14 December 2022 Spatial frequency domain Mueller matrix imaging
Author Affiliations +
Abstract

Significance

Mueller matrix polarimetry (MMP) and spatial frequency domain imaging (SFDI) are wide-field optical imaging modalities that differentiate tissue primarily by structure alignment and photon transport coefficient, respectively. Because these effects can be related, combining MMP and SFDI may enhance tissue differentiation beyond the capability of each modality alone.

Aim

An instrument was developed to combine MMP and SFDI with the goal of testing whether it enhances contrast of features in reflection mode.

Approach

The instrument was constructed using liquid crystal elements for polarization control, a digital light processing projector for generating sinusoidal illumination patterns, and a digital camera for imaging. A theoretical analysis shows that the SFD Mueller matrix is complex-valued and does not follow the same behavior as a regular Mueller matrix. Images were acquired from an anisotropic tissue phantom, an optical fiber bundle, and cerebellum, thalamus, and cerebrum tissues.

Results

The measurement results suggest that singly scattered, few scattered, and diffusely scattered photon paths can be distinguished in some of the samples investigated. The combined imaging modality yields additional spatial frequency phase information, which highlights paths having only a few scattering events.

Conclusions

The combination of MMP and SFDI offers contrast mechanisms inaccessible by each modality used alone.

1.

Introduction

The Mueller matrix characterizes the polarization-sensitive reflection or transmission properties of a material for a given optical path.1,2 From the Mueller matrix, specific polarimetric properties, such as retardance, diattenuation, and depolarization, can be determined. Highly scattering materials, such as biological tissues and optical phantoms intended to mimic them, generally exhibit a strong depolarizing response.3 Retardance and diattenuation can be observed in anisotropic media, such as connective tissue, muscle, and brain tissue.49 Mueller matrix measurements in transmission can be interpreted in terms of a differential formalism yielding birefringence or diattenuation.6,1012 Coherent Stokes vector and Mueller matrix methods in transmission have shown differentiation between structures in different biological tissues.1318 Changes in retardance and diattenuation can reflect disruptions in connective tissue, and Mueller matrix polarimetry (MMP) has been shown to be useful for detecting changes in collagen alignment, such as in cervixes during pregnancy or cancer.4,5,19 The MMP is commonly conducted using full field illumination, allowing it to characterize a large sample region. The measured Mueller matrix results from all the paths the light takes, including those that have penetrated deeply into the sample. Biological tissues, however, are rarely uniform in their optical properties throughout a volume,20 and discrimination between contributions from surface and deep tissues can be important. Organization of structures is important in distinguishing healthy and diseased tissue, and MMP is particularly sensitive to changes in structural organization.19 One issue with MMP is that deeply penetrating photons can impede detection of those photons that have only interacted with shallow tissues. Confocal MMP imaging has been one technique that has been attempted to circumvent this issue.21,22

Spatial frequency domain imaging (SFDI) has the potential for controlling the penetration depth of investigating photons.23 In SFDI, an incoherent sinusoidal spatially periodic pattern is projected onto the material, and the amplitude of the reflected modulation is recorded as the phase of that pattern is varied. In this manner, only photons whose path lengths are shorter than the period of the pattern are sensed, making depth sensitivity possible. A high spatial frequency can mitigate the amount of noise in an image by reducing the multiply scattered photons detected in favor of singly scattered photons and providing better image contrast.23,24 Like MMP, SFDI provides full field imaging, allowing for a macroscopic sample area to be investigated. Measurements of the polarization of light scattered a distance from a point of illumination have been performed in the past, but the method is less amenable to full field imaging.25 Because SFDI is typically operated in reflection mode, it can be employed noninvasively.

In this paper, we describe the combination of SFDI and MMP imaging and present some example measurements. Whereas polarized SFDI has been studied in the past,24,26 we expand on this idea by measuring full Mueller matrices in the SFD. In terms of the instrumentation, combining SFDI with MMP is relatively straightforward, as they share a number of components, and each modality only needs the addition of those specific to the other. Both modalities are also inexpensive compared with some other imaging modalities, such as confocal imaging or optical coherence tomography. By blending these methods, we combine the structural sensitivity of MMP with the photon path length selectivity of SFDI.

This work expands on two previous studies,27,28 which first explored this combination of modalities, by improving upon the data acquisition and analysis and expanding the range of materials studied. For example, a number of details need to be addressed to ensure that unwanted artifacts do not appear. In addition, it is found that a SFD Mueller matrix differs from a regular Mueller matrix, in that it does not follow the same rules and is complex-valued.

In Sec. 2, we develop the theory associated with the combined SFDI and MMP modalities. In Sec. 3, we describe the instrumentation for our measurements. Our data reduction methods are then outlined in Sec. 4. The samples we use to demonstrate the measurements are described in Sec. 5. In Sec. 6, we show the results of the measurements. We discuss the combined method in Sec. 7 and make some conclusions in Sec. 8.

2.

Theory

The SFDI is often interpreted using an approach based upon the diffusion equation.23 However, this approach does not account for singly scattered radiation and is not valid when the spatial frequency is on order of or greater than the transport coefficient. Polarization imaging often is particularly sensitive to the contributions of singly scattered radiation, and as a result, we find that a different approach is needed. In the theory of spectrophotometry,29,30 the bidirectional scattering-surface reflectance distribution function (BSSRDF) S expresses the relationship between the radiance dLr reflected from a surface at a location x,y and propagating in a direction θ^, when radiant power dΦi=Eidxdy (Ei is the incident irradiance) is incident from a direction θ^ onto the surface at a location x,y. When considering polarization, the radiation properties are naturally represented as 4-element Stokes vectors, and the BSSRDF as a 4×4 Mueller matrix. Thus, we make the transformation to Stokes vector quantities, dLrdLr, dΦidΦi, and EiEi, the Mueller matrix SS, and express this relationship as

Eq. (1)

dLr=SdΦi=SEidxdy.

In Eq. (1), we abbreviate dLr=dLr(x,y), dΦi=dΦi(x,y), Ei=Ei(x,y), and S=S(x,y;x,y). We also drop the directional coordinates θ^ and θ^, as well as the wavelength dependency, as they are generally fixed in our measurements. Whereas the phenomenological BSSRDF does not describe the physics of the scattering process, it successfully describes the radiation redistribution caused by the material under conditions of incoherent illumination. It is also straightforward to model using Monte Carlo methods.3133 It should be borne in mind that regular Stokes vectors obey the inequality, for example

Eq. (2)

Lr,1(Lr,22+Lr,32+Lr,42)1/2,
where Lr,j is the j’th element of Lr. Stokes vectors that obey Eq. (2) are either fully polarized (in the case of equality) or the sum of two (in the case of inequality) fully polarized Stokes vectors. The Mueller matrix S obeys the requirement that it must map the space of valid Stokes vector radiant powers onto the space of valid Stokes vector radiances. In addition, S must satisfy the Cloude requirement that it be the convex sum of up to four nondepolarizing Jones–Mueller matrices.34,35

In SFD Mueller matrix imaging, we illuminate the surface with an incoherent Stokes vector irradiance

Eq. (3)

Ei(x,y)=E0[1+cos(2πfx+ϕ)],
where f is the spatial frequency, ϕ is a phase, and E0 is a Stokes vector irradiance. Note that we could generalize f to include the direction along which illumination is oriented, and in some of our measurements, we modulated the irradiance along y. The radiance emitted by the surface is determined by integrating Eq. (1)

Eq. (4)

Lr=SEidxdy.

