusion–reaction models were also able to reliably predict sensor behavior and were used to further conclude that there is no anticipated chemical crosstalk between reaction regions of the barcoded biosensor. Due to the complexity of the computational modeling within this work, experimental validation and verification should be pursued both to validate the findings presented herein and to better understand how factors such as optical property variability affects these results. Whereas this work features the exploration of a geometry for a fully insertable biosensor, the design of the associated wearable should also be explored. Specifically, the layout of LEDs and photodiodes will affect the amount of luminescent output that is received by the wearable device.

Disclosures

The authors have no relevant financial interests in this article and no potential conflicts of interest to disclose.

Acknowledgements

This work was supported by the National Science Foundation Engineering Research Center for Precise Advanced Technologies and Health Systems for Underserved Populations (PATHS-UP) (Grant No. 1648451) and the National Institutes of Health (R01EB024601, MJM). The authors acknowledge assistance and advice from T.J. Falohun and Brian Ko.

Code and Data Availability

The code utilized and data generated for the completion of this work can be supplied by the authors upon reasonable request.

References

1. 

Jr. L. C. Clark and C. Lyons, “Electrode systems for continuous monitoring in cardiovascular surgery,” Ann. N.Y. Acad. Sci., 102 (1), 29 –45 https://doi.org/10.1111/j.1749-6632.1962.tb13623.x ANYAA9 0077-8923 (1962). Google Scholar

2. 

V. Zucolotto, “Specialty grand challenges in biosensors,” Front. Sens., 3 (2020). Google Scholar

3. 

R. Vashistha et al., “Futuristic biosensors for cardiac health care: an artificial intelligence approach,” 3 Biotech, 8 (8), 358 https://doi.org/10.1007/s13205-018-1368-y (2018). Google Scholar

4. 

E. V. Karpova et al., “Noninvasive diabetes monitoring through continuous analysis of sweat using flow-through glucose biosensor,” Anal. Chem., 91 (6), 3778 –3783 https://doi.org/10.1021/acs.analchem.8b05928 ANCHAM 0003-2700 (2019). Google Scholar

5. 

D. Rodrigues et al., “Skin-Integrated wearable systems and implantable biosensors: a comprehensive review,” Biosensors, 10 (7), 79 https://doi.org/10.3390/bios10070079 BISSED 0265-928X (2020). Google Scholar

6. 

R. Rebelo et al., “An outlook on implantable biosensors for personalized medicine,” Engineering, 7 (12), 1696 –1699 https://doi.org/10.1016/j.eng.2021.08.010 ENGNA2 0013-7782 (2021). Google Scholar

7. 

K. Scholten and E. Meng, “A review of implantable biosensors for closed-loop glucose control and other drug delivery applications,” Int. J. Pharm., 544 (2), 319 –334 https://doi.org/10.1016/j.ijpharm.2018.02.022 IJPHDE 0378-5173 (2018). Google Scholar

8. 

W. K. Chung et al., “Precision medicine in diabetes: a consensus report from the American Diabetes Association (ADA) and the European Association for the Study of Diabetes (EASD),” Diabetes Care, 43 (7), 1617 –1635 https://doi.org/10.2337/dci20-0022 DICAD2 0149-5992 (2020). Google Scholar

9. 

T. Bobrowski and W. Schuhmann, “Long-term implantable glucose biosensors,” Curr. Opin. Electrochem., 10 112 –119 https://doi.org/10.1016/j.coelec.2018.05.004 (2018). Google Scholar

10. 

G. Cappon et al., “Wearable continuous glucose monitoring sensors: a revolution in diabetes treatment,” Electronics, 6 (3), 65 https://doi.org/10.3390/electronics6030065 ELECAD 0013-5070 (2017). Google Scholar

11. 

C. f. D. Control, and Prevention, National Diabetes Statistics Report, 2020, Centers for Disease Control and Prevention, US Department of Health and Human Services, Atlanta, GA (2020). Google Scholar

12. 

A. Das et al., “Predicting the macronutrient composition of mixed meals from dietary biomarkers in blood,” IEEE J. Biomed. Health. Inf., 26 2726 –2736 https://doi.org/10.1109/JBHI.2021.3134193 (2021). Google Scholar

13. 

M. J. Tierney et al., “The GlucoWatch® biographer: a frequent, automatic and noninvasive glucose monitor,” Ann. Med., 32 (9), 632 –641 https://doi.org/10.3109/07853890009002034 ANMDEU 0785-3890 (2000). Google Scholar

14. 

H. Thabit et al., “Accuracy of continuous glucose monitoring during three closed-loop home studies under free-living conditions,” Diabetes Technol. Ther., 17 (11), 801 –807 https://doi.org/10.1089/dia.2015.0062 (2015). Google Scholar

15. 