If we insert the irradiance Eq. (3) into Eq. (4), and assume that the material is uniform [so that S(x,y;x,y)=S(xx,yy)], the radiance is given as

Eq. (5)

Lr=frE0+S1(xx)E0cos(2πfx+ϕ)dx,
where the line spread function is given as

Eq. (6)

S1(xx)=S(xx,yy)dy,
and fr is the Mueller matrix BRDF.

In SFDI, we vary the phase ϕ of the irradiance and measure the modulation of the radiance. To demodulate the radiance, we evaluate

Eq. (7)

L˜r(f)=1π02πLrexp(2πifx+iϕ)dϕ.
If we insert Eq. (5) into Eq. (7), and perform the integral, we obtain

Eq. (8)

L˜r(f)=S1(xx)E0exp[2πif(xx)]dx.
The SFD Stokes radiance L˜r(f) does not obey the same inequality Eq. (2) as the regular Stokes vector Lr, because it is not a convex sum of regular Stokes vectors. Furthermore, it can be complex-valued. Negative radiance can be interpreted as radiance emitted from locations in the troughs of the irradiance, whereas complex parameters result from the polarization of radiation emitted to one side being different from that emitted from the other. At f=0, the integral in Eq. (8) evaluates to

Eq. (9)

L˜r(0)=frE0.

We thus define the SFD BRDF

Eq. (10)

Fr(f)=S1(xx)exp[2πif(xx)]dx,
so that

Eq. (11)

L˜r(f)=Fr(f)E0.

We notice a couple interesting properties of the matrix Fr(f). Because the quantity is a Fourier transform of a regular Mueller matrix, it is not necessarily physically realizable as one. That is, Mueller matrices can only be safely added with positive coefficients. Matrices that are physically unrealizable have the property that they can overpolarize or yield negative power. If radiation is incident at one location and exits from another location, the SFD Stokes radiance and the SFD BRDF may not follow the customary rules that apply to regular Stokes vectors or Mueller matrices. We use the term SFD Mueller matrix, or SFD BRDF to represent the generalization of the Mueller matrix and emphasize that its properties at nonzero spatial frequency differ from those of regular Mueller matrices.

A regular Mueller matrix can be expressed34,35 as a covariance matrix H, whose ordered eigenvalues obey λ0λ1λ2λ30. The polarimetric purity indices,36

Eq. (12)

P1=λ0λ1iλi,P2=λ0+λ12λ2iλi,P3=λ0+λ1+λ23λ3iλi,
obey the inequality

Eq. (13)

0P1P2P31.
In the SFD, some of the eigenvalues λi can be negative or even complex-valued, and the polarimetric purity indices no longer obey the inequality in Eq. (13). In particular, P3 can be greater than unity.

If the BSSRDF is asymmetric, i.e., S1(xx)S1(xx), then the matrix Fr(f) will be complex-valued. Note that the absolute phase of the signal is not necessarily known in an experiment. For example, for nonnormal illumination and viewing, the surface topography and the depth of the scatterer can affect the overall phase of the image compared with the illumination. As a result, we correct the phase by requiring that the upper-left element (M11) be real-valued.

3.

Instrumentation

Figure 1(a) shows an illustration of the SFD Mueller matrix imaging system. Structured illumination is generated using a digital light processing (DLP) projector (Texas Instruments, model Lightcrafter 4500), which uses three light emitting diodes (LEDs) with wavelengths centered at 455, 520, and 630 nm as well as digital micromirror devices to pulse-width modulate the light source according to the image and settings sent to the device. All images in this study were taken using only the 630-nm LED, because using a single wavelength simplifies the operation of the instrument. This LED’s spectrum was measured to have a full width at half maximum (FWHM) of 16 nm centered at 628 nm. We used a 10-nm bandpass filter centered at 633 nm to narrow the spectral width. The spectrum measured after applying the filter had a FWHM of 6.5 nm centered at 632 nm. The illumination is 20  deg from the sample surface normal. The DLP has 1140 pixels in the vertical direction and 912 pixels in the horizontal direction arranged in a diamond pattern [see Fig. 1(b)] with the rows being separated by half the distance of the columns and offset by half a pixel horizontally. To ensure that a horizontal or vertical sinusoidal projected pattern is commensurate with the array, to have a common number of phases, and to have a smoothly modulating pattern, spatial frequencies were restricted to those having periods with multiples of three pixels and N=6 phases were used, twice that necessary,37 providing some oversampling and improving complex-valued demodulation. The periods , 24, 18, 12, 9, and 6 pixels on the projector correspond to f=(0,1.2,1.7,2.5,3.3,5.0)  cm1 at the sample plane. Because three pixels are not truly commensurate with the array, demodulation was often poor and results from that spatial frequency are not presented. The projected sinusoidal pattern can be displayed in either the horizontal or vertical direction, both of which are used in this manuscript. An example image can be seen in Fig. 1(c), where a vertical sinusoidal pattern is used to illuminate a section of bovine thalamus at f=5  cm1. The total area illuminated by the projector at the sample plane (>17  cm diameter) is much larger than the imaged area (1.1  cm×1.1  cm) and illuminates surrounding objects and the support structure. To avoid the effects of stray light, especially for f=0 illumination, the projected pattern was cropped in software to illuminate only the area close to the sample.

Fig. 1

(a) SFD Mueller matrix imaging system: PSG with polarizer (P1) and two LC retarders (LC1 and LC2), DLP projector, PSA with polarizer (P2) and two LC retarders (LC3 and LC4), objective (OBJ), and sCMOS camera (CAMERA). (b) A representation of how the sinusoid pattern for f=5  cm1 is displayed by the DLP projector pixels. (c) One of the 96 images acquired of bovine thalamus for vertical sinusoidal illumination at f=5  cm1.

JBO_27_12_126003_f001.png

The sample is imaged normal to the sample with a 50-mm focal length objective (Edmund Optics, model 59-873) onto a monochrome 16-bit 2048×2048 scientific complementary metal oxide semiconductor (sCMOS) camera with (6.5×6.5)  μm2 pixel area (PCO, model pco.panda 4.2). The system has a full field of view at the sample plane of 7.5  cm×7.5  cm, though all images presented here are cropped to 1.1  cm×1.1  cm (300  pixels×300  pixels). The camera is synchronized to the projector and its exposure time is set as a multiple number of milliseconds. Thus, to avoid interference of the camera exposure time with the pulse width modulation of the projector (frame rate 120  s1), the camera exposure time was chosen to be a multiple of 25 ms (3  ms×8.33  ms).

The polarization state generator (PSG) and the polarization state analyzer (PSA) each consist of a polarizer and a pair of nematic liquid crystal (LC) retarders (Meadowlark Optics). The axes of the LCs are aligned nominally at 27.4 deg and 72.4 deg with respect to the polarizer, which is the optimum configuration.38 Each of the four LCs were switched between two retardance states (determined by applied voltage) providing the 16 different combinations needed to measure a full Mueller matrix. The polarimeter was calibrated with the eigenvalue method described by Compain et al. using three samples:39 air (no sample, in transmittance), a Glan–Thompson polarizer (in transmittance), and a Si wafer with an 1000-nm-thick SiO2 layer acting as a mirror with retardance and diattenuation (in reflection with an incident angle of 60 deg). This eigenvalue method yields a calibration that is valid for both transmission and reflection measurement geometries. The 10-nm spectral bandwidth filter, mentioned earlier, was necessary to achieve an acceptable calibration due to the native bandwidth of the light source producing a reduction matrix with high dispersion. The standard deviation of all the reference measurements across all elements of the normalized Mueller matrix was 0.014. We believe we can use this standard deviation as the standard uncertainty in subsequent measurements. Control of the projector, the LCs, and the camera was performed using a program written in MATLAB (MathWorks) software.

The final array of 672 images (7 spatial frequencies × 6 phases × 16 PSG/PSA combinations) took 4.2 min to acquire, if only imaged with one spatial frequency orientation. This time was significantly improved from the previous iteration of the instrument, mostly due to the use of a faster camera and synchronization of it with the projector.27 The response and settling time of the LC retarders is the predominant factor limiting acquisition time.

Figure 1(c) shows one acquired image, having a specific phase and PSG/PSA combination, for one of the samples (bovine thalamus, described later) for vertical sinusoidal illumination at f=5  cm1. Most of the phase variation observed in Fig. 1(c) is a result of the topography and shape of the specimen. The lack of sharp optical contrast is a result of diffuse scatter in the medium.

4.

Data Reduction

In this work, we perform measurements of the radiance (an image) Lr,jk for a set of 16 combinations of PSG and analyzer configurations (denoted by index) for each of six illumination phases (denoted by index j). We begin by demodulating the signal by substituting the integral in Eq. (7) with a sum, ignoring the overall phase factor exp(2πifx)

Eq. (14)

L^r,k(f)=1Nj=1NLr,jk(f)exp(i2πjN),
where N is the number of phases. Note that this demodulation scheme preserves phase, unlike that typically used for SFDI, which only measures amplitude.23 We then apply a data reduction matrix Rk to the L^r,k(f) to yield the complex SFD Mueller matrix

Eq. (15)

M(f)=k=116RkL^r,k(f).

The data reduction matrix is determined from the eigenvalue calibration procedure.39 The SFD Mueller matrix carries with it the phase factor exp(2πifx), in addition to phase that results from topography and varying depth. We thus remove the overall phase by forcing M11 to be real-valued by applying the transformation

Eq. (16)

M(f)M(f)|M11(f)|M11(f).

Finally, the SFD Mueller matrix BRDF f˜r(f) can be related to the measured matrix27 M, as follows:

Eq. (17)

Fr(f)=M(f)×frdπM11d(0)×M11r(0)M11r(f),
where the first correction factor uses a measurement Md(0) of a diffuse reflector of known BRDF frd in the same incident and viewing geometry to provide a scale in inverse steradians, and the second correction factor uses a measurement of Mr of a nondiffusive surface, such as a rough metal, to account for the modulation transfer function of the instrument.

Further analysis of the SFD Mueller matrix is complicated by the fact that the matrix is not a convex sum of Jones–Mueller matrices and that it can be complex-valued. In those cases where the imaginary part of the matrix is negligible and the matrix is realizable, we can apply the Lu–Chipman decomposition,40 expressing the matrix as the ordered product of a diattenuator MD, a phase retarder MR, and a depolarizer MΔ

Eq. (18)

M=MΔMRMD.

From these individual elements, we can extract the linear diattenuation, diattenuation orientation, linear retardance, retardance orientation, and depolarization.1

The diattenuation D shown in this paper is on a scale from 0 to 1, where D=0 indicates that the material has no selectivity of the orientation of the polarization plane of linearly polarized light, whereas D=1 indicates one polarization state is entirely attenuated. Anisotropic materials often display some degree of diattenuation. In addition, the 20 deg incident angle geometry can impose some effective diattenuation.

5.

Samples

We performed measurements on five samples. The first sample was a phantom with an anisotropic scatterer embedded at a graded depth in an isotropic scattering medium. The phantom fabrication procedure is detailed elsewhere,41 but in brief, polydimethylsiloxane (PDMS) (Corning, model Sylgard 184 Elastomer) was used as a base to create the scattering and absorbing media. Titanium dioxide (Atlantic Equipment Engineers, TI-602, size 0.3 to 1.0  μm) and carbon black particles (Atlantic Equipment Engineers, FE-603, size 1 to 2  μm) were mixed within the PDMS to act as scatterers and absorbers, respectively. The scatterer and absorber were kept in stock suspensions of 1% mass ratio TiO2 powder in pure PDMS and 0.1% mass ratio carbon black powder in pure PDMS, respectively. The absorption coefficient (0.4  cm1) and the reduced scattering coefficient (7.5  cm1) were determined from diffuse reflectance and transmittance measurements, together with an inverse adding-doubling algorithm.42 Para-aramid fibers obtained from cladding of furcation cables (Thorlabs, FT030) were used as anisotropic scatterers. The fibers were fixed to a glass microscope slide with adhesive. The slide was then inclined in a Petri dish acting as a mold and filled with the PDMS solution. The para-aramid fibers were set so that one end of the fibers was exposed while the other end was submerged. In this way, the thickness of the scattering/absorbing PDMS media above the fibers gradually increases from 0 to 4 mm from one end of the slide to the other. Only a portion of the slide was imaged, so the results we present had fibers embedded less than about 1 mm. A glossy white ceramic tile was used as the background when this sample was imaged. The use of the tile was chosen to demonstrate the reduced impact of a highly depolarizing background.

Our second sample was an optic fiber bundle (image conduit). The fiber bundle had a diameter of 3.2 mm, contained 50,419 individual fibers having 12-μm cores, and was approximately 25 mm long (Edmund Optics, model 53-839). The individual fibers are expected to act as anisotropic scatterers and scatter primarily perpendicular to the fiber axes. The fibers may also exhibit some form birefringence or diattenuation. Furthermore, light entering the bundle may internally reflect and be reemitted from a different location. We imaged the fiber bundle resting on its side upon the glossy white ceramic tile.

The remainder of our samples were mammalian brain tissues: a caprine (goat) cerebellum, a bovine (cow) thalamus, and a bovine cerebrum. The brains were acquired intact from a local Halal abattoir. Sections of brain were then excised and placed in a Petri dish for imaging within an hour of acquiring the brains. The samples were all greater than 1.5 cm thick.

6.

Results

6.1.