I. L. Jernelv et al., “A review of optical methods for continuous glucose monitoring,” Appl. Spectrosc. Rev., 54 (7), 543 –572 https://doi.org/10.1080/05704928.2018.1486324 APSRBB 0570-4928 (2019). Google Scholar

16. 

B. M. Cummins, “A rationally-designed fluorescence competitive binding assay for continuous glucose monitoring applications,” (2014). Google Scholar

17. 

J. Fine, M. J. McShane and G. L. Coté, “Monte Carlo method for assessment of a multimodal insertable biosensor,” J. Biomed. Opt., 27 (8), 083017 https://doi.org/10.1117/1.JBO.27.8.083017 JBOPFO 1083-3668 (2022). Google Scholar

18. 

T. Pan et al., “Biophotonic probes for bio-detection and imaging,” Light Sci. Appl., 10 (1), 124 https://doi.org/10.1038/s41377-021-00561-2 (2021). Google Scholar

19. 

S. L. Jacques et al., “Developing an optical fiber reflectance spectrometer to monitor bilirubinemia in neonates,” Proc. SPIE, 2975 115 –124 https://doi.org/10.1117/12.275465 PSISDG 0277-786X (1997). Google Scholar

20. 

V. V. Dremin et al., “Multimodal optical measurement for study of lower limb tissue viability in patients with diabetes mellitus,” J. Biomed. Opt., 22 (8), 085003 https://doi.org/10.1117/1.JBO.22.8.085003 JBOPFO 1083-3668 (2017). Google Scholar

21. 

D. Marti et al., “MCmatlab: an open-source user-friendly MATLAB-integrated 3D Monte Carlo light transport solver with heat diffusion and tissue damage,” Proc. SPIE, 70876 108760T https://doi.org/10.1117/12.2507754 PSISDG 0277-786X (2019). Google Scholar

22. 

E. M. Dede, T. Nomura and J. Lee, Multiphysics Simulation, Springer( (2014). Google Scholar

23. 

K. Goh, H. Li and K. Y. Lam, “Development of a multiphysics model to characterize the responsive behavior of urea-sensitive hydrogel as biosensor,” Biosens. Bioelectron., 91 673 –679 https://doi.org/10.1016/j.bios.2017.01.023 BBIOE4 0956-5663 (2017). Google Scholar

24. 

A. E. Davies et al., “In vitro and computational modelling of drug delivery across the outer blood–retinal barrier,” Interface Focus, 10 (2), 20190132 https://doi.org/10.1098/rsfs.2019.0132 (2020). Google Scholar

25. 

J. Fine, M. McShane and G. Coté, “Design selection of a fully-implantable and optically-transduced glucose biosensor via Monte Carlo Modeling,” Proc. SPIE, 11651 116510U https://doi.org/10.1117/12.2577737 PSISDG 0277-786X (2021). Google Scholar

26. 

S. L. Jacques, “Optical properties of biological tissues: a review,” Phys. Med. Biol., 58 (11), R37 https://doi.org/10.1088/0031-9155/58/11/R37 (2013). Google Scholar

27. 

X. Chen et al., “In vivo real-time imaging of cutaneous hemoglobin concentration, oxygen saturation, scattering properties, melanin content, and epidermal thickness with visible spatially modulated light,” Biomed. Opt. Express, 8 (12), 5468 –5482 https://doi.org/10.1364/BOE.8.005468 BOEICL 2156-7085 (2017). Google Scholar

28. 

J. Sandby-Moller, T. Poulsen and H. C. Wulf, “Epidermal thickness at different body sites: relationship to age, gender, pigmentation, blood content, skin type and smoking habits,” Acta Derm. Venereol., 83 (6), 410 –413 https://doi.org/10.1080/00015550310015419 (2003). Google Scholar

29. 

T. Van Mulder et al., “High frequency ultrasound to assess skin thickness in healthy adults,” Vaccine, 35 (14), 1810 –1815 https://doi.org/10.1016/j.vaccine.2016.07.039 VACCDE 0264-410X (2017). Google Scholar

31. 

J. R. Lakowicz, Principles of Fluorescence Spectroscopy, Springer( (2006). Google Scholar

32. 

L. P. Andrus et al., “Characterization of lactate sensors based on lactate oxidase and palladium benzoporphyrin immobilized in hydrogels,” Biosensors, 5 (3), 398 –416 https://doi.org/10.3390/bios5030398 BISSED 0265-928X (2015). Google Scholar

33. 

A. Biswas et al., “Cross-linked nanofilms for tunable permeability control in a composite microdomain system,” RSC Adv., 6 (75), 71781 –71790 https://doi.org/10.1039/C6RA13507B (2016). Google Scholar

34. 