Para-Aramid Fiber Embedded in PDMS

Figure 2 shows a subset of the SFD Mueller matrix images for the para-aramid fiber embedded in the PDMS phantom (full datasets are available online, see Data Availability below). Figure 2(a) shows the Mueller matrix BRDF at f=0  cm1, whereas Figs. 2(b) and 2(c) show the real and imaginary parts, respectively, of the SFD Mueller matrix BRDF illuminated with a horizontal sinusoidal pattern at f=5  cm1. All but the M11 terms are shown normalized to their respective M11, which are forced to be real-valued. As M11 is defined to be real, it is zero in the imaginary part of the SFD Mueller matrix [see Fig. 2(c)]. The thickness of the PDMS is very thin at the bottom of the images and increases to about 1 mm at the top of the image. The para-aramid fibers, which are embedded within the PDMS phantom, stretch vertically through the center of the image and are most visible at the bottom, where the PDMS is thinnest and some of the fibers are exposed at the edge. The M11 at f=5  cm1 has lower intensity than that at f=0  cm1, which is typical for SFDI and is due to the filtering of diffusing photons. The imaginary part of the f=5  cm1 SFD Mueller matrix is negligible for this sample, suggesting that the scattering is primarily symmetric. The real part of the f=5  cm1 SFD Mueller matrix has somewhat higher contrast than its f=0  cm1 counterpart. Most noticeably, the vertical extent to which the fibers can be observed in the SFD Mueller matrix is increased at the higher spatial frequency. The spatial frequency imaging filters the diffusing photons, which are more depolarizing, yielding the increased image contrast in the SFD Mueller matrix elements. Because the scatter from the fibers is more dependent upon polarization than the scattering from the PDMS, the single scattering from the fibers can be observed from deeper in the material.

Fig. 2

SFD Mueller matrix BRDFs from the para-aramid fiber embedded in a PDMS-based scattering and absorbing phantom: (a) the Mueller matrix BRDF at f=0  cm1, (b) the real part of the SFD Mueller matrix BRDF at f=5  cm1, and (c) the imaginary part of the SFD Mueller matrix BRDF at f=5  cm1. The maximum value of the black and white scale used for Fr,11 at f=0  cm1 is twice the maximum value for Fr,11 of f=5  cm1. The scale for the normalized SFD Mueller matrix elements is shown at the bottom of the figure. The spatial frequency sinusoidal pattern for f=5  cm1 was oriented horizontally. The field of view of the images is 1.1  cm×1.1  cm.

JBO_27_12_126003_f002.png

The SFD Mueller matrix at f=5  cm1 in Fig. 2(b) is mostly, but not entirely, a physically realizable regular Mueller matrix (as determined by evaluating the eigenvalues of the covariance matrix). As pointed out earlier in Sec. 2, nonrealizability occurs when there is significant radiance emitted from one location when irradiance is applied at another. Because most of the image contains realizable matrices, we can apply Lu-Chipman decomposition to them and determine local diattenuation. Figure 3 shows the local diattenuation as a function of spatial frequency. The background regions shows some diattenuation, which we attribute to the off-axis geometry of the measurement, whereas the fibers show additional diattenuation, which we attribute to their alignment. Included in Fig. 3 is a blue mask showing those areas of the image which are not realizable. As one can see, the area of the sample for which the PDMS layer is thinnest yields nonrealizable behavior. We suspect that this occurs progressively in this region because the PDMS lies above a microscope slide, which separates the PDMS scattering medium from the diffusely scattering white tile under the slide. Note that the matrix is realizable in the region with the fibers. We believe that the fibers are more local scatterers, so that the signal from them is less dependent upon spatial frequency than that of the more diffusely scattering PDMS material.

Fig. 3

Diattenuation of the para-aramid fiber embedded in a PDMS-based scattering and absorbing phantom calculated using Lu-Chipman decomposition from the real part of the SFD Mueller matrix. Spatial frequency is listed above each image. Nonrealizable Mueller matrix pixels are depicted as blue. The spatial frequency sinusoidal pattern was oriented horizontally. The field of view of the images is 1.1  cm×1.1  cm.

JBO_27_12_126003_f003.png

The length of the fiber that is visible in the diattenuation images extends deeper into the PDMS as the spatial frequency is increased. Whereas SFDI is often used to reduce the probing depth, our observations suggests that the polarization information yields the opposite trend. Polarization preserving paths favor those which scatter the fewest times, and the combination of SFDI and MMP tends to enhance those paths.

6.2.

Optical Fiber Bundle

Figures 4 and 5 show the SFD Mueller matrix BRDF measured for the optical fiber bundle at f=0  cm1 and f=5  cm1, shown in the same manner as in Fig. 2, illuminated by the horizontal and vertical sinusoidal patterns, respectively. The image background is a glossy white ceramic. The results differ from those shown in Fig. 2 in that the SFD Mueller matrix has a significant imaginary component at the fiber bundle, especially when the spatial frequency pattern is oriented horizontally (see Fig. 4). The imaginary component arises, for example, because light incident on one side of the fiber bundle is emitted from the other side, both from scattering and from reflection inside the bundle. This behavior is less evident for the vertically oriented sinusoidal pattern (see Fig. 5), because this pattern selects light that is scattered and reflected horizontally, the opposite of the fiber bundle’s preferred scattering direction.

Fig. 4

SFD Mueller matrix BRDFs from the optic fiber bundle: (a) the Mueller matrix BRDF at f=0  cm1, (b) the real part of the SFD Mueller matrix BRDF at f=5  cm1, and (c) the imaginary part of the SFD Mueller matrix BRDF at f=5  cm1. The maximum value of the black and white scale used for Fr,11 at f=0  cm1 is six times the maximum value for Fr,11 of f=5  cm1. The scale for the normalized SFD Mueller matrix elements is shown at the bottom of the figure. The spatial frequency sinusoidal pattern for f=5  cm1 was oriented horizontally. The field of view of the images is 1.1 cm × 1.1 cm.

JBO_27_12_126003_f004.png

Fig. 5

SFD Mueller matrix BRDFs from the optic fiber bundle: (a) the Mueller matrix BRDF at f=0  cm1, (b) the real part of the SFD Mueller matrix BRDF at f=5  cm1, and (c) the imaginary part of the SFD Mueller matrix BRDF at f=5  cm1. The maximum value of the black and white scale used for Fr,11 at f=0  cm1 is seven times the maximum value for Fr,11 of f=5  cm1. The scale for the normalized SFD Mueller matrix elements is shown at the bottom of the figure. The spatial frequency sinusoidal pattern for f=5  cm1 was oriented vertically. The field of view of the images is 1.1  cm×1.1  cm.

JBO_27_12_126003_f005.png

Figures 6 and 7 show the polarimetric purity indices (P1,P2, and P3, described in Sec. 2) calculated using only the real part of the matrices at the different spatial frequencies and the horizontally and vertically patterned illumination, respectively. The color scale for Figs. 6 and 7 was chosen to highlight those values that are less than one, compared with those greater than one. Values greater than one signify that the SFD Mueller matrix is not a valid regular Mueller matrix. Whereas the Mueller matrices at f=0  cm1 are entirely realizable, as expected, the matrices in the location of the fiber bundle are not realizable for even the smallest nonzero spatial frequency for horizontal sinusoidal illumination (see Fig. 6). Starting from P1, values greater than one can be seen in the pixels that characterize the fiber bundle. The number of the nonrealizable pixels can be seen to increase with P2 and P3, where a majority of the fiber bundle is greater than one. The fraction of pixels within the fiber bundle having polarization purity indices greater than one also increases with spatial frequency for all three polarimetric purity indices. For vertical sinusoidal illumination, the imaginary parts are much weaker, and the fiber bundle remains mostly realizable until about f=3.3  cm1, where P3 becomes greater than one along the fiber bundle’s core (see Fig. 7). P3 of the fiber bundle is entirely greater than one by f=5  cm1. The background ceramic tile does not show nonrealizable behavior at the spatial frequencies shown. Due to the strong presence of nonrealizable Mueller matrices measured from the fiber bundle itself and uncertainty as to how such matrices should be interpreted, further Mueller matrix decomposition was not performed. These results show that SFD Mueller matrix imaging has contrast mechanisms that are very different than those observed in either SFDI or MMP alone.

Fig. 6

Polarimetric purity indices (rows) for each spatial frequency (columns) of an optic fiber bundle. The color scale is shown with a discontinuity at unity, differentiating those regions where the matrix is a valid regular Mueller matrix from those that are not. The spatial frequency sinusoidal patterns were oriented horizontally. The field of view of the images is 1.1  cm×1.1  cm.

JBO_27_12_126003_f006.png

Fig. 7

Polarimetric purity indices (rows) for each spatial frequency (columns) of an optic fiber bundle. The color scale is shown with a discontinuity at unity, differentiating those regions where the matrix is a valid regular Mueller matrix from those that are not. The spatial frequency sinusoidal patterns were oriented vertically. The field of view of the images is 1.1  cm×1.1  cm.

JBO_27_12_126003_f007.png

6.3.

Caprine Cerebellum

Figure 8 shows the SFD Mueller matrix BRDF images of the excised caprine cerebellum at f=0  cm1 and f=5  cm1, shown in the same manner as in Fig. 2. The imaginary part of the normalized SFD Mueller matrix elements were multiplied by a factor of 5 so that their features would be more visible on the same color scale used for the real part of the SFD Mueller matrices. Due to the moisture present, the unevenness, and curvature of the surfaces in the excised tissue, some areas of specular reflectance could not be avoided during the measurement. These localized and saturated pixels were masked and are seen as white pixels in the SFD Mueller matrix images. Specular reflection was less of an issue in our previous samples, where we were better able to mitigate it.

Fig. 8

SFD Mueller matrix BRDF images of a caprine cerebellum: (a) the Mueller matrix BRDF at f=0  cm1, (b) the real part of the SFD Mueller matrix BRDF at f=5  cm1, and (c) the imaginary part of the SFD Mueller matrix BRDF multiplied by a factor of 5 at f=5  cm1. The maximum value of the black and white scale used for Fr,11 at f=0  cm1 is four times the maximum value for Fr,11 of f=5  cm1. The scale for the normalized SFD Mueller matrix elements is shown at the bottom of the figure. The spatial frequency sinusoidal pattern for f=5  cm1 was oriented horizontally. The field of view of the images is 1.1  cm×1.1  cm.

JBO_27_12_126003_f008.png

The branching structure observed in the images in Fig. 8 is white matter. Myelin sheaths that surround nerve fibers give those regions a different color and higher reflectance than the surrounding gray matter, which consists of cell bodies. The Mueller matrix at f=0  cm1 [see Fig. 8(a)] shows less contrast between the white and gray matters compared to the SFD Mueller matrix at f=5  cm1 [see Fig. 8(b)]. This difference in contrast is strongest between the Fr,11 elements of both spatial frequencies where the delineation between the two tissue types improves with greater spatial frequency.43 The outline of the white matter can be seen to various degrees in the other Mueller matrix elements at f=0  cm1 but with less clarity compared to the higher SFD Mueller matrix elements. The gray matter in the real part of the SFD Mueller matrix at f=5  cm1 shows large magnitudes particularly along the matrix diagonal where the white matter maintains values close to zero.

Perhaps the most interesting feature for the cerebellum data is the imaginary component at f=5  cm1. Whereas the white matter exhibits only real SFD Mueller matrix values (imaginary part close to zero), the gray matter shows a striking pattern in contrast. The SFD Mueller matrix becomes complex-valued when the scattering is directionally unbalanced. The striped patterns seen in the imaginary part of the SFD Mueller matrix data appear to wrap around the borders of the white matter and are believed to be related to anisotropic scattering from the highly reflective white matter towards the gray matter. As stated previously the white matter is comprised of nerve fibers wrapped in myelin sheathes, making the white matter much more anisotropic in structure compared to the gray matter. The patterns observed in Fig. 8(c) show opposite signs above and below the white matter. For vertically oriented sinusoidal illumination, the pattern of opposite signs shows up left and right (not shown, but available in the online data).

6.4.

Bovine Thalamus

Figure 9 shows SFD Mueller matrix BRDF images from the excised bovine thalamus at f=0 and f=5  cm1. The change in texture seen at the bottom of the images is the white background beneath the Petri dish that the brain tissue is resting on. Similar to the caprine cerebellum, both white and gray matter are present in the image area. Structurally, the thalamus is a mass of gray matter at the center of the mammalian brain with myelinated axons branching internally and externally from it to reach the surrounding brain areas. In Fig. 9, the contrast between the white and gray matter is significantly greater at f=5  cm1 compared with that at f=0  cm1. As with the cerebellum data, the white matter maintains a mostly real-valued SFD Mueller matrix (imaginary part negligible) at f=5  cm1. Interestingly, a marble pattern of white matter can be seen in the center of the gray matter of the Fr,11 elements at f=5  cm1 that is difficult to observe at f=0  cm1. This white matter also maintains near-zero values in the SFD Mueller matrix element at f=5  cm1. This is likely part of the external medullary laminae, which are white matter structures that cover the lateral surface of the thalamus.44 Perhaps because the white matter is not so uniformly aligned in this sample, and because the spatial frequency is not high enough, there is not enough directional scattering to demonstrate as strong a pattern in the imaginary part of the SFD Mueller matrix compared with what was observed in the cerebellum data.

Fig. 9

SFD Mueller matrix BRDF images of a bovine thalamus: (a) the Mueller matrix BRDF at f=0  cm1, (b) the real part of the SFD Mueller matrix BRDF at f=5  cm1, and (c) the imaginary part of the SFD Mueller matrix BRDF multiplied by a factor of 5 at f=5  cm1. The maximum value of the black and white scale used for Fr,11 at f=0  cm1 is 3.5 times the maximum value for Fr,11 of f=5  cm1. The scale for the normalized SFD Mueller matrix elements is shown at the bottom of the figure. The spatial frequency sinusoidal pattern for f=5  cm1 was oriented horizontally. The field of view of the images is 1.1  cm×1.1  cm.

JBO_27_12_126003_f009.png

6.5.

Bovine Cerebrum

Figure 10 shows SFD Mueller matrix images from the inner surface area of the excised bovine cerebrum at f=0  cm1 and f=5  cm1. Like the caprine cerebellum and the bovine thalamus, white matter and gray matter are distinguishable. At f=0  cm1, the Fr,11 image is dominated by white matter, except for the lower right section of the image. The SFD Mueller matrix at f=5  cm1 shows areas of gray matter beneath the less dense sections of white matter that were not visible at f=0  cm1. The borders between the white and gray matter are stronger for this SFD Mueller matrix compared with the f=0  cm1 Mueller matrix. This delineation is most apparent in the diagonal of the real part of the SFD Mueller matrix at f=5  cm1, where the white matter retains a value close to zero and the gray matter makes up most of the imaginary component.