J. Park and M. J. McShane, “Dual-function nanofilm coatings with diffusion control and protein resistance,” ACS Appl. Mater. Interfaces, 2 (4), 991 –997 https://doi.org/10.1021/am900673r AAMICK 1944-8244 (2010). Google Scholar

35. 

J. Q. Brown, Modeling, Design, and Validation of fluoRescent Spherical Enzymatic Glucose Microsensors Using Nanoengineered Polyelectrolyte Coatings, Louisiana Tech University( (2005). Google Scholar

36. 

O. G. Jepps et al., “Modeling the human skin barrier—towards a better understanding of dermal absorption,” Adv. Drug Delivery Rev., 65 (2), 152 –168 https://doi.org/10.1016/j.addr.2012.04.003 ADDREP 0169-409X (2013). Google Scholar

37. 

U. Grossmann, “Simulation of combined transfer of oxygen and heat through the skin using a capillary-loop model,” Math. Biosci., 61 (2), 205 –236 https://doi.org/10.1016/0025-5564(82)90004-9 MABIAR 0025-5564 (1982). Google Scholar

38. 

F. Uettwiller et al., “Real-time continuous glucose monitoring reduces the duration of hypoglycemia episodes: a randomized trial in very low birth weight neonates,” PLoS One, 10 (1), e0116255 https://doi.org/10.1371/journal.pone.0116255 POLNCL 1932-6203 (2015). Google Scholar

39. 

S. C. Kanick et al., “Continuous monitoring of interstitial tissue oxygen using subcutaneous oxygen microsensors: in vivo characterization in healthy volunteers,” Microvasc. Res., 124 6 –18 https://doi.org/10.1016/j.mvr.2019.02.002 MIVRA6 0026-2862 (2019). Google Scholar

40. 

N. Longo et al., “Increased glucose transport by human fibroblasts with a heritable defect in insulin binding,” Metabolism, 38 (7), 690 –697 https://doi.org/10.1016/0026-0495(89)90109-1 METAAJ 0026-0495 (1989). Google Scholar

41. 

N. Longo et al., “Glucose transport by cultured human fibroblasts: regulation by phorbol esters and insulin,” Biochim. Biophys. Acta (BBA)-Biomembr., 1104 (1), 24 –30 https://doi.org/10.1016/0005-2736(92)90127-8 (1992). Google Scholar

42. 

D. von Heimburg et al., “Oxygen consumption in undifferentiated versus differentiated adipogenic mesenchymal precursor cells,” Respir. Physiol. Neurobiol., 146 (2-3), 107 –116 https://doi.org/10.1016/j.resp.2004.12.013 (2005). Google Scholar

43. 

A. J. White-Cotsmire and A. M. Healy, “Ketogenic diet as a trigger for diabetic ketoacidosis in a misdiagnosis of diabetes: a case report,” Clin. Diabetes, 38 (3), 318 –321 https://doi.org/10.2337/cd20-0001 CLDIE8 0891-8929 (2020). Google Scholar

44. 

American Diabetes Association, “Standards of medical care in diabetes-2006,” Diabetes Care, 29 (1), S4 DICAD2 0149-5992 (2006). Google Scholar

Biography

Jesse Fine received his bachelor’s degree in biomedical engineering from Ohio State University in 2018 and joined the Biomedical Engineering Department at Texas A&M University as a fellow within PATHS-UP to use computational modeling as a tool to design wearable and implantable optical biosensors toward continuous monitoring. Currently, he is a PhD candidate in biomedical engineering at Texas A&M University.

Gerard L. Coté, PhD, is currently a regents professor at Texas A&M University. He holds the James J. Cain Professorship I in biomedical engineering at Texas A&M University. His research interest includes biomedical sensing for diagnostic and monitoring applications with a recent focus on the development of wearable and handheld medical devices for underserved populations. He is a fellow of four societies (IEEE, SPIE, BMES, and AIMBE) and an entrepreneur who founded several startup companies. Also, he is an elected fellow of the National Academy of Inventors.

Michael J. McShane, PhD, is currently a professor and department head of biomedical engineering at Texas A&M University. He holds the James J. Cain Professorship II in biomedical engineering. He has 28 years of experience in optical sensor technology development, including new concepts and applications of micro/nanotechnology for implantable optical sensors, related materials, and instrumentation. He has published more than 300 journal articles, proceedings, and abstracts on these topics. He is a senior member of SPIE and IEEE as well as a fellow of AIMBE.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Jesse N. Fine, Gerard L. Coté, and Michael J. McShane "Geometry design for a fully insertable glucose biosensor with multimodal optical readout," Journal of Biomedical Optics 27(11), 117001 (18 November 2022). https://doi.org/10.1117/1.JBO.27.11.117001
Received: 8 June 2022; Accepted: 19 October 2022; Published: 18 November 2022