Fig. 10

SFD Mueller matrix BRDF images of a bovine cerebrum: (a) the Mueller matrix BRDF at f=0  cm1, (b) the real part of the SFD Mueller matrix BRDF at f=5  cm1, and (c) the imaginary part of the SFD Mueller matrix BRDF multiplied by a factor of 5 at f=5  cm1. The maximum value of the black and white scale used for Fr,11 at f=0  cm1 is three times the maximum value for Fr,11 of f=5  cm1. The scale for the normalized SFD Mueller matrix elements is shown at the bottom of the figure. The spatial frequency sinusoidal pattern for f=5  cm1 was oriented horizontally. The field of view of the images is 1.1  cm×1.1  cm.

JBO_27_12_126003_f010.png

7.

Discussion

The SFD Mueller matrix images shown in Sec. 6 demonstrate contrast mechanisms that differ from those observed in either SFDI or regular Mueller matrix imaging. In most of our results, the SFD Mueller matrix has higher contrast as the spatial frequency is increased. This finding highlights how combining MMP with SFDI can facilitate measurements of the polarimetric properties of the surfaces of diffusively scattering materials, which is difficult to measure with MMP alone.

There are other aspects of the results which we can point out and qualitatively understand. Scattering can be treated generally as the sum of contributions from single scatter, double scatter, and diffusive scatter (those paths that scatter three or more times). Single scatter should exhibit a flat SFD response, because the point spread function associated with it is highly localized. In addition, single scatter should yield a real-valued SFD Mueller matrix that obeys the realizability rules for regular Mueller matrices. We observe this in the PDMS-based scattering phantom in the region where the para-aramid fibers are only shallowly embedded (see Figs. 2 and 3).

Diffusive scatter is expected to be highly depolarizing. Some polarizing behavior can arise from the polarization dependence of Fresnel transmission in and out of the material and the polarization dependence of the first and last scattering events.45 We can see this in the data, especially for the background materials (PDMS in Fig. 2 or ceramic in Figs. 4 and 5) in many of the images. Because the incident radiation is incident at a 20 deg angle and viewed from the surface normal, we observe diattenuation (as evident by negative values for M12) consistent with this geometry, where p-polarization has higher transmittance into the material.

Double scatter involves the spatial separation between two scattering events. For horizontal sinusoidal irradiance, if the two scatter events are vertically displaced, there would be a phase imparted in the SFD image if there is more scattering upwards than downwards or vice versa. In unpolarized SFD imaging, it is impossible to separate phase shifts that occur due to topography from those that occur due to the directionality of the scatter. However, with SFD Mueller matrix imaging, when one polarization scatters more than another in a geometrically unbalanced fashion, these phase shifts can be distinguished. For example, consider the fiber bundle data shown in Fig. 5(c), where there is a strong imaginary component to the SFD Mueller matrix. From Fig. 6, the SFD Mueller matrix (as reduced to polarization purity indices) varies significantly with spatial frequency. This strong spatial frequency dependence results from the matching of the irradiance period with the dimensions of the fiber bundle and the multiple internal reflections in it. This effect does not occur as strongly when the sinusoidal irradiation is perpendicular to the fiber bundle (compare Figs. 6 with 7). Thus, SFD Mueller matrix imaging allows the directional scattering within a medium to be probed.

These directional effects of the scattering can also be observed in the imaginary part of the SFD Mueller matrix measured from the various brain specimens, especially the caprine cerebellum shown in Fig. 8(c). Whereas the imaginary part of the SFD Mueller matrix is weak in the regions containing white matter, it tends to have significant values in neighboring gray matter. For example, in ImFr,22, in areas immediately above the white matter, the values are negative, while in those below, the values are positive. The change in the sign above and below the white matter results from the displacement of the photon transfer being in opposite directions, with the polarization behavior otherwise being the same. These effects can be observed to a lesser extent in the bovine thalamus data in Fig. 9. However, because the length scales for the white matter features in the thalamus are smaller and the length associated with the spatial frequency is long, the effect is relatively weak.

At this time, the analysis is hampered by the lack of a framework for characterizing the resulting matrices. As mentioned in Sec. 2, the properties of a SFD Mueller matrix are not the same as a regular Mueller matrix, in that it does not express the relationship between a Stokes vector irradiance and a Stokes vector radiance. Because a SFD Stokes vector radiance does not follow Eq. (2) and because it is complex-valued, the SFD Mueller matrix does not have the same requirements as a regular Mueller matrix. Thus, it is not straightforward to apply decomposition methods, such as Eq. (18), to the results. It is expected that diattenuation angles and retardance angles could be extracted from data using these traditional methods, because rotation transformations are not expected to change in the SFD. However, the current understanding of the mathematics of depolarization is intimately related to the requirement that the eigenvalues of the covariance matrix H be positive. Thus, further work needs to go into developing the analysis infrastructure to fully interpret the results.

8.

Summary

We demonstrated an imaging modality that combines MMP and SFDI. We demonstrated theoretically that the combined SFD Mueller matrix imaging at nonzero frequency is best described as a complex-valued matrix, having phase and amplitude information and that this phase information can be differentiated from that caused by topography or shape. Whereas this measurement method is very much in its infancy, we show that it has potential for distinguishing the polarization behavior for different photon path lengths and scattering directions in a material. This imaging modality may have applications in the field of biomedical imaging where thick and structurally complex tissues are commonplace. Depending on the tissue type, selecting for photon path length can potentially highlight different structures such as how the contrast between the white matter and gray matter improved with increased spatial frequency in the brain samples shown. Brain tissue has been the subject of recent interest in the field of MMP.8 Mueller matrix imaging has been used to observe changes that occur during pregnancy or disease in cervical collagen anisotropy.46 One difficulty in imaging cervical collagen is that it is surrounded by other tissues of lesser interest. Imaging modalities that allow for selection criteria for what is imaged may have potential in similar settings. Complicating the analysis and interpretation is the finding that the SFD Mueller matrix is inherently complex-valued and does not represent a realizable regular Mueller matrix. However, it is observed that contrast to different features in the image can be enhanced and that single and multiply scattered paths can be distinguished. Future work will continue to explore the utility of this imaging method for biomedical applications.

Disclosures

The authors have no conflicts of interest to report.

Acknowledgments

The authors would like to thank Dr. Jeeseong C. Hwang for fabricating the para-aramid fiber PDMS phantom. Certain commercial equipment, instruments, or materials are identified in this paper to specify the experimental procedure adequately. Such identification is not intended to imply recommendation or endorsement by the National Institute of Standards and Technology (NIST), nor is it intended to imply that the materials or equipment identified are necessarily the best available for the purpose.

Data Availability

The data presented in this manuscript are available for download online at the NIST Public Data Repository, https://doi.org/10.18434/mds2-2694.

References

1. 

N. Ghosh, M. F. Wood and I. A. Vitkin, “Mueller matrix decomposition for extraction of individual polarization parameters from complex turbid media exhibiting multiple scattering, optical activity, and linear birefringence,” J. Biomed. Opt., 13 (4), 044036 https://doi.org/10.1117/1.2960934 JBOPFO 1083-3668 (2008). Google Scholar

2. 

A. Pierangelo et al., “Multispectral Mueller polarimetric imaging detecting residual cancer and cancer regression after neoadjuvant treatment for colorectal carcinomas,” J. Biomed. Opt., 18 (4), 046014 https://doi.org/10.1117/1.JBO.18.4.046014 JBOPFO 1083-3668 (2013). Google Scholar

3. 

G. Jarry et al., “Coherence and polarization of light propagating through scattering media and biological tissues,” Appl. Opt., 37 (31), 7357 –7367 https://doi.org/10.1364/AO.37.007357 APOPAI 0003-6935 (1998). Google Scholar

4. 

J. Vizet et al., “In vivo imaging of uterine cervix with a Mueller polarimetric colposcope,” Sci. Rep., 7 2471 https://doi.org/10.1038/s41598-017-02645-9 SRCEC3 2045-2322 (2017). Google Scholar

5. 

J. Chue-Sang et al., “Use of Mueller matrix colposcopy in the characterization of cervical collagen anisotropy,” J. Biomed. Opt., 23 (12), 121605 https://doi.org/10.1117/1.JBO.23.12.121605 JBOPFO 1083-3668 (2018). Google Scholar

6. 

M. Menzel et al., “Diattenuation imaging reveals different brain tissue properties,” Sci Rep-UK, 9 1939 https://doi.org/10.1038/s41598-019-38506-w (2019). Google Scholar

7. 

H. R. Lee et al., “Digital histology with Mueller microscopy: how to mitigate an impact of tissue cut thickness fluctuations,” J. Biomed. Opt., 24 (7), 076004 https://doi.org/10.1117/1.JBO.24.7.076004 JBOPFO 1083-3668 (2019). Google Scholar

8. 

O. Rodríguez-Núñez and T. Novikova, “Polarimetric techniques for the structural studies and diagnosis of brain,” Adv. Opt. Technol., 11 (5-6), 157 –171 https://doi.org/10.1515/aot-2022-0015 1687-6393 (2022). Google Scholar

9. 

O. Rodríguez-Núñez et al., “Polarimetric visualization of healthy brain fiber tracts under adverse conditions: ex vivo studies,” Biomed. Opt. Express, 12 (10), 6674 –6685 https://doi.org/10.1364/BOE.439754 BOEICL 2156-7085 (2021). Google Scholar

10. 

N. Agarwal et al., “Spatial evolution of depolarization in homogeneous turbid media within the differential Mueller matrix formalism,” Opt. Lett., 40 (23), 5634 –5637 https://doi.org/10.1364/OL.40.005634 OPLEDP 0146-9592 (2015). Google Scholar

11. 

H. R. Lee et al., “Mueller microscopy of anisotropic scattering media: theory and experiments,” Proc. SPIE, 10677 1067718 https://doi.org/10.1117/12.2306943 PSISDG 0277-786X (2018). Google Scholar

12. 

L. Trifonyuk et al., “Differential Mueller matrix imaging of partially depolarizing optically anisotropic biological tissues,” Laser Med. Sci., 35 (4), 877 –891 https://doi.org/10.1007/s10103-019-02878-2 (2020). Google Scholar

13. 

M. Peyvasteh et al., “Two-point Stokes vector diagnostic approach for characterization of optically anisotropic biological tissues,” J. Phys. D Appl. Phys., 53 (39), 395401 https://doi.org/10.1088/1361-6463/ab9571 (2020). Google Scholar

14. 

M. Peyvasteh et al., “3D Mueller-matrix-based azimuthal invariant tomography of polycrystalline structure within benign and malignant soft-tissue tumours,” Laser Phys. Lett., 17 (11), 115606 https://doi.org/10.1088/1612-202X/abbee0 1612-2011 (2020). Google Scholar

15. 

A. Ushenko et al., “Stokes-correlometry analysis of biological tissues with polycrystalline structure,” IEEE J. Sel. Top. Quantum Electron., 25 (1), 7101612 https://doi.org/10.1109/JSTQE.2018.2865443 IJSQEN 1077-260X (2019). Google Scholar

16. 

V. Ushenko et al., “3D Mueller-matrix diffusive tomography of polycrystalline blood films for cancer diagnosis,” Photonics-Basel, 5 (4), 54 https://doi.org/10.3390/photonics5040054 (2018). Google Scholar

17. 

V. A. Ushenko et al., “Biomedical applications of Jones-matrix tomography to polycrystalline films of biological fluids,” J. Innov. Opt. Health Sci., 12 (6), 1950017 https://doi.org/10.1142/S1793545819500172 (2019). Google Scholar

18. 

M. Borovkova et al., “Mueller-matrix-based polarization imaging and quantitative assessment of optically anisotropic polycrystalline networks,” PLoS One, 14 (5), https://doi.org/10.1371/journal.pone.0214494 POLNCL 1932-6203 (2019). Google Scholar

19. 

J. Rehbinder et al., “Ex vivo Mueller polarimetric imaging of the uterine cervix: a first statistical evaluation,” J. Biomed. Opt., 21 (7), 071113 https://doi.org/10.1117/1.JBO.21.7.071113 JBOPFO 1083-3668 (2016). Google Scholar

20. 

J. Chue-Sang et al., “Optical phantoms for biomedical polarimetry: a review,” J. Biomed. Opt., 24 (3), 030901 https://doi.org/10.1117/1.JBO.24.3.030901 JBOPFO 1083-3668 (2019). Google Scholar

21. 

J. M. Bueno et al., “Enhancement of confocal microscopy images using Mueller-matrix polarimetry,” J. Microsc., 235 (1), 84 –93 https://doi.org/10.1111/j.1365-2818.2009.03185.x JMICAR 0022-2720 (2009). Google Scholar

22. 

I. Saytashev et al., “Self validating Mueller matrix micro-mesoscope (SAMMM) for the characterization of biological media,” Opt. Lett., 45 (8), 2168 –2171 https://doi.org/10.1364/OL.387747 OPLEDP 0146-9592 (2020). Google Scholar

23. 

J. Angelo et al., “Review of structured light in diffuse optical imaging,” J. Biomed. Opt., 24 (7), 071602 https://doi.org/10.1117/1.JBO.24.7.071602 JBOPFO 1083-3668 (2018). Google Scholar

24. 

B. Yang et al., “Polarized light spatial frequency domain imaging for non-destructive quantification of soft tissue fibrous structures,” Biomed. Opt. Express, 6 (4), 1520 –1533 https://doi.org/10.1364/BOE.6.001520 BOEICL 2156-7085 (2015). Google Scholar

25. 

B. Kunnen et al., “Application of circularly polarized light for non-invasive diagnosis of cancerous tissues and turbid tissue-like scattering media,” J. Biophotonics, 8 (4), 317 –323 https://doi.org/10.1002/jbio.201400104 (2015). Google Scholar

26. 

S. V. Jett et al., “Integration of polarized spatial frequency domain imaging (pSFDI) with a biaxial mechanical testing system for quantification of load-dependent collagen architecture in soft collagenous tissues,” Acta Biomater., 102 149 –168 https://doi.org/10.1016/j.actbio.2019.11.028 (2020). Google Scholar

27. 

J. P. Angelo, T. A. Germer and M. Litorja, “Structured illumination Mueller matrix imaging,” Biomed. Opt. Express, 10 (6), 2861 –2868 https://doi.org/10.1364/BOE.10.002861 BOEICL 2156-7085 (2019). Google Scholar

28. 

J. Chue-Sang et al., “Spatial frequency domain Mueller matrix imaging,” Proc. SPIE, 11646 116460E https://doi.org/10.1117/12.2576350 PSISDG 0277-786X (2021). Google Scholar

29. 

F. E. Nicodemus et al., Geometrical Considerations and Nomenclature for Reflectance, National Bureau of Standards, Gaithersburg (1977). Google Scholar

30. 

T. A. Germer, J. C. Zwinkels and B. K. Tsai, “Theoretical concepts in spectrophotometric measurements,” Spectrophotometry: Accurate Measurements of Optical Properties of Materials, Elsevier, Amsterdam (2014). Google Scholar

31. 

M. Wood, X. Guo and A. Vitkin, “Polarized light propagation in multiply scattering media exhibiting both linear birefringence and optical activity: Monte Carlo model and experimental methodology,” J. Biomed. Opt., 12 (1), 014029 https://doi.org/10.1117/1.2434980 JBOPFO 1083-3668 (2007). Google Scholar

32. 

M. R. Antonelli et al., “Mueller matrix imaging of human colon tissue for cancer diagnostics: how Monte Carlo modeling can help in the interpretation of experimental data,” Opt. Express, 18 (10), 10200 –10208 https://doi.org/10.1364/OE.18.010200 OPEXFF 1094-4087 (2010). Google Scholar

33. 

V. V. Tuchin, Tissue Optics: Light Scattering Methods and Instruments for Medical Diagnostics, 3rd ed.SPIE Press, Bellingham, Washington (2015). Google Scholar

34. 

R. Simon, “The Connection between Mueller and Jones matrices of polarization optics,” Opt. Commun., 42 (5), 293 –297 https://doi.org/10.1016/0030-4018(82)90234-6 OPCOB8 0030-4018 (1982). Google Scholar

35. 

S. R. Cloude, “Group theory and polarisation algebra,” Optik, 75 26 –36 OTIKAJ 0030-4026 (1986). Google Scholar

36. 

A. Van Eeckhout et al., “Synthesis and characterization of depolarizing samples based on the indices of polarimetric purity,” Opt. Lett., 42 (20), 4155 –4158 https://doi.org/10.1364/OL.42.004155 OPLEDP 0146-9592 (2017). Google Scholar

37. 

D. J. Cuccia et al., “Quantitation and mapping of tissue optical properties using modulated imaging,” J. Biomed. Opt., 14 (2), 024012 https://doi.org/10.1117/1.3088140 JBOPFO 1083-3668 (2009). Google Scholar

38. 

B. Laude-Boulesteix et al., “Mueller polarimetric imaging system with liquid crystals,” Appl. Opt., 43 (14), 2824 –2832 https://doi.org/10.1364/AO.43.002824 APOPAI 0003-6935 (2004). Google Scholar

39. 

E. Compain, S. Poirier and B. Drevillon, “General and self-consistent method for the calibration of polarization modulators, polarimeters, and Mueller-matrix ellipsometers,” Appl. Opt., 38 (16), 3490 –3502 https://doi.org/10.1364/AO.38.003490 APOPAI 0003-6935 (1999). Google Scholar

40. 

S. Y. Lu and R. A. Chipman, “Interpretation of Mueller matrices based on polar decomposition,” J. Opt. Soc. Am. A, 13 (5), 1106 –1113 https://doi.org/10.1364/JOSAA.13.001106 JOAOD6 0740-3232 (1996). Google Scholar

41. 

A. M. Goldfain et al., “Polydimethylsiloxane tissue-mimicking phantoms with tunable optical properties,” J. Biomed. Opt., 27 (7), 074706 https://doi.org/10.1117/1.JBO.27.7.074706 JBOPFO 1083-3668 (2021). Google Scholar

42. 

P. Lemaillet et al., “Correction of an adding-doubling inversion algorithm for the measurement of the optical parameters of turbid media,” Biomed. Opt. Express, 9 (1), 55 –71 https://doi.org/10.1364/BOE.9.000055 BOEICL 2156-7085 (2018). Google Scholar

43. 

D. J. Cuccia et al., “Modulated imaging: quantitative analysis and tomography of turbid media in the spatial-frequency domain,” Opt. Lett., 30 (11), 1354 –1356 https://doi.org/10.1364/OL.30.001354 OPLEDP 0146-9592 (2005). Google Scholar

44. 

T. J. Torrico and S. Munakomi, “Neuroanatomy, Thalamus,” StatPearls Publishing, Treasure Island, Florida (2021). Google Scholar

45. 

T. Germer, “Polarized light diffusely scattered under smooth and rough interfaces,” Proc. SPIE, 5158 193 –204 https://doi.org/10.1117/12.505979 PSISDG 0277-786X (2003). Google Scholar

46. 

H. R. Lee et al., “Mueller matrix imaging for collagen scoring in mice model of pregnancy,” Sci. Rep., 11 15621 https://doi.org/10.1038/s41598-021-95020-8 SRCEC3 2045-2322 (2021). Google Scholar

Biography

Joseph Chue-Sang received his BS and PhD degrees in biomedical engineering from Florida International University in 2013 and 2019, respectively. He was working as a postdoctoral associate employed by the University of Maryland under the National Institute of Standards and Technology Professional Research Experience Program. His dissertation focused on the integration of polarization imaging in a clinical setting for the detection of preterm birth. He also specializes in optical coherence tomography.

Maritoni Litorja received her PhD in physical/analytical chemistry from Northwestern University in 1996 and was working as a postdoctoral fellow at Argonne National Labs before joining NIST in 1999, where she has been a research chemist ever since. Her research interest includes finding ways to establish SI traceability of image-derived quantities, especially for biomedical applications. She is a Member of SPIE.

Aaron M. Goldfain received his BA degree from the University of Colorado in 2012 and his PhD from Harvard University in 2018. Currently, he is working as a physicist in the Sensor Science Division of the National Institute of Standards and Technology. His research interests include characterizing light scatter from surfaces and particles.

Thomas A. Germer received his BA degree from the University of California, Berkeley, in 1985, and MS and PhD degrees from Cornell University in 1988 and 1992, respectively. He has been working as a physicist at the National Institute of Standards and Technology since 1992. He specializes in the measurement and modeling of polarized light scattering from surfaces. He is a Fellow Member of SPIE and Optica.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Joseph Chue-Sang, Maritoni Litorja, Aaron M. Goldfain, and Thomas A. Germer "Spatial frequency domain Mueller matrix imaging," Journal of Biomedical Optics 27(12), 126003 (14 December 2022). https://doi.org/10.1117/1.JBO.27.12.126003
Received: 22 August 2022; Accepted: 28 November 2022; Published: 14 December 2